K Number
K183319

Validate with FDA (Live)

Device Name
Triangle System
Manufacturer
Date Cleared
2019-05-10

(161 days)

Product Code
Regulation Number
870.2920
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Triangle System is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, the Triangle System can record Lead-I. In two channel mode, the Triangle System can record Lead-I and Lead-II simultaneously and derive Lead-II and unipolar limb leads aVR, aVF and aVL. The Triangle System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAl platform including detecting the presence of normal sinus rhythm, atral fibrillation, bradycardia, and others. The Triangle System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.

Device Description

The Triangle System is a trans-telephonic (transmission by smartphone) electrocardiogram (ECG) event recorder that records, stores, transfers, and analyzes single-channel or two-channel ECG rhythm recordings. The Triangle System provides output of one or six ECG leads by recording Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF, and aVL for identifying cardiac arrhythmias. The device utilizes the computing power of iOS-based or Android-based smartphones (referred to as "Mobile Computing Platforms" (MCP) within this submission) to obtain and analyze ECG signals. The device consists of the Triangle Hardware (portable small wireless hardware with electrodes) and the Kardia app, which is installed on an MCP (i.e., iPhone or Android phone). The Triangle System is a new addition to AliveCor's ambulatory ECG devices that are intended to be used by lay users to record their ECG and obtain ECG analysis of Atrial Fibrillation. Normal Sinus Rhythm, Bradycardia, Tachycardia, or unclassified file. The Triangle System hardware transmits the ECG signal from the electrodes to the Kardia phone app on the MCP to be analyzed and presented to the user. All ECGs are synced with the user's account. The device is available for Over-the-Counter (OTC) purchase and for purchase with clinician prescription.

AI/ML Overview

Here's an analysis of the provided text, outlining the acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a numerical or target format for the qualitative and quantitative analyses. Instead, it describes the results of the assessments that demonstrate equivalence.

Acceptance Criteria (Implied/Derived from Results)Reported Device Performance (Triangle System)
Qualitative Equivalence for Cardiac Arrhythmia Assessment: Diagnostic equivalence with a gold standard 12-lead ECG device.All paired recordings (100%, n=44 subjects) were deemed equivalent for assessing cardiac arrhythmias by both electrophysiologists when comparing 6-lead ECG rhythm strips from the Triangle device to corresponding leads from the reference standard 12-lead ECG device.
Quantitative Equivalence (Lead I & II): High beat cross-correlation with a gold standard 12-lead ECG device.Minimum beat cross-correlation of 0.96 for Lead I and Lead II compared to the corresponding lead of the 12-lead ECG.
Quantitative Equivalence (All 6 Limb Leads): Low RMS error with a gold standard 12-lead ECG device.Maximum RMS error of 47 uV for all 6 limb leads compared to the corresponding lead of the 12-lead ECG.
Safety: No adverse events during clinical study.No adverse events were observed during the clinical study.

2. Sample Size for the Test Set and Data Provenance:

  • Sample Size: 44 subjects
  • Data Provenance: The text states, "Overall, 44 subjects participated in the study, that included nearly equal numbers of healthy volunteers and arrhythmia patients." It does not explicitly state the country of origin or if the study was prospective or retrospective, but the clinical study setup (simultaneous ECG recordings) strongly suggests a prospective design. The subjects appear to be a mix of healthy individuals and those with arrhythmias.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • Number of Experts: Two
  • Qualifications of Experts: "Two board-certified electrophysiologists."

4. Adjudication Method for the Test Set:

  • Adjudication Method: The text states, "two board-certified electrophysiologists compared 6-lead ECG rhythm strips acquired from the Triangle device and the corresponding leads from the reference standard 12-lead ECG device for diagnostic equivalence." It then concludes, "All paired recordings (100%, n=44 subjects), were deemed equivalent for assessing cardiac arrhythmias by both electrophysiologists." This implies that both experts had to agree for a recording to be deemed equivalent. There is no mention of a tie-breaking third expert, suggesting unanimosity was required for the reported 100% equivalence.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Was an MRMC study done? No. The study described is a comparison of the device's output to a gold standard, not a comparison of human readers' performance with and without AI assistance. While the device utilizes the KardiaAI platform, this specific clinical study focused on the hardware's ability to record equivalent ECG signals, not the AI's impact on human interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Was a standalone study done? Yes, to a degree outlined in the "Validation of KardiaAI integration" under non-clinical testing. However, the specific clinical study described does involve human experts (electrophysiologists) for the qualitative assessment, indicating a human-in-the-loop component for validation of the output. The quantitative assessment is purely algorithmic comparison (cross-correlation and RMS error). The statement, "The Triangle System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others," suggests the AI provides its own analysis which would be a standalone capability, but the clinical study discussed here is primarily about the signal equivalence of the hardware. The KardiaAI itself would have had standalone performance studies, but those are not detailed here for this K183319 submission.

7. Type of Ground Truth Used:

  • Type of Ground Truth: The ground truth for the clinical study was established using a gold standard 12-lead ECG device (for both qualitative and quantitative comparison of ECG signals) and the expert consensus of two board-certified electrophysiologists (for qualitative diagnostic equivalence).

8. Sample Size for the Training Set:

  • The document does not provide information regarding the sample size for the training set. The clinical study described involves performance validation of the Triangle System hardware, not training of the KardiaAI algorithms.

9. How the Ground Truth for the Training Set Was Established:

  • The document does not provide information on how the ground truth for the training set of the KardiaAI platform was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

AliveCor, Inc. % Prabhu Raghavan Vice President of Regulatory Quality AliveCor 224 Airport Parkway, Suite 250 San Jose, California 95110

Re: K183319

Trade/Device Name: Triangle System Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter And Receiver Regulatory Class: Class II Product Code: DXH, DPS, QDA Dated: April 9, 2019 Received: April 10, 2019

Dear Prabhu Raghavan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Acting Assistant Director External Heart Rhythm and Rate Devices Team Division of Cardiac. Electrophysiology. Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183319

Device Name Triangle System

Indications for Use (Describe)

The Triangle System is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, the Triangle System can record Lead-I. In two channel mode, the Triangle System can record Lead-I and Lead-II simultaneously and derive Lead-II and unipolar limb leads aVR, aVF and aVL. The Triangle System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAl platform including detecting the presence of normal sinus rhythm, atral fibrillation, bradycardia, and others. The Triangle System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use. and and in the

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Notification K183319

GENERAL INFORMATION [807.92(a)(1)]

Applicant:

AliveCor, Inc. 444 Castro Street, Suite 600 Mountain View, CA 94041 Phone: 650-396-8553 Fax: 650-282-7932

Contact Person:

Prabhu Raghavan Vice President of Regulatory and Quality Phone: 650-396-8553 Email: RA@alivecor.com.

Date Prepared:

April 8, 2019

DEVICE INFORMATION [807.92(a)(2)]

Trade Name:

Triangle System

Generic/Common Name:

Telephone electrocardiograph transmitter and receiver

Classification:

21 CFR$870.2920, Transmitters and Receivers, Electrocardiograph, Telephone, Class II

Product Code:

DXH, DPS, QDA

{4}------------------------------------------------

PREDICATE DEVICE(S) [807.92(a)(3)]

K142672 - AliveCor Heart Monitor (currently marketed as "KardiaMobile")

K181823 - AliveCor KardiaAI

DEVICE DESCRIPTION [807.92(a)(4)]

The Triangle System is a trans-telephonic (transmission by smartphone) electrocardiogram (ECG) event recorder that records, stores, transfers, and analyzes single-channel or two-channel ECG rhythm recordings. The Triangle System provides output of one or six ECG leads by recording Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF, and aVL for identifying cardiac arrhythmias. The device utilizes the computing power of iOS-based or Android-based smartphones (referred to as "Mobile Computing Platforms" (MCP) within this submission) to obtain and analyze ECG signals. The device consists of the Triangle Hardware (portable small wireless hardware with electrodes) and the Kardia app, which is installed on an MCP (i.e., iPhone or Android phone). The Triangle System is a new addition to AliveCor's ambulatory ECG devices that are intended to be used by lay users to record their ECG and obtain ECG analysis of Atrial Fibrillation. Normal Sinus Rhythm, Bradycardia, Tachycardia, or unclassified file. The Triangle System hardware transmits the ECG signal from the electrodes to the Kardia phone app on the MCP to be analyzed and presented to the user. All ECGs are synced with the user's account. The device is available for Over-the-Counter (OTC) purchase and for purchase with clinician prescription.

INDICATIONS FOR USE [807.92(a)(5)]

The Triangle System is intended to record, store and transfer one- and two-channel electrocardiogram (ECG) rhythms. In single channel mode, the Triangle System can record Lead-I. In two channel mode, the Triangle System can record Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF and aVL. The Triangle System also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, tachycardia, and others. The Triangle System is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.

{5}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE[807.92(a)(6)]

FeatureTriangle System(subject device)AliveCor Heart MonitorK142672
Product CodeDXH, DPS QDANo difference
Mechanism of ActionUser completes circuit with skin contactand hardware transmits ECG signal toMCP to convert and display ECGwaveformNo difference
Where used (intended use)Mobile/active users at rest (ambulatory)No difference
Anatomical sitesLeft hand fingers to right hand fingersLeft hand fingers to right hand fingersand to left leg/kneeLeft hand fingers to right hand fingersRight hand fingers to left leg/knee
Available AlgorithmsAtrial FibrillationNoise AlgorithmNormal Sinus RhythmTachycardiaBradycardiaAtrial Fibrillation
Data Acquisition:Frequency Response0.5Hz - 40HzNo difference
No. of ECG electrodesThree (3) dry electrodesTwo (2) dry electrodes
No. of ECG channelsSingle-channel and two-channelSingle channel only
Resolution16 bitsNo difference
Sample Rate300 Samples/SecondNo difference
Power Supply:BatteryBattery Life (typical)1 Lithium Manganese Dioxide CoinCells100 hours operationalNo difference
LeadsLead-I, or Lead-I, Lead-II, Lead-III,aVL, aVR and aVF.Lead-I or Lead-II
Hardware interfaceThree-electrode sensorTwo-electrode sensor
Software interfaceApple iOS-based or GoogleAndroid-based softwareNo difference
CommunicationsBluetooth Low EnergyUltrasonic Acoustics acquired by phone
Physical Specs:Dimensions90 x 30 x 7mm82 x 32 x 3.5mm
Weight24 grams15 grams

SUBSTANTIAL EQUIVALENCE

The Triangle System subject device has the same intended use and similar technological characteristics as the AliveCor Heart Monitor predicate device. The differences in technological characteristics have been analyzed and addressed through performance testing. The testing results showed that differences between the subject and predicate device do not raise different questions of safety or effectiveness.

{6}------------------------------------------------

PERFORMANCE DATA [807.92(b)]

All necessary non-clinical and clinical performance testing was conducted on the Triangle System to support a determination of substantial equivalence to the predicate device.

[807.92(b)(1)] Nonclinical Testing Summary:

All necessary performance testing was conducted on the Triangle System with passing results supporting the determination of substantial equivalence to the predicate device. This testing included the following:

  • Verification of the device's specification -
  • Testing to software level of concern requirements -
  • Validation of KardiaAI integration -
  • Validation of BLE ECG transmission -
  • -Human factors evaluation
  • Evaluation to the following standards: -
    • ISO 10993-1:2009 (CORR 2010), Biological Evaluation of Medical O Devices - Part 1: Evaluation and Testing Within a Risk Management Process
    • IEC 60601-1:2012, Medical Electrical Equipment Part 1: General O Requirements For Basic Safety And Essential Performance,
    • IEC 60601-1-2:2014, Medical Electrical Equipment Part 1-2: General о Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests,
    • IEC 60601-1-11:2015, Medical electrical equipment -- Part 1-11: General O requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, and
    • IEC 60601-2-47:2012, Medical Electrical Equipment -- Part 2-47: O Particular Requirements For The Basic Safety And Essential Performance Of Ambulatory Electrocardiographic Systems.

[807.92(b)(2)] Clinical Study Summary:

The performance of the Triangle System for recording a 6-lead ECG was validated in a clinical study. Overall, 44 subjects participated in the study, that included nearly equal numbers of healthy volunteers and arrhythmia patients. ECG recordings were taken simultaneously with both the Triangle device and gold standard 12-lead ECG device for comparison. Qualitative and quantitative analyses of equivalence were performed on the 44 pairs of ECGs.

For qualitative assessment, two board-certified electrophysiologists compared 6-lead ECG rhythm strips acquired from the Triangle device and the corresponding leads from the reference standard 12-lead ECG device for diagnostic equivalence. All paired recordings (100%, n=44 subjects), were deemed equivalent for assessing cardiac arrhythmias by both electrophysiologists. The results of the assessment determined that

{7}------------------------------------------------

ALIVECOR, INC.

the subject device records a 6-lead ECG that is qualitatively equivalent to the recordings of corresponding leads from a commercial gold standard 12-lead ECG device.

For quantitative equivalence, median beat cross correlation for Lead I and II and RMS error for all 6 limb leads were computed between the paired ECGs for each subject. This analysis was conducted on the unfiltered ECG output as well as the enhanced filtered (EF) ECG output. Triangle ECGs had a minimum correlation of 0.96 and a maximum RMS error of 47 uV as compared to the corresponding lead of the 12-lead ECG. The results of the quantitative analysis of the ECG recordings further confirmed that the Triangle device ECG has equivalent output to that of the gold standard 12-lead ECG device.

During this clinical study, no adverse events were observed.

SUMMARY

The Triangle System has the same intended use and similar technological characteristics as the AliveCor Heart Monitor predicate device. The differences in technological characteristics have been analyzed and addressed through performance testing which included non-clinical and clinical testing. The testing results show that differences between the subject and predicate device do not raise different questions of safety or effectiveness. The Triangle System is substantially equivalent to the predicate device.

§ 870.2920 Telephone electrocardiograph transmitter and receiver.

(a)
Identification. A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.(b)
Classification. Class II (performance standards).