K Number
K183630

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-06-10

(166 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SPINEART® Navigation Instrument System reusable instruments are intended to be used during the preparation and placement of Spineart screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SPINEART® Navigation Instrument System reusable instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

Device Description

The SPINEART® Navigation Instrument System reusable instruments are surgical instruments for use with the Medtronic® StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants.

The SPINEART® Navigation Instrument System includes the following: Screwdrivers, Taps, Drills, and Drill Guides.

The SPINEART® Navigation Instrument System are to be used with the following Spineart Systems:

  • ■Romeo® 2
    ▪Romeo® 2 MIS

▪Perla® Cervico-thoracic Fixation System

All instruments are made of stainless steel per ASTM F899. Taps range in size from Ø4mm to Ø7.5mm for the Romeo® 2 systems and from Ø3mm to Ø4mm for the Perla system.

The SPINEART® Navigation Instruments are not compatible with implants from other manufacturers.

The SPINEART® Navigation Instrument System are designed for use only with Medtronic StealthStation Navigation System hardware and software.

AI/ML Overview

The provided text describes the regulatory clearance for the "SPINEART® Navigation Instrument System" and includes information about its intended use, technological characteristics, and performance data. However, it does not describe a study to prove the device meets acceptance criteria in the format requested, particularly for an AI/algorithm-driven device with detailed metrics like sensitivity, specificity, or AUC based on expert reads.

The device described is a surgical navigation instrument system, which is a physical device used during surgery to assist in precisely locating anatomical structures. The performance data section refers to non-clinical testing such as connection, registration, simulated use, and accuracy testing, which are relevant to the physical instrument's functionality and compatibility, rather than the performance of an AI/algorithm in interpreting medical data.

Therefore, I cannot fulfill the request as framed, as the provided document details the clearance of a physical medical instrument, not an AI/algorithm, and the information required for AI acceptance criteria and study design (e.g., sample size for test/training sets, expert consensus for ground truth, MRMC studies) is not present.

The document primarily focuses on establishing substantial equivalence to predicate devices based on:

  • Identical Technological Characteristics: Indications for use, materials, principle of operation, device technology, sizes, mechanical performance, packaging.
  • Performance Data: Connection, registration, simulated use, and accuracy testing (per ASTM F2554-10) to ensure functionality and compatibility with the Medtronic StealthStation®.

To illustrate why I cannot answer the specific questions:

  • 1. Table of acceptance criteria and reported performance: The document only mentions "accuracy testing per ASTM F2554-10" and states results "show that the performance...is sufficient for its intended use and is substantially equivalent." It doesn't provide specific numerical acceptance criteria (e.g., "accuracy > 95%") or reported performance values.
  • 2. Sample sized for the test set and data provenance: Not applicable in the context of physical instrument testing. "Test set" here refers to the instruments themselves and their interaction, not a dataset for an AI.
  • 3. Number of experts and qualifications: Not applicable. Ground truth for a physical instrument's accuracy is typically measured engineeringly, not through human expert interpretation of images.
  • 4. Adjudication method: Not applicable.
  • 5. MRMC comparative effectiveness study: Not applicable, as this is a physical instrument, not an AI assisting human readers.
  • 6. Standalone (algorithm only) performance: Not applicable, as there is no standalone algorithm being evaluated in the document.
  • 7. Type of ground truth: For a physical instrument, ground truth for accuracy testing would be engineering measurements against known standards, not expert consensus, pathology, or outcomes data in the sense of an diagnostic AI.
  • 8. Sample size for the training set: Not applicable, as this is not an AI model.
  • 9. How ground truth for training set was established: Not applicable.

In conclusion, the provided text describes the regulatory submission for a physical surgical instrument system, not an AI or algorithm. Therefore, the specific criteria and study design elements typically associated with AI/algorithm acceptance (e.g., sensitivity, specificity, expert consensus, MRMC studies) are not applicable and are not found in the document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

June 06, 2019

Spineart SA % Meredith May Vice President Empirical Consulting LLC 4628 Northpark Dr Colorado Springs, Colorado 80918

Re: K183630

Trade/Device Name: SPINEART Navigation Instrument System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: May 3, 2019 Received: May 6, 2019

Dear Meredith May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For: Shumaya Ali Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.

510(k) Number (if known) K183630

Device Name

SPINEART® Navigation Instrument System

Indications for Use (Describe)

The SPINEART® Navigation Instrument System reusable instruments are intended to be used during the preparation and placement of Spineart screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SPINEART® Navigation Instrument System reusable instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, includingthe time to review instructions, search existing data sources, gather and maintain the data needed andcomplete and review the collection of information. Send comments regarding this burden estimate or anyother aspect of this information collection, including suggestions for reducing this burden, to:Department of Health and Human ServicesFood and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff@fda.hhs.gov"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of informationunless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

5.510(K) SUMMARY

Submitter's Name:Spineart SA
Submitter's Address:Chemin du Pré- Fleuri 31228 Plan-les-OuatesSwitzerland
Submitter's Telephone:+41 22 570 1200
Contact Person:Meredith Lee May MS, RACEmpirical Consulting719.337.7579MMay@EmpiricalConsulting.com
Date Summary was Prepared:21-Dec-18
Trade or Proprietary Name:SPINEART® Navigation Instrument System
Common or Usual Name:Orthopedic Stereotaxic Instrument
Classification:Class II per 21 CFR §882.4260
Product Code:OLO
Classification Panel:Division of Orthopedic Devices

Description of the Device Subject to Premarket Notification:

The SPINEART® Navigation Instrument System reusable instruments are surgical instruments for use with the Medtronic® StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants.

The SPINEART® Navigation Instrument System includes the following: Screwdrivers, Taps, Drills, and Drill Guides.

The SPINEART® Navigation Instrument System are to be used with the following Spineart Systems:

  • ■Romeo® 2
    ▪Romeo® 2 MIS

▪Perla® Cervico-thoracic Fixation System

All instruments are made of stainless steel per ASTM F899. Taps range in size from Ø4mm to Ø7.5mm for the Romeo® 2 systems and from Ø3mm to Ø4mm for the Perla system.

The SPINEART® Navigation Instruments are not compatible with implants from other manufacturers.

The SPINEART® Navigation Instrument System are designed for use only with Medtronic StealthStation Navigation System hardware and software.

{4}------------------------------------------------

INDICATIONS FOR USE

The SPINEART® Navigation Instrument System reusable instruments are intended to be used during the preparation and placement of Spineart screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SPINEART® Navigation Instrument System reusable instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:

  • Indications for use
  • Materials of manufacture
  • Principle of operation
  • Device technology ●
  • Sizes ●
  • Mechanical performance
  • Packaging (materials and processes) ●
510k NumberTrade or Proprietary orModel NameManufacturerPredicateType
K143628/K140454/K143375Medtronic NavigatedInstrumentsMedtronic SofamorDanek USA, Inc.Primary
K172517SeaSpine NavigationSystemSeaSpine OrthopedicsCorporationAdditional
K153203Navigation InstrumentsGlobus Medical, Inc.Additional
K140948ROMEO® PosteriorOsteosynthesis SystemSpineartReference
K153386PERLA® PosteriorCervico-Thoracic FixationSystemSpineartReference

Performance Data

Design validation testing including connection, registration, simulated use, and accuracy was conducted to ensure that the SPINEART® Navigation Instrument System instruments are acceptable for their intended use, to ensure functionality and compatibility with the Medtronic StealthStation®, and to demonstrate substantial equivalence to the predicate instruments.

{5}------------------------------------------------

Connection testing evaluated the connection between the NavLock Tracker and the instruments. Registration testing was performed to ensure that the instruments can be registered to the StealthStation®. Simulated use testing was completed to verify successful implantation of Spineart screws by the SPINEART® Navigation Instrument System. Accuracy testing per ASTM F2554-10, "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems," was completed for comparison to the predicate instruments. The results of this non-clinical testing show that the performance of the SPINEART® Navigation Instrument System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the SPINEART® Navigation Instrument System is substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).