K Number
K183628

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-03-26

(90 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex TensionLocTM System is indicated for fixation of soft tissue, such as tendons and ligaments to bone in orthopedic procedures.

Device Description

The Arthrex TensionLocTM System is a two-component system comprised of a polyetheretherketone (PEEK) plug and a mating collar for suture fixation within a bone tunnel.

The primary differences between the Arthrex TensionLocTM System and the predicate device are the geometry and material. The proposed Arthrex TensionLocTM System contains a ribbed outer edge collar and plug as opposed to the predicate geometry of a double spiked plate and cannulated post. The Arthrex TensionLocTM System is also manufactured of PEEK material whereas the predicate is manufactured from titanium. The Arthrex Tension LocTM System will offer external diameters of 11, 14, and 16 as opposed to the 13-22 mm range of the predicate. The Arthrex TensionLocTM System is 9.1 mm in length as opposed to the various lengths of the predicate (24-40 mm).

AI/ML Overview

This document outlines the FDA's 510(k) clearance for the Arthrex TensionLocTM System, a medical device for orthopedic procedures. It focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed study of an AI/ML product's performance against acceptance criteria. Therefore, most of the requested information regarding AI/ML study design, acceptance criteria, multi-reader studies, and ground truth establishment is not applicable or cannot be extracted from this document.

However, I can extract information related to the device's performance testing from the "Performance Data" section of the 510(k) Summary.

Here's the relevant information that can be extracted:

  • Device Name: Arthrex TensionLocTM System
  • Purpose: Fixation of soft tissue (tendons and ligaments) to bone in orthopedic procedures.
  • Predicate Device: K994202: Acufex® Washer System

Since this is a mechanical device clearance and not an AI/ML product, the questions related to AI/ML specific study design (e.g., sample size for test/training set, number of experts, MRMC studies, ground truth establishment for AI) are not directly addressed.

The "Performance Data" section details the studies conducted to demonstrate safety and effectiveness for this type of medical device:

1. Acceptance Criteria and Reported Device Performance (Table):

Acceptance Criteria / Test TypeReported Device Performance (Summary)
Mechanical Testing
Cyclic Load TestingConducted and compared to the predicate device.
Ultimate Load TestingConducted and compared to the predicate device.
Biocompatibility Testing
ISO 10993-1Conducted in accordance with requirements.
Sterilization Validation
ISO 11135:2014Conducted in accordance with requirements.
Bacterial Endotoxin Testing
EP 2.6.14/USP <85>Conducted to demonstrate device meets pyrogen limit specifications.

Performance Conclusion: "The submitted mechanical testing data demonstrates that the performance of the proposed devices meets or exceeds the predicate device for the desired indications."

2. Sample size used for the test set and the data provenance: Not explicitly stated for each test, but typically, these tests involve a defined number of device samples manufactured under controlled conditions. The data provenance is internal to the manufacturer (Arthrex Inc., Naples, FL). The study is prospective in the sense that the tests are designed and executed specifically for this regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For mechanical and biocompatibility testing of a physical medical device, "ground truth" is established through standardized testing protocols, material specifications, and comparison to predicate device performance, not expert human interpretation.

4. Adjudication method for the test set: Not applicable. Testing results are quantitative and objective, not subject to human adjudication in the context of diagnostic interpretation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No, this is not an AI/ML diagnostic device, so MRMC studies are not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical medical device, not a software algorithm.

7. The type of ground truth used:
* Mechanical Performance: Comparison to "predicate device performance" (implied "gold standard" of a previously cleared device) and meeting specified mechanical properties.
* Biocompatibility: Adherence to ISO 10993-1 standards, which involve in vitro and in vivo tests for cytotoxicity, sensitization, irritation, etc.
* Sterilization: Adherence to ISO 11135:2014 standards, ensuring a specified sterility assurance level (SAL).
* Pyrogenicity: Meeting bacterial endotoxin limits specified by pharmacopeial standards (EP/USP).

8. The sample size for the training set: Not applicable. This is not an AI/ML product that undergoes a "training" phase with data.

9. How the ground truth for the training set was established: Not applicable. No training set is used for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 26, 2019

Arthrex Inc. Stacy Valdez Senior Regulatory Affairs Associate 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K183628

Trade/Device Name: Arthrex TensionLocTM System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: December 18. 2018 Received: December 26, 2018

Dear Stacy Valdez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laurence D. Coyne -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K183628

Device Name Arthrex TensionLocTM System

Indications for Use (Describe)

The Arthrex TensionLoc™ System is indicated for fixation of soft tissue, such as tendons and ligaments to bone in orthopedic procedures.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date PreparedDecember 18, 2018
SubmitterArthrex Inc.1370 Creekside BoulevardNaples, FL 34108-1945
Contact PersonStacy ValdezSenior Regulatory Affairs Associate1-239-643-5553, ext. 72010Stacy.valdez@arthrex.com
Name of DeviceArthrex TensionLocTM System
Common NameScrew, Fixation, Bone
Product CodeHWC
Classification Name21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener
Regulatory ClassII
Predicate DeviceK994202: Acufex® Washer SystemK151092: Arthrex Short Suture Anchors (Reference)K062466: Interference Screws (Reference)K062747: Arthrex RetroButton (Reference)
Purpose of SubmissionThis Traditional 510(k) premarket notification is submitted to obtain clearance forthe Arthrex TensionLocTM System
Device DescriptionThe Arthrex TensionLocTM System is a two-component system comprised of apolyetheretherketone (PEEK) plug and a mating collar for suture fixation within abone tunnel.The primary differences between the Arthrex TensionLocTM System and thepredicate device are the geometry and material. The proposed ArthrexTensionLocTM System contains a ribbed outer edge collar and plug as opposed tothe predicate geometry of a double spiked plate and cannulated post. TheArthrex TensionLocTM System is also manufactured of PEEK material whereas thepredicate is manufactured from titanium. The Arthrex Tension LocTM System willoffer external diameters of 11, 14, and 16 as opposed to the 13-22 mm range ofthe predicate. The Arthrex TensionLocTM System is 9.1 mm in length as opposedto the various lengths of the predicate (24-40 mm).
Indications for UseThe Arthrex TensionLocTM System is indicated for fixation of soft tissue, such astendons and ligaments to bone in orthopedic procedures.
Performance DataCyclic and ultimate load testing was conducted and compared to the predicatedevice.Biocompatibility testing was conducted in accordance with the requirements ofISO 10993-1 Biological Evaluation of Medical Devices: Evaluation and testingwithin a risk management process.Sterilization validation has been conducted in accordance with ISO 11135:2014Sterilization of health-care products – Ethylene Oxide – Requirements for thedevelopment, validation, and routine control of a sterilization process for medicaldevices.
Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate thatthe device meets pyrogen limit specifications.
ConclusionThe Arthrex TensionLoc™ System is substantially equivalent to the predicatedevice in which the intended use, sterility, and fundamental scientific technologyare the same. Any differences between the proposed device and the predicatedevice are considered minor and do not raise questions concerning safety oreffectiveness.
The submitted mechanical testing data demonstrates that the performance ofthe proposed devices meets or exceeds the predicate device for the desiredindications.
Based on the indications for use, technological characteristics, and the summaryof data submitted, Arthrex Inc. has determined that the Arthrex TensionLoc™ System is substantially equivalent to the currently marketed predicate device.

{4}------------------------------------------------

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.