K Number
K183462

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-10-01

(291 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Applied Biosystems Bacillus anthracis Detection Kit is a real-time polymerase chain reaction (PCR) test kit intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences for Bacillus anthracis, or BA). The Applied Biosystems Bacillus anthracis Detection Kit is intended to test human whole blood (EDTA) specimens and blood culture specimens with growth detected by a continuous monitoring blood culture system. Blood culture specimens must be determined to contain gram-positive bacilli by Gram stain prior to testing of whole blood specimens must be performed concomitantly with standard of care blood culture.

The Applied Biosystems Bacillus anthracis Detection Kit is indicated for use in CLIA-certified high-complexity laboratories in response to a confirmed Bacillus anthracis event only in accordance with the guidelines provided by public health authorities prior to or during a public health emergency. Testing with the Applied Biosystems Bacillus anthracis Detection Kit must only be performed when public health authorities have determined the need for this test must only be used with specimens from individuals with clinical signs and symptoms of B. anthracis infection and who have either been exposed to B. anthracis or may have been exposed to B. anthracis.

The Applied Biosystems Bacillus anthracis Detection Kit is intended for use as an aid in the diagnosis of anthrax infection and results are for the presumptive identification of Bacillus anthracis. The diagnosis of B. anthracis infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of B. anthracis from cultures or directly from clinical specimens. The definitive identification of B. anthracis requires additional testing and confirmation procedures in consultation with the appropriate public health authorities for whom reports may be required.

The Applied Biosystems Bacillus anthracis Detection Kit has not been clinically evaluated with specimens collected from individuals with B. anthracis infection or those presumed to B. anthracis. 'B. anthracis Not detected' results do not preclude infection with Bacillus anthracis and should not be used as for diagnosis, treatment, or other patient management decisions.

Laboratories implementing this test must have the appropriate biosafety equipment, personal protective equipment (PPE), containment facilities and personnel trained in the safe handling of diagnostic clinical specimens potentially containing B. anthracis. Anthrax is a nationally notifiable disease caused by a biothreat microbial agent and must be reported to public health authorities.

The distribution of in vitro diagnostic devices for Bacillus spp. detection is limited to laboratories that follow public health guidelines that address appropriate biosafety conditions, interpretation of test results, and coordination of findings with public health authorities.

The Applied Biosystems Bacillus anthracis Detection Kit is intended for use with the AB1 7500 Fast Dx Real-Time PCR Instrument with analysis using the Applied Biosystems Bacillus anthracis Interpretive Software (BaIS).

Device Description

The Applied Biosystems™ Bacillus anthracis Detection Kit is a multiplexed real-time polymerase chain reaction (PCR) test kit intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences for B. anthracis. Reagents are lyophilized in a 96-well plate format as a fully formulated Mastermix and are stable at room temperature for up to one year. The kit is specifically designed for performing real-time PCR using the Applied Biosystems (ABI) 7500 Fast Dx instrument and software, with nucleic acids extracted from clinical specimens using a Qiagen manual extraction method or Roche MagNA Pure automated extraction methods. An automated interpretative software component (BalS) is included in the kit but supplied separately and operates on a computer(s) that is separate from the ABI 7500 Fast Dx computer.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly present a formal "Acceptance Criteria" table with pass/fail thresholds for all aspects. Instead, it describes performance studies and their outcomes. Based on the performance data provided, we can infer the de facto acceptance associated with demonstrating strong performance indicators.

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
Limit of Detection (LOD)Achieve >95% detection when averaged across three reagent lots for each matrix and extraction method.Qiagen DSP DNA Blood Mini Kit: - Whole Blood: 150 CFU/mL - Blood Culture, Aerobic: 10,000 CFU/mL - Blood Culture, Anaerobic: 10,000 CFU/mLRoche MagNA Pure: - Whole Blood: 50 CFU/mL - Blood Culture, Aerobic: 2270 CFU/mL - Blood Culture, Anaerobic: 3040 CFU/mL (All values are presumptively "met" the >95% detection criteria given they are reported as LODs).
Analytical Inclusivity100% detection rate for B. anthracis strains carrying both plasmid targets. Correctly identify 'Bacillus anthracis suspected' for strains with only one plasmid target.100% detection rate for B. anthracis strains that carry both plasmid targets. For three strains known to carry only one plasmid target, the assay generated 'Bacillus anthracis suspected' results as expected.
Analytical ExclusivityHigh no detection rate (close to 100%) for non-target organisms. No false positives.Initial no detection rate for exclusivity testing was 95.68%. Repeat testing was conducted, resulting in a final no detection rate of 100%. Some known Bacillus cereus strains (BAG1X1-1, 03BB102, and G-9241) that contain a pXO1-like plasmid were correctly identified as 'Bacillus anthracis suspected'.
Interfering SubstancesNo significant interference from common substances unless noted with appropriate labeling limitations.Identified specific technique-specific substances that interfere (e.g., MagNA Pure Wash Buffer, 10% Bleach, Ethanol >5%). Appropriate limitations were added to product labeling.
Microbial Interference100% detection rate of B. anthracis in the presence of other clinically-relevant organisms.100% detection rate of B. anthracis in the presence of potentially interfering organisms in whole blood and blood culture.
ReproducibilityConsistent results across operators, instruments, and reagent lots. Limited false negatives/positives in control samples. Positive percent agreement metrics indicate reproducibility.One low positive sample returned negative. One high positive sample returned 'Bacillus anthracis suspected' due to pXO1 assay failure, but returned 'Bacillus anthracis detected' on repeat. No false positive events out of 270 negative PCR tests. 24 sample replicates (across all three lots) returned initial indeterminate results, all determined negative by supervisor review. No invalid results.
Carry-Over/Cross-contaminationMinimal to no false positive events due to carry-over/cross-contamination.Automated Method (MagNA Pure): 2 false positive events out of 135 negative samples (Specificity: 98.5%). Manual Method (Qiagen DSP): No false positive events out of 139 negative samples (Specificity: 100%).Contamination events identified on surfaces during testing, highlighting areas of risk.
Clinical Specificity100% Negative Percent Agreement (NPA) for B. anthracis negative samples.For 401 blood culture specimens and 439 whole blood specimens: 100% Negative Percent Agreement (NPA) when compared to the expected negative result.
Clinical SensitivityHigh Positive Percent Agreement (PPA) for B. anthracis positive samples.96% Positive Percent Agreement (PPA) (CI: 90.4-96.5%) for 87 low-positive whole blood specimens. Three specimens initially generated 'Bacillus anthracis suspected' results.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Limit of Detection (LOD): 1612 technical replicates were tested. Data provenance is not specified by country, but it's analytical data likely generated in-house at MRIGlobal (developer). Prospective (controlled spike-in experiments).
  • Analytical Inclusivity: 24 different Bacillus anthracis strains tested in triplicate (24 strains * 2 extraction methods * 3 replicates = 144 technical replicates)
  • Analytical Exclusivity: 154 non-target organisms tested in wet lab.
  • Interfering Substances: 50 potentially interfering substances. Each substance tested in triplicate-paired samples (with and without B. anthracis). 50 * 2 * 3 = 300 samples (assuming paired means 3 concentrations per, and triplicate-paired means 3 replicates per substance per condition).
  • Reproducibility: 7 panel members tested twice a day by three teams on five non-consecutive days. Details not fully specified for total sample count, but likely included multiple replicates for each panel member/condition. (e.g., 7 panel members * 2 times/day * 3 teams * 5 days = 210 runs, each run likely involving multiple replicates). "No false positive events occurred out of the 270 PCR tests of negative samples."
  • Carry-Over/Cross-contamination:
    • Automated method (MagNA Pure): 9 runs, each with 31 samples, for a total of 279 samples. 135 negative samples.
    • Manual method (Qiagen DSP): 12 batches, total of 279 samples. 139 negative samples.
  • Clinical Specificity: 401 blood culture specimens and 439 whole blood specimens. Samples were "left-over fresh and frozen blood culture samples," "randomly accessed, residual blood," and "febrile whole blood samples." Collected prospectively and serially from three point-of-care collection sites within the US.
  • Clinical Sensitivity: 87 low-positive whole blood specimens (simulated/contrived). Each aliquot was spiked with 1 of 18 Bacillus anthracis strains. "Febrile whole blood specimens ... determined to be negative for Bacillus anthracis" were used as the matrix for spiking. Samples were contrived (prospective experimental design using collected negative clinical matrix). Clinical performance testing was conducted at three laboratories within the US.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number or qualifications of experts used to establish ground truth.

  • For analytical studies (LOD, Inclusivity, Exclusivity, Interference, Reproducibility, Carry-Over), ground truth is established by the known concentration/presence of cultured organisms or purified DNA, or by the known absence of the target.
  • For clinical specificity, samples were "assumed to be negative for Bacillus anthracis" based on their collection context (e.g., routine CBC, fever of unknown origin that would not indicate B. anthracis exposure, or confirmed to be positive for other bacterial species but not B. anthracis). No independent expert assessment or gold standard positive confirmation was used for these negative samples.
  • For clinical sensitivity, the ground truth was established by artificial spiking of known concentrations of Bacillus anthracis into negative clinical samples.

4. Adjudication Method for the Test Set

The document mentions adjudication in two instances:

  • Reproducibility: "24 sample replicates (across all three reagent lots) that returned initial indeterminate results. All were determined to be negative by supervisor review." This suggests a form of expert review for indeterminate results.
  • Analytical Exclusivity: "Some known Bacillus cereus strains... Detection of these strains was indicated by a test result of ‘Bacillus anthracis suspected’ with amplification of the plasmid manually confirmed by the supervisor." This also indicates supervisor review for specific expected "suspected" results.

No broad 2+1 or 3+1 adjudication method for clinical test sets is described. For the clinical performance, the results are directly compared to the assumed/contrived ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic (IVD) PCR test kit, not an imaging AI designed to assist human readers. Its performance is assessed as a standalone diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, the studies presented (Analytical Sensitivity, Inclusivity, Exclusivity, Interference, Microbial Interference, Reproducibility, Carry-Over/Cross-contamination, Clinical Specificity, Clinical Sensitivity) represent the standalone performance of the "Applied Biosystems™ Bacillus anthracis Detection Kit" which includes the AB1 7500 Fast Dx Real-Time PCR Instrument and the Applied Biosystems Bacillus anthracis Interpretive Software (BaIS). It is an automated test with automated interpretation and report generation, so its performance is inherently a "standalone" or "algorithm only" type of assessment.

7. The Type of Ground Truth Used

The ground truth varied by study type:

  • Analytical Studies (LOD, Inclusivity, Exclusivity, Interference, Microbial Interference, Reproducibility, Carry-Over): Controlled laboratory settings with known concentrations of purified DNA or cultured organisms. For negative controls, known absence of target.
  • Clinical Specificity: "Assumed to be negative for Bacillus anthracis" for clinical samples collected from individuals not suspected of anthrax, or confirmed to have other bacteria. This is a clinical "expected negative" based on the patient population and standard clinical practice, rather than an independent gold standard for B. anthracis negativity.
  • Clinical Sensitivity: Contrived samples where known concentrations of Bacillus anthracis were spiked into clinical samples previously determined to be negative for B. anthracis. This is a simulated positive ground truth.

8. The Sample Size for the Training Set

The document describes premarket validation studies for a diagnostic kit. It does not provide information about a "training set" size for model development, as this device is a PCR assay with interpretive software, not a machine learning model that undergoes a separate training phase with a distinct dataset of this nature. The "training" in this context refers to the development and optimization process, which is not typically quantified in terms of a "training dataset size" in the same way as for AI/ML algorithms.

9. How the Ground Truth for the Training Set was Established

As above, the concept of a "training set" and its ground truth establishment, as it applies to AI/ML, is not directly applicable to this PCR diagnostic kit. The "ground truth" during the development phase would have involved known positive and negative controls, spiked samples, and characterized bacterial strains, similar to the analytical studies performed for validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 1, 2019

MRIGlobal % Fran White President MDC Associates, LLC 180 Cabot Street Beverly, Massachusetts 01915

Re: K183462

Trade/Device Name: Applied Biosystems Bacillus anthracis Detection Kit Regulation Number: 21 CFR 866.4000 Regulation Name: Device to detect and identify biothreat microbial agents in human clinical specimens Regulatory Class: Class II Product Code: QIF, OOI Dated: December 13, 2018 Received: December 14, 2018

Dear Fran White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kristian Roth, Ph.D. Branch Chief Bacterial Multiplex and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183462

Device Name

Applied Biosystems(TM) Bacillus anthracis Detection Kit

Indications for Use (Describe)

The Applied Biosystems Bacillus anthracis Detection Kit is a real-time polymerase chain reaction (PCR) test kit intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences for Bacillus anthracis, or BA). The Applied Biosystems Bacillus anthracis Detection Kit is intended to test human whole blood (EDTA) specimens and blood culture specimens with growth detected by a continuous monitoring blood culture system. Blood culture specimens must be determined to contain gram-positive bacilli by Gram stain prior to testing of whole blood specimens must be performed concomitantly with standard of care blood culture.

The Applied Biosystems Bacillus anthracis Detection Kit is indicated for use in CLIA-certified high-complexity laboratories in response to a confirmed Bacillus anthracis event only in accordance with the guidelines provided by public health authorities prior to or during a public health emergency. Testing with the Applied Biosystems Bacillus anthracis Detection Kit must only be performed when public health authorities have determined the need for this test must only be used with specimens from individuals with clinical signs and symptoms of B. anthracis infection and who have either been exposed to B. anthracis or may have been exposed to B. anthracis.

The Applied Biosystems Bacillus anthracis Detection Kit is intended for use as an aid in the diagnosis of anthrax infection and results are for the presumptive identification of Bacillus anthracis. The diagnosis of B. anthracis infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of B. anthracis from cultures or directly from clinical specimens. The definitive identification of B. anthracis requires additional testing and confirmation procedures in consultation with the appropriate public health authorities for whom reports may be required.

The Applied Biosystems Bacillus anthracis Detection Kit has not been clinically evaluated with specimens collected from individuals with B. anthracis infection or those presumed to B. anthracis. 'B. anthracis Not detected' results do not preclude infection with Bacillus anthracis and should not be used as for diagnosis, treatment, or other patient management decisions.

Laboratories implementing this test must have the appropriate biosafety equipment, personal protective equipment (PPE), containment facilities and personnel trained in the safe handling of diagnostic clinical specimens potentially containing B. anthracis. Anthrax is a nationally notifiable disease caused by a biothreat microbial agent and must be reported to public health authorities.

The distribution of in vitro diagnostic devices for Bacillus spp. detection is limited to laboratories that follow public health guidelines that address appropriate biosafety conditions, interpretation of test results, and coordination of findings with public health authorities.

The Applied Biosystems Bacillus anthracis Detection Kit is intended for use with the AB1 7500 Fast Dx Real-Time PCR Instrument with analysis using the Applied Biosystems Bacillus anthracis Interpretive Software (BaIS).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

Date of Summary:

September 27, 2019

Product Name

Applied Biosystems™ Bacillus anthracis Detection Kit

Sponsor

MRIGlobal 425 Volker Boulevard Kansas City, Missouri 64110

Correspondent

MDC Associates, Inc. Fran White, President 180 Cabot Street Beverly, MA 01915 Phone: (978) 705 5011 Fax: (866) 540 3448 Email: regulatory@mdcassoc.com

Common Name

Device to detect and identify biothreat microbial agents in human clinical specimens

Product Classification

866.4000

Classification

QIF, Class II

Substantial Equivalency

CharacteristicMRIGlobalApplied Biosystems™ Bacillus anthracisDetection Kit(New Device)BioFire Defense, Inc.FilmArray® NGDS Warrior PanelK170883(Predicate Device)
Similarities
Product CodeQIFPRD
Intended UseThe Applied Biosystems Bacillus anthracisDetection Kit is a real-time polymerasechain reaction (PCR) test kit intended forthe qualitative in vitro diagnostic (IVD)detection of target DNA sequences forBacillus anthracis ( B. anthracis , or BA). TheApplied Biosystems Bacillus anthracisDetection Kit is intended to test humanwhole blood (EDTA) specimens and bloodculture specimens with growth detected bya continuous monitoring blood culturesystem. Blood culture specimens must bedetermined to contain gram-positive bacilliThe FilmArray® NGDS Warrior Panel isa qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intendedfor use with the FilmArray® 2.0 system.The FilmArray® NGDS Warrior Paneldetects and identifies Bacillusanthracis, Yersinia pestis, Francisellatularensis, Coxiella burnetii, Ebolavirus, and Marburg virus nucleic acidsdirectly from human whole blood(EDTA).The FilmArray® NGDS Warrior Panel isalso intended to be used to test for
CharacteristicMRIGlobalApplied Biosystems™ Bacillus anthracisDetection Kit(New Device)BioFire Defense, Inc.FilmArray NGDS Warrior PanelK170883(Predicate Device)
by Gram stain prior to testing. Testing ofwhole blood specimens must be performedconcomitantly with standard of care bloodculture.Bacillus anthracis or Yersinia pestisnucleic acids in blood cultures that aredetermined to be positive either by anautomated system, by turbidity, or bydaily Gram stain. In addition, theFilmArray® NGDS Warrior Panel mayalso be used to detect and identifyYersinia pestis and Francisellatularensis nucleic acids directly fromsputum specimens.
The Applied Biosystems Bacillus anthracisDetection Kit is indicated for use in CLIA-certified high-complexity laboratories inresponse to a confirmed Bacillus anthracisevent only in accordance with theguidelines provided by public healthauthorities prior to or during a public healthemergency. Testing with the AppliedBiosystems Bacillus anthracis Detection Kitmust only be performed when public healthauthorities have determined the need forthis test. The test must only be used withspecimens from individuals with clinicalsigns and symptoms of B. anthracisinfection and who have either beenexposed to B. anthracis or may have beenexposed to B anthracis.The Applied Biosystems Bacillus anthracisDetection Kit is intended for use as an aid inthe diagnosis of anthrax infection andresults are for the presumptiveidentification of Bacillus anthracis. Thediagnosis of B. anthracis infection must bemade based on history, signs, symptoms,exposure likelihood, and other laboratoryevidence, in addition to the identification ofB. anthracis from cultures or directly fromclinical specimens. The definitiveidentification of B. anthracis requiresadditional testing and confirmationThe FilmArray® NGDS Warrior Panel isintended to test individuals with signsand symptoms of infection frombiothreat agents and/or individualswho are at risk for exposure or mayhave been exposed to these agents.The FilmArray® NGDS Warrior Panel isindicated as an aid in the diagnosis ofanthrax, plague, tularemia, Q fever,and the hemorrhagic fevers caused byEbola and Marburg viruses, inresponse to a suspected or confirmedbioterrorism event or outbreaks. It isfor diagnostic use in conjunction withother clinical, epidemiologic, andlaboratory data, in accordance withthe guidelines provided by theappropriate Department of Defenseand public health authorities.Results are for the presumptiveidentification of Bacillus anthracis,Yersinia pestis, Francisella tularensis,Coxiella burnetii, Ebola virus, andMarburg virus. The definitiveidentification of Bacillus anthracis,Yersinia pestis, Francisella tularensis,
procedures in consultation with theappropriate public health authorities forwhom reports may be required.Coxiella burnetii, Ebola virus, andMarburg virus requires additionaltesting and confirmation procedures inconsultation with the appropriate
The Applied Biosystems Bacillus anthracisDetection Kit has not been clinicallyevaluated with specimens collected fromindividuals with B. anthracis infection orthose presumed to be exposed to B.anthracis. 'Bacillus anthracis not detected'results do not preclude infection withBacillus anthracis and should not be used asthe sole basis for diagnosis, treatment, orother patient management decisions.Department of Defense and publichealth authorities for whom reportsmay be necessary.
CharacteristicMRIGlobalApplied Biosystems™ Bacillus anthracisDetection Kit(New Device)BioFire Defense, Inc.FilmArray NGDS Warrior PanelK170883(Predicate Device)
Methodology/TechnologicalPrincipleReal-time-PCRNested PCR with melt curve analysis
Specimen typesWhole Blood or Blood CultureWhole Blood; Blood Culture (forBacillus anthracis and Yersinia pestis),Sputum (for Yersinia pestis andFrancisella tularensis)
Assay TargetsDNA sequences unique to B. anthracisNucleic acid sequences unique toBacillus anthracis, Yersinia pestis,Francisella tularensis, Coxiellaburnetii, Ebola virus, and Marburgvirus
Result OutputQualitative detection of B. anthracis DNAQualitative detection of Bacillusanthracis, Yersinia pestis, Francisellatularensis, Coxiella burnetii, Ebolavirus, and Marburg virus nucleic acids
Test InterpretationAutomated test, interpretation, and reportgenerationAutomated test, interpretation, andreport generation
Differences
MultiplexCapabilityMultiplexed assay. Two gene targets andinternal process control run in singlereaction.Multiplexed assay for detection ofBacillus anthracis, Yersinia pestis,Francisella tularensis, Coxiella
CharacteristicMRIGlobalApplied Biosystems™ Bacillus anthracisDetection Kit(New Device)BioFire Defense, Inc.FilmArray NGDS Warrior PanelK170883(Predicate Device)
Sample ExtractionQiagen QIAamp™ DNA Mini Blood Kit(manual) and Roche MagNA PureAutomated Nucleic acid Extraction(automated)Automated in FilmArray pouch
PCR ReagentsLyophilized reagents are formulated as acomplete Mastermix and are reconstitutedupon sample addition with no requirementto add water or buffer.Reagents contained within FilmArraypouch
InstrumentationApplied BioSystems® 7500 Fast Dx[K082562]FilmArray 2.0
Throughput96-well format allows for simultaneousanalysis of up to 93 patient samples (+controls) for detection of two Bacillusanthracis targets and an internal processcontrol on a single plateOne sample at a time.
User ComplexityEnd UserHigh ComplexityTargeted CLIA-certified, sentinellaboratories (commercial diagnosticlaboratories, hospital-based laboratoriesand clinical institutions) in the event of apublic health emergencyLow/Moderate ComplexityThe FilmArray® NGDS Warrior Panel issolely for use by United StatesDepartment of Defense laboratories,and laboratories designated by theDepartment of Defense.

{5}------------------------------------------------

Page 2 of 12

{6}------------------------------------------------

{7}------------------------------------------------

Page 4 of 12

THIS SPACE INTENTIONALLY LEFT BLANK.

{8}------------------------------------------------

Page 5 of 12

Intended Use

The Applied Biosystems Bacillus anthracis Detection Kit is a real-time polymerase chain reaction (PCR) test kit intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences for Bacillus anthracis, or BA). The Applied Biosystems Bacillus anthracis Detection Kit is intended to test human whole blood (EDTA) specimens and blood culture specimens with growth detected by a continuous monitoring blood culture system. Blood culture specimens must be determined to contain gram-positive bacilli by Gram stain prior to testing. Testing of whole blood specimens must be performed concomitantly with standard of care blood culture.

The Applied Biosystems Bacillus anthracis Detection Kit is indicated for use in CLIA-certified highcomplexity laboratories in response to a confirmed Bacillus anthracis event only in accordance with the guidelines provided by public health authorities prior to or during a public health emergency. Testing with the Applied Biosystems Bacillus anthracis Detection Kit must only be performed when public health authorities have determined the need for this test. The test must only be used with specimens from individuals with clinical signs and symptoms of B. anthracis infection and who have either been exposed to B. anthracis or may have been exposed to B anthracis.

The Applied Biosystems Bacillus anthracis Detection Kit is intended for use as an aid in the diagnosis of anthrax infection and results are for the presumptive identification of Bacillus anthracis. The diagnosis of B. anthracis infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of B. anthracis from cultures or directly from clinical specimens. The definitive identification of B. anthracis requires additional testing and confirmation procedures in consultation with the appropriate public health authorities for whom reports may be required.

The Applied Biosystems Bacillus anthracis Detection Kit has not been clinically evaluated with specimens collected from individuals with B. anthracis infection or those presumed to be exposed to B. anthracis. 'Bacillus anthracis not detected' results do not preclude infection with Bacillus anthracis and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Laboratories implementing this test must have the appropriate biosafety equipment, personal protective equipment (PPE), containment facilities and personnel trained in the safe handling of diagnostic clinical specimens potentially containing B. anthrax is a nationally notifiable disease caused by a biothreat microbial agent and must be reported to public health authorities.

The distribution of in vitro diagnostic devices for Bacillus spp. detection is limited to laboratories that follow public health guidelines that address appropriate biosafety conditions, interpretation of test results, and coordination of findings with public health authorities.

{9}------------------------------------------------

Page 6 of 12

The Applied Biosystems Bacillus anthracis Detection Kit is intended for use with the ABI 7500 Fast Dx Real-Time PCR Instrument with analysis using the Applied Biosystems Bacillus anthracis Interpretive Software (BaIS).

Methodology

The Applied Biosystems™ Bacillus anthracis Detection Kit is a multiplexed real-time polymerase chain reaction (PCR) test kit intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences for B. anthracis. Reagents are lyophilized in a 96-well plate format as a fully formulated Mastermix and are stable at room temperature for up to one year. The kit is specifically designed for performing real-time PCR using the Applied Biosystems (ABI) 7500 Fast Dx instrument and software, with nucleic acids extracted from clinical specimens using a Qiagen manual extraction method or Roche MagNA Pure automated extraction methods. An automated interpretative software component (BalS) is included in the kit but supplied separately and operates on a computer(s) that is separate from the ABI 7500 Fast Dx computer.

THIS SPACE INTENTIONALLY LEFT BLANK.

{10}------------------------------------------------

Performance Data

Analytical Sensitivity

Limit of Detection (expressed as CFU/mL) was determined using Bacillus anthracis Ames, tested with twenty technical replicates per reagent lot of the Applied Biosystems™ Bacillus anthracis Detection Kit to achieve >95% detection when averaged across three reagent lots. Limit of Detection (LOD) was independently confirmed for whole blood, aerobic blood culture, and anaerobic blood culture for each extraction method, Qiagen DSP DNA Blood Mini Kit and Roche MagNA Pure using the Roche MagNA Pure LC 2.0 Robot. One thousand six hundred and twelve technical replicates were tested to determine LOD. See Table 1 below.

Table 1. Limit of Detection (expressed as CFU/mL) determined for Whole Blood, Blood Culture, Aerobic and Anaerobic per Extraction Method

Extraction MethodMatrixLimit of Detection(CFU/mL)
Qiagen DSP DNABlood Mini KitWhole Blood150
Blood Culture,Aerobic10,000
Blood Culture,Anaerobic10,000
Roche MagNA PureWhole Blood50
Blood Culture,Aerobic2270
Blood Culture,Anaerobic3040

Analytical Inclusivity

Quantified stocks of 24 different Bacillus anthracis strains representing geographic, temporal, genotypic, and phenotypic diversity of the species were prepared in culture under ATCCrecommended conditions and quantitated by enumeration on agar media to evaluate inclusivity using the Applied Biosystems™ Bacillus anthracis Detection Kit. Each BA inclusivity panel member was spiked into whole blood (K2 EDTA), at 3X the established LoD (expressed as CFU/mL) and processed using both extraction methods in triplicate – Roche MagNa Pure (MNP) extraction robot and Qiagen DSP DNA Blood Mini Kit per the kit instructions. The detection rate for both extraction methods was 100% for B. anthracis strains that carry both plasmid targets. For three strains that are known to carry only one plasmid target, the assay generated 'Bacillus anthracis suspected' results as expected.

{11}------------------------------------------------

Inclusivity in silico Analysis Summary

All available Bacillus anthracis genome assemblies were downloaded from NCBI and imported into the CLC Genomics Workbench (Qiagen) for a total of 144 panel members. A local BLAST within the software was then performed with imported assemblies against the amplicon sequence for each assay. A list of assemblies without amplicon hits was generated for each assay to further investigate assemblies not producing hits. The results indicated that assemblies that did not produce a hit contained only chromosome sequences or only the alternate plasmid. Several assemblies showed very low coverage of plasmids, likely due to absence of the plasmid but presence of conserved regions in the genome, which have homology to small regions of plasmid sequences.

Analytical Exclusivity

Specificity of the Applied Biosystems™ Bacillus anthracis Detection Kit was evaluated by testing a panel of 154 non-target organisms which included near-neighbors and other bacteria, virus and fungi that might present with similar clinical presentation to those patients suspected of anthrax. For instances where bacteria, virus or fungi were not available as isolates, purified nucleic acid was used. Some known Bacillus cereus strains (BAG1X1-1 NR28575, 03BB102, and G-9241 NR-9564) contain a pXO1-like plasmid. Detection of these strains was indicated by a test result of 'Bacillus anthracis suspected' with amplification of the plasmid manually confirmed by the supervisor. Initial no detection rate for exclusivity testing was 95.68%. Repeat testing was conducted for three organisms. Final no detection rate for exclusivity testing was 100%.

In addition, in silico exclusivity analysis was conducted with assay primers and probe positions aligned against additional non-target organisms. For in silico analysis, the results were filtered for: Exclusion of Bacillus anthracis hits; ≥80% for % Identity; and ≥80% for high-scoring segment pair (HSP) length (of the sequence for that primer or probe BLAST query).

Exclusivity in silico Analysis Summary

In silico exclusivity analysis was conducted with each of the assay primers and probes to identify non-target organisms with the potential to cross react with the assays. A BLAST search was performed against the NCBI nr, wgs, and genome databases. Organisms identified with high and reduced risk of amplification by in silico analysis were then evaluated by wet laboratory testing. Additionally, a specific set of organisms were screened by in silico analysis. A BLAST search of the primer and probe sequences was conducted. A primer and probe list was created and a BLAST search against each of the three databases performed for "all organisms" within the CLC Genomics Workbench software. For each primer or probe BLAST, the results were filtered for: (1) exclusion of Bacillus anthracis hits; (2) ≥80% for % identity; and (3)≥80% for HSP length (of the sequence for that primer or probe BLAST query). The results of each BLAST were transferred to a Microsoft Excel spreadsheet and reviewed to identify organisms with "hits" for all three primer and probe sequences for each assay.

{12}------------------------------------------------

Page 9 of 12

For the pXO1 assay, one organism (Bacillus cereus m1550) was identified as a match with 100% identity to the primer and probe sequences as high potential to support amplification and was evaluated by wet laboratory testing. For the pXO2 assay, two organisms (B. cereus AH1272 and AH1273) were identified as matches with 100% identity to the probe sequence. However, through a secondary NCBI Primer BLAST, each strain also contains three SNPs in both primer binding sequences, reducing the likelihood to support amplification.

Interfering Substances

The Applied Biosystems™ Bacillus anthracis Detection Kit's potential to detect Bacillus anthracis in the presence of potentially interfering substances was evaluated. Fifty potentially interfering substances consisting of endogenous, exogenous, and technique-specific substances were tested for potential interference with the assay. Each substance was tested in whole blood with triplicate-paired samples in the presence and absence of B. anthracis Ames, spiked at 3X LoD. Endogenous substances were spiked at concentrations typically found in human whole blood or at the highest achievable concentration.

Technique-specific substances were spiked directly into PCR reactions at 2.5%, 5% and 10% v/v test concentrations.

The following technique-specific substances were found to interfere with the assay and appropriate limitation was added to the product labeling.

  • MagNA Pure Wash Buffer I MagNA Pure Lysis/Binding Buffer MagNA Pure Proteinase K 10% Bleach Fisherbrand™ DNase Displace QIAGEN Protease QIAGEN Buffer AL QIAGEN Buffer AW1 (with and without EtOH) EtOH at greater than 5%

Microbial Interference

Detection of Bacillus anthracis in the presence of other clinically-relevant organisms, e.g. pathogens or normal flora that may be present in whole blood or positive blood culture specimens, was tested with the Applied Biosystems™ Bacillus anthracis Detection Kit. The detection rate of B. anthracis was 100% in the presence of potentially interfering organisms in whole blood and blood culture.

Reproducibility

Seven panel members were tested twice a day by three teams of two different operators on five non-consecutive days using two Applied Biosystems™ 7500 Fast Dx Real-Time PCR Instruments across three different reagent lots to determine the reproducibility rate of detecting Bacillus

{13}------------------------------------------------

anthracis Ames with the Applied Biosystems™ Bacillus anthracis Detection Kit. Two teams performed extraction using the Qiagen QlAamp DSP DNA Blood Mini Kit and one team performed extraction using the Roche MagNA Pure LC. A breakdown of the concentrations tested is below:

HP = High Positive: organism spiked at 10 × LOD; 1.5e3 CFU/mL for DSP and 5e2 CFU/mL for MNP

LP = Low Positive: organism spiked at 3 × LOD; 4.5e2 CFU/mL for DSP and 1.5e2 CFU/mL for MNP

NEG = Negative: non-Ba organism spiked at 3 x Ba LOD; 4.5e2 CFU/mL for DSP and 1.5e2 CFU/mL for MNP

One low positive sample extracted by one operator returned negative results when tested against three reagent lots indicating an error during the sample preparation procedure. In a separate instance a high positive sample returned 'Bacillus anthracis suspected' results when the pXO1 assay failed to amplify. The PCR for this sample was repeated and returned 'Bocillus anthracis detected' results as expected. No false positive events occurred out of the 270 PCR tests of negative samples.

There were a total of 24 sample replicates (across all three reagent lots) that returned initial indeterminate results. All were determined to be negative by supervisor review. There were no invalid results for this study.

Carry-Over/Cross-contamination

The Applied Biosystems™ Bacillus anthracis Detection Kit workflow was evaluated for the potential for cross-contamination or sample carry-over during extraction using the automated system (Roche MagNA Pure) and the manual DSP (Qiagen) methods. For the automated method nine full MagNA Pure runs were performed with each run of 31 samples placed in a checkerboard fashion alternating high positive, spiked whole blood (K2 EDTA), blood culture aerobic and blood culture anaerobic and negative samples throughout the sample cartridge. Blood culture spiked samples were spiked with Bacillus anthracis Ames at 1 x 10° CFU/mL. Whole blood spiked samples were spiked with Bacillus anthracis Ames at 500 CFU/mL (10X LOD). For the manual method 12 batches of 15-24 samples each were performed using the DSP method with samples batched such that high positive samples were placed next to negative whole blood, aerobic blood culture and anaerobic blood culture. Blood culture spiked samples were spiked with Bacillus anthracis Ames at 1 x 10° CFU/mL. Whole blood spiked samples were spiked with Bacillus anthracis Ames at 1500 CFU/mL (10X LOD).

Two hundred and seventy-nine samples were extracted on the MagNA Pure in nine separate runs and resulting extracts were analyzed in singlet PCR reactions. A total of two false positive events occurred out of a total of 135 negative samples analyzed, resulting in a specificity rate of 98.5%.

Two hundred and seventy-nine samples were manually extracted using the Qiagen DSP blood mini kit in twelve separate extraction batches and resulting extracts were analyzed in singlet PCR

{14}------------------------------------------------

Page 11 of 12

reactions. No false positive events occurred out of a total of 139 negative samples analyzed, resulting in a specificity rate of 100%.

Test area surface monitoring control swabs were collected at various steps in the workflow during execution of the Carryover/Cross-Contamination study to support a detailed evaluation of sample handling and processing steps that have the potential to introduce contamination.

One hundred and sixty-two test area surface monitoring control swabs were collected during nine separate MagNA Pure runs for the Carryover/Cross-Contamination study. Swabs were processed and analyzed in singlet PCR. Results were used to determine areas of the instrument and steps in the process that are at high risk of cross-contamination. A total of nine swab locations were tested to include three swab locations in the BSC and six surfaces inside the MagNA Pure work deck. Contamination events occurred on one BSC surface location and five surfaces of the MagNA Pure during testing.

Two hundred and sixteen test area surface monitoring control swabs were collected during twelve separate DSP extraction batches for the Carryover/Cross-Contamination study. Swabs were processed and analyzed in singlet PCR. Results were used to determine areas of the BSC and steps in the process that are at high risk of cross-contamination. One location (BSC - left surface - after loading DSP columns) was identified as having a higher rate of contamination during the workflow.

Clinical Specificity

To demonstrate clinical specificity, the Applied Biosystems™ Bacillus anthracis Detection Test System was tested using left-over fresh and frozen blood culture samples, (confirmed by Gram stain to be positive for any bacterial species and selectively positive for Gram Positive rods), and blood culture samples spiked with any bacterial species and a subset spiked with Gram positive rod bacteria. Randomly accessed, residual blood - blood collected in K2EDTA (e.g., CBC, tested in hematology) and with no correlation to positive blood culture specimens, for which standard of care testing was performed (patients may or may not have fever, but specimens were obtained for other laboratory testing) was collected from Hematology laboratories for immediate testing with the Applied Biosystems™ Bacillus anthracis Detection Kit. All samples (blood culture and whole blood) were collected prospectively and serially without further selection criteria. Additionally, febrile whole blood samples were collected from consented patients presenting with fever and flu-like symptoms. Samples were collected from three (3) point-of-care collection sites including physician offices, urgent care centers or hospital clinics/ER. Clinical performance testing was conducted at three laboratories within the US. Febrile whole blood samples were shipped to one laboratory on the day of collection and were processed within 24 hours of receipt.

No reference method testing was performed.

All samples were assumed to be negative for Bacillus anthracis.

{15}------------------------------------------------

The study generated 'Bacillus anthracis not detected' results for the 401 blood culture specimens and 439 whole blood specimens included in the performance analyses, demonstrating 100% negative percent agreement when compared to the expected negative result.

Clinical Sensitivity

To demonstrate product sensitivity, the Applied Biosystems™ Bacillus anthracis Detection Test System was tested using whole blood specimens collected from febrile patients presenting with fever of unknown origin and flu-like symptoms. Following testing of febrile whole blood specimens determined to be negative for Bacillus anthracis, simulated samples positive for Bacillus anthracis were contrived using unique febrile whole blood specimens each spiked with Bacillus anthracis at or near LOD. Simulated positive samples were prepared and tested in BSL-3 laboratory facilities. Each febrile whole blood aliquot was spiked with 1 of 18 strains of Bacillus anthracis. Contrived samples were randomized with negative, unspiked whole blood samples and blinded to the operators. Two different operators processed samples using either the Qiagen DSP manual method or with the Roche MagNA Pure automated robot. Two different instruments were used to analyze samples. A minimum of 25 samples were spiked with Bacillus anthracis Ames at LOD and additional samples were spiked at 3× LOD with 1 of 17 other Bacillus anthracis strains.

The study generated 'Bacillus anthracis detected' results for 84 of 87 low-positive whole blood specimens. Three specimens initially generated 'Bacillus anthracis suspected' results due to detection only one plasmid target. Final positive percent agreement was 96% (Cl: 90.4-96,5%).

N/A