K Number
K190337

Validate with FDA (Live)

Device Name
ReWalk Restore
Date Cleared
2019-06-03

(109 days)

Product Code
Regulation Number
890.3480
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ReWalk ReStore is intended to be used to assist ambulatory functions in rehabilitation under the supervision of a trained therapist for people with hemiplegia/hemiparesis due to stroke who can ambulate at least 1.5m (5ft) with no more than minimal to moderate levels of assistance. The trained therapist must successfully complete a training program prior to operating the device. The device is not intended for sports or stair climbing.

Device Description

The ReWalk ReStore™ ("ReStore™") device is a lightweight, wearable, battery-powered soft orthotic exosuit is intended to be used to assist ambulatory functions in rehabilitation institutions under the supervision of a trained therapist for people with hemiplegia/hemiparesis due to stroke.

The ReStore™ device is designed to provide plantarflexion (PF) and dorsiflexion (DF) assistance to the paretic ankle during walking by transferring mechanical forces from two motors mounted on a Waistpack through Bowden cables to attachment points worn on the calf and insole. Through a set of gears, the motors actuate the Bowden cables to provide plantarflexion and dorsiflexion forces. Forces are transferred through the Bowden cables by pulling an inner cable within an outer sheath. The distal end of the outer sheath is fixed at an attachment point worn on the Calf Wrap, and the distal end of the inner cable is fixed at an attachment point worn on an insole made of flexible plastic that is worn within the shoe on the paretic foot. The relative motion that occurs between these two points when the inner cable is moved results in torques being applied across the ankle joint. The placement of the attachment points anterior and posterior to the ankle joint result in corresponding torques to provide assistance during dorsiflexion and plantarflexion, respectively. Load cells attached to the cables provide force feedback allowing control of the level of assistance provided to the patient. Sensors attached to the shoes are used to detect gait events during walking to inform timing of the assistance from the ReStore™ device. Specifically, inertial measurement units (IMUs) are attached to shoes on both legs to provide signals corresponding to events within the gait cycle, such as heel-strike or toeoff, in order to determine the appropriate timing of assistance from ReStore™.

The ReStore™ is designed to provide adjustable levels of assistance to the patient's paretic leg. It does not constrain or provide assistance to the patient's non-paretic leg. The device enables the therapist to individually modify the level of assistance provided during plantarand dorsiflexion as appropriate for each individual patient. The timing of assistance is based on data collected from the motion sensors on both the paretic and non-paretic feet, which determines the patient's orientation within their gait cycle.

Plantarflexion assistance occurs during the late stance and toe-off portion of the paretic leg gait cycle. This level of assistance targets a desired cable force, which can be adjusted from 0% to 25% of the patient's body weight. Dorsiflexion assistance occurs during the swing and heel-strike phase of the paretic gait cycle. This level of assistance targets a desired cable travel, which can be adjusted from 0 cm to 5 cm. Plantarflexion and dorsiflexion assistance levels are determined by the therapist based on visual evaluation of the patient's gait symmetry, the therapeutic goals of the walking session or the ability of the patient to achieve consistent ground clearance or heel-strike landing patterns with their paretic foot. The ReStore™ device includes apparel-like components, mechanical components, electrical components, cables, sensors and a handheld controller.

AI/ML Overview

The ReWalk Restore device is a soft orthotic exosuit intended to assist ambulatory functions in rehabilitation for individuals with hemiplegia/hemiparesis due to stroke.

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily focuses on demonstrating the safety and effectiveness of the ReWalk Restore device against a predicate device (Indego®) through clinical and non-clinical studies. While explicit "acceptance criteria" in a quantitative sense for performance metrics are not clearly defined as pass/fail thresholds in the provided text, the successful completion of the clinical trial demonstrating safety and efficacy, alongside meeting all design requirements and compliance with standards, serves as the basis for acceptance. The performance is reported in terms of safety outcomes and improvements in walking parameters.

Acceptance Criteria (Implied)Reported Device Performance (ReWalk Restore)
Device Safety (no device-related serious adverse events or falls)PASS: Zero device-related serious adverse events and zero device-related falls in the 44-subject multi-site clinical trial and engineering studies.
Manageable Adverse EventsPASS: Initial adverse event rate of 13.5% in early study phase, reduced to 1.3% after retraining on donning/fitting. Majority were minor.
Improved Walking Speed (Comfortable)PASS: In subjects completing all 7 visits (n=36), 86.1% increased comfortable walking speed with ReStore™.
Improved Walking Speed (Maximal)PASS: In subjects completing all 7 visits (n=36), 80.6% increased maximal walking speed with ReStore™.
Increased Walking Distance (2-minute walk test)PASS: Average walking distance increased by 47.9 ft. during walking with assistance (Assist mode) compared to without assistance (Slack mode).
Device ReliabilityPASS: Not explicitly quantified with a threshold, but assessed as a secondary endpoint and implied by successful study completion.
Device Safety for Physical TherapistPASS: Not explicitly quantified with a threshold, but assessed as a secondary endpoint and implied by successful study completion.
Compliance with Design Requirements/StandardsPASS: Met all design inputs and specifications, and complied with applicable standards (e.g., ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 10993-1/5/10).

2. Sample Size and Data Provenance for Test Set:

  • Sample Size for Clinical Trial: 44 subjects for the multi-site clinical trial.
  • Sample Size for Engineering Studies: 35 unique participants across two ongoing engineering studies (with 10 participants enrolled in both).
  • Sample Size for Clinical Validation Substudy: 9 participants.
  • Total Unique Participants: 79 individuals.
  • Data Provenance: The multi-site clinical trial was conducted across 5 study sites, and the engineering studies were conducted at the Wyss Institute at Harvard University (Cambridge, MA). The country of origin is the USA for the engineering and clinical validation studies, and likely includes multiple sites/countries for the "multi-site clinical trial" but specific country details are not provided beyond the Wyss Institute location for some studies. All studies were prospective as they were IRB-approved and conducted according to Good Clinical Practice (GCP) principles.

3. Number of Experts and Qualifications for Ground Truth:

The document does not detail the number of experts used to establish the ground truth for specific performance metrics in the way one might see in an image classification study. Instead, the "ground truth" for the device's safety and effectiveness is established through a clinical trial involving trained therapists supervising patients. Their observations and assessments during the trials, along with objective measurements of walking parameters, form the basis of evaluating the device's performance. The document states, "The trained therapist must successfully complete a training program prior to operating the device," indicating a level of qualification for the individuals involved in using and assessing the device. Specific qualifications like "radiologist with 10 years of experience" are not applicable to this type of device and study.

4. Adjudication Method for the Test Set:

The document does not describe "adjudication methods" in the context of expert consensus (e.g., 2+1, 3+1). The primary endpoints for the clinical trial are device safety (frequency and severity of device-related adverse events, falls) and exploratory endpoints for efficacy (walking speed, distance). These are assessed through direct observation and measurement during the study, rather than through expert consensus on qualitative events. Adverse event reporting and management would involve a clinical team review rather than an adjudication panel for "ground truth."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study is more relevant to diagnostic imaging interpretation (where multiple readers interpret cases). The ReWalk Restore is a therapeutic and assistive device, and its effectiveness is measured by direct patient outcomes and performance improvements in ambulation, not by human reader interpretation.

6. Standalone (Algorithm Only) Performance:

No, a standalone (algorithm only) performance study was not done. The ReWalk Restore device is inherently human-in-the-loop, designed to provide assistance to a person with hemiparesis/hemiplegia during ambulation. Its function is to augment human movement, not to perform independently. The performance is always assessed with the human user.

7. Type of Ground Truth Used:

The ground truth for device safety was based on the occurrence and severity of device-related adverse events and falls as reported and managed during the clinical trials. For efficacy, the ground truth was based on objective physiological and functional outcome measures such as 10-meter walk tests (comfortable and maximal speeds) and 2-minute walk distances, directly measured from the subjects during their walking assessments.

8. Sample Size for the Training Set:

The document does not explicitly mention a "training set" in the context of machine learning. The studies described are for clinical validation and engineering development. If any machine learning components are part of the device's control system, the data used for their development/training is not detailed in this summary. The 79 individuals who participated in the various studies represent the subjects on whom the device was tested and refined, rather than a distinct "training set" for an AI algorithm in the traditional sense of a diagnostic or predictive model.

9. How the Ground Truth for the Training Set Was Established:

As there is no clearly defined "training set" for an AI algorithm in the provided text, the method for establishing ground truth for such a set is not described. The studies focused on validating the device's safety and effectiveness through direct clinical assessment and objective measures in human participants.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 3, 2019

ReWalk Robotics Ltd. Ofir Koren General Manager, VP R&D and Regulatory Affairs 3 Hetnufa St., POB Box- 161 Yokneam, 2069203 Israel

Re: K190337

Trade/Device Name: ReWalk Restore Regulation Number: 21 CFR 890.3480 Regulation Name: Powered Lower Extremity Exoskeleton Regulatory Class: Class II Product Code: PHL Dated: March 4, 2019 Received: March 5, 2019

Dear Ofir Koren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vivek Pinto, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190337

Device Name ReWalk Restore

Indications for Use (Describe)

The ReWalk ReStore is intended to be used to assist ambulatory functions in rehabilitation under the supervision of a trained therapist for people with hemiplegia/hemiparesis due to stroke who can ambulate at least 1.5m (5ft) with no more than minimal to moderate levels of assistance. The trained therapist must successfully complete a training program prior to operating the device. The device is not intended for sports or stair climbing.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for ReWalk Robotics. The logo has the word "ReWalk" in a combination of gray and yellow. Below the word "ReWalk" is the word "Robotics" in gray with a series of horizontal lines to the left of the word.

ReStore™ 510(k) Summary

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the ReWalk Robotics logo. The logo features the word "ReWalk" in a bold, stylized font, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in a smaller, gray font with lines to the left of the word. The logo is clean and modern, reflecting the company's focus on innovative robotic solutions.

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92

1. Submission Sponsor

ReWalk Robotics Ltd. 3 Hatnufa St. Yokneam, Israel 2069203. POB: 161 Establishment Registration: 3007615665

2. Submission Correspondent

Ofir Koren, General Manager, VP R&D and Regulatory Affairs

Phone: (+972) 54-52-25-792 Fax: (+972) 4-95-90-125 E-mail: Ofir@rewalk.com

3. Date Prepared:

June 3, 2019

4. Device Identification

Name of Device:ReWalk ReStore™
Classification Name:Powered Exoskeleton
Regulation:21 CFR §890.3480
Regulatory Class:Class II
Product Classification Code:PHL
Classification panel:Neurology

ട്. Legally Marketed Predicate Device

Predicate Manufacturer:Parker Hannifin Corporation
Predicate Trade Name:Indego®
Predicate 510(k):K173530

6. Device Description

The ReWalk ReStore™ ("ReStore™") device is a lightweight, wearable, battery-powered soft orthotic exosuit is intended to be used to assist ambulatory functions in rehabilitation institutions under the supervision of a trained therapist for people with hemiplegia/hemiparesis due to stroke.

The ReStore™ device is designed to provide plantarflexion (PF) and dorsiflexion (DF) assistance to the paretic ankle during walking by transferring mechanical forces from two motors mounted on a Waistpack through Bowden cables to attachment points worn on the calf and insole. Through a set of gears, the motors actuate the Bowden cables to provide plantarflexion and dorsiflexion forces. Forces are transferred through the Bowden cables by pulling an inner cable within an outer sheath. The distal end of the outer sheath is fixed at an attachment point worn on the Calf Wrap, and the distal end of the inner cable is fixed at an attachment point worn on an insole made of flexible plastic that is worn within the shoe on the paretic foot. The relative motion that occurs between these two points when the inner cable is moved results in torques being applied across the ankle joint. The

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for ReWalk Robotics. The logo features the word "ReWalk" in a combination of gray and yellow colors, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in a smaller font size and gray color. There are also some gray lines below the word "ReWalk".

placement of the attachment points anterior and posterior to the ankle joint result in corresponding torques to provide assistance during dorsiflexion and plantarflexion, respectively. Load cells attached to the cables provide force feedback allowing control of the level of assistance provided to the patient. Sensors attached to the shoes are used to detect gait events during walking to inform timing of the assistance from the ReStore™ device. Specifically, inertial measurement units (IMUs) are attached to shoes on both legs to provide signals corresponding to events within the gait cycle, such as heel-strike or toeoff, in order to determine the appropriate timing of assistance from ReStore™.

The ReStore™ is designed to provide adjustable levels of assistance to the patient's paretic leg. It does not constrain or provide assistance to the patient's non-paretic leg. The device enables the therapist to individually modify the level of assistance provided during plantarand dorsiflexion as appropriate for each individual patient. The timing of assistance is based on data collected from the motion sensors on both the paretic and non-paretic feet, which determines the patient's orientation within their gait cycle.

Plantarflexion assistance occurs during the late stance and toe-off portion of the paretic leg gait cycle. This level of assistance targets a desired cable force, which can be adjusted from 0% to 25% of the patient's body weight. Dorsiflexion assistance occurs during the swing and heel-strike phase of the paretic gait cycle. This level of assistance targets a desired cable travel, which can be adjusted from 0 cm to 5 cm. Plantarflexion and dorsiflexion assistance levels are determined by the therapist based on visual evaluation of the patient's gait symmetry, the therapeutic goals of the walking session or the ability of the patient to achieve consistent ground clearance or heel-strike landing patterns with their paretic foot. The ReStore™ device includes apparel-like components, mechanical components, electrical components, cables, sensors and a handheld controller.

Mode NameMode Description
Assist Mode(dynamic ankleassistance)A mode in which the cables provide assistance for ankle plantar and dorsiflexion during forward walking on level ground (treadmill or over ground). The level of assistance provided by the device can be adjusted by the therapist.
Slack Mode(no assistance)A mode in which the cables are positioned so as to not provide any resistance to movement by the subject during his/her full range of movement.
Brace Mode(fixed ankleposition)A mode in which the cables are retracted in a configuration that restricts the ankle movement in the plantar and/or dorsiflexion directions to maintain ankle position.

Table 1: (System Modes)

During treadmill walking with ReStore™ in Slack Mode, no assistance is provided by ReStore. Patients should be evaluated for their baseline treadmill walking ability, and overhead harnessing, body weight support, or the integrated ReStore™ Support Handle should be used for safety if determined appropriate by the therapist.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo has the word "ReWalk" in large letters, with the "Re" in gray and the "Walk" in yellow. Below the word "ReWalk" is the word "Robotics" in smaller letters. There are also some gray lines below the word "ReWalk".

7. Intended Use / Indication for Use

The ReWalk ReStore is intended to be used to assist ambulatory functions in rehabilitation institutions under the supervision of a trained therapist for people with hemiplegia/ hemiparesis due to stroke who can ambulate at least 1.5m (5ft) with no more than minimal to moderate levels of assistance. The trained therapist must successfully complete a training program prior to operating the device. The device is not intended for sports or stair climbing.

8. Substantial Equivalence Discussion

The following table compares the ReStore™ to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance. The comparison of the devices in Table 2 below provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new or different questions of safety or effectiveness based on the similarities to the predicate device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo has the word "ReWalk" in large letters, with the "Re" in gray and the "Walk" in yellow. Below the word "ReWalk" is the word "Robotics" in smaller letters. There are also some gray lines below the word "ReWalk".

Table 2: (Substantial Equivalence Discussion)
CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
510(k) numberK173530K190337N/A
Product CodePHLPHLSame
Regulation NamePowered ExoskeletonPowered ExoskeletonSame
Regulation No.890.3480890.3480Same
Indication for UseThe Indego® orthotically fits to thelower limbs and trunk. The device isintended to enable individuals withspinal cord injury at levels T3 to L5to perform ambulatory functions withsupervision of a specially trainedcompanion in accordance with theuser assessment and trainingcertification program. The device isalso intended to enable individualswith spinal cord injury at levels C7 toL5 to perform ambulatory functionsin rehabilitation institutions inaccordance with the user assessmentand training certification program.Finally, the Indego® is also intendedto enable individuals with hemiplegia(with motor function of 4/5 in leastone upper extremity) due tocerebrovascular accident (CVA) toperform ambulatory functions inrehabilitation institutions inaccordance with the user assessmentand training certification program.The Indego is not intended for sportsor stair climbing.The ReWalk ReStore is intended to be used toassist ambulatory functions in rehabilitationinstitutions under the supervision of a trainedtherapist for people with hemiplegia/hemiparesis due to stroke who can ambulate atleast 1.5m (5ft) with no more than minimal tomoderate levels of assistance. The trainedtherapist must successfully complete a trainingprogram prior to operating the device. Thedevice is not intended for sports or stairclimbing.Comparable -The intended use of enabling individuals to performambulatory functions in rehabilitation institutions isthe same.Indego® has broader indications, namely forindividuals with spinal cord injury in addition toindividuals with hemiplegia (with motor function of4/5 in least one upper extremity) due tocerebrovascular accident (CVA).The ReStore™ indications are a subset of theseindications (individuals with hemiplegia due tostroke).This difference does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
Patient PopulationIndividuals with hemiplegia (withmotor function of 4/5 in at least oneupper extremity) due tocerebrovascular accident (CVA)Individuals with hemiplegia / hemiparesis dueto strokeSimilar –Both devices are indicated for use by individualswith hemiplegia (hemiparesis which is subset ofHemiplegia is also included in ReStore™).ReStore™ does not require use of upper extremitymobility aids. Indego® inherently requires use ofmobility aides like crutches for support, and thusrequires motor function of 4/5 in at least one upperextremity.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Deviceconstruction andmaterialsThe Indego® device is composed ofrigid frame and textile strappingproviding body weight support to thepatientReStore™ device is composed of textile frameand strapping worn by patientDifferent –ReStore™ provides the ankle joint with assistancewithout supporting the patient, while the Indego®provides assistance within a rigid body supportingframe.Both frames enable transfer of mechanical forcesfrom motors and gears (based on data obtainedfrom sensors) to the wearer.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Body CoverageWorn over legs and around hips andlower torsoWorn over legs and around lower torsoSimilar –ReStore™ provides assistance to paretic anklejoint, while Indego® provides assistance to kneeand hip joints, thus the difference in body coverage.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
Size ofComponentsModular small, medium and largeupper leg, lower leg and hipcomponents and control unitintegrated into hip unitModular small, medium, large and extra-largeparetic leg component, control unit(waistpack) integrated into adjustable waistbeltSimilar -Both devices utilize similar component sizes.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Mobility AidWalker, cane or crutchesWalker, cane, crutches, harness or overheadsupportSimilar -Mobility aides are required for use of Indego®. Useof mobility aides for ReStore™ is per therapistdiscretion.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Ability of UserMobilitySit, stand, walk and turnSit, stand, walk and turnSame
Walking Speed~2 km/hr.Up to 5km/hr.Different -Both devices utilize upper limit for walking speed.The ReStore™ maximum supported speed ishigher, however the gait algorithm providesconsistent force and timing profiles to supportwearer's natural walking speed.This difference does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Type of SurfaceSmooth, grass, cement, carpet,transitions, thresholdsIn the clinical trial ReStore™ was tested underthe following conditions:Treadmill: Assist ModeLevel Indoor Floor: All modes (Assist, Brace,and Slack mode)Different -ReStore™ can be used on treadmills as well.The device can be used on indoor only, on surfacesincluding carpeting and hard floors or evensurfaces.This difference does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
Control MethodUses postural cues and user motion totrigger all transitionsUses patient gait motion to trigger assistanceSimilar -Both devices are activated by user motion.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Patientheight/weight61" to 75" (155 to 191 cm)Up to 250 lbs. (113kg)56" to 75.5" (142 to 192cm)Up to 264 lbs. (120kg)Similar -Both devices support similar range of height/weightof patient.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Range of MotionHip: 110° flexion to 30° extensionKnee: 110° flexion to 10° extensionAxial forces applied to the ankle jointcorresponding to wearer's ankle joint range ofmotionSimilar -Both devices utilize assistance to the patient gaitcycle by allowing movement within the joint'srange of motion.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
Use environmentRehabilitation institutionsRehabilitation institutionsSame
Device weight26 lbs. (11.6 kg)11 lbs. (5 kg)Different -ReStore™ weighs less than the Indego® device.The relatively low weight of the ReStore™minimally affects the patient's balance.This difference does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by clinical data.
RechargeableBatteryRechargeable lithium ion.33.3 V, 36A peak current, 12Acontinuous current.159Wh fully charged; 1.5 hours ofcontinuous walking fully chargedTwo off-the-shelf rechargeable lithium ionbatteries: 15.0 VDC x 2(total voltage = 30 VDC)3.2 AH, 48Wh. Minimum 2 hours continuouswalking fully chargedSimilar -Both devices utilize the same battery technologyand provide similar use duration.
CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
Battery ChargeTimeMaximum of 4 hoursMaximum of 2 hoursSimilar -Charge time difference between the devices doesnot raise questions of safety and effectiveness
Training andCertificationProgram(Clinical Use)A thorough training program thatprovides certification is required forclinicians before using Indego® withpatientsA thorough training program providingcertification is required for clinicians beforeusing ReStore™ with patientsSame
Expected UseableLife5 years with proper servicing3 years with proper servicingSimilar -Expected useable life difference between thedevices does not raise questions of safety andeffectiveness
User FeedbackProvides vibratory feedback and LEDindicators on top of hip unit, visibleto wearerProvides audible feedback to therapist andwearer in addition to visual feedback viaHandheld Device and LED indicators onWaistpack unit visible to therapist.Similar -Both devices provide visual feedback to thetherapist, audible or vibratory feedback to thewearer.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by usability and clinical data.
Fall Detection andMitigationDetects forward, backward andsideways falling as it occurs. Thedevice makes adjustments during thecourse of the fall to position the userfor minimal risk of injury or allowthe user to attempt unassistedrecoveryTherapist support handle on Waistbelt fortherapist to provide support in case of patientloss of balance. In addition, non-paretic leg isunconstrained and free to move at all times toallow the user to attempt unassisted recovery.Different -ReStore™ usage is accompanied by a therapistwhich when observes the patient loss of balancecan support the patient using the Therapist supporthandle.This difference does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by usability and clinical data.
CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
Failsafe FeatureIn the event of power failure knees become locked and hips free (similar to typical passive leg braces)In the event of most error statuses, ReStore™ defaults to brace mode and holds the cables with the ankle in a neutral position (similar to typical passive ankle foot orthoses).In the event of sudden, unexpected power failure, Bowden cables loosen and ankle joint can move freelySimilar -Similar to the predicate device, the ReStore™ defaults to a fixed position at the relevant joint during most error conditions. In the specific event of an unexpected power failure, this controlled cable movement is not possible, so the system instead releases the cables to allow the patient full range of motion for recovery.Minor differences between failsafe features are considered within ReWalk's risk management and mitigations have been demonstrated by usability and clinical data. The ReStore™ Waist Belt is designed with an integrated support handle, and a precaution regarding the use of overhead harnessing with patients requiring significant assistance or support from the Physical Therapist to ambulate short distances is intended to mitigate this risk and ensure safe and appropriate use of ReStore™.
OperatingTemperature32° to 104°F (0° to 40° C)41° to 86°F (+5° to 30°C)Similar -ReStore™ fits operating conditions expected in its use environment (indoors).
OperatingHumidity30% - 75% relative humidity15% to 90% non-condensingSimilar -ReStore™ fits a wider range humidity conditions
Electrical SafetyTestingPassed ANSI/AAMIES60601-1:2005/(R)2012Passed ANSI/AAMIES60601-1:05+A1:12:C1:09+A2:10Same
ElectromagneticCompatibilityTestingPassed IEC 60601-1-2:2014Passed IEC 60601-1-2:2014Same
Handheld DeviceHandheld Device is used forcontrolling and configuring throughBT communicationHandheld Device is used for controlling andconfiguring through BT communicationSame
WirelesscommunicationBluetooth Low Energy using 2.4GHz industrial, scientific, andmedical radio bandBluetooth Low Energy using 2.4 GHzindustrial, scientific, and medical radio bandSame
CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
DrivemechanismsBatteries, motors, gears, frame andstrappingBatteries, motors, gears, strapping andBowden cablesSimilar -Operating principle is performed by cables orstructure applying appropriate force to achieve jointrotation.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by usability and clinical data.
Alignment(between user anddevice)Alignment of hip, knee and anklejoints with strap mechanism attachedto device frameAlignment of user to device with soft shellcalf strap and attachment points within theshoeSimilar -Alignment between user and device achieved bystraps preventing sliding or misalignment of wearerand maintaining tightness.ReStore™ straps are attached to a soft structure andIndego® straps to a rigid structure.This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by usability and clinical data.
Mechanical userinterfaceDevice attached to user withstrapping mechanismDevice attached to user with strappingmechanismSame
AssistancemethodMotors/gears moving rigid frame torotates/moves hip and knee jointsMotors/gears pulling Bowden cables torotate/move paretic ankle jointSimilar -Both devices use mechanical assistance to facilitatejoint rotation. ReStore™ delivers assistance to theankle joint while the Indego® delivers assistance tothe knee and hip joints (ankle is supported byAFO).This similarity does not raise any new or differentquestions of safety or effectiveness, as isdemonstrated by usability and clinical data.

Table 2: (Substantial Equivalence Discussion)

ReStore™ Traditional 510(k) Section 005 510(k) Summary

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for ReWalk Robotics. The word "ReWalk" is in large, bold letters, with "Re" in gray and "Walk" in yellow. Below "ReWalk" is the word "Robotics" in smaller, gray letters with a series of gray lines to the left of the word.

8 Rev 03

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for ReWalk Robotics. The word "ReWalk" is in large, bold letters, with "Re" in gray and "Walk" in yellow. Below "ReWalk" is the word "Robotics" in smaller, gray letters with a series of gray lines to the left of the word.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for ReWalk Robotics. The word "ReWalk" is written in a bold, sans-serif font, with the "Re" in gray and the "Walk" in yellow. Below the word "ReWalk" is the word "Robotics" in a smaller, sans-serif font. The word "Robotics" is also in gray, and there are several horizontal lines to the left of the word.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo features the word "ReWalk" in a combination of gray and yellow colors, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in gray, with a series of horizontal lines above it, giving the impression of a road or path.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for ReWalk Robotics. The word "ReWalk" is in large, bold letters, with "Re" in gray and "Walk" in yellow. Below "ReWalk" is the word "Robotics" in smaller, gray letters with a series of horizontal lines to the left of the word.

ReStore™ Traditional 510(k) Section 005 510(k) Summary Page 10 of 18 Rev 03

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo features the word "ReWalk" in a bold, two-tone color scheme, with "Re" in gray and "Walk" in yellow. Below "ReWalk" is the word "Robotics" in a smaller, sans-serif font, with a gradient effect that fades from dark gray to light gray.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo features the word "ReWalk" in a combination of gray and yellow colors, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in gray, with a series of horizontal lines above it, giving the impression of a road or path.

CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
BiocompatibilityNoneCalf Wrap Liner:Surface device - contact with intact skin (calf)for limited time duration ≤ 24 hours).Incompliance with ISO 10993-1, 5 and 10Different -ReStore™ device utilizes Calf Wrap liner applieddirectly on skin of wearer's paretic leg calf.This difference does not raise any new or differentquestions of safety and effectiveness, asdemonstrated by a 3rd party Biocompatibilitytesting & risk assessment."
System modesSit, Stand and WalkAssist, Slack and BraceDifferent-Differences reflect that Restore™ is intended for asubset of the Indego® population, i.e., strokepatients only.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo features the word "ReWalk" in a combination of gray and yellow colors, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in gray, with a series of horizontal lines above it, giving the impression of a road or path.

CharacteristicParker Hannifin CorporationIndego®ReWalk Robotics Ltd. ReStore™SE Justification
Clinical studygeneralinformation30 subjectsMultisite (7) clinical trialDuration: approx. 2 weeksNumber of visits: 544 subjectsMultisite (5) clinical trialDuration: approx. 4 weeks.Number of visits: 7Similar -Both studies apply to acute, sub-acute and chronicas well as mild, moderate and severe strokepopulations. Both studies have same primaryoutcome - safety measure, and similar secondary/exploratory outcome measures concerning howwell each device supports ambulatory functions.
Primary endpoint:Assess the safety of the device duringgait rehabilitation activities in post-stroke individuals undergoingphysical therapy.Primary endpoint:Assess device safety during gait rehabilitationactivities in post-stroke individualsundergoing physical therapy.
Secondary endpoints:1. Functional ambulation category;2. Gait Speed, as measured via 10meter walk test3. Spasticity of bilateral upper andlower extremities as measuredwith Modified Ashworth Scale(MAS)4. Pain measured with Face, Legs,Activity, Cry, ConsolabilityScale (FLACC)Secondary endpoints:1. Device reliability2. Device safety for Physical Therapist.
Exploratory endpoints1. 10m walk assessments.2. Temporal/spatial measures of gait3. 2 minute walk tests in 3 modes.4. Subject satisfaction5. Physical Therapist satisfaction

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo features the word "ReWalk" in a combination of gray and yellow colors. Below "ReWalk" is the word "Robotics" in gray with horizontal lines to the left of the word.

9. Non-Clinical Performance data

Comprehensive bench testing has been performed on the ReStore™ device to confirm that it meets design inputs and specifications for the device, to demonstrate substantial equivalence to the predicate device and to demonstrate safety and effectiveness. These tests include nonIclinical performance testing, EMC/EMI testing, biocompatibility testing and electrical safety testing.

The test results show that the ReStore™ device meets all design requirements and is in compliance to all applicable standards and regulations including special controls in 21CFR890.3480.

These tests include the following:

TestTest descriptionResult
Mechanical performanceEvaluate mechanical components ability to withstand applied loads (static and cyclic) throughout device lifecycle while ensuring system accuracy and precision performance is maintained during operation and post-fatiguePASS
Mechanical analysisEvaluate adequate safety margin between clinically expected loads and failure loads on system load-bearing components throughout device lifecyclePASS
Battery performance and safety testingEvaluate battery performance and safety under maximum load conditions, charging time and high current safety disconnectionPASS
Apparel performanceEvaluate apparel component ability to withstand applied loads static and cyclic (Durability) and cleanability throughout device lifecyclePASS
Motion performance simulation and testingMotion performance simulation and testing to evaluate frequency response motion latency and accuracy of the device.PASS
Weight distribution and center of mass analysisEvaluate Center of Mass of the Restore™ waist pack and related implications on the Restore™ patientPASS
Wireless performance and coexistence using FDA guidance AAMI TIR 69Evaluate wireless technology, service quality, security, and electromagnetic compatibility (EMC)PASS
Environmental using ASTM D4169 ASTM D4332Environmental testing conformance to standards, accompanied by certificatePASS
Usability using FDA guidance and IEC 62366-1Demonstrate that when used after training by intended users in the appropriate use environment, the device can be used safely and effectively without producing failure pattern that could result in negativePASS

Table 3: (Non-clinical Performance Data)

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image is a logo for ReWalk Robotics. The word "ReWalk" is written in a combination of gray and yellow. The word "Robotics" is written in gray below the word "ReWalk" with a series of gray lines to the left of the word.

TestTest descriptionResult
clinical impact, injury to patients andusers or device damage
Software verification andvalidation using FDA GuidancesIEC 62304 and AAMI TIR 57The device meets design andcybersecurity requirementsPASS
Electrical safety usingANSI AAMI ES 60601-1Conformance to electrical safetystandards, accompanied by certificatePASS
Electromagnetic compatibilityusing IEC 60601-1-2Conformance to electromagneticcompatibility standards, accompaniedby certificatePASS
Biocompatibility usingFDA guidance,ISO 10993-1/5 and 10GLP, Conformance to biocompatibilitystandardsPASS

10. Clinical Performance Data

Three IRB-Approved protocols have been conducted in individuals with hemiplegia / hemiparesis resulting from Stroke. These studies consist of one 44 subject multi-site clinical trial and two engineering studies. The primary endpoint for the multi-site clinical trial was to demonstrate the safety of the ReStore™ device when used during gait training activities for individuals with hemiparesis / hemiplegia resulting from Stroke under the supervision of a trained therapist in rehabilitation institutions. Additionally, two engineering studies are ongoing at the Wyss Institute at Harvard University to develop and evaluate exosuit designs. The ReStore™ clinical validation testing was also conducted under the second engineering protocol.

10.1 44- Subject Multisite Clinical Trial

44 subjects were enrolled in a multi-site clinical trial conducted across 5 study sites. Participation consisted of 7 study visits with the ReStore™ device. All 44 subjects completed at least one study visit with the ReStore™ device, and 36 subjects completed all 7 study visits. This study was IRB-approved and conducted in accordance with principles of Good Clinical Practice (GCP). The primary endpoint for this study was to demonstrate the safety of the ReStore™ device when used during activities of gait training in rehabilitation institutions, as well as exploratory endpoints to demonstrate substantial equivalence with respect to efficacy when compared to the predicate device.

Safety was evaluated on the basis of frequency and severity of device-related adverse events, including serious adverse events and falls. There were zero device-related serious adverse events and zero device-related falls in this study. During the initial phase of the study, adverse events were reported in 13.5% of study visits. The majority of these adverse events were minor in severity and did not require intervention or follow up. The cause was determined to be related to improper donning and fitting techniques and a mandatory re-training was conducted for all study sites. Following the retraining, just one (1) device-related adverse event was reported in the remaining 75 study visits, reducing the frequency to 1.3% of study visits.

Several measures were used to evaluate subjects' walking in the study. Walking speed was measured using 10m walk tests at comfortable and maximal walking speeds, and assessments were performed during walking with and without the ReStore™ device. Additionally, subjects' 2-minute walk distances were measured in all three operational modes of the ReStore™ device: Slack (no assistance), Brace (fixed ankle position), and Assist (dynamic ankle assistance).

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image is a logo for ReWalk Robotics. The logo features the word "ReWalk" in a combination of gray and yellow colors. Below the word "ReWalk" is the word "Robotics" in gray. The logo is simple and modern.

All subjects were able to complete some or all of the walking assessments in the ReStore™ device on their first study visit. The average duration of walking in Assist mode during gait training visits increased by 12.4% between Visit 2 (the first gait training visit) and Visit 6 (the last gait training visit). Of the 36 subjects who completed all 7 study visits, the majority of subjects increased walking speed from Visit 1 to Visit 7. During walking with the ReStore™ device, 31 subjects (86.1%) increased comfortable walking speed and 29 subjects (80.6%) increased maximal walking speed. During walking assessments independent of the ReStore™ device, 23 subjects (63.9%) increased comfortable walking speed and 28 subjects (77.7%) increased maximal walking speed. Average walking distance during the 2-minute walk test increased by 47.9 ft. during walking with assistance from the ReStore™ device (Assist mode) compared to walking without assistance from the ReStore™ device (slack mode).

10.2 Engineering studies

Two ongoing protocols are being conducted at the Wyss Institute for Biologically Inspired Engineering at Harvard University (Cambridge, MA). These studies involve the development and testing of the ReStore™ device and other similar prototypes. Both studies are IRB approved and are being conducted according to principles of Good Clinical Practice (GCP). 35 unique participants have been enrolled in these two studies (10 participants are enrolled in both studies), and a total of 472 study visits have been conducted since the first study began enrolling in 2014. Device-related adverse events have been reported with low frequency and low severity. There are zero device-related serious adverse events and zero device-related falls reported in either study.

10.3 Clinical Validation Study

Nine (9) participants were enrolled in a ReStore™ validation substudy conducted under the Engineering Protocols above. The purpose of this study was to validate the ReStore™ design as well as to assess the safety and the use of the ReStore™ device during common activities of Physical Therapy. One device-related adverse event was encountered in this substudy which was resolved on its own without intervention or complications. There were zero device-related serious adverse events and zero device related falls reported in this substudy.

Summary of Clinical Performance Data 10.4

A total of 79 individuals with hemiparesis / hemiplegia resulting from stroke have completed a combined 740 study visits with the ReStore™ device or a similar prototype. All three studies have demonstrated that the ReStore™ is safe and well tolerated by individuals post-stroke and is appropriate for use during ambulatory functions in rehabilitation settings under the supervision of a trained therapist for people with hemiplegia/hemiparesis due to stroke.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for ReWalk Robotics. The logo features the word "ReWalk" in a bold, sans-serif font, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in a smaller, gray font with a series of horizontal lines to the left of the word.

Table 4: Summary of IRB-Approved ReStore™ Studies
TrialNumber ofSubjectsNumber of Sessionsper SubjectTotal Sessions
44-Subject Multisite ClinicalTrial447268
Clinical Validation Study¹919
Engineering Studies (1 & 2)35Variable472
Total79Variable740

' Clinical Validation Testing was conducted as a subset of Engineering Study 2 (Wyss Study IRB161845) so totals from this study do not contribute to overall totals.

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for ReWalk Robotics. The word "ReWalk" is in a bold, sans-serif font, with "Re" in gray and "Walk" in yellow. Below "ReWalk" is the word "Robotics" in a smaller, gray, sans-serif font. There are also some gray lines to the left of the word "Robotics".

11. Statement of Substantial Equivalence

The ReStore™ and the predicate legally marketed device (Indego®) have the same intended use and similar indications for use (a subset of the predicate indications). In addition, both devices have similar technological characteristics. Finally, clinical and non-clinical performance data are similar as well.

The differences in the above mentioned characteristics do not raise any new or different questions of safety or effectiveness, thus the ReStore™ is substantially equivalent to its predicate device.

§ 890.3480 Powered lower extremity exoskeleton.

(a)
Identification. A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) Design characteristics must ensure geometry and materials composition are consistent with intended use.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include:
(i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use;
(ii) Simulated use testing (
i.e., cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present;
(iv) The accuracy of device features and safeguards; and
(v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance.
(6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for:
(i) Level of supervision necessary, and
(ii) Environment of use (
e.g., indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use.
(8) Labeling for the Physician and User must include the following:
(i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk.
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations;
(B) Instructions on fitting the patient;
(C) Instructions and explanations of all available programs and how to program the device;
(D) Instructions and explanation of all controls, input, and outputs;
(E) Instructions on all available modes or states of the device;
(F) Instructions on all safety features of the device; and
(G) Instructions for properly maintaining the device.
(iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness.
(iv) Pertinent non-clinical testing information (
e.g., EMC, battery longevity).(v) A detailed summary of the clinical testing including:
(A) Adverse events encountered under use conditions,
(B) Summary of study outcomes and endpoints, and
(C) Information pertinent to use of the device including the conditions under which the device was studied (
e.g., level of supervision or assistance, and environment of use (e.g., indoors and/or outdoors) including obstacles and terrain).