K Number
K183400

Validate with FDA (Live)

Date Cleared
2019-03-22

(105 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SmartClip™ Soft Tissue Marker is intended to be placed percutaneously in soft tissue to permanently mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (EnVisio™ Navigation System) the SmartClip™ is located and surgically removed with the target tissue. The EnVisio™ Navigation System is intended only for the non-imaging detection and localization (by navigation) of the SmartClip™ that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.

Device Description

The EnVisio™ Navigation System is a medical device that detects the presence of the SmartClip™ Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery. The EnVisio™ Navigation System acquires and analyzes electromagnetic signatures that display the relative distance, depth, and direction from the tip of the electrosurgical tool to the implanted SmartClips™ within the surgeon's field of view during the procedure. The EnVisio™ Navigation System will be utilized in an operating room where surgical excision of soft tissue is performed. The EnVisio™ system can detect and provide real-time three-dimensional navigation for up to three separate SmartClips™ each programmed to emit a unique electromagnetic signature. The EnVisio™ Navigation System equipment components are the Console, Heads Up Display, Patient Pad and Foot Pedal. The EnVisio™ Navigation System Navigator and Calibration Disk are sterile, non-patient contacting single use devices which are available separately.

AI/ML Overview

The document provided does not contain a detailed study that proves the device meets specific acceptance criteria with quantifiable metrics such as accuracy, sensitivity, or specificity. Instead, it states that "Performance testing was conducted to evaluate and characterize the performance of the EnVisio Navigation System" and that "The EnVisio Navigation System met all specified criteria and did not raise new safety or performance questions."

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, data provenance, expert involvement, adjudication methods, MRMC studies, or standalone performance.

Based on the information available in the document:

1. A table of acceptance criteria and the reported device performance
Not explicitly provided with quantifiable metrics in the document. The document states that the system "met all specified criteria."

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
Not provided in the document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not provided in the document.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable, as this device (EnVisio Navigation System) is described as an "aid" for non-imaging detection and localization. It is not an AI-assisted diagnostic tool for "human readers." Therefore, no MRMC study comparing human readers with and without AI assistance is mentioned or relevant in the provided text.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
The device is a "Navigation System" intended to aid in detection and localization. The description "non-imaging detection and localization (by navigation)" implies human-in-the-loop operation, as it guides a surgeon. There is no mention of a standalone algorithm-only performance study.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not provided in the document.

8. The sample size for the training set
Not provided in the document.

9. How the ground truth for the training set was established
Not provided in the document.

The document focuses on establishing substantial equivalence to a predicate device (Cianna Medical SAVI Scout Reflector and SAVI Scout System K181007) based on indications for use, intended use, technological characteristics, and principles of operation, rather than detailing a specific clinical performance study with statistical endpoints and acceptance criteria for a novel AI or diagnostic system. The performance testing mentioned ("System Design Verification," "Navigation Performance Verification," "Navigation Performance Validation," "EMC/Safety Testing," "Packaging/Shelf Life," "Sterilization Validation") are engineering and regulatory compliance tests, not typically clinical efficacy studies that would report accuracy metrics against a ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 22, 2019

Elucent Medical, Inc Ms. Debra Kridner Vice President of Regulatory and Quality 7480 Flying Cloud Drive, Suite 110 Eden Prairie, Minnesota 55344

Re: K183400

Trade/Device Name: EnVisio Navigation System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: February 20, 2019 Received: February 22, 2019

Dear Ms. Kridner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by David Krause David Krause -S Date: 2019.03.22 14:41:33 -04'00' for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183400

Device Name EnVisio™ Navigation System

Indications for Use (Describe)

The SmartClip Soft Tissue Marker is intended to be placed percutaneously in soft tissue to permanently mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (EnVisio Navigation System); the SmartClip is located and surgically removed with the target tissue. The EnVisio Navigation System is intended only for the non-imaging detection and localization (by navigation) of the SmartClip that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared:March 20, 2019
1. APPLICANTElucent Medical, Inc.7480 Flying Cloud Drive, Suite 110Eden Prairie, MN 55344 USA
Official Correspondent:Debra KridnerTel: (844) 417 1700 / Fax: (952) 314-7105debra.kridner@elucent.com
2. DEVICE
Trade Name:EnVisio™ Navigation System
510(k)K183400
Common Name:Implanted, Radiographic, Tissue Marker
Classification Number:21 CFR 878.4300
Classification Name:Implantable Clip
Device Class:Class II
Product Code:NEU
Review Panel:General and Plastic Surgery

3. PREDICATE DEVICE (PRIMARY)

Trade Name:Cianna Medical SAVI Scout Reflector and SAVI Scout System
510(k)K181007
Common Name:Implanted, Radiographic, Tissue Marker
Classification Number:21 CFR 878.4300
Classification Name:Implantable Clip
Device Class:Class II
Product Code:NEU
Review Panel:General and Plastic Surgery

4. DEVICE DESCRIPTION (for the device subject to this 510(k) notification)

The EnVisio™ Navigation System is a medical device that detects the presence of the SmartClip™ Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery. The EnVisio™ Navigation System acquires and analyzes electromagnetic signatures that display the relative distance, depth, and direction from the tip of the electrosurgical tool to the implanted SmartClips™ within the surgeon's field of view

{4}------------------------------------------------

during the procedure. The EnVisio™ Navigation System will be utilized in an operating room where surgical excision of soft tissue is performed. The EnVisio™ system can detect and provide real-time three-dimensional navigation for up to three separate SmartClips™ each programmed to emit a unique electromagnetic signature.

The EnVisio™ Navigation System equipment components are the Console, Heads Up Display, Patient Pad and Foot Pedal. The EnVisio™ Navigation System Navigator and Calibration Disk are sterile, non-patient contacting single use devices which are available separately.

5. INDICATIONS FOR USE (for the Device Subject to this 510(k) Notification)

The SmartClip™ Soft Tissue Marker is intended to be placed percutaneously in soft tissue to permanently mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (EnVisio™ Navigation System) the SmartClip™ is located and surgically removed with the target tissue. The EnVisio™ Navigation System is intended only for the non-imaging detection and localization (by navigation) of the SmartClip™ that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.

Comparison ofSubject Deviceto PredicateDeviceSubject DeviceEnVisio™ Navigation System(K183400)Predicate Device:Cianna Medical SAVI Scout Reflectorand SAVI Scout System (K181007)
Indications forUseThe SmartClip™ Soft Tissue Marker isintended to be placed percutaneously insoft tissue to permanently mark a biopsysite or a soft tissue site intended forsurgical removal. Using imaging guidance(such as ultrasound, MRI, or radiography)or aided by non-imaging guidance(EnVisio™ Navigation System) theSmartClip™ is located and surgicallyremoved with the target tissue. TheEnVisio™ Navigation System is intendedonly for the non-imaging detection andlocalization (by navigation) of theSmartClip™ that has been implanted in asoft tissue biopsy site or a soft tissue siteintended for surgical removal.The SAVI Scout Reflector is intended to beplaced percutaneously in soft tissue to mark(>30 days) a biopsy site or a soft tissue siteintended for surgical removal. Usingimaging guidance (such as ultrasound, MRI,or radiography) or aided by non-imagingguidance (SAVI Scout System) the SAVIScout Reflector is located and surgicallyremoved with the target tissue. The SAVIScout System is intended only for the non-imaging detection and localization of theSAVI Scout Reflector that has beenimplanted in a soft tissue biopsy site or a softtissue site intended for surgical removal.
Intended UseDetect and Navigate to the implantedSmartClip - Soft Tissue MarkerDetect and Navigate to the implantedReflector - Soft Tissue Marker
UseEnvironmentIntra-operative, guided surgical procedureIntra-operative, guided surgical procedure
Anatomical SiteSoft TissueSoft Tissue
PatientPopulationAdultAdult
Comparison of SubjectDevice to PredicateDeviceSubject DeviceEnVisio™ Navigation System(K183400)Predicate Device:Cianna Medical SAVI ScoutReflector and SAVI Scout System(K181007)
ManufacturerElucent Medical, Inc.Cianna Medical, Inc.
Device NameEnVisio™ Navigation SystemSAVI Scout Reflector and Scout System
510(k) NumberK183400K181007
Principles of OperationElectromagnetic wave technology todetect and navigate to the soft tissuemarkerElectromagnetic wave technology todetect and navigate to the soft tissuemarker
IntraoperativeLocalization EnergyElectromagnetic, passive markerElectromagnetic, passive marker
Audible LocationIndicationYesYes
Visual LocationIndicationYesYes
VisualizationCompatibilityUsing imaging guidance (such asultrasound, MRI, or radiography) oraided by non-imaging guidance(EnVisio Navigation System)Using imaging guidance (such asultrasound, MRI, or radiography) oraided by non-imaging guidance (SAVIScout System)
System Components(Equipment)Console• Heads-Up Display• Patient Pad• Foot Pedal (optional)• Cart• Console
System Components(Sterile)Navigator• Calibration Disk• Hand piece• Sterile Sheath applied tononsterile Handpiece
Sterilization MethodEthylene oxideEthylene oxide
Implanted Soft TissueMarkerYes - SmartClipYes - Reflector
BiocompatibilityNot applicable for Navigator andCalibration Disk - no direct orindirect patient contactYes -Handpiece has direct patientcontact

Indications for Use Comparison:

{5}------------------------------------------------

6. TECHNOLOGICAL AND PERFORMANCE CHARACTERISTICS COMPARISON

Performance Testing:

Performance testing was conducted to evaluate and characterize the performance of the EnVisio Navigation System. The following performance testing was conducted on the EnVisio Navigation System to support a determination of substantial equivalence to the predicate device.

  • System Design Verification
  • Navigation Performance Verification ●
  • Navigation Performance Validation ●

{6}------------------------------------------------

  • EMC/Safety Testing ●
  • Packaging/Shelf Life ●
  • . Sterilization Validation

Conclusion:

The EnVisio Navigation System met all specified criteria and did not raise new safety or performance questions.

Basis for Determination of Substantial Equivalence:

The Indications for Use, Intended Use, Fundamental Scientific Technology and Principles of Operation for the EnVisio Navigation System are the same as those described for the predicate device.

In summary, the EnVisio Navigation System has the following similarities to the predicate device which has previously received 510(k) clearance:

  • Has substantially equivalent indications for use ●
  • Has substantially equivalent intended use ●
  • Used in substantially equivalent anatomical sites ●
  • Uses substantially equivalent technological characteristics ●
  • Uses substantially equivalent principles of operation
  • Uses substantially equivalent sterilization methodology ●

Therefore, the conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device predicate. The EnVisio Navigation System is substantially equivalent to the predicate device.

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.