K Number
K183220

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-06-21

(213 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OrthoScan TAU Mini C-arm is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatic populations when imaging limbs/extremittes, shoulders; at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.

Device Description

The OrthoScan TAU Mini C-Arm is a mobile fluoroscopic mini Carm system that provides fluoroscopic images of patients of all ages during diagnostic, treatment and surgical procedures involving anatomical regions such as but not limited to that of extremities, limbs, shoulders, knees, and Hips. The system consists of C-arm support attached to the image workstation. The proposed device provides the option of three CMOS flat panel detector sizes and identical X-ray source HVPS monoblock generator assembly with continuous or pulsed operation for image acquisition. The C-arm supports the CMOS FPD, X-Ray controls, collimator, high voltage generator with a fixed SID imaging. The C-arm and support arm which is connected to the mobile workstation platform are mechanically balanced allowing the operator precise positioning and locking of the vertical, horizontal, orbital and rotational movements at various angles and distances when imaging the patient's anatomical structures. The main workstation platform that supports the C-arm assembly contains the power control system, image processing system, system software, monitor display control and main user interface controls. The combination of C-Arm and workstation provides the clinician with a stable platform to obtain precise angles for localizing the patient's anatomical structures and visualization of pathology during live fluoroscopic imaging. The touch screen interface and keyboard provide user concise selectable imaging, X-ray technique control, entry of patient demographics and related procedural information. The workstation supports both an optional wired or wireless fluoroscopic footswitch allowing optimal positioning for the clinician. The optional connector interface panel of the OrthoScan TAU Mini C-Arm provides convenient connection of peripheral devices such as thermal video printers, image storage devices (USB) and DICOM fixed wire and wireless network interfaces.

AI/ML Overview

The provided text is a 510(k) Summary for the OrthoScan TAU Mini C-Arm, which is a premarket notification to the FDA to demonstrate that the new device is substantially equivalent to a legally marketed predicate device. This type of submission focuses on non-clinical testing to support the claim of substantial equivalence, rather than a full clinical study with specific acceptance criteria and performance metrics typically seen for novel devices or AI/software as a medical device (SaMD) where performance improvement is a key claim.

Therefore, many of the requested details about acceptance criteria, specific performance metrics, sample sizes for test/training sets, expert qualifications, and ground truth establishment, which are standard for AI/SaMD studies, are not explicitly provided in this document as it pertains to a traditional medical imaging device (C-arm) that primarily demonstrates substantial equivalence to existing technology.

However, I can extract the information that is present regarding device performance and the "study" conducted to support substantial equivalence.

Here's a breakdown of what can be inferred or directly stated from the document, and what is missing due to the nature of this 510(k) submission for an imaging device, not an AI algorithm:

Acceptance Criteria and Reported Device Performance

The document doesn't present a table of "acceptance criteria" in the traditional sense of specific numerical thresholds for diagnostic performance (e.g., sensitivity, specificity, AUC) that an AI algorithm would be tested against. Instead, it aims to demonstrate "substantial equivalence" to a predicate device. The performance is assessed through "image comparison" and "dose assessment" to show that the new device performs "as intended" and provides "similar image quality with new IDR filter" at "lower entrance dose level" compared to the predicate.

The table below summarizes the comparative technological characteristics which are used to argue substantial equivalence, and indirectly imply performance. The primary "performance" studied here is image quality and dose reduction, not diagnostic accuracy of an AI.

Table of Performance Comparison (Excerpted and Reinterpreted from the Provided Document)

CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (OrthoScan TAU Mini C-Arm)Comparison to Predicate, Comments to Differences (Why it's "Acceptable")
Image Quality / Detector
Detector Technologymedical grade GadOx (T1)/CMOS solid state X-ray detectormedical grade CsI(T1)/CMOS solid state X-ray detectorAll Detectors of the TAU Mini C-Arm are of similar design Technology and Scientific principal to that of the Predicate (K133174) They share the advantages of SSXI image receptors.
Detector Resolution1.5 k x 1.5 kTAU 2020 = 2.0k x 2.2 kTAU 1515 = 1.5 k x 1.5 kTAU 1512 = 2.0 k x 1.5 kSubstantially Equivalent. The proposed device has added the ability of a larger FOV for Physician. These changes do not raise new safety or effectiveness concerns.
Field of View (Full)5.5" x 5.5"TAU 2020 = 8" x 8"TAU 1515 = 5.5" x 5.5"TAU 1512 = 5.5" x 4.3"Substantially Equivalent. The proposed device has added the ability of a larger FOV for Physician. These changes do not raise new safety or effectiveness concerns.
Field of View (Collimated Mag)4.3" x 4.3"TAU 2020 = 4" x 4"TAU 1515 = 4.3" x 4.3"TAU 1512 = 4.3" x 3.3"Substantially Equivalent. The proposed device has added the ability of a larger Field of view for the Physician. These changes do not raise new safety or effectiveness concerns.
Detector Size15.0 x 15.0 (cm)TAU 2020 = 20 x 20 cmTAU 1515 = 15 x 15 cmTAU 1512 = 15 x 12 cmSubstantially Equivalent. The proposed device has added the ability of a larger Field of view for the Physician. The difference does not affect the safety or efficacy of the device.
DQE70% (implied for predicate, not explicitly stated as a value)TAU 2020 = 70%TAU 1515 = 70%TAU 1512 = 70%Identical. The difference does not affect the safety or efficacy of the device.
Grayscale Resolution16 bit (65,536 shades of gray)16 bit (65,536 shades of gray)Identical.
Dose Reduction
Pediatric Dose Reduction IDRNOYESIDR unique set of features and functions. Pediatric Dose reduction and special features for pediatric population. Dose assessment and image comparison of dose reduction for pediatric confirmed similar image quality with new IDR filter. The difference does not affect the safety or efficacy of the device. See Substantially Equivalent.
Adult Dose Reduction IDRNOYESIDR unique set of features and functions for Adult population. Dose assessment and image comparison of dose reduction confirmed similar image quality with new IDR filter. The difference does not affect the safety or efficacy of the device. Substantially Equivalent.
Beam Pre-filterPredicate has 2.5mm AL0.1mm Cu (Yes)Although not identical, New Device X-ray beam pre-filter helps reduce skin entrance dose by adding additional filtration of Cu.
CollimatorFixed Aperture @ Fixed SID (Normal, Mag)TAU 1512/1515 Fixed Aperture @ Fixed SID (Normal, Mag)TAU 2020 Stepless Collimator with Fixed SID (4 Leaf, 2 Axis)Although not Identical both have similar intended use of collimating X-ray beam providing compliance with the regulations. Substantially Equivalent.

Study Details (as inferable from the 510(k) Summary)

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Images: 330 individual images arranged in 20 groups of image sets.
    • Data Provenance: The study involved images taken from "anthropomorphic (PMMA material) phantoms and anatomical simulation phantoms." This means the data is synthetic/phantom-based, not from human patients. The country of origin is not specified, but given the FDA submission, it's presumed to be a controlled laboratory setting. The study is inherently non-clinical (not retrospective or prospective on human subjects).
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Number of Experts: 1 Radiologist.
    • Qualifications: "a board-certified Radiologist." No specific years of experience or sub-specialty are explicitly mentioned beyond board certification.
  3. Adjudication Method for the Test Set:

    • There is no mention of an adjudication method (like 2+1 or 3+1). The evaluation was "conducted by a board-certified Radiologist." This implies a single reader assessment for comparison.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. The document explicitly states: "OrthoScan TAU mobile fluoroscopic mini C-arm system did not require live human clinical studies to support substantial equivalence... Therefore, OrthoScan conducted a lab test image comparison study employing the use of anthropomorphic phantoms..." The study was a "lab test image comparison study" and involved a "Radiologist performed an assessment of 330 individual images." This is not an MRMC study.
    • Effect Size of Human Readers Improve with AI vs. without AI assistance: Not applicable, as this is a device clearance, not an AI algorithm. The device aims to provide better image quality at lower dose, which indirectly can improve human interpretation, but this was not quantified in an MRMC study.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable as this is a medical imaging device (C-arm), not an AI algorithm. The "performance" being evaluated is the direct image output of the device itself and its dose characteristics, not a diagnostic output from an automated algorithm.
  6. The Type of Ground Truth Used:

    • Phantom-based comparison with expert assessment. The "ground truth" for image quality and dose reduction in this context is established by the comparative assessment of images generated using standardized phantoms and evaluated by a qualified radiologist in conjunction with laboratory performance data (e.g., on dose). There's no "pathology" or "outcomes data" ground truth as this is a technical assessment of an imaging device.
  7. The Sample Size for the Training Set:

    • Not applicable. This is a medical device, not an AI/machine learning algorithm, so there is no "training set" in the computational sense. The device's design and software are developed through engineering and quality processes, not through autonomous learning from a dataset.
  8. How the Ground Truth for the Training Set was Established:

    • Not applicable due to the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Below the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

June 21, 2019

Orthoscan, Inc. % Mr. Steve D. Seeman Director of Quality Assurance and Regulatory Affairs 14555 N. 82nd St. SCOTTSDALE AZ 85260

Re: K183220

Trade/Device Name: OrthoScan TAU Mini C-Arm Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OXO, JAA and MQB Dated: June 3, 2019 Received: June 4, 2019

Dear Mr. Seeman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183220

Device Name OrthoScan TAU Mini C-Arm

Indications for Use (Describe)

The OrthoScan TAU Mini C-arm is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatic populations when imaging limbs/extremittes, shoulders; at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for OrthoScan Imaging Excellence. The logo features a blue square on the left with a white abstract design inside. To the right of the square is the word "ORTHOSCAN" in large, bold, black letters. Below the word "ORTHOSCAN" is the phrase "IMAGING EXCELLENCE" in smaller, blue letters.

510(k) Summary

510 (k) Premarket Notification Submission- OrthoScan TAU Mini C-Arm

In accordance with the requirements of 21 CFR §807.92 the following 510(k) summary of information is provided:

Submitter Address:OrthoScan14555 N 82nd St.Scottsdale, AZ 85260Phone: (480) 503-8010Fax: (480) 503-8011
Primary Contact Person:Richard Westrich14555 N 82nd St.Scottsdale, AZ 85260Cell: (909) 262-9930Phone: (480) 503-8010Fax: (480) 503-8011
Secondary Contact Person:Steve Seeman14555 N 82nd St.Scottsdale, AZ 85260Phone: (480) 503-8010Fax: (480) 503-8011

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for OrthoScan Imaging Excellence. The logo features a blue square on the left with a white abstract image inside. To the right of the square is the word "OrthoScan" in a bold, black font. Below "OrthoScan" is the phrase "Imaging Excellence" in a smaller, blue font.

Proposed Device:

Device Trade Name: 510(k) Number: Common Names:

Device Class: Regulation Number: Regulation Name: Product Code:

Primary Predicate Device:

Device Identification: Requlation Number: Device Class: Requlation Name: Product Codes:

Secondary Predicate Device:

Device Identification: Regulation Number: Device Class: Requlation Name: Product Codes:

General Description:

21CFR 807.92(a)(2)

OrthoScan TAU Mini C-Arm K183220 Fluoroscopic X-Ray System, Mobile Mini Mobile C-arm, Mini C-arm Class II 21CFR 892.1650 image-intensified fluoroscopic x-ray system OXO, JAA, MQB

21 CFR 807.92(a)(3)

OrthoScan FD Mini C-Arm (K133174) 21CFR 892.1650 Class II Image-intensified fluoroscopic x-ray system OXO, JAA, MQB

21 CFR 807.92(a)(3)

Ziehm Solo FD (K161976) 21CFR 892.1650 Class II Image-intensified fluoroscopic x-ray system OXO, JAA, MQB

The OrthoScan TAU Mini C-Arm is a mobile fluoroscopic mini Carm system that provides fluoroscopic images of patients of all ages during diagnostic, treatment and surgical procedures involving anatomical regions such as but not limited to that of extremities, limbs, shoulders, knees, and Hips. The system consists of C-arm support attached to the image workstation. The proposed device provides the option of three CMOS flat panel detector sizes and identical X-ray source HVPS monoblock generator assembly with continuous or pulsed operation for image acquisition.

The C-arm supports the CMOS FPD, X-Ray controls, collimator, high voltage generator with a fixed SID imaging. The C-arm and support arm which is connected to the mobile workstation platform are mechanically balanced allowing the operator precise positioning and locking of the vertical, horizontal, orbital and

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for OrthoScan Imaging Excellence. The logo features a blue square on the left with a white abstract design inside. To the right of the square is the word "ORTHOSCAN" in large, bold, black letters. Below the word "ORTHOSCAN" is the phrase "IMAGING EXCELLENCE" in smaller, blue letters.

rotational movements at various angles and distances when imaging the patient's anatomical structures.

The main workstation platform that supports the C-arm assembly contains the power control system, image processing system, system software, monitor display control and main user interface controls. The combination of C-Arm and workstation provides the clinician with a stable platform to obtain precise angles for localizing the patient's anatomical structures and visualization of pathology during live fluoroscopic imaging.

The touch screen interface and keyboard provide user concise selectable imaging, X-ray technique control, entry of patient demographics and related procedural information. The workstation supports both an optional wired or wireless fluoroscopic footswitch allowing optimal positioning for the clinician. The optional connector interface panel of the OrthoScan TAU Mini C-Arm provides convenient connection of peripheral devices such as thermal video printers, image storage devices (USB) and DICOM fixed wire and wireless network interfaces.

OrthoScan TAU Mini C-Arm is intended to provide fluoroscopic images of the patient including but not limited to, diagnostic, surgical, and critical emergency care procedures during diagnostic or therapeutic treatment/surgical procedures of the upper and lower extremities.

Indications for Use: The OrthoScan TAU Mini C-Arm is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders; at locations including but not limited to, hospitals, ambulatory surgery, eemergency, traumatology, orthopedic, critical care, or physician office environments.

Intended Use:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for OrthoScan. The logo consists of a blue square on the left with a white abstract design inside. To the right of the square is the word "OrthoScan" in a large, bold, serif font. Below "OrthoScan" is the phrase "IMAGING EXCELLENCE" in a smaller, sans-serif font.

Technology:The proposed modified device OrthoScan TAU Mini C-armemploys the same fundamental control, and scientific technologyas that of our predicate devices OrthoScan FD Mini C-arm(K133174) and the Ziehm Solo FD (K161976).
The radiation control, HVPS X-Ray monoblock generator, powersupplies as well as our advanced imaging system are identical tothe predicate OrthoScan FD Mini C-arm (K133174).
Software architecture design is identical to that of the predicatedevice OrthoScan FD Mini C-arm (K133174) with modification ofthe software to support pediatric indication of use, low dosefunctionality, processing applications related to the optionalrange of FPD sizes, Variable beam limiting device, and devicespecific features.
The primary modifications of the C-Arm include a larger butvirtually the same medical grade CsI(T1)/CMOS solid state X-raydetector as that of the predicate OrthoScan FD Mini C-Arm(K133174) and the Ziehm Solo FD (K161976). A variable beamlimiting device for precise collimating to anatomical structures,new pre-filter for pediatric imaging, touch control monitor,optional UPS power supply, incorporation of mechanical designimprovements in the C-Arm and mobile workstation balancing,locks, and maneuverability improving operator workflow duringextended procedures while keeping the essential smaller profileof our predicate OrthoScan FD Mini C-arm (K133174).
Summary of TechnologicalCharacteristics:The following table provides a comparison of the technologicalcharacteristics of the proposed device OrthoScan TAU Mini C-Armto that of the predicate device demonstrates that the scientificand technology characteristics are substantial equivalence to thepredicate device OrthoScan FD mini C-Arm (K133174).
DifferencesFeatures/Technology:New DeviceOrthoScan TAU Mini C-ArmTAU 1512 -1000-0015,TAU 1515- 1000-0016,TAU 2020- 1000-0017Predicate DeviceOrthoScan FD 1000-0004(K133174)Comparison to Predicate,Comments to Differences
Product Codes
Device Classification Nameimage-intensifiedfluoroscopic x-ray system,mobileimage-intensifiedfluoroscopic x-raysystem, mobileIdentical
Regulation DescriptionImage-intensifiedfluoroscopic x-ray system.Image-intensifiedfluoroscopic x-raysystem.Identical
Classification Product CodeOXOOXOIdentical
Subsequent Product CodeJAAJAAIdentical
Regulation Number892.165892.165Identical
Device ClassIIIIIdentical
Non-Contact DeviceNon-ContactNon-ContactIdentical
510(k) Panel ReviewRadiologyRadiologyIdentical
510(K) NumberOrthoScan TAU unknownat this timeK133174OrthoScan TAU unknownat this time
Detector Specifications
CMOS Flat PanelDetector/Image Receptormedical gradeCsI(T1)/CMOS solid stateX-ray detectormedical gradeGadOx (T1)/CMOS solidstate X-ray detectorAll Detectors of the TAUMini C-Arm are of similardesign Technology andScientific principal to thatof the Predicate(K133174) They share theadvantages of SSXI imagereceptors.
Detector ResolutionTAU 2020 = 2.0k x 2.2 k1.5 k x 1.5 kSubstantially Equivalent.The proposed device hasadded the ability of a
TAU 1515 = 1.5 k x 1.5 klarger FOV for Physician.These changes do notraise new safety or
TAU 1512 = 2.0 k x 1.5 keffectiveness concerns.
Field of View (Full)TAU 2020 = 8" x 8"5.5" × 5.5"Substantially Equivalent.The proposed device has
TAU 1515 = 5.5" x 5.5"added the ability of alarger FOV for Physician.These changes do not
TAU 1512 = 5.5" x 4.3"raise new safety oreffectiveness concerns.
Field of View (Collimated MagMode)TAU 2020 = 4" x 4"4.3" x 4.3"Substantially Equivalent.The proposed device has
TAU 1515 = 4.3" x 4.3"added the ability of alarger Field of view for thePhysician.
TAU 1512 = 4.3" x 3.3"These changes do notraise new safety oreffectiveness concerns.
Detector SizeTAU 2020 = 20 x 20 cm15.0 x 15.0 (cm)Substantially Equivalent.The proposed device hasadded the ability of a
TAU 1515 = 15 x 15 cmlarger Field of view for thePhysician.
TAU 1512 = 15 x 12 cmThe difference does notaffect the safety orefficacy of the device.
Useful ArrayTAU 2020 = 20 x 20 cm15.0 x 15.0 (cm)Substantially Equivalent.
TAU 1515 = 15 x 15 cmThe proposed device hasadded the ability of alarger Field of view for thePhysician.
TAU 1512 = 15 x 12 cmThe difference does notaffect the safety orefficacy of the device.
TAU 2020 = 99 micronsSubstantially Equivalent
Pixel SpacingTAU 1515 = 100 microns100 micronsThe difference does notaffect the safety or
TAU 1512 = 75 micronsefficacy of the device.
Dynamic RangeTAU 2020 = 71 dBSubstantially Equivalent
TAU 1515 = 71 dBThe difference does notaffect the safety or
TAU 1512 = 70 dBefficacy of the device.
DQETAU 2020 = 70%Identical
TAU 1515 = 70%The difference does notaffect the safety or
TAU 1512 = 70%efficacy of the device.
Grayscale Resolution16 bit (65,536 shades ofgray)16 bit (65,536 shades ofgray)Identical
Image Processing Features
Startup time30 sec30 secIdentical
Cine Loop ExportYesYesIdentical
Fluoroscopy Frame Rate30/15/7.5/2 fps30/15/7.5/2 fpsIdentical
Edge EnhancementYesYesIdentical
Post ProcessBrightness/ContrastYesYesIdentical
Adaptive Noise SuppressionAutomaticAutomaticIdentical
Manual Noise Suppression3 Modes3 ModesIdentical
AERC Automatic X-RayTechnique ControlYESYESIdentical
Adaptive Noise FilterNoise reductionNoise reductionIdentical
Save and Auto StoreYESYESIdentical
Last image holdYESYESIdentical
Edge EnhancementYESYESIdentical
Cine Loop Frame Rate30 fps30 fpsIdentical
Snapshot CapabilitiesYESYESIdentical
Post Processing (B/C)YESYESIdentical
Image invertYESYESIdentical
Image ZoomYESYESIdentical
Manual Noise Suppression4 modes4 modesIdentical
Image Documentation:
Wireless Communication (Wi-Fi)/(WLAN)Capable IEEE 802.11Capable IEEE 802.11Identical
DICOM 3 CompliantYesYesIdentical
MPPSCapableCapableIdentical
RDSRYESYESIdentical
Image Capacity26, 00026, 000Identical
Video Capacity14.4 min14.4 minIdentical
Cine Loop ExportYesYesIdentical
TP Link High Gain Wireless USBOptionOptionIdentical
USB Ports22Identical
Printer option22Identical
Dose Measurement
Air Kerma (US-Standard)YESYESIdentical
DAP (Optional in the US)OptionalOptionalIdentical
Pediatric Features
Pediatric Dose Reduction IDRYESNOIDR unique set of featuresand functions. PediatricDose reduction andspecial features forpediatric population. Doseassessment and imagecomparison of dosereduction for pediatricconfirmed similar imagequality with new IDR filterThe difference does notaffect the safety orefficacy of the device. SeeSubstantially Equivalent
Adult Dose Reduction IDRYESNOIDR unique set of featuresand functions for Adultpopulation. Doseassessment and imagecomparison of dosereduction confirmedsimilar image quality withnew IDR filterThe difference does notaffect the safety orefficacy of the device.Substantially Equivalent
Software ArchitectureOrthoMini ApplicationOrthoMini ApplicationSoftware architecture design is Substantially Equivalent to that of the predicate device OrthoScan FD Mini C-arm (K133174) with modification of the software to support pediatric indication of use, low dose functionality, processing applications related to the optional range of FPD sizes, Variable beam limiting device, and device specific features.
Graphical User Interface(GUI)OrthoMini ApplicationOrthoMini ApplicationSimilar operationhowever, with new device TAU the software is configured to support pediatric functionality not available be on OrthoScanFD.The difference does not affect the safety or efficacy of the device.Substantially Equivalent
Operating systemWindows 8.1 EmbeddedWindows 8.1 EmbeddedIdentical
Pediatric Workflow SupportYesNoNot available on predicateThe changes to the proposed device do not raise new safety or effectiveness concerns
MeasurementYesNoNot available on predicateThe changes to the proposed device do not raise new safety or effectiveness concerns
X-Ray Generator Specifications
Focal Spot42.5 microns42.5 micronsIdentical
kV Range40 - 78 kVp40 - 78 kVpIdentical
mA Range0.04 - 0.160 mA0.04 - 0.160 mAIdentical
Operating ModePulse/ ContinuousPulse/ ContinuousIdentical
Pulse Rate2 to 30 pps2 to 30 ppsIdentical
Beam Pre-filter 0.1mm CuYesNoPredicate has 2.5mm ALand complies with the regulations for HVL in x-ray beam.New Device X-ray beam pre-filter helps reduce
skin entrance dose byadding additional filtrationof Cu.
HVL Filter2.5mm(Al Equivalent)2.5mm(Al Equivalent)Identical
Magnification ModeYesYesBoth support mag modesof operation nearlyIdentical
CollimatorTAU 1512 Fixed Aperture@ Fixed SID(Normal, Mag)TAU 1515 Fixed Aperture@ Fixed SID(Normal, Mag)TAU 2020Stepless Collimator withFixed SID(4 Leaf, 2 Axis)Fixed Aperture @ FixedSID (Normal, Mag)Although not Identicalboth have similar indenteduse of collimating X-raybeam providingcompliance with theregulations.Substantially Equivalent
Physical Dimensions:
Source to Image17.7" (45cm)17.7" (45cm)Identical
Free space13.8"13.8"Identical
Arc Depth20"19"Substantially EquivalentProvides more depth forprocedures in clinicaluse. The change to theproposed device doesnot raise new safety oreffectiveness concerns
Pivot430°430°Identical
Lateral RotationB + Y axis (Wig-Wag)320°320°Identical
Orbital Rotation160°150°Substantially EquivalentThe change to theproposed device doesnot raise new safety oreffectiveness concerns
Vertical Range26.5"26.5"Identical
Distance to Cabinetmax 68"max 68"Identical
Distance to Wheel basemax 45"max 45"Identical
Weight475lb400lbSubstantially EquivalentThe change to theproposed device doesnot raise new safety oreffectiveness concerns
Height48"44.5"Substantially EquivalentThe change to theproposed device doesnot raise new safety oreffectiveness concerns
Footprint28" x 33"28" x 33"Identical
Power System
Input Power90-253 VAC @ 47-63 Hz90-253 VAC @ 47-63 HzIdentical
EMI FilterFN2060B-6-06FN2060B-6-06Identical
AC Power CordRetractable (25ft)Retractable (25ft)Identical
Isolation TransformerYesYesIdentical
EMI FilterYesYesIdentical
UPS (Battery Backup)TAU 2020 OptionalTAU 1515 OptionalTAU 1512 OptionalNoPower system of thePredicate device and thenew Device TAU areSubstantially EquivalentThe proposed new deviceTAU Mini C-arm addedan optional backupbattery (UPS) to protectthe data integrity of thesystem if the deviceexperiences a suddenloss of power e.g. powerLine outage, accidentaldisconnect of power cordfrom main power etc.The UPS allows forlimited additional powerfor controllable exit or tocomplete a shortduration of in-progressprocedure. Safetytesting confirms thischange did not raise anynew safety andeffectiveness concerns.
Laser Alignment:
Laser position indicatorYesYesIdentical
Surgical Lights:
Light bar assembliesYesYesIdentical

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for OrthoScan Imaging Excellence. The logo features a blue square on the left side with a white abstract design inside. To the right of the square, the word "OrthoScan" is written in a bold, black font. Below "OrthoScan", the words "Imaging Excellence" are written in a smaller, blue font.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for OrthoScan. The logo consists of a blue square on the left with a white abstract image inside. To the right of the square is the word "OrthoScan" in large, black, sans-serif font. Below the word "OrthoScan" is the phrase "IMAGING EXCELLENCE" in a smaller, blue, sans-serif font.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for OrthoScan Imaging Excellence. The logo features a blue square on the left with a white abstract image of a face. To the right of the square is the word "OrthoScan" in large, black, sans-serif font. Below "OrthoScan" is the phrase "IMAGING EXCELLENCE" in a smaller, blue, sans-serif font.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image is a logo for OrthoScan. The logo consists of a blue square on the left with a white abstract design inside, followed by the word "ORTHOSCAN" in large, bold, black letters. Below "ORTHOSCAN" is the phrase "IMAGING EXCELLENCE" in smaller, blue letters.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image is a logo for OrthoScan Imaging Excellence. The logo features a blue square on the left with a stylized white image of a face. To the right of the square is the word "OrthoScan" in a large, bold, black font. Below "OrthoScan" is the phrase "IMAGING EXCELLENCE" in a smaller, blue font.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for OrthoScan Imaging Excellence. The logo features a blue square on the left with a white abstract image of a face. To the right of the square is the word "ORTHOSCAN" in black, with the words "IMAGING EXCELLENCE" in blue underneath.

Conclusion:

The changes and differences of the proposed OrthoScan TAU Mini C-Arm described in the table do not change the control mechanism, operating principle, energy type, or intended

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for OrthoScan Imaging Excellence. The logo features a blue square on the left side with a white abstract design inside. To the right of the square is the word "OrthoScan" in large, bold, black letters. Below "OrthoScan" is the phrase "IMAGING EXCELLENCE" in smaller, blue letters.

use found on predicate device OrthoScan FD Mini C-Arm (K133174). In addition, given the high similarity between the legally marketed Solo FD K161976 and the OrthoScan TAU mini C-Arm K183220 we believe that this additional predicate demonstrates that the OrthoScan TAU mini C-Arm K183220 is intended for the same pediatric populations and provide the same intended use and indication for use when imaging all pediatric patient populations

Adverse Effects on Health: The proposed OrthoScan TAU Mini C-arm's potential radiation, mechanical, and electrical hazards are identified and analyzed as part of risk management, and controlled by meeting the applicable CDRH 21CFR subchapter J performance requirements, Recognized Consensus Standards, designing and manufacturing under OrthoScan Quality System, and system verification and validation testing ensure the device performs to the product specifications and its intended use. The adherence to these applicable regulations and certification to Recognized Consensus Standards that apply to this product provides the assurance of device safety and effectiveness.

Applicable Standards:

MDD 93/42/EEC Annex II of the European Medical Devices Directive (MDD) 93/42/EEC.

EN ISO 13485

Medical devices - Quality management systems - Requirements for regulatory purposes

IEC 60601-1 Medical Electrical Equipment, General Requirements for Safety Edition 3.0 (+Corr.1+Corr.2), Date: 2005-12-15

Applicable Standards:

Medical Electrical Equipment, General Requirements for Safety, Electromagnetic Compatibility Edition 3.0, Date: 2007-03-30, Conformance Standard #19-1

IEC 60601-1-3

IEC 60601-1-2

Medical Electrical Equipment, Radiation Protection in Diagnostic X-ray Equipment Edition 2.1, Date: 2013-04 Conformance Standard #12-269

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image is a logo for OrthoScan. The logo consists of a blue square on the left side with a white abstract image inside. To the right of the square is the word "ORTHOSCAN" in large, black, sans-serif font. Below the word "ORTHOSCAN" is the phrase "IMAGING EXCELLENCE" in a smaller, blue, sans-serif font.

60601-2-28 Edition 2 (2010/03/10) Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis Conformance Standard #12-309

Applicable Standards:

IEC 60601-2-54

Medical electrical equipment, Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy Edition 1.0, Date: 2009-06-29 Conformance Standard #12-274

IEC 60825-1

Safety of laser products, Equipment Safety, requirements, and user quide Edition 2.0. Date: 2007-03-30 Conformance Standard #12-273

ISO 14971

Medical devices - Application of risk management to medical devices Edition 2.0, Date: 2007-03-01 Conformance Standard #5-40

Determination of Substantial Equivalence:

Summary Bench Testing

Verification and Validation including hazard mitigations executed resulted in demonstrated system met Design Input and user needs.

The device was tested by the notified test laboratory resulting in device being certified compliant with 6060-1-1 ED 3 series, including IEC 60601-2-54. Further device met all applicable sections of 21 CFR Subchapter J performance standards.

The OrthoScan TAU Mini C-Arm development occurred under our design control processes, software development processes, and overall quality management system. They included but are not limited to,

  • Risk Analysis
  • Required reviews
  • Design reviews ●
  • Component testing
  • Integration testing ●
  • Performance testing ●
  • Safety testing ●
  • . Product use testing

Performance bench testing included:

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image is a logo for OrthoScan Imaging Excellence. The logo features a blue square on the left with a white abstract image of a face. To the right of the square is the word "OrthoScan" in large, bold, black letters. Below "OrthoScan" is the phrase "IMAGING EXCELLENCE" in smaller, blue letters.

Non-clinical testing methods specific to guidance for submission of 510(k)s for Solid State X-Ray Imaging Devices (SSXI); demonstrating system, and imaging performance. Further in line with UCM089742-Premarket Assessment of Pediatric Medical Devices May 24, 2014 and UCM 302938- Pediatric Information for Xray Imaging Device Premarket Notifications Nov 28, 2017. Non-clinical image and dose Lab testing, were employed. Anthropomorphic (PMMA material) phantoms and anatomical simulation phantoms were employed, images were taken by both the new TAU Mini C-arm and predicate device OrthoScan FD (K133174) and the Ziehm Solo FD (K161976). Image comparison sets taken were representative of both the adult and pediatric populations. A Radiologist performed an assessment of 330 individual images arranged in 20 groups of image sets. His conclusion was the image quality combined with a reduced patient dosage will result in a significant improvement in patient care for the TAU device over the Predicate device. Therefore, OrthoScan believes the TAU Mini C-arm image quality, safety and effectiveness to be substantially equivalent to that of the predicate devices at a lower entrance dose level.

OrthoScan TAU mobile fluoroscopic mini C-arm system did not require live human clinical studies to support substantial equivalence in accordance with the FDA quidance Documents, UCM089742- Premarket Assessment of Pediatric Medical Devices May 24, 2014 and UCM 302938- Pediatric Information for X-ray Imaging Device Premarket Notifications Nov 28, 2017. Therefore, OrthoScan conducted a lab test image comparison study employing the use of anthropomorphic phantoms in establishing substantial equivalence based on the modifications to the proposed device and the bench data taken with OrthoScan TAU Mini C-Arm in comparison to the predicate OrthoScan FD Mini C-Arm K133174. Evaluation of the 330 individual images arranged in 20 groups of image sets was conducted by a board-certified Radiologist. His conclusion was the image quality combined with a reduced patient dosage will result in a significant improvement in patient care for the TAU device. His comparison of the dose and images provided further evidence in addition to the laboratory performance data that the complete system works as intended and is substantially equivalent to the predicate device.

Summary of Clinical Test Data:

K183220 510(K) Summary

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.