K Number
K190225

Validate with FDA (Live)

Date Cleared
2019-11-05

(273 days)

Product Code
Regulation Number
862.1675
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TAP Blood Collection® Device is a lithium heparin coated single use device intended to collect capillary blood from the upper arm of adults (21 years of age or older). The TAP Blood Collection® Device is for measurement of HbAlc on blood specimens which can be collected by self-administration of the TAP Device by a layperson or by a healthcare worker in a healthcare setting. The collected sample is then transported for analysis in a clinical laboratory for determination of Hemoglobin Alc (HbA1c) using tests intended for monitoring glycemic control.

Device Description

The TAP Blood Collection® Device (herein “TAP Device”) is a single-use, sterilized whole blood specimen collection and transportation device that uses a combination of two mechanisms, capillary action and vacuum extraction, to obtain a capillary blood sample from the upper arm. The device contains lithium heparin as an anticoagulant. The device is intended for self-administration by a layperson or by a healthcare worker. When the TAP Device is actuated, it collects the sample in an integrated reservoir and provides a visual indicator (fill indicator window) to the end user to confirm that the collection is complete and sufficient blood has been collected to conduct HbA1c testing. The sample collection time is 7 minutes or less and typically takes 2-3 minutes. The TAP Device is then sent to the laboratory for testing. The sample must be tested within 6 hours from time of collection or as indicated in the HbA1c test system package insert (whichever is less).

AI/ML Overview

The provided text describes the TAP Blood Collection® Device, focusing on its re-submission for an expanded indication of use, specifically for self-administration by laypersons for HbA1c testing. The device itself (K161521) was previously cleared, and this submission (K190225) aims to add layperson use.

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA 510(k) summary does not explicitly list "acceptance criteria" in a table format with specific numerical targets. Instead, it describes general performance goals for the additional testing supporting the layperson use indication. The reported device performance is presented as conclusions from these studies.

Acceptance Criteria (Implied)Reported Device Performance
Usability: Laypeople can successfully use the device to self-collectThe usability study demonstrated that laypeople can successfully use the TAP Device to self-collect a blood sample according to the TAP Device written instructions for use.
Analytical Performance (Comparison): Self-collected samples are adequate for HbA1c testing and provide similar results to venous blood.The analytical performance testing demonstrated that self-collected TAP Device samples provide similar HbA1c test results when compared to venous blood samples collected by a healthcare worker.
Analytical Performance (Repeatability/Precision): Sufficient repeatability for HbA1c testing with multiple self-collected samples.The study also demonstrated sufficient repeatability for HbA1c testing when multiple samples are self-collected with the TAP Device.

2. Sample Size Used for the Test Set and Data Provenance

The provided text does not specify the sample size for either the usability study or the analytical performance study. It also does not mention the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the given text.

4. Adjudication Method for the Test Set

This information is not provided in the given text.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This device is a blood collection device, not an AI-powered diagnostic tool requiring human reader interpretation. Therefore, an MRMC comparative effectiveness study regarding human readers and AI assistance is not applicable and not mentioned.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This device is a blood collection device. Its primary function is to collect a blood sample. The "performance" being evaluated here is the ability to collect a suitable sample for laboratory analysis and the comparability of the collected sample to a standard collection method. There is no "algorithm only" performance that would be applicable to this type of device.

7. The Type of Ground Truth Used

  • For the usability study, the "ground truth" was likely defined by successful completion of specified tasks by laypersons according to instructions. This would be assessed through observation and potentially questionnaires.
  • For the analytical performance testing, the "ground truth" for HbA1c values was established by venous blood samples collected by a healthcare worker, considered the reference standard for HbA1c measurement.

8. The Sample Size for the Training Set

This information is not provided in the given text. The studies described are performance validation studies, not AI model training.

9. How the Ground Truth for the Training Set Was Established

As this is not an AI algorithm, there is no "training set" in the context of machine learning. The studies are evaluating the device's ability to facilitate sample collection and the quality of those samples.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

November 5, 2019

Seventh Sense Biosystems, Inc. Tim Richards Chief Operating Officer 200 Boston Avenue, Suite 3700 Medford, MA 02155

Re: K190225

Trade/Device Name: TAP Blood Collection® Device Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: PRJ Dated: September 26, 2019 Received: September 27, 2019

Dear Tim Richards:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190225

Device Name TAP Blood Collection® Device

Indications for Use (Describe)

The TAP Blood Collection® Device is a lithium heparin coated single use device intended to collect capillary blood from the upper arm of adults (21 years of age or older). The TAP Blood Collection® Device is for measurement of HbAlc on blood specimens which can be collected by self-administration of the TAP Device by a layperson or by a healthcare worker in a healthcare setting. The collected sample is then transported for analysis in a clinical laboratory for determination of Hemoglobin Alc (HbA1c) using tests intended for monitoring glycemic control.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.0 510(K) SUMMARY – K190225

Submitted by:Seventh Sense Biosystems, Inc.
200 Boston Avenue, Suite 3700
Medford, MA 02155
Phone:(617) 547-7246
Contact:Tim Richards
Date of Summary:November 4, 2019
Device Trade Name:TAP Blood Collection® Device
Common or Usual Name:Blood Specimen Collection Device
Regulation Number:21 CFR 862.1675
Device Class:II
Product Code:PRJ
Panel:Clinical Chemistry
Predicate Device:TAP Blood Collection® Device (K161521)
Device Description:The TAP Blood Collection® Device (herein “TAP Device”) is a single-use, sterilized whole blood specimen collection and transportationdevice that uses a combination of two mechanisms, capillaryaction and vacuum extraction, to obtain a capillary blood samplefrom the upper arm. The device contains lithium heparin as ananticoagulant.
The device is intended for self-administration by a layperson or bya healthcare worker.
When the TAP Device is actuated, it collects the sample in anintegrated reservoir and provides a visual indicator (fill indicatorwindow) to the end user to confirm that the collection is completeand sufficient blood has been collected to conduct HbA1c testing.The sample collection time is 7 minutes or less and typically takes2-3 minutes. The TAP Device is then sent to the laboratory fortesting. The sample must be tested within 6 hours from time ofcollection or as indicated in the HbA1c test system package insert(whichever is less).
Intended Use:Intended for use as a blood specimen collection device.
Indications for UseThe TAP Blood Collection® Device is a lithium heparin coated single use device intended to be used to collect capillary blood from the upper arm of adults (21 years of age or older). The TAP Blood Collection Device is for measurement of HbA1c on blood specimens which can be collected by self-administration of the TAP Device by a layperson or by a healthcare worker in a healthcare setting. The collected sample is then transported for analysis in a clinical laboratory for determination of Hemoglobin A1c (HbA1c) using tests intended for monitoring glycemic control.

{4}------------------------------------------------

Technological Characteristics / Substantial Equivalence 5.1.

The candidate TAP Device is substantially equivalent in design, function, and intended use to the predicate TAP Device, based on the information presented in the table below. Additional detailed information is provided in Section 12.0 - Substantial Equivalence Discussion.

CharacteristicCandidate DeviceTAP Blood Collection DevicePredicate DeviceTAP Blood Collection Device(K161521)
Similarities
Intended UseDevice is intended for use as a blood specimen collection deviceSame
Prescription UseYesSame
Number of UsesSingle use / DisposableSame
AnticoagulantAvailable with lithium heparinSame
SterilityProvided Sterile by gamma radiation; SAL of 10-6Same
Collection MethodBlood access via microneedlesSame
Mechanism of Blood DrawSample obtained by capillary action and vacuumSame
Sample TypeCapillary whole bloodSame
Puncture SiteUpper ArmSame
Sample SizeUp to 100 $μ$ LSame
Specimen ContainerDrawn blood stored in plastic internal chamberSame
Storage18-28°C (64-82°F)Same
Differences
Indicated UserLayperson or Healthcare workerHealthcare worker

5.2. Performance Data

The purpose of this submission is to request a change in indications to add layperson use for the collection of the capillary blood sample.

{5}------------------------------------------------

Seventh Sense previously conducted bench studies to confirm the overall functional performance of the TAP Device against its design specifications and intended use as part of the original (predicate) TAP Device Premarket Notification (K161521). The previous performance testing remains valid as those studies are independent of the user and were specific to the design and functional performance requirements of the TAP Device.

In addition, Seventh Sense previously evaluated product performance attributes by conducting extensive clinical studies to confirm the overall functional performance of the TAP Device against its design specifications and intended use. These studies, detailed in the Premarket Notification for K161521, were conducted to support the design and functional properties of the TAP Device and remain valid for the candidate TAP Device.

The following additional testing was conducted in support of the proposed change in indications:

  • . Usability testing to demonstrate that laypeople can successfully use the TAP Device to self-collect a blood sample according to the product's written instructions for use.
  • . Analytical performance testing to confirm a blood sample collected by the layperson is adequate for HbA1c testing in a clinical laboratory. This study includes matrix comparison of TAP Device samples that are self-collected by laypeople compared to venous blood samples collected by a healthcare worker as well as evaluation of intraoperator repeatability (precision).

Performance Data Conclusions

The usability study demonstrated that laypeople can successfully use the TAP Device to selfcollect a blood sample according to the TAP Device written instructions for use.

The analytical performance testing demonstrated that self-collected TAP Device samples provide similar HbA1c test results when compared to venous blood samples collected by a healthcare worker. The study also demonstrated sufficient repeatability for HbA1c testing when multiple samples are self-collected with the TAP Device.

5.3. Standards/Guidance Documents Referenced

The following Standards/Guidance Documents relevant to the proposed change in indications are listed below.

  • . Guidance for Industry and FDA Staff – Deciding When to Submit a 510(k) for a Change to an Existing Device, dated October 25, 2017
  • Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline, CLSI Document EP09-A3. (2013).
  • . Validation and Verification of Tubes for Venous and Capillary Blood Specimen Collection; Approved Guideline, CLSI document GP-34-A. CLSI (2010).
  • User Verification of Precision and Estimation of Bias; Approved Guideline, CLSI document EP15-A3. CLSI (2014).

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.