(29 days)
The BIOK 7600W is intended for use by a qualified doctor or technologist on both adult and pediativ for taking diagnostic radiographic exposures of all body parts of the patients. The BIOK 7600W provides digital image capture and is intended to replace radiographic film/screen. The x-ray generator, x-ray tube and associated equipment are not provided with the proposed sensor. Prescription use only.
The BIOK7600W is not intended for mammography.
The BIOK7600W sensor is a wireless digital radiography sensor which automatically collects x-ray images from an x-ray source. The BIOK7600W sensor (flat panel type) collects x-rays and digitizes the images for their transfer and display to a computer. The sensor does not have an x-ray source, which is provided by independent manufacturers. The sensor includes a flat panel for x-ray acquisition and a computer (including proprietary processing software) for processing, annotating and storing x-ray images.
An X-ray generator with a minimum voltage output range between 55 to 120 kVp and a minimum power of 30 KW can be used with the BIOK7600. The BIOK7600 does not connect to the generator. It automatically detects x-rays and enables the image acquisition (AEC). No changes to the generator hardware or software are required to operate the BIOK7600.
The ChiroSight software runs on a PC workstation and provides the user interface to control the BIOK7600 flat panel. ChiroSight initiates the taking of X-ray images from the BIOK7600. through the Sensor Driver application. Once an X-ray is taken. Sensor Driver reads it from the flat panel and transfers it to ChiroSight. After it receives an image from the BIOK7600, ChiroSight:
Automatically optimizes the image quality
Permanently stores the image to its database
Displays the image to the user
Allows the user to perform image manipulations and to add annotations
Allows the user store and transfer the images in DICOM format.
The provided text describes the Biokinemetrics BIOK7600W Wireless Flat Panel Detector and its 510(k) submission. Here's a breakdown of the acceptance criteria and the study conducted:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria/Tests | Reported Device Performance |
|---|---|---|
| Non-Clinical Performance | Equivalent performance to predicate device | The non-clinical considerations document demonstrates that the BIOK7600W sensor offers equivalent performance (including biocompatibility, effectiveness, thermal, electrical, and mechanical safety) to the predicate device. The design, development, and production of the sensor conforms to 892.1680 and ISO 13485 quality systems. |
| Software Functionality | Equivalent functionality to predicate reference device (BIOK4600) | The BIOK7600W software functionality is equivalent to the original BIOK4600 reference device, with the exception of the Imaray interface application, which now reads images from the new BIOK7600W sensor instead of the BIOK4600 sensor. |
| Image Quality | Production of images of similar quality and characteristics to predicate device | The BIOK7600W produces images of similar quality and characteristics that are equivalent to those of the predicate device, as demonstrated in the clinical considerations - concurrence study. A set of test images have been submitted along with the equivalent images from the predicate device. |
Study Details
-
Sample size used for the test set and the data provenance:
- Test Set Size: "A set of test images" was used. The exact number of images is not specified in the provided text.
- Data Provenance: Not explicitly stated, but the "clinical considerations - concurrence study" involved comparing images from the BIOK7600W with images from the predicate device. This implies retrospective comparison of clinical image data or a controlled collection for this specific study. The country of origin of the data is not mentioned.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number or qualifications of experts used. It refers to a "clinical considerations concurrence study" where images were compared, but details on ground truth establishment or expert involvement for image evaluation are missing.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not specify any adjudication method for the test set's ground truth.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study is mentioned. This device is a flat panel detector, not an AI-assisted diagnostic tool. The study focuses on demonstrating equivalence in image output rather than user performance with or without AI.
-
If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:
- Yes, implicitly. The performance of the BIOK7600W (including its software for image optimization and processing) was evaluated directly by comparing its output images and functionalities to a predicate device. This is a standalone performance evaluation of the device as a system.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the clinical study appears to be the image quality and characteristics of the predicate device. The study aims to show that the BIOK7600W produces images "equivalent to those of the predicate device." This is a comparative equivalence ground truth, rather than an absolute diagnostic ground truth (like pathology).
-
The sample size for the training set:
- The document does not provide information about a "training set" in the context of machine learning or AI. The BIOK7600W is a digital radiography sensor, and its "software" primarily handles image processing, annotation, and storage. If any internal parameters of the image optimization algorithms were "trained," this is not disclosed.
-
How the ground truth for the training set was established:
- As no training set is mentioned in the context of an AI/ML algorithm requiring ground truth, this information is not applicable from the provided text.
{0}------------------------------------------------
March 13, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" on the first line and "U.S. FOOD & DRUG ADMINISTRATION" on the second line.
Biokinemetrics Inc. % Steven Kraus President 211 East 4th Street CARROLL IOWA 51401
Re: K190300
Trade/Device Name: BIOK7600W Wireless Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: February 9, 2019 Received: February 12, 2019
Dear Steven Kraus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination. product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K190300
Device Name BIOK7600W Wireless Flat Panel Detector
Indications for Use (Describe)
The BIOK 7600W is intended for use by a qualified doctor or technologist on both adult and pediativ for taking diagnostic radiographic exposures of all body parts of the patients. The BIOK 7600W provides digital image capture and is intended to replace radiographic film/screen. The x-ray generator, x-ray tube and associated equipment are not provided with the proposed sensor. Prescription use only.
The BIOK7600W is not intended for mammography.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Biokinemetrics. The logo is in blue and green, with the word "BIOKINE" in blue and the "ME" in green. Below the logo is the tagline "ALIGNING CHIROPRACTIC AND TECHNOLOGY" in a smaller font.
TRADITIONAL 510K - SUMMARY (K190300)
-
- Device type and name: BIOK7600W Wireless Flat Panel Detector
-
- Submitter Biokinemetrics Inc. 211 East 4th Street Carroll lowa 51401
- Dr. Steven Kraus Contact person: President Tel: (712) 210-4750 Fax: (888) 800-2149 e-mail: skraus@biokinemetrics.com
Date prepared: March 7, 2019
-
- Device name: BIOK7600W Wireless Flat Panel Detector BIOK7600W Common name:
- Class II, 21 CFR 892.1680, Solid-State X-ray Imager (Flat panel/ 4. Device classification: Digital Imager)
-
- Product Code: 90MQB
-
- Requlation Name: Stationary x-ray system
-
- Basis for the submission: Addition of a sensor to the list of medical devices compatible with the Biokinemetrics devices
-
- Device description: The BIOK7600W sensor is a wireless digital radiography sensor which automatically collects x-ray images from an x-ray source. The BIOK7600W sensor (flat panel type) collects x-rays and digitizes the images for their transfer and display to a computer. The sensor does not have an x-ray source, which is provided by independent manufacturers. The sensor includes a flat panel for x-ray acquisition and a computer (including proprietary processing software) for processing, annotating and storing x-ray images.
An X-ray generator with a minimum voltage output range between 55 to 120 kVp and a minimum power of 30 KW can be used with the BIOK7600. The BIOK7600 does not connect to the generator. It automatically detects x-rays and enables the image acquisition (AEC). No changes to the generator hardware or software are required to operate the BIOK7600.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Biokinemetrics. The logo is blue and green, with the word "BIOKINETRICS" in large, bold letters. Below the word is the tagline "ALIGNING CHIROPRACTIC AND TECHNOLOGY" in smaller letters. The logo is simple and modern, and it conveys the company's focus on chiropractic and technology.
The ChiroSight software runs on a PC workstation and provides the user interface to control the BIOK7600 flat panel. ChiroSight initiates the taking of X-ray images from the BIOK7600. through the Sensor Driver application. Once an X-ray is taken. Sensor Driver reads it from the flat panel and transfers it to ChiroSight. After it receives an image from the BIOK7600, ChiroSight:
Automatically optimizes the image quality Permanently stores the image to its database Displays the image to the user Allows the user to perform image manipulations and to add annotations Allows the user store and transfer the images in DICOM format.
The software level of concern for the BIOK7600 has been determined to be moderate based on the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
The cybersecurity risks of the BIOK7600 have been addressed to assure that no new or increased cybersecurity risks were introduced as a part of device risk analysis. These risks are defined as sequence of events leading to a hazardous situation, and the controls for these risks were treated and implemented as proposed in the risk analysis (e.g., requirements, verification).
- The BIOK7600W is intended for use by a qualified doctor or 10. Indications for use: technologist on both adult and paediatric patients for taking diagnostic radiographic exposures of all body parts of the patients. The BIOK7600W is not intended for mammography.
-
- Comparison with predicate devices: The BIOK7600W sensor is composed of a flat panel detector very similar to the technology used in the Wireless Digital Flat Panel Detector (K182551, predicate device hardware) and the BIOK4600 (K092307) software on the PC. The hardware and firmware of the BIOK7600W sensor are considered to be substantially equivalent to the predicate device hardware, as demonstrated in the non-clinical considerations document, based on the SSXI Guidance from FDA. The BIOK7600W software functionality is equivalent to the original BIOK4600 reference device, except for the Imaray interface application which reads the images from the new BIOK7600W sensor instead of the BIOK 4600 sensor. The BIOK7600W produces images of similar quality and characteristics that are equivalent to those of the predicate device, as demonstrated in the clinical considerations - concurrence study provided in this submission.
- a. Non-clinical considerations: the non-clinical considerations document demonstrates that the BIOK7600W sensor offers equivalent performance, biocompatibility, effectiveness, thermal, electrical and mechanical safety and is equivalent to the predicate device. The design, development and production of the sensor conforms to 892.1680 and ISO 13485 quality systems.
- b. Clinical considerations concurrence study: A set of test images have been submitted along with the equivalent images from the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Biokinemetrics. The logo is in blue and green, with the word "BIOKINE" in blue and "METRICS" in green. Below the logo is the tagline "ALIGNING CHIROPRACTIC AND TECHNOLOGY" in a smaller font.
- c. Conclusion: the device was evaluated against the predicate device (Wireless Digital Flat Panel Detector, K182551) and was found to be equivalent to the predicate device.
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.