K Number
K183674

Validate with FDA (Live)

Date Cleared
2019-10-11

(288 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mode 1 (PMS): To be used for stimulating healthy muscles in order to improve and facilitate muscle performance.

Mode 2 (TENS): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities

Device Description

Electronic Pulse Stimulator is a product that adopts modern electronic science and technology to delivers electric pulses generated to the user's skin through the electrodes. It has two modes of different pulse frequencies, mode 1 uses Electrical Muscle Stimulation (EMS) technology for muscle training and mode 2 uses Transcutaneous Electrical Nerve Stimulation (TENS) technology to temporary relieve pain. The device includes operating elements of Power ON/OFF button, intensity increase"+" button, and intensity decrease"-" button and Mode selection button, and could be attached and detached to the electrode pad through the snap-on connector.

In additional, according to different simulation-needed bodies, users can choose the four types of electrode pad based on their own situation. Electronic Pulse Stimulator is suitable for the simulation-needed bodies including hips, arms and legs, abdomen, and feet.

AI/ML Overview

The provided document is a 510(k) Premarket Notification summary for an Electronic Pulse Stimulator (Model S3). It does not contain information about acceptance criteria for a study proving the device meets the acceptance criteria.

Instead, this document focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and performance data related to existing safety and electrical standards. The "performance data" section primarily discusses compliance with various IEC standards (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10, IEC 60601-1-2) and biocompatibility testing (ISO 10993-1). There is no mention of a clinical study with specific acceptance criteria that would involve human subjects, expert readers, or a ground truth process as typically defined for performance evaluation of diagnostic or AI-driven medical devices.

Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the provided text does not contain this type of information. It is a regulatory submission for a simple TENS/EMS device for pain relief and muscle stimulation, not a complex diagnostic system that would involve the kind of performance study you are asking about (e.g., MRMC studies, expert ground truth adjudication).

To directly answer your numbered points based on the absence of such information in this document:

  1. A table of acceptance criteria and the reported device performance: Not provided. The document outlines technological characteristics and compliance with electrical safety and biocompatibility standards, not performance metrics from a study with defined acceptance criteria.
  2. Sample sizes used for the test set and the data provenance: Not applicable. No "test set" in the context of a performance study is described. The "testing" refers to bench testing against industry standards.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No ground truth establishment is described.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No test set adjudication is described.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-driven or diagnostic device, and no MRMC study is mentioned.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device does not involve an algorithm with standalone performance in the sense of AI.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. Ground truth for a performance study is not established or discussed.
  8. The sample size for the training set: Not applicable. This device is not described as involving a training set for an algorithm.
  9. How the ground truth for the training set was established: Not applicable. No training set or ground truth establishment for it is described.

The document states that "The electronic pulse stimulator is substantially equivalent to the predicate devices based on intended use, design, specifications and performance." The "performance" here refers to meeting established electrical safety and biocompatibility standards, not a clinical performance validation against a pre-defined ground truth in a study with human subjects.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

Shenzhen Leading Perfection Technology Co., Ltd % Rain Yip Registered Engineer Feiying Drug & Medical Consulting Technical Service Group Rm. 3005. Area B. Bldg.1. Southward Ruifeng Business Center. Guimiao Road Shenzhen, Guangdong 518000 CN

Re: K183674

Trade/Device Name: Electronic Pulse Stimulator, Model S3 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH. NGX Dated: March 30, 2019 Received: September 9, 2019

Dear Rain Yip:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Vivek Pinto, PhD Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183674

Device Name Electronic Pulse Stimulator (Model: S3)

Indications for Use (Describe) Mode 1 (PMS): To be used for stimulating healthy muscles in order to improve and facilitate muscle performance.

Mode 2 (TENS):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

"510(K) Summary" as required by 21CFR 807.92.

Date: 2019-01-25

Submitter I.

Shenzhen Leading Perfection Technology Co., Ltd 4th Floor, J Building, San Yi Dui Industrial Park, Zhoushi Road, Jiuwei, Hangcheng Street, Bao'an District, Shenzhen, China

Tel.: +86 755 2999 1449 Tax: +86 755 2371 3490

Simon Hou (General Manager) Tel: +86 134 8075 8801 Email: info@leadingind.com

II. Device

Type of 510(k): Traditional Common Name: Powered muscle stimulator Trade Name: Electronic Pulse Stimulator Model: S3 Classification Name: Transcutaneous electrical nerve stimulator for pain relief Review Panel: Neurology Regulatory Class: II Product Code: NUH, NGX Regulation Number: 21 CFR 882.5890

III. Predicate Device

ApplicantPredicate Device510(k) NumberApproval Date
PhilipsLifstyle(Primary):PulseReliefK151035July 21,2015
HIVOX BIOTEK INC.HIVOX OTC ElectricalStimulator /SEM44K171803November 29, 2017
ShenzhenOSTOTechnology CompanyLimitedHealth Expert ElectronicStimulator, Model AST-300C and AST-300DK133929November 12, 2014

{4}------------------------------------------------

IV. Device Description

Electronic Pulse Stimulator is a product that adopts modern electronic science and technology to delivers electric pulses generated to the user's skin through the electrodes. It has two modes of different pulse frequencies, mode 1 uses Electrical Muscle Stimulation (EMS) technology for muscle training and mode 2 uses Transcutaneous Electrical Nerve Stimulation (TENS) technology to temporary relieve pain. The device includes operating elements of Power ON/OFF button, intensity increase"+" button, and intensity decrease"-" button and Mode selection button, and could be attached and detached to the electrode pad through the snap-on connector.

In additional, according to different simulation-needed bodies, users can choose the four types of electrode pad based on their own situation. Electronic Pulse Stimulator is suitable for the simulation-needed bodies including hips, arms and legs, abdomen, and feet.

V. Indications for Use

Mode 1 (PMS): To be used for stimulating healthy muscles in order to improve and facilitate muscle performance.

Mode 2 (TENS): To be used for temporary relief of pain associate with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.

VI. Comparison of Technological Characteristics With the Predicate Devices

The Electronic Pulse Stimulator is substantially equivalent to the predicate devices based on intended use, design, specifications and performance.

The Electronic Pulse Stimulator raises no safety or efficacy concerns when compared to the following predicate devices.

Information for predicate device was obtained from publicly available sources, including the 510(k) Summary and device instruction manual. A technical comparison to the predicate is provided below:

{5}------------------------------------------------

Comparison ElementsSubiect Device(Primary) Predicate Device K151035Predicate Device K171803Predicate Device K133929Comparison
Trade Name/ModelElectronic Stimulator/S3PulseReliefHIVOX OTC Electrical Stimulator/SEM44Health Expert Electronic Stimulator, Model AST-/300C and AST-300D/
510(k) NumberTo be assignedK151035K171803K133929/
ManufacturerShenzhen Leading Perfection Technology Co., LtdPhilips LifstyHIVOX BIOTEK INC.Shenzhen OSTO Technology Company Limited/
Regulation Number21CFR 882.589021CFR 882.5890 and 890.585021CFR 882.589021 CFR 882.5890SE
Device ClassIIIIIIIISE
Product CodeNUH, NGXNUH, NGXNUH, NGXNUH, NGXSE NOTE 1
Indication for Use/Intended UseMode 1(PMS): To be used for stimulating healthy muscles in order to improve and facilitate muscle performance.Mode 2(TENS): To be used for temporary relief of pain associate with sore and aching muscles in the shoulder, waist, back, neck, upper extremities(arm) andThe OTC TENS/EMS stimulator PulseRelief is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities(arm) andSEM44 (EMS): The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.SEM44 (TENS): The device is designed to be used for temporary reliefPMS (Mode 18) It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.TENS (Mode 925) To be used for temporary relief of pain associated with sore and aching muscles inSE NOTE 1
extremities (leg),abdomen and bottomdue to strain fromexercise or normalhousehold workactivities
ComparisonElementsSubject Device(Primary) PredicateDevice K151035Predicate DeviceK171803Predicate DeviceK133929Comparison
lower extremities (leg)due to strain fromexercise or normalhousehold workactivities.It should be applied tonormal, healthy, dryand clean skin of adultpatients, and is to beused for stimulatehealthy muscles inorder to improve andfacilitate muscleperformance.of pain associated withsore and aching musclesin the shoulder, waist,back, neck, upperextremities (arm), lowerextremities (leg),abdomen and bottom dueto strain from exercise ornormal household workactivities.the shoulder, waist, back,back of the neck, arm, leg,and foot due to strain fromexercise or normal householdwork activities by applyingcurrent to stimulate nerve.
Shape of electrodesRectangleSquareUnknownRectangleSE
Anatomical applicationArm, leg, abdomen,hip, feetShoulder, waist, back,upper extremities(arm), and lowerextremities (leg)Shoulder, waist, back,neck, upper extremities(arm), lower extremities(leg), abdomen andbottomShoulder, waist, back, backof the neck, arm, leg, andfootSE
Applicable PeopleAdultsAdultsAdultsAdultsSE
Location for UseOTCOTCOTCOTCSE
Basic Unit Characteristics
ComparisonElementsSubiect Device(Primary)Predicate DeviceK151035Predicate DeviceK171803Predicate DeviceK133929Comparison
Power SupplyUSB rechargeablebattery, 3.7VLi-ion 3.7V 500mAH4.5V (batteries, 3×1.5V AAA)Adaptor Input: 100-240Vac, 50-60Hz, 0.1AOutput: 5Vdc, 1AUnit Input: 5Vdc, 1ASENOTE 2
Method of LineCurrent IsolationN/AN/AN/AType BF Applied PartSE
PatientLeakageCurrent< 0.1<10μAN/AAC: 54.5μA, DC: 0.5μASE
Signal FaultCurrentConditionN/A<50μAN/AAC:120.0μA,DC: 0.6μASE
Number of OutputModes2 (1 TENS and 1 EMS)EMS: 5TENS:15EMS: 35TENS:15EMS: 8TENS:17SENOTE 2
Number of OutputChannels1122SENOTE 2
Synchronous orAlternating?N/AN/ASynchronousSynchronousSE
Method of ChannelIsolationN/AN/ABy electrical circuit andsoftwareVoltage TransformIsolation"BODY▼" and" BODY▼" buttons forbody channel," SOLE ▲" and"SOLE▼" buttons forfeet channelSE
ComparisonElementsSubiect Device(Primary) PredicateDevice K151035Predicate DeviceK171803Predicate DeviceK133929Comparison
Software /Firmware/MicroprocessorControl?YesYesYesYesSE
Automatic OverloadTrip?YesYesYesNoSE
YesYesYesNoSE
Automatic No-LoadTrip?YesYesYesNoSE
Automatic Shut OFF?YesYesYesYesSE
Patient OverrideControl?YesYesYesYesSE
IndicatorDisplayOn/OffStatus?YesYesYesYesSE
LowBattery?YesYesYesNoSE
Voltage/CurrentLevel?NoYesYesYesSE
Timer Range25 minutes1~59minutes andcontinuous5~100minutes25 minutesSENOTE 2
CompliancewithVoluntary Standards?IEC60601-1-2IEC60601-1IEC60601-11IEC60601-1-2IEC60601-1IEC60601-2-10IEC60601-1-2IEC60601-1IEC60601-2-10IEC60601-1-2IEC60601-1IEC60601-2-10SE
ComparisonElementsSubiect Device(Primary)PredicateDevice K151035Predicate DeviceK171803Predicate DeviceK133929Comparison
IEC60601-2-10ISO10993-5/10ISO10993-5/10ISO10993-5/10ISO10993-5/10
Compliancewith21CFR898YesYesYesYesSE
Weight14.5g(with battery)62g (excl. electrodes)89g (including belt clip,without batteries)123g (including belt clip,and batteries)2Kg (Without accessories)SENOTE 3
DimensionsФ4.2cm×1.38cm54×54×14mm (unit)132 x 63 x29.5mm(includingclip)428mm x 428.8mm x185mmSENOTE 3
Housing Materials andConstruction(Mainunit)ABSABS, PCABSABSSE
Output Specifications
WaveformSymmetricalrectangularBiphasic, rectangularBiphasic, squarePulsed, symmetric,Biphasic, Rectangular,SENOTE 4
MaximumOutput$@500Ω$40±20%31±20%100±10 % (50±10 %(Vp))44±10%SENOTE 4
Voltage(Volts,$@2kΩ$64±20%69±20%180±10 % (90±10 %(Vp))80±10%SENOTE 4
ComparisonElementsSubject Device(Primary)PredicateDevice K151035Predicate DeviceK171803Predicate DeviceK133929Comparison
Vp)@10kΩ76±20%70±20%250±10 % (125±10 %(Vp))112±10%SENOTE 4
MaximumOutputCurrent(mA)@500Ω80±20%62±20%200±10 % (100±10 %(Ip))88±10%SENOTE 4
@2kΩ32±20%34±20%90±10 % (45±10 %(Ip))40±10%SENOTE 4
@10kΩ7.6±20%7±20%25±10 % (12.5±10 %(Ip))11.2±10%SENOTE 4
Pulse Width200μsTENS: 60350μsEMS: 150350μs50-450μs120μsSENOTE 4
Frequency (Hz)Mode 1: 4-35HzMode 2: 4-41HzTENS: 1100HzEMS: 4565Hz1-150Hz77.3HzSENOTE 4
MaximumPhaseCharge (μC) @500Ω22.31TENS: 1.66.8EMS: 4.710.90.04512.78μC @ 500ΩSENOTE 4
MaximumDensity@500ΩCurrent(mA/cm²)Hips pad: 0.079Arms and legs pad:0.104Abdomen pad: 0.111Feet pad: 0.016TENS: 0.0020.045EMS: 0.0190.0370.6670.235mA/cm² @ 500ΩSENOTE 4
MaximumDensity@500ΩPower(W/cm²)Hips pad: 0.0001Arms and legs pad:0.0002Abdomen pad: 0.0002(average)TENS: 0.241.69EMS: 0.621.150.00461.38mW/cm² @ 500ΩSENOTE 4
ComparisonElementsSubject Device(Primary) PredicateDevice K151035Predicate DeviceK171803Predicate DeviceK133929Comparison
Feet pad: 0.00003
BurstModePulses perburstN/ATENS: 5, 7EMS: N/A3--SE
Bursts persecondN/ATENS: 1, 2, 3EMS: N/A2/60Hz--SE
Burstduration(seconds)N/ATENS: 62.5~87.5EMS: N/A36ms--SE
Duty Cycle[Line(b) xLine (c)]N/ATENS: 6.3%~19%EMS: N/A36ms/390ms--SE
ON Time (seconds)1sTENS: N/AEMS:--intensity ramp-up:24sec--constant intensity:56sec--intensity ramp-down:1~2sec2s0.6sSE
OFF Time (seconds)4sTENS: N/AEMS: N/A2s0.6sSE

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

Comnarison in details:

NOTE 1: The "Intended Use" of the subject device is the substantially equivalent to the predicate devices selected, all have adopted TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscle Stimulation) technologies to the treatment of temporary relief of pain associate with sore and aching muscle performance improvement of healthy muscles respectively. Meanwhile, the

{13}------------------------------------------------

"Product Code" of the subject device is the same as the predicate devices. This define does not affect the intended use of the subject device.

NOTE 2: Although the "Power Supply", "Number of Output Channels", and "Timer Range" of the subject device are a little different from the predicate devices, they are all complaint with the requirements of IEC 60601-1 standard. So these differences will not raise any safety or effectiveness issue.

NOTE 3: Although the "Weight" and "Dimensions" of the subject device are different from the predicate devices are insignificant in the terms of safety or effectiveness.

NOTE 4: Although the "Waveform", "Maximum Output Current", "Pulse Duration", "Frequency", "Maximum Phase Charge", "Maximum Current Density", and "Maximum Power Density" of the subject device are a little different from the predicate devices, they all comply with the requirements of IEC 60601-1 and IEC 60601-2-10 standard, as well as the Guidance for Powered Muscle Stimulator for Muscle Conditioning. So these differences will not raise any safety or effectiveness issue.

CONCLUSION: The subject device Electronic Pulse Stimulator is substantially equivalent to the predicate devices.

{14}------------------------------------------------

VII.Performance Data

The following performance data were provided in support of the substantial equivalence determination.

1) Biocomnatibility Testing

The direct components of applied parts contacting with the users in the Electronic Pulse Stimulator is as the following item:

Component NameMaterial of ComponentBody Contact CategoryContact Duration
Electrode pads used for hips, arms and legs, and abdomen (hips pad, arms and legs pad, and abdomen pad)Conductive hydrogelSurface-contacting skinMax. 25 minutes, less than 24 hours
Electrode pads used for feet (feet pad)PU fabric (contained sliver paste and conductive oil)Surface-contacting skinMax. 25 minutes, less than 24 hours

The electrode pads are directly purchased from qualified supplier which has been tested and passed the In Vitro Cytotoxicity, Skin Sensitization test in accordance with the ISO10993-1 standards. So we have reason to believe that the electrode pads are safe for the users. For details, please refer to "Biocompatibility Discussion" in this application.

2) Electrical and EMC Safety

Electrical safety and EMC safety testing was performed to, and passed, the following standards:

  • IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-11 Medical electrical equipment -Part 1-11: General requirements for basic safety and essential performance -Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

  • IEC 60601-2-10 Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators

  • IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility -Requirements and tests

In addition to the compliance of voluntary standards:

  • The software verification has been carries out according to the FDA Guidance for the Content of Premarket Submission for Software Contained in Medical Devices.

  • The waveform test report has also been conducted to verify the output specifications of the subject device according to the FDA Guidance for Powered Muscle Stimulator 510(k)s

  • The adhesion test report has been conducted to verify the maximum duration of use for the electrodes used by the subject device according to the requirements of AAMI EC 12_2000(R) 2010-Section 5.4.

{15}------------------------------------------------

  • The dispersion and shelf life test report has been conducted to verify the current dispersion and shelf-life of the electrodes used by the subject device in the expiration date according to the requirements of the FDA guidance –Shelf Life of Medical Device and ASTM F 1980-07 standard.

Summary

Based on the above performance as documented in this application, Electronic Pulse Stimulator was found to have a safety and effectiveness profile that is similar to the predicate devices.

VIII. Conclusions

The subject device Electronic Pulse Stimulator is to be concluded substantial equivalent to its predicate devices.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).