K Number
K190127

Validate with FDA (Live)

Date Cleared
2019-08-23

(207 days)

Product Code
Regulation Number
870.1220
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

Device Description

The Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter is a flexible electrode catheter constructed of a polyurethane insulation/shaft and incorporates platinum electrodes. The active tip may be manipulated by remote means located at the proximal end of the catheter.

AI/ML Overview

The provided document is a 510(k) premarket notification letter and summary for a Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter. This document describes the device's technical characteristics, predicate devices, and the testing performed to demonstrate its safety and effectiveness.

Crucially, this document pertains to a medical device that is reprocessed (reused after cleaning and sterilization), not an AI/ML-enabled medical device. Therefore, the information requested in your prompt regarding acceptance criteria, study methodologies (e.g., MRMC studies, standalone AI performance), ground truth establishment, expert adjudication, and sample sizes for AI training/test sets is not applicable to this specific document.

The document focuses on demonstrating substantial equivalence to a predicate device through:

  • Technological Characteristics: Stating that the reprocessed device has identical purpose, design, materials, function, and intended use as the predicate devices, with no changes to claims, clinical applications, patient population, performance specifications, or method of operation.
  • Functional and Safety Testing: A list of bench and laboratory tests performed, including biocompatibility, cleaning validation, sterilization validation, functional testing (visual inspection, dimensional verification, electrical continuity and resistance, simulated use, mechanical characteristics), electrical safety testing (dielectric and current leakage), and packaging validation.

Since this is not an AI/ML device, I cannot fill in the requested table and study information as it's designed for AI/ML performance evaluation. The "acceptance criteria" for this device would be its ability to meet the performance and safety standards of the predicate device after reprocessing, as demonstrated by the functional and safety testing listed.

In summary, based on the provided text, I cannot describe acceptance criteria and study proving device meets acceptance criteria related to AI/ML performance, because the document is about a reprocessed medical catheter, not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 23, 2019

Innovative Health, LLC. Amanda Babcock Principal Regulatory Affairs Specialist 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257

Re: K190127

Trade/Device Name: Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: July 23, 2019 Received: July 24, 2019

Dear Amanda Babcock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark Fellman Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

The following device models are included in the scope of this 510(k) submission:
Item NumberDevice DescriptionElectrodesSpacing (mm)Curve TypeFrenchSizeUsableLength(cm)
401582Livewire SteerableDiagnostic EP Catheter102-5-2Med Sweep7F115
401932Livewire SteerableDiagnostic EP Catheter202-8-2-(60)-2-8-2Super Lrg Curl7F95
401918Livewire SteerableDiagnostic EP Catheter202-20-2-2-2-2-2-2-2-2-2-2-2-25-2-25-2-25-2Super Lrg Curl7F95
401580Livewire SteerableDiagnostic EP Catheter62-5-2Med Sweep7F115
401581Livewire SteerableDiagnostic EP Catheter82-5-2Med Sweep7F115
401938Livewire SteerableDiagnostic EP Catheter105-5-5Med Sweep5F115
401939Livewire SteerableDiagnostic EP Catheter102-2-2Med Sweep5F115
401940Livewire SteerableDiagnostic EP Catheter102-5-2Med Sweep5F115
401941Livewire SteerableDiagnostic EP Catheter102-8-2Extra Lrg Curl5F115
401990Livewire SteerableDiagnostic EP Catheter102-2-2Extra Lrg Curl5F115
401991Livewire SteerableDiagnostic EP Catheter102-5-2Extra Lrg Curl5F115
401575Livewire SteerableDiagnostic EP Catheter102-5-2Med Sweep6F115
401915Livewire SteerableDiagnostic EP Catheter102-5-2Extra Lrg Curl6F115
401923Livewire SteerableDiagnostic EP Catheter102-2-2Extra Lrg Curl6F115
401926Livewire SteerableDiagnostic EP Catheter102-8-2Extra Lrg SweepCSL6F115
401904Livewire SteerableDiagnostic EP Catheter202-10-2Super Lrg Curl7F95
401905Livewire SteerableDiagnostic EP Catheter205-5-5Super Lrg Curl7F95
401914Livewire SteerableDiagnostic EP Catheter202-5-2Super Lrg Curl7F95
401908Livewire SteerableDiagnostic EP Catheter202-2-2Med Curl7F115
401917Livewire SteerableDiagnostic EP Catheter82-2-2Med Sweep6F115
401949Livewire SteerableDiagnostic EP Catheter82-2-2Lrg Sweep6F115
401600Livewire SteerableDiagnostic EP Catheter42-5-2Med Sweep6F115
401603Livewire SteerableDiagnostic EP Catheter42-5-2Med Curl6F115
401572Livewire SteerableDiagnostic EP Catheter42-5-2Med Sweep6F115
401606Livewire SteerableDiagnostic EP Catheter42-5-2Lrg Sweep6F115
401933Livewire SteerableDiagnostic EP Catheter45-5-5Med Sweep6F115
Item NumberDevice DescriptionElectrodesSpacing (mm)Curve TypeFrenchSizeUsableLength(cm)
401934Livewire SteerableDiagnostic EP Catheter45-5-5Lrg Sweep6F115
401576Livewire SteerableDiagnostic EP Catheter42-5-2Sml Sweep7F115
401577Livewire SteerableDiagnostic EP Catheter42-5-2Med Sweep7F115
401578Livewire SteerableDiagnostic EP Catheter42-5-2Lrg Sweep7F115
401586Livewire SteerableDiagnostic EP Catheter42-5-2Sml Curl7F115
401587Livewire SteerableDiagnostic EP Catheter42-5-2Med Curl7F115
401588Livewire SteerableDiagnostic EP Catheter42-5-2Lrg Curl7F115
401652Livewire SteerableDiagnostic EP Catheter82-2-2Med SweepCRD-25F115
401653Livewire SteerableDiagnostic EP Catheter62-5-2Med SweepCRD25F115
401654Livewire SteerableDiagnostic EP Catheter65Med SweepCRD-25F115

(Continued on the following page)

{3}------------------------------------------------

The following device models are included in the scope of this 510(k) submission (continued):

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K190127

Device Name

Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter

Indications for Use (Describe)

The Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

SECTION 5: 510(k) SUMMARY

As required by 21 CFR 807.92(c)

Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

Contact Name and Information:

Amanda Babcock Requlatory Affairs Manager Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7705 (fax) ababcock@innovative-health.com

Date prepared:

January 25, 2019

Device Information:

Trade/Proprietary Name:

Common Name: Classification Name: Classification Number: Product Code:

Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters Diagnostic Electrophysiology Catheter Catheter, Recording, Electrode, Reprocessed Class II, 21 CFR 870.1220 NLH

Predicate Device:

510(k) Number510(k) DeviceManufacturer
K160242Reprocessed Livewire SteerableDiagnostic EP CatheterInnovative Health, LLC.
K151622Livewire Electrophysiology CatheterSt. Jude Medical

Device Description:

The Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter is a flexible electrode catheter constructed of a polyurethane insulation/shaft and incorporates platinum electrodes. The active tip may be manipulated by remote means located at the proximal end of the catheter.

Indications for Use:

The Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

{6}------------------------------------------------

Item NumberDevice DescriptionElectrodesSpacing (mm)Curve TypeFrenchSizeUsableLength(cm)
401582Livewire SteerableDiagnostic EP Catheter102-5-2Med Sweep7F115
401932Livewire SteerableDiagnostic EP Catheter202-8-2-(60)-2-8-2Super Lrg Curl7F95
401918Livewire SteerableDiagnostic EP Catheter202-20-2-2-2-2-2-2-2-2-2-2-25-2-25-2-25-2Super Lrg Curl7F95
401580Livewire SteerableDiagnostic EP Catheter62-5-2Med Sweep7F115
401581Livewire SteerableDiagnostic EP Catheter82-5-2Med Sweep7F115
401938Livewire SteerableDiagnostic EP Catheter105-5-5Med Sweep5F115
401939Livewire SteerableDiagnostic EP Catheter102-2-2Med Sweep5F115
401940Livewire SteerableDiagnostic EP Catheter102-5-2Med Sweep5F115
401941Livewire SteerableDiagnostic EP Catheter102-8-2Extra Lrg Curl5F115
401990Livewire SteerableDiagnostic EP Catheter102-2-2Extra Lrg Curl5F115
401991Livewire SteerableDiagnostic EP Catheter102-5-2Extra Lrg Curl5F115
401575Livewire SteerableDiagnostic EP Catheter102-5-2Med Sweep6F115
401915Livewire SteerableDiagnostic EP Catheter102-5-2Extra Lrg Curl6F115
401923Livewire SteerableDiagnostic EP Catheter102-2-2Extra Lrg Curl6F115
401926Livewire SteerableDiagnostic EP Catheter102-8-2Extra Lrg SweepCSL6F115
401904Livewire SteerableDiagnostic EP Catheter202-10-2Super Lrg Curl7F95
401905Livewire SteerableDiagnostic EP Catheter205-5-5Super Lrg Curl7F95
401914Livewire SteerableDiagnostic EP Catheter202-5-2Super Lrg Curl7F95
401908Livewire SteerableDiagnostic EP Catheter202-2-2Med Curl7F115
401917Livewire SteerableDiagnostic EP Catheter82-2-2Med Sweep6F115
401949Livewire SteerableDiagnostic EP Catheter82-2-2Lrg Sweep6F115
401600Livewire SteerableDiagnostic EP Catheter42-5-2Med Sweep6F115
401603Livewire SteerableDiagnostic EP Catheter42-5-2Med Curl6F115
401572Livewire SteerableDiagnostic EP Catheter42-5-2Med Sweep6F115
401606Livewire SteerableDiagnostic EP Catheter42-5-2Lrg Sweep6F115
401933Livewire SteerableDiagnostic EP Catheter45-5-5Med Sweep6F115
Item NumberDevice DescriptionElectrodesSpacing (mm)Curve TypeFrenchSizeUsableLength(cm)
401934Livewire SteerableDiagnostic EP Catheter45-5-5Lrg Sweep6F115
401576Livewire SteerableDiagnostic EP Catheter42-5-2Sml Sweep7F115
401577Livewire SteerableDiagnostic EP Catheter42-5-2Med Sweep7F115
401578Livewire SteerableDiagnostic EP Catheter42-5-2Lrg Sweep7F115
401586Livewire SteerableDiagnostic EP Catheter42-5-2Sml Curl7F115
401587Livewire SteerableDiagnostic EP Catheter42-5-2Med Curl7F115
401588Livewire SteerableDiagnostic EP Catheter42-5-2Lrg Curl7F115
401652Livewire SteerableDiagnostic EP Catheter82-2-2Med SweepCRD-25F115
401653Livewire SteerableDiagnostic EP Catheter62-5-2Med SweepCRD25F115
401654Livewire SteerableDiagnostic EP Catheter65Med SweepCRD-25F115

The item numbers in scope of this submission are as follows:

{7}------------------------------------------------

Table 5.1: Device Scope

Technological Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed Livewire Steerable Diagnostic Electrophysiology (EP) Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Livewire Steerable Diagnostic EP Catheter. This included the following:

  • Biocompatibility ●
  • Cleaning Validation ●
  • Sterilization Validation ●
  • Functional Testing ●
    • Visual Inspection ●
    • Dimensional Verification .
    • Electrical Continuity and Resistance .
    • . Simulated Use
    • Mechanical Characteristics ●
  • Electrical Safety Testing ●
    • Dielectric and Current Leakage .
  • Packaging Validation

{8}------------------------------------------------

The Reprocessed Livewire Steerable Diagnostic EP Catheters are reprocessed no more than three (3) times. Each device is marked, serialized and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

Conclusion:

Innovative Health concludes that the Reprocessed Livewire Steerable Diagnostic EP Catheters are as safe and effective as the predicate devices described herein.

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).