K Number
K183248

Validate with FDA (Live)

Date Cleared
2019-04-25

(155 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Synapse Enterprise Viewer (EV) is a software-based enterprise image viewer used with general purpose servers and dedicated reading workstations. It allows for communication, rendering on the server component and communication and display of DICOM and non-DICOM images as well as reports.

The device is intended for use as a diagnostic, review and analysis tool by trained professionals such as radiologists, physicians and technologist. Synapse EV is not intended for diagnostic use on mobile devices.

Synapse Enterprise Viewer is not to be used for primary mammography diagnosis.

Device Description

The Synapse Enterprise Viewer Software Version 1.0 (Hereafter referred to as Synapse EV), the subject of this Premarket Notification Submission, is intended for communication, storage, display, manipulation, measurement, printing, and processing of radiological data including images, reports, patient status and clinical information acquired from various medical imaging and information systems. This software device provides post processing of existing medical images; it does not control how the medical images are acquired.

Synapse EV is a multipurpose, enterprise wide application used for Radiologist interpretation, in-house clinical review, and physician image and information access. It is a web-based, zero-download viewer that uses Internet technology at the foundation of its design. This design allows for a user interface that takes advantage of familiar functions of modern browsers and can be used as a web application on a PC or Mobile device.

As a zero-download viewer application, Synapse is accessible by anyone connected to a Synapse web server via an intranet or the Internet over standard HTTP or a secure HTTPS (SSL) connection. Synapse users may be assigned different security levels to access patient and study folders. All tools are available to all users regardless of their physical location or health provider function.

Compatible with both Intranet and Internet network connections, an installed Synapse EV Software will typically be comprised of one or more Synapse EV Servers. As the client a standard browser is being used.

AI/ML Overview

The provided text is a 510(k) summary for the FUJIFILM Synapse Enterprise Viewer Version 1.0. It aims to demonstrate substantial equivalence to a predicate device rather than providing a study proving specific performance metrics against defined acceptance criteria for a diagnostic medical device per se.

The document indicates that no clinical studies were required to support the safety and effectiveness of the software. Therefore, many of the requested details regarding acceptance criteria and a study proving those criteria are not applicable in this context. The submission focuses on software validation and verification to ensure the device performs as intended and is safe for its stated use as an image viewer.

Here's an analysis based on the provided text, addressing the points where information is available and noting where it is not:

1. A table of acceptance criteria and the reported device performance

No explicit "acceptance criteria" and "reported device performance" in the typical sense for a diagnostic medical device (e.g., sensitivity, specificity, AUC) are provided. This submission is for an image viewer, not an AI diagnostic algorithm. The acceptance criteria would primarily relate to software functionality, display accuracy, and adherence to standards.

Acceptance Criteria (Implied from text)Reported Device Performance (Summary from text)
Compliance with IEC 62304 standardComplies with IEC 62304 "Medical device software - Software life cycle processes" standard.
Functionality (display, reformatting, communication, measurements, etc.)Software verification includes integration testing for all components. Selected simulated workflow validation and new feature validation performed. All tools (including image loading, viewing, measurement, annotation) are available and function.
User Guide validationPerformed by the test engineering team.
No defects impacting safety/effectivenessAny defects logged as unresolved anomalies do not impact device safety or effectiveness.
Substantial Equivalence to PredicateDetermined to be substantially equivalent to K161130 (ResolutionMD®).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: Not explicitly stated. The text mentions "integration testing for all components" and "selected simulated workflow validation." This suggests a comprehensive test suite covering various functionalities, but a specific number of cases or images is not provided.
  • Data Provenance: Not specified. Given it's a viewer for various radiological data, it would likely involve diverse image types and formats, but no origin or retrospective/prospective nature is detailed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable in the context of image viewer software validation. The text states:

  • "Selected simulated workflow validation and new feature validation was performed by product owners who have comprehensive understanding of user requirements as well as clinical expertise to validate the design and behavior of the implemented features and functionality."
  • "User Guide validation was performed by the test engineering team."

These individuals are verifying the software's functionality and adherence to requirements, not establishing a diagnostic ground truth for an AI algorithm.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. There is no mention of expert consensus or adjudication for diagnostic accuracy. The validation described is for software functionality.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was done or mentioned. The device is an enterprise image viewer, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a viewer, intended for human-in-the-loop review and analysis. It is not a standalone algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

Not applicable in the diagnostic sense. The "ground truth" for this software validation would be the functional specifications and requirements. The software is verified to correctly display, manipulate, and communicate medical images and data according to its design.

8. The sample size for the training set

Not applicable. The Synapse Enterprise Viewer is a traditional software application, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

April 25, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

FUJIFILM Corporation % Mr. Jeffrey Naveda Site Director, Regulatory and Quality Affairs FUJIFILM Medical Systems U.S.A., Inc. 3020 Carrington Mill Blvd. MORRISVILLE NC 27560

Re: K183248

Trade/Device Name: Synapse Enterprise Viewer Version 1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: April 15, 2019 Received: April 16, 2019

Dear Mr. Naveda:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm ); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael D. O'Hara

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183248

Device Name Synapse Enterprise Viewer Version 1.0

Indications for Use (Describe)

Synapse Enterprise Viewer (EV) is a software-based enterprise image viewer used with general purpose servers and dedicated reading workstations. It allows for communication, rendering on the server component and communication and display of DICOM and non-DICOM images as well as reports.

The device is intended for use as a diagnostic, review and analysis tool by trained professionals such as radiologists, physicians and technologist. Synapse EV is not intended for diagnostic use on mobile devices.

Synapse Enterprise Viewer is not to be used for primary mammography diagnosis.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5) 510(k) Summary

This summary of safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.

5a) Submitter

FUJIFILM Corporation 26-30, NISHIAZABU 2-CHOME MINATO-KU TOKYO 106-8620 JAPAN FDA Establishment Registration Number: 3005930550

Contact Person

Jeffrey Naveda Site Director, Regulatory and Quality Affairs FUJIFILM Medical Systems U.S.A., Inc. Telephone: (919) 948-1803 E-Mail: jnaveda@fujifilm.com Date Prepared: November 14, 2018

5b) Device

Trade Name:Synapse Enterprise Viewer Version 1.0
Common Name:Medical Information and Image Management Software
Regulatory Class:II
Classification Panel:Radiology
Regulation Name:Picture archiving and communications system
Regulation Number:892.2050
Product Code:LLZ
  • 5c) Predicate Device

K161130 - Calgary Scientific Inc. ResolutionMD®

No reference devices were used in this submission.

{4}------------------------------------------------

FUJIFILM Medical Systems, U.S.A., Inc. Synapse Enterprise Viewer v1.0 (Medical Information and Image Management Software) Section 5 - 510(k) Summary

5d) Device Description

The Synapse Enterprise Viewer Software Version 1.0 (Hereafter referred to as Synapse EV), the subject of this Premarket Notification Submission, is intended for communication, storage, display, manipulation, measurement, printing, and processing of radiological data including images, reports, patient status and clinical information acquired from various medical imaging and information systems. This software device provides post processing of existing medical images; it does not control how the medical images are acquired.

Synapse EV is a multipurpose, enterprise wide application used for Radiologist interpretation, in-house clinical review, and physician image and information access. It is a web-based, zero-download viewer that uses Internet technology at the foundation of its design. This design allows for a user interface that takes advantage of familiar functions of modern browsers and can be used as a web application on a PC or Mobile device.

As a zero-download viewer application, Synapse is accessible by anyone connected to a Synapse web server via an intranet or the Internet over standard HTTP or a secure HTTPS (SSL) connection. Synapse users may be assigned different security levels to access patient and study folders. All tools are available to all users regardless of their physical location or health provider function.

Compatible with both Intranet and Internet network connections, an installed Synapse EV Software will typically be comprised of one or more Synapse EV Servers. As the client a standard browser is being used.

5e) Indications for use

Synapse Enterprise Viewer (EV) is a software-based enterprise image viewer used with general purpose servers and dedicated reading workstations. It allows for communication, reformatting, rendering on the server component and communication and display of DICOM and non-DICOM images as well as reports.

The device is intended for use as a diagnostic, review and analysis tool by trained professionals such as radiologists, physicians and technologists. Synapse EV is not intended for diagnostic use on mobile devices.

{5}------------------------------------------------

Synapse Enterprise Viewer is not to be used for primary mammography diagnosis.

5f) Comparison of Technological Characteristics

Both Synapse Enterprise Viewer v1.0 and the predicate device, ResolutionMD® are enterprise viewers used with standard browsers for visualization of radiological images on web and mobile devices. Both are software-only devices for post processing of existing images and do not control how the medical images are acquired. Both Synapse Enterprise Viewer v1.0 and ResolutionMD® provide display, reformatting, and communication of DICOM and common non-DICOM file types.

Platforms for both Synapse Enterprise Viewer v1.0 and the predicate device, ResolutionMD® are general servers, web browsers on PC and specific mobile devices. Features which are the same for Synapse Enterprise Viewer v1.0 and the predicate device are that both run within a standard browser and do not require user installation, have data encryption and security in transit, use standard TCP/IP network protocol and support commonly used modalities. Both have the same search and results display features as well as viewer features such as image loading, viewing, measurement and the availability of the User Guide. Image scrolling, layouts, freehand annotation and measurement features are the same for Synapse Enterprise Viewer v1.0 and the predicate device.

The primary differences in features are that the predicate device offers collaboration and conferencing functionalities and allows multitenancy, while the current device does not. The predicate device offers more 2D viewer functionality as well as 3D viewer features, maximum intensity projection and multi-planar reformatting which the current device does not. There are additional minor differences in features with regard to involve reporting, image manipulation and printing functionality as well as differences in a few supported modalities, supported video formats, and search and results display features.

The subject of this 510(k) submission, Synapse Enterprise Viewer v1.0, did not require clinical studies to support safety and effectiveness of the software. Software development and testing for both the current and predicate devices is in compliance with the IEC 62304 "Medical device software - Software life cycle processes" standard. Software verification on the system level includes integration testing for all components. Selected simulated workflow validation and new feature validation was performed by product owners who have

{6}------------------------------------------------

FUJIFILM Medical Systems, U.S.A., Inc. Synapse Enterprise Viewer v1.0 (Medical Information and Image Management Software) Section 5 - 510(k) Summary

comprehensive understanding of user requirements as well as clinical expertise to validate the design and behavior of the implemented features and functionality. User Guide validation was performed by the test engineering team. Any defects loqged as unresolved anomalies do not impact device safety or effectiveness.

5g) Conclusions

The Synapse Enterprise Viewer v1.0 has the same general intended use, and similar indications for use, principles of operation, and technological characteristics as the predicate device ResolutionMD.® (K161130). Both are intended for use as a diagnostic, review and analysis tool by trained professionals such as radiologists, physicians and technologists and are software-only devices for post processing of existing images. The implemented design controls, risk management activities, labeling, and verification and validation testing demonstrate the safety and efficacy of the current device. Software development and testing for both the current and predicate devices complies with the IEC 62304 "Medical device software - Software life cycle processes" standard. Differences in features or functionality do not raise issues of safety or effectiveness. Thus, Synapse Enterprise Viewer Version 1.0 is substantially equivalent to the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).