K Number
K190053

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-02-01

(21 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SECULOK™ ACP System is intended for anterior vertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
· Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration disc confirmed by patient history and radiographic studies):

  • · Spondylolisthesis
  • · Trauma (including fractures, dislocation)
  • · Spinal stenosis
  • Tumors
  • · Deformity (defined as scoliosis, kyphosis, or lordosis
  • · Pseudoarthrosis
  • · Failed previous fusion

WARNING: The device is not approved for screw attachment or the posterior elements (nedicles) of the cervical, thoracic or lumbar spine.

Device Description

The SECULOK™ ACP System consists of a variety of shapes and sizes of bone plates, screws and associated instruments. Fixation is provided by bone screws inserted into the vertebral body of the cervical spine using an anterior approach. All implant components are made from titanium alloy (Ti-6AI-4V ELI) in accordance with ASTM F136. All implants are single use only. The width of plates of to the proposed SECULOK™ ACP System was modified compared to the prior devices (K182055).

AI/ML Overview

This document is a 510(k) summary for the SECULOK™ ACP System, a spinal intervertebral body fixation orthosis. It describes the device and its indications for use, and it demonstrates substantial equivalence to predicate devices based on non-clinical (bench) testing.

Here's an analysis of the provided information concerning acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance Criteria (from Predicate Device K182055)Reported Device Performance (SECULOK™ ACP System)
Static compression bending testMeets acceptance criteria of prior devices (K182055)Met the acceptance criteria of the predicate devices (K182055).
Static torsion testMeets acceptance criteria of prior devices (K182055)Met the acceptance criteria of the predicate devices (K182055).
Compression bending fatigue testMeets acceptance criteria of prior devices (K182055)Met the acceptance criteria of the predicate devices (K182055).

Note: The document explicitly states that the acceptance criteria for the SECULOK™ ACP System were "the acceptance criteria of the prior devices (K182055)." However, the specific quantitative values for these criteria are not provided in this document. It only states that the device met these unstated criteria.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for the Test Set: Not explicitly stated. The document refers to "the proposed devices" and "the data of the proposed devices," implying that multiple devices were tested for each performance evaluation. However, the exact number of units tested per test type is not provided.
  • Data Provenance: The testing was "non-clinical setting (bench testing)." No information about country of origin of data in terms of patient data is relevant as it's bench testing. It's prospective in the sense that the testing was performed specifically to support this 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This question is not applicable as the study involved bench testing (mechanical performance evaluation of a physical device), not an evaluation of diagnostic accuracy or clinical outcomes requiring expert interpretation of medical data. Therefore, there was no "ground truth" established by experts in this context.

4. Adjudication Method for the Test Set:

  • This question is not applicable for the same reason as point 3. Bench testing results are typically quantitative measurements against defined engineering specifications, not subjective interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC comparative effectiveness study was not done. The presented study is a non-clinical (bench) performance evaluation of the device itself, not an assessment of human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • This question is not applicable. The device (SECULOK™ ACP System) is a physical spinal fixation system, not a software algorithm or an AI-powered diagnostic tool. Therefore, a "standalone algorithm performance" study is irrelevant to this device.

7. The Type of Ground Truth Used:

  • The "ground truth" in this context refers to the established mechanical performance specifications and safety profiles derived from recognized industry standards (ASTM F1717) and the prior predicate device (K182055). The performance of the new device was compared against these pre-defined, acceptable mechanical limits and characteristics.

8. The Sample Size for the Training Set:

  • This question is not applicable. There is no "training set" as this is a physical medical device, not a machine learning algorithm that requires training data.

9. How the Ground Truth for the Training Set Was Established:

  • This question is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 1, 2019

U&I Corporation Kwang-Eun Song Regulatory Affairs Specialist 20. Sandan-ro 76beon-gil(Rd) Uijeongbu-si, Gyeonggi-do 11781 KOREA

Re: K190053

Trade/Device Name: SECULOK™ ACP System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: January 8, 2019 Received: January 11, 2019

Dear Kwang-Eun Song:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ronald P. Jean -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190053

111,000

Device Name SECULOK™ ACP System

Indications for Use (Describe)

The SECULOK™ ACP System is intended for anterior vertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:

· Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration disc confirmed by patient history and radiographic studies):

  • · Spondylolisthesis
  • · Trauma (including fractures, dislocation)
  • · Spinal stenosis
  • Tumors
  • · Deformity (defined as scoliosis, kyphosis, or lordosis
  • · Pseudoarthrosis
  • · Failed previous fusion

WARNING: The device is not approved for screw attachment or the posterior elements (nedicles) of the cervical, thoracic or lumbar spine.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

6. 510(k) Summary

Manufacturer:U & I Corporation20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,11781, Korea,
Sponsor:U & I Corporation20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,11781, Korea,
Sponsor Contact:Kwang-Eun Song, Regulatory Affairs Specialist+82 31 860 6837kesong@youic.com
Date Prepared:January 08, 2019
Device Name:Trade Name: SECULOKTM ACP System
Classification Name:Spinal Intervertebral Body Fixation Orthosis
Common Name:Anterior Cervical Plate
Product Code:KWQ
Predicate Devices:
Primary - SECULOKTM ACP System (K182055)Secondary - ASPIRONTM ACP System (K131200, K140234)

Description of Device:

The SECULOK™ ACP System consists of a variety of shapes and sizes of bone plates, screws and associated instruments. Fixation is provided by bone screws inserted into the vertebral body of the cervical spine using an anterior approach. All implant components are made from titanium alloy (Ti-6AI-4V ELI) in accordance with ASTM F136. All implants are single use only. The width of plates of to the proposed SECULOK™ ACP System was modified compared to the prior devices (K182055).

{4}------------------------------------------------

Indications For Use:

The SECULOK™ ACP System is intended for anterior vertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:

  • . Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration disc confirmed by patient history and radiographic studies);
  • Spondylolisthesis O
  • Trauma (including fractures, dislocation)
  • Spinal stenosis O
  • O Tumors
  • O Deformity (defined as scoliosis, kyphosis, or lordosis)
  • O Pseudoarthrosis
  • Failed previous fusion

WARNING: The device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.

Substantial Equivalence:

SECULOK™ ACP System is substantially equivalent to SECULOK™ ACP System (K182055) and ASPIRON™ ACP System (K131200, K140234) in design, material, mechanical performance, function and intended use.

    1. Comparison Technological Characteristics
      The predicate and proposed devices have the similar intended use and basic fundamental scientific technology and share the following similarities;
  • The similar indications for use

  • O Similar design features

  • Incorporate the same materials O

  • The equivalent mechanical performance

{5}------------------------------------------------

2. Performance evaluation

The SECULOK™ ACP System was evaluated in a non clinical setting (bench testing) to verify that the data of the proposed devices meet the acceptance criteria of the prior devices (K182055) and no new safety and efficiency issues were raised with this device.

The following tests were performed same as test methods and parameters of the prior system (K182055):

  • (1) Static compression bending test according to ASTM F1717
  • (2) Static torsion test according to ASTM F1717
  • (3) Compression bending fatigue test according to ASTM F1717

The mechanical performance of the proposed devices met the acceptance criteria of the predicate devices (K182055).

The SECULOK™ ACP System is substantially equivalent to predicate devices.

3. Conclusion

The data and information provided in this submission support the conclusion that the SECULOK™ ACP System is substantially equivalent to predicate devices with respect to indications for use and technological characteristics.

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.