K Number
K183581

Validate with FDA (Live)

Date Cleared
2019-01-20

(30 days)

Product Code
Regulation Number
882.4305
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Codman Disposable Perforator is for use in perforating the cranium. When properly used, it is designed to automatically disengage once perforation is accomplished and when pressure is removed from the drill point.

Device Description

The Codman Disposable Perforators are pre-assembled, single-use, sterile devices that are designed to perforate the cranium. When properly used, the Disposable Perforators are designed to automatically disengage once perforation is accomplished and when pressure is removed from the drill point. They are designed with a Hudson end and are available in three color-coded sizes:

  • 14mm (blue ABS sleeve) – product number: 26-1221
  • 11mm (green ABS sleeve) product number: 26-1222 ●
  • 9mm (yellow ABS sleeve) product number: 26-1223 ●
AI/ML Overview

This document describes a 510(k) premarket notification for "Codman Disposable Perforators," which are medical devices used to perforate the cranium. The submission aims to demonstrate substantial equivalence to previously cleared devices.

Here's an analysis of the acceptance criteria and study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

TestTest Method / PurposeAcceptance CriteriaResult
Full Dose Mapping ValidationEstablish a new gamma radiation sterilization cycle in compliance with ISO 11137-1Dose Uniformity Ratio: <2 (Implicit for compliance with ISO 11137-1)Pass

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions a "full dose mapping study" but does not explicitly state the sample size used for this study. The data provenance is not specified, but given it's a regulatory submission for a medical device manufacturer (Integra LifeSciences Corp. in Mansfield, MA, USA), it's highly likely to be prospective data generated internally for the purpose of this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable to this type of submission. The "ground truth" here is the adherence to sterilization standards (ISO 11137-1) for a physical device. Expert clinical or image-based review is not part of this evaluation. The expertise would lie in sterilization validation specialists who conducted and analyzed the dose mapping study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is a technical validation of a manufacturing process (sterilization), not a subjective assessment requiring adjudication. The study results are objectively measured against established standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This submission is for a disposable perforator, a physical surgical tool. It does not involve AI, image analysis, or human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the dose mapping study is the regulatory standard for sterilization (ISO 11137-1), which defines requirements for achieving a Sterility Assurance Level (SAL) of 10^-6. The study aimed to demonstrate that the new gamma radiation sterilization cycle meets this standard and achieves the required dose uniformity.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device; therefore, there is no training set. The "training" in the context of device manufacturing refers to process validation.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for an AI/ML model. The "ground truth" for the overall device's safety and effectiveness relies on its previous 510(k) clearances and the demonstration that the only change (the sterilization cycle) does not compromise those established safety and effectiveness parameters, as verified by adherence to international sterilization standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

January 20, 2019

Integra LifeSciences Corp. Jocelyn Raposo Associate Director, Regulatory Affairs 11 Cabot Blvd Mansfield, Massachusetts 02048

Re: K183581

Trade/Device Name: Codman Disposable Perforators Regulation Number: 21 CFR 882.4305 Regulation Name: Powered Compound Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBF Dated: December 20, 2018 Received: December 21, 2018

Dear Jocelyn Raposo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew C. Krueger -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: 06/30/2020See PRA Statement below.
Indications for Use
510(k) Number (if known)
K183581
Device Name

Codman Disposable Perforator

Indications for Use (Describe)

The Codman Disposable Perforator is for use in perforating the cranium. When properly used, it is designed to automatically disengage once perforation is accomplished and when pressure is removed from the drill point.

Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Sevices (301) 443-6740 EF

{3}------------------------------------------------

Section 5 - 510(k) Summary

I. SubmitterIntegra LifeSciences Corp.11 Cabot Blvd.Mansfield, MA 02048
Establishment Registration Number: 1226348
Primary Contact:Phone:Jocelyn Raposo(781) 971-5688
Secondary Contact:Phone:Jennifer Siu(781) 971-5692

Date of Preparation: December 19, 2018

II. Device(s)

Device Proprietary NameCodman Disposable Perforators
Common NamePerforators
Classification NameDrills, Burrs, Trephines & Accessories(Compound, Powered) (21 CFR 882.4305)
Regulatory ClassificationII
Product CodeHBF

III. Predicate

The predicate devices for this submission are:

Device(s)

The predicate devices for this submission are:

510(k) NumberDate ClearedTitle
K791101Aug. 28, 1979Codman Disposable Perforator(14mm)
K071931Oct. 22, 2007Codman Disposable Perforators(11mm & 9mm)

IV. Device Description(s) The Codman Disposable Perforators are pre-assembled, single-use, sterile devices that are designed to perforate the cranium. When properly used, the Disposable Perforators are designed to automatically disengage once perforation is accomplished and when pressure is removed from the drill point. They are designed with a Hudson end and are available in three color-coded sizes:

  • 14mm (blue ABS sleeve) – product number: 26-1221
  • 11mm (green ABS sleeve) product number: 26-1222 ●
  • 9mm (yellow ABS sleeve) product number: 26-1223 ●

{4}------------------------------------------------

V. Indications for Use

The Indications for Use statement of the proposed device remains identical to the predicate devices'.

EquivalenceComparisonCodman DisposablePerforators (Predicates)Codman DisposablePerforators (Subject of ThisSubmission)
Indications for UseThe Codman DisposablePerforator is for use inperforating the cranium.When properly used, it isdesigned to automaticallydisengage once perforation isaccomplished and whenpressure is removed from thedrill point.The Codman DisposablePerforator is for use inperforating the cranium.When properly used, it isdesigned to automaticallydisengage once perforation isaccomplished and whenpressure is removed from thedrill point.

VI. Comparison to Predicate Device

The proposed Codman Disposable Perforators are identical to the currently marketed Codman Disposable Perforators (K791101 & K071931) except for the alternate gamma irradiation sterilization cycle proposed in this 510(k). The indications for use, design, principle of operation, performance, packaging, and shelf life remain identical to the predicate devices.

Substantial Equivalence Comparison
CharacteristicCodman DisposablePerforators (Predicates)Codman DisposablePerforators (Subject ofThis Submission)
ManufacturerCodman & Shurtleff, Inc.Identical
Classification PanelNeurologyIdentical
Classification NameDrills, Burrs, Trephines &Accessories (Compound,Powered) (21 CFR882.4305)Identical
Indications for UseThe Codman DisposablePerforator is for use inperforating the cranium.When properly used, it isdesigned to automaticallydisengage onceperforation isaccomplished and whenpressure is removed fromthe drill point.Identical
Operating PrincipleWith adequate pressureapplied on the perforatorpoint, engagement shouldoccur as the Hudson Endis rotated in a driver. Oncepressure is removed fromthe drill point the device isdesigned to disengage.Identical
ImplantNoIdentical
Single Use OnlyYesIdentical

{5}------------------------------------------------

VII.

Data

Performance

Special 510(k) Premarket Notification
Codman Disposable Perforators
Substantial Equivalence Comparison
CharacteristicCodman DisposablePerforators (Predicates)Codman DisposablePerforators (Subject ofThis Submission)
Design9 mm Perforatorconsisting of outer andinner body drill, drilldriver, compressionspring, drive pin, moldedsleeve, snap bearing,flanged nyliner11mm Perforatorconsisting of outer andinner body drill, drilldriver, compressionspring, drive pin, moldedsleeve, snap bearing,flanged nyliner14mm Perforatorconsisting of outer andinner body drill, drilldriver, compressionspring, drive pin, moldedsleeve, snap bearing,flanged nylinerIdentical
Shelf Life5 yearsIdentical
Sterilization MethodGammaIdentical
Sterility Assurance Level(SAL)10-6Identical
PackagingPETG Blister Tray withheat-sealed lid placed intoa unit boxIdentical

There were no changes made that affect the Codman Disposable Perforators' indications for use, principle of operation, performance, packaging, or shelf life. The only difference between the predicate and proposed devices is the use of an alternate gamma irradiation sterilization cycle to sterilize the products. There is no change to the sterilization dose parameters (25kGy-50kGy). The products continue to meet a SAL of 10-6 with the new cycle. A full dose mapping study has been performed to ensure there are no new issues relating to the safety or effectiveness of the products.

Please see the Summary of Testing table below for acceptance criteria and testing results.

Table Summary of Testing
TestTest Method /PurposeAcceptance CriteriaResult
Full Dose MappingValidationEstablish a newgamma radiationsterilization cycle incompliance withISO 11137-1Dose Uniformity Ratio: <2Pass

{6}------------------------------------------------

Integra LifeSciences Codman Specialty Surgical Special 510(k) Premarket Notification Codman Disposable Perforators

Animal Testing:

No animal studies were required based on similarities of the proposed device to the predicate device and from the results of the sterilization validation.

Clinical Testing:

No clinical studies were required based on similarities of the proposed device to the predicate device and from the results of the sterilization validation.

Conclusion:

Results from the dose mapping study have demonstrated that the alternate sterilization cycle is suitable and does not raise new issues of safety and effectiveness for the finished devices. Based on the indications for use, fundamental scientific technology and a comparison to the predicate devices, the subject Codman Disposable Perforators are substantially equivalent to the predicate devices.

§ 882.4305 Powered compound cranial drills, burrs, trephines, and their accessories.

(a)
Identification. Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.(b)
Classification. Class II (performance standards).