K Number
K183215

Validate with FDA (Live)

Date Cleared
2019-09-07

(292 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for the relief of pain associated with sore or aching muscles of the lower extremities (leg) due to strain from exercise or normal household work activities. It is also intended for the use of symptomatic relief and management of chronic, intractable pain associated with arthritis. Environments of Use: Clinics, hospital and home environments Patient Population: Adult

Device Description

The PM710 is a wearable electrotherapy device that is designed to alleviate chronic, acute and arthritic muscle leg pain. It delivers TENS (Transcutaneous Electrical Nerve Stimulation) technology through the simple, convenient control on the main unit. The reusable, self-adhesive contouring pads allow for discreet and convenient placement on the pain locations below the knee. The device contains one main TENS unit which is rechargeable and can be attached to a single sized gel and a medium or large band (model number follow by "-M" / "-L") which helps to attach the device to the leg. The pad with attached TENS unit can then be applied to intact skin at the desired location below the knee for therapy and pain relief. Control of the PM710 TENS system is through the control buttons on the main unit and the intensity of the therapy is displayed by the Intensity Level Indicators during treatment. The PM710 will be packaged with an Instruction Manual which provides details on setting up the device for use, setting and controlling intensity levels, and troubleshooting. The system accessories include one charger, one AC adapter for the charger, and one pad holder and storage case.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the Omron Focus TENS Therapy, Model PM710-M/-L, based on the provided FDA 510(k) submission.

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not present a formal table of "acceptance criteria" against which a clinical study's performance metrics are directly compared. Instead, it demonstrates substantial equivalence through extensive non-clinical testing and comparison of technological characteristics with predicate and reference devices. The "performance data" provided refers exclusively to these non-clinical tests.

However, based on the comparison tables, we can infer some "performance criteria" that the device implicitly meets by matching or being within acceptable ranges of the predicate and reference devices.

Feature/ParameterAcceptance Criteria (Inferred from Predicate/Reference)Reported Device Performance (Omron Focus TENS PM710-M/-L)
Classification Regulation21 CFR§882.5890, Transcutaneous electrical nerve stimulator for pain relief21 CFR§882.5890, Transcutaneous electrical nerve stimulator for pain relief
Product CodeNUH and NYN (Transcutaneous electrical nerve stimulator for pain relief)Primary: NUH, Secondary: NYN
Indications for UseRelief of pain associated with sore or aching muscles (lower back, arms, legs, shoulders, or feet) due to strain from exercise or normal household work activities; symptomatic relief and management of chronic, intractable pain, and relief of pain associated with arthritis. The Quell reference device also includes temporary relief of pain associated with sore and aching muscles in the lower extremities due to strain from exercise or normal household and work activities, and symptomatic relief and management of chronic intractable pain.Intended for the relief of pain associated with sore or aching muscles of the lower extremities (leg) due to strain from exercise or normal household work activities. Also intended for the use of symptomatic relief and management of chronic, intractable pain associated with arthritis. (Narrower than Avail but similar to Quell for lower extremities and chronic pain/arthritis).
Environments of UseClinics, hospital, and home environments (for predicate Avail and implied for other TENS devices)Clinics, hospital, and home environments
Patient PopulationAdultAdult
ContraindicationsAvoid use with cardiac pacemakers, implanted defibrillators, or other implanted metallic or electronic devices.Same as predicate and reference devices regarding cardiac pacemakers, implanted defibrillators, or other implanted metallic or electronic devices.
Electrical ParametersOutput parameters within the range of predicate Avail device, and within IEC60601-2-10:2012 (Clause 201.4.2) limit of <2mA/cm² for current density. Waveforms should achieve similar therapeutic outcomes.Max current density: 0.97 mA/cm² (well below 2 mA/cm² limit). Max average power density: 7.59x10⁻³ W/cm². Waveform: Biphasic Rectangular. Pulse Duration: 60 µsec. Frequency: 1-250 Hz. Net Charge: 0 µC per pulse.
Treatment DurationSelectable time ranges for TENS therapy, commonly 5-60 minutes, or 30-180 minutes, with reference devices offering 30 or 60 minutes.Fixed timer duration for 30 minutes. (Matches R-T1 reference, within predicate range).
Safety StandardsCompliance with ES 60601-1, IEC60601-1-2, IEC60601-2-10, IEC 60601-1-11.ES 60601-1, IEC60601-1-2, IEC60601-2-10, IEC 60601-1-11.
BiocompatibilityPatient-contacting components must demonstrate biocompatibility.Tested to demonstrate biocompatibility.
Software/FirmwareVerification and validation to ensure the device performs as intended.Software verification and validation completed.

2. Sample Size for Test Set and Data Provenance

The submission explicitly states: "Clinical testing was not performed to support this 510(k) submission." Therefore, there is no "test set" in the context of clinical data, no sample size, and no data provenance for a clinical test set. The evaluation relies entirely on non-clinical (bench) testing and comparison to legally marketed predicate devices.

3. Number of Experts and Qualifications for Ground Truth of Test Set

Not applicable, as no clinical test set was used to establish ground truth in this submission.

4. Adjudication Method for Test Set

Not applicable.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done, as traditional clinical studies were not performed. The submission does not discuss human reader improvement with or without AI assistance as this is a TENS device, not an AI diagnostic tool.

6. Standalone (Algorithm Only) Performance Study

Not applicable. This is a medical device, not an AI algorithm. Its performance is evaluated through its physical and electrical characteristics as a TENS unit.

7. Type of Ground Truth Used

The "ground truth" for this submission is based on engineering specifications, adherence to international and national standards (e.g., IEC 60601 series), and the demonstrated performance of legally marketed predicate and reference TENS devices. The "truth" is that the device's electrical outputs and functional characteristics are safe and effective by being substantially equivalent to these established devices.

8. Sample Size for the Training Set

Not applicable, as no AI model or learning algorithm is described, and no clinical "training set" was used. The device is a TENS unit, not a machine learning system.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set mentioned in the submission.

{0}------------------------------------------------

September 7, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a stylized image of a human figure. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".

Omron Healthcare, Inc. % Ronald Warren Vice President, Regulatory Affairs Experien Group 224 Airport Parkway, Suite 250 San Jose, California 95110

Re: K183215

Trade/Device Name: Focus TENS Therapy, Model PM710-M/-L Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: August 5, 2019 Received: August 6, 2019

Dear Ronald Warren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

Page

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Robert Kang, PharmD Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183215

Device Name Omron Focus TENS Model PM710-M/-L

Indications for Use (Describe)

The device is intended for the relief of pain associated with sore or aching muscles of the lower extremities (leg) due to strain from exercise or normal household work activities.

It is also intended for the use of symptomatic relief and management of chronic, intractable pain associated with arthritis. Environments of Use: Clinics, hospital and home environments

Patient Population: Adult

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Notification K183215

GENERAL INFORMATION [807.92(a)(1)]

Applicant:

Omron Healthcare, Inc. 1925 West Field Court Lake Forest, IL 60045 USA Phone: 847-247-5626 FAX: 847-680-6269

Contact Person:

Ronald S. Warren Vice President, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 408-505-3926 FAX: 408-400-0856

Date Prepared: December 11, 2018

DEVICE INFORMATION [807.92(a)(2)]

Trade Name:

Focus TENS Model PM710-M/-L

Generic/Common Name:

Transcutaneous Electrical Nerve Stimulator for Pain Relief

Classification:

Class II per 21CFR882.5890

Product Code:

NUH and NYN

{4}------------------------------------------------

PREDICATE DEVICE(S) [807.92(a)(3)]

The Omron Focus TENS Model PM710-M/-L ("PM710") is substantially equivalent to the predicate device, the Omron Avail Wireless Dual Channel TENS PM601 (K172079), with regard to product labeling, intended use, anatomical sites, patient population, performance testing, technological characteristics and safety characteristics. In addition, the PM710 device is also similar to two other reference devices, the Neurometrix, Inc. Quell (K152954) and the Shenzhen Roundwhale Technology Co., Ltd Model R-T1 TENS stimulator (K180956) in regard to intended use and features.

DEVICE DESCRIPTION [807.92(a)(4)]

The PM710 is a wearable electrotherapy device that is designed to alleviate chronic, acute and arthritic muscle leg pain. It delivers TENS (Transcutaneous Electrical Nerve Stimulation) technology through the simple, convenient control on the main unit. The reusable, self-adhesive contouring pads allow for discreet and convenient placement on the pain locations below the knee.

The device contains one main TENS unit which is rechargeable and can be attached to a single sized gel and a medium or large band (model number follow by "-M" / "-L") which helps to attach the device to the leg. The pad with attached TENS unit can then be applied to intact skin at the desired location below the knee for therapy and pain relief. Control of the PM710 TENS system is through the control buttons on the main unit and the intensity of the therapy is displayed by the Intensity Level Indicators during treatment. The PM710 will be packaged with an Instruction Manual which provides details on setting up the device for use, setting and controlling intensity levels, and troubleshooting. The system accessories include one charger, one AC adapter for the charger, and one pad holder and storage case.

INDICATIONS FOR USE [807.92(a)(5)]

The device is intended for the relief of pain associated with sore or aching muscles of the lower extremities (leg) due to strain from exercise or normal household work activities.

It is also intended for the use of symptomatic relief and management of chronic, intractable pain associated with arthritis.

Environments of Use: Clinics, hospital and home environments.

Patient Population: Adult.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]

In regard to technological characteristics, the PM710 device is similar to the predicate Avail TENS devices in respect to the intended use and indication for use. PM710 is specifically intended for relieving pain of sore or aching muscles of the leg, whereas the Avail device is intended to relieve pain with sore or aching muscles for a wider range of body sites including the lower back, arms, legs, shoulders or feet. Both PM710 and Avail are also intended for the use of symptomatic relief and management of chronic, intractable pain associated with arthritis, and can be used in clinics, hospitals and home for adults.

{5}------------------------------------------------

510(k) SUMMARY (CONT.)

The PM710 device offers one TENS treatment mode similar to the reference device Quell, whereas the predicate Avail device offers nine TENS treatment modes and one microcurrent mode. The waveform generated by PM710 is also different from the waveforms of the Avail system, however the electrical parameters of both are very similar. As such, the waveforms produced by PM710 and Avail can both achieve the same therapeutic outcome of "symptomatic relief and management of chronic, intractable pain as well as the discomfort associated with arthritis". This is due to the presence of:

  • o High frequency (50+) Hz for chronic pain that requires non-contracting stimulation modes for a comfortable smoother therapy for chronic pain therapy.
  • Modulation via different pulse trains to prevent muscle accommodation for both . chronic pain and arthritis.
  • Variation of both high and low rate pulses within the mode to address both muscle o and joint pain from arthritis and various chronic pain.

The composite action of the low and high rate Hz stimulation produces rhythmic muscle contractions which then produces improved local blood vessel vasodilation thus reducing edema and retained metabolites commonly occurring from excessive muscle strain to support weak/arthritic joints. In addition, it enhances the supply of fresh oxygen and nutrients that can further result in decreased discomfort.

In regard to other output characteristics, summary tables of output parameters for the PM710, predicate and reference devices are provided in this submission. In general, the PM710 output parameters fall within the range of output parameters for the predicate Avail device. For example, the maximum current density (mA/cm2) range for PM710 is 0.97, while the Avail is 0.0008 ~ 0.17 which are both well below the IEC60601-2-10:2012 (Clause 201.4.2) limit of less than 2mA/cm². The maximum average power density (W/cm²) range is 7.59x103 for PM710, whereas the Avail maximum average power density range is 1.4x10-8 ~ 6.8x10-4, and the reference device R-T1 TENS is 1.08x10-4~1.48x10-2, which has a closer match to PM710.

In regard to treatment duration, the PM710 offers a specified timer duration for 30 minutes. For Avail, there are two selectable time ranges for TENS therapy; 5 to 60 minutes and 30 to180 minutes, depending on the selected mode. The same time duration of 30 minutes is also offered by the reference device of R-T1. In this regard, PM710 should be considered at least as the predicate and reference devices. A comparison of the PM710 output specification to the predicate device Avail and reference device R-T1 TENS are also provided in this submission.

SUBSTANTIAL EQUIVALENCE

The proposed device, predicate device and reference devices are all intended for use as transcutaneous electrical nerve stimulation in adult populations for use in clinic, hospital or home settings. Omron has completed comprehensive design verification testing, electrical safety and electromagnetic compatibility testing, software verification and validation, and usability testing to ensure that the PM710 device performs as intended. The PM710 also passed testing requirements for electrical safety and EMC, and the device patient-contacting were tested to demonstrate biocompatibility. The minor differences in labeling and technological characteristics between the proposed device and the predicate device have been evaluated and determined to not raise different questions

{6}------------------------------------------------

510(k) SUMMARY (CONT.)

of safety or effectiveness. As such, the proposed PM710 is substantially equivalent to the predicate device. A comparison table summarizing the specifications and features of the proposed PM710 device, the predicate and reference devices is included in the Substantial Equivalence Table (Table 1) below.

In regard to other technological characteristics, summary tables (Tables 2, 3 and 4) of output parameters for the PM710, predicate and reference devices are provided below.

PERFORMANCE DATA [807.92(b)]

All necessary bench and nonclinical testing were conducted on the PM710 to support a determination of substantial equivalence to the predicate device.

[807.92(b)(1)|Nonclinical Testing Summary:

All necessary non-clinical and usability testing was conducted on the PM710 to confirm that the device performs as intended. The nonclinical, bench testing included:

  • Current Consumption
  • Output Waveform
  • PAD Peeling-off Detection
  • Function Check
  • AD Voltage Detection Accuracy
  • Charging Current ●
  • Voltage Endurance, Insulation Resistance ●
  • Battery Characteristics
  • Charger:
    • 0 Current Consumption
    • o Voltage Endurance, Insulation Resistance
  • Durability test for the connector (between the unit and pad) ●
  • Dispersion ●
  • Adhesion (EC12 and Cutaneous Electrode) ●
  • Adhesion on Human Skin .
  • Impedance for PAD
  • Battery Characteristics after Drop/Vibration Testing ●

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the PM710 meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the PM710 does not raise different questions of safety or effectiveness for TENS therapy when compared to the predicate device.

[807.92(b)(2)|Clinical Testing Summarv:

Not applicable. Clinical testing was not performed to support this 510(k) submission.

{7}------------------------------------------------

510(k) SUMMARY (CONT.)

CONCLUSIONS [807.92(b)(3)]

Based on the results from the nonclinical and usability tests performed in support of PM710, it is concluded that the proposed device is safe, is effective, and performs at least as safely and effectively as the legally marketed predicate device.

SUMMARY

The proposed device and predicate device are both intended for use as transcutaneous electrical nerve stimulation in adult populations for use in clinic, hospital or home settings. Omron has completed comprehensive design verification testing, electrical safety and electromagnetic compatibility testing, software verification and validation, and usability testing to ensure that the PM710 device performs as intended. The PM710 also passed testing requirements for electrical safety and EMC, and the device patientcontacting components were tested to demonstrate biocompatibility. The minor differences in technological characteristics between the proposed device and the predicate device have been evaluated and determined to not raise different questions of safety or effectiveness. As such, PM710 is substantially equivalent to the predicate device.

{8}------------------------------------------------

5
For the may be and the be do
1
production and state of the program and any andSCA WANAA1
A 41 Ch WAW 1
And Real Property of Children1
o4
200811
No. of Concession, North Street, Income of the pro-1when a need soon coupe some and whenA Microsoft
1of Children of Children1
1BW PS P B B B W W TNAI A1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
P11111
FeatureProposed DeviceFocus TENSPM710-M/PM710-LPrimary Predicate DeviceOmron Avail Wireless DualChannel TENS PM601K172079Reference DeviceQuellK152954Reference DeviceModel R-T1 TENSStimulatorK180956Analysis of TechnologicalDifferences
Classification-Regulation21 CFR§882.5890,Transcutaneous electricalnerve stimulator for painrelief21 CFR§882.5890,Transcutaneous electricalnerve stimulator for painrelief21 CFR§882.5890,Transcutaneous electricalnerve stimulator for painrelief21 CFR§882.5890,Transcutaneous electricalnerve stimulator for painreliefNo difference. Proposeddevice and predicate devicehave the same medical deviceclassification number.
Classification-Product CodePrimary: NUH -Transcutaneous electricalnerve stimulator for painrelief. Secondary: NYN-Transcutaneous electricalnerve stimulator for painrelief.Primary: NUH -Transcutaneous electricalnerve stimulator for painrelief. Secondary: NYN-Transcutaneous electricalnerve stimulator for painrelief.NUH - Transcutaneouselectrical nerve stimulatorfor pain reliefNUH - Transcutaneouselectrical nerve stimulator forpain reliefNo difference. Proposeddevice and predicate devicehave the same medical deviceproduct code.
Indications for UseThe Focus TENS isintended for the relief ofpain associated with soreor aching muscles of thelower extremities (leg) dueto strain from exercise ornormal household workactivities.It is also intended for theuse of symptomatic reliefand management ofchronic, intractable painassociated with arthritis.The Avail Wireless DualChannel TENS is intendedfor the relief of painassociated with sore oraching muscles of the lowerback, arms, legs, shoulders orfeet due to strain fromexercise or normal householdwork activities.When used for thesymptomatic relief andmanagement of chronic,intractable pain and relief ofpain associated with arthritis,choose Tap, Shoulder, Armor Leg mode.Quell is intended for use asa transcutaneous electricalnerve stimulation device fortemporary relief of painassociated with sore andaching muscles in the lowerextremities due to strainfrom exercise or normalhousehold and workactivities.Quell is intended for use asa transcutaneous electricalnerve stimulation device forthe symptomatic relief andmanagement of chronicintractable pain.The device may be usedduring sleep. The device islabeled for use only withThis device is designed to beused for temporary relief ofpain associated with sore andaching muscles in theshoulder, waist, back, neck,upper extremities (arm), andlower extremities (leg) due tostrain from exercise ornormal household workactivities.Similar to Avail but withnarrower use for knee ratherthan the leg generally.
Environments of Use:Clinics, hospital and homeenvironmentsClinics, hospital and homeenvironments
Patient Population:Adult

{9}------------------------------------------------

100 1 1000
the was an any and the the more------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
1
111300
1-1rmatic11/2 10/2 10/20170
IoAWA 40PR 2661Change than the coupleﺍ ﺍﮮ1tCPB
IC1Callery of Children1410.001BB
001. 1. 1.e100CILIVA.
11And Income of the later of the later of the later of the later of the later of the later of the later of the later of the later of the later of the later of the later of theI1I
tianAnd AddresswaSI281
ni ki------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ﺮ1
1S
FeatureProposed DevicePM710-M/PM710-LPrimary Predicate DeviceOmron Avail Wireless DualChannel TENS PM601K172079Patient Population: AdultReference DeviceQuellK152954compatible NeuroMetrixelectrodes.Reference DeviceModel R-T1 TENSStimulatorK180956Analysis of TechnologicalDifferences
Environment of UseClinics, hospitals and homeenvironmentsClinics, hospitals and homeenvironmentsUnknownUnknownSame as Avail
Patient PopulationAdultsAdultsAdultsAdultsNo difference. Intended forsame population.
Contraindications/Warnings/PrecautionsContraindicationsDo not use this device ifyou have a cardiacpacemaker, implanteddefibrillator, or otherimplanted metallic orelectronic device. Such usecould cause electric shock,burns, electricalinterference, or death.ContraindicationsDo not use this device if youhave a cardiac pacemaker,implanted defibrillator, orother implanted metallic orelectronic device. Such usecould cause electric shock,burns, electrical interference,or death.ContraindicationsDo not use this device if youhave a cardiac pacemaker,implanted defibrillator, orother implanted metallic orelectronic device. Such usecould cause electric shock,burns, electricalinterference, or death.Contraindications1) Do not use this device ifyou have a cardiacpacemaker, implanteddefibrillator, or otherimplanted metallic orelectronic device. Such usecould cause electric shock,burns, electrical interference,or death.2) The device should not beused when cancerous lesionsor other lesions are present inthe treatment area.3) Stimulation should not beapplied over swollen,infected, inflamed areas orskin eruptions (e.g. phlebitis,thrombophlebitis, varicoseveins, etc.).4) Electrode placements mustbe avoided in the carotidsinus area (anterior neck) ortran-scerebrally (through thehead).5) This device should not beused in overly enervatedSame as Avail and Quell forContraindications
Table 1: Substantial Equivalence Table – Regulatory Information (cont.)
FeatureProposed DevicePM710-M/PM710-LPrimary Predicate DeviceOmron Avail Wireless DualChannel TENS PM601,K172079Reference DeviceQuellK152954Reference DeviceModel R-T1 TENSStimulatorK180956Analysis of TechnologicalDifferences
areas.6) Inguinal hernia.7) Do not use on scarredareas following a surgery forat least 10 months after theoperation.8) Do not use with seriousarterial circulatory problemsin the lower limbs.
Single UsePads are for single patient usePads are for single patient usePads are for single patient usePads are for single patient useNo difference. Proposed andpredicate devices are durablemedical equipment intendedfor multiple uses.
SterilityExternal contacting device, nonsterileExternal contacting device, nonsterileUnknownUnknownNo difference. Proposed andpredicate device are providednonsterile.
Specifications/FeaturesOver-the-Counter(OTC)YesYesYesYesNo difference
Power Source(s)Rechargeable Lithium-ionbatteryRechargeable Lithium-ionbatteryRechargeable Lithium-ionbattery6.0 V D.C., 4 x AAAbatteriesSame battery type as Availand QuellNo difference
Method of LineCurrent IsolationN/A (internal powersource)N/A (internal power source)N/A (internal power source)N/A (internal power source)
Patient LeakageCurrent---------------

{10}------------------------------------------------

{11}------------------------------------------------

has a le have " starts and the in an
1
O VAVAA O AUV OARA CA WARREN WARREN WARREN BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE BE1
WAWA 40Party of the Br------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1
C111
A A WANo. of Conce
when at need sound second coups and subscription of the contraction of the contribution and subject of the county of the county of the county of the county of the county of t
of Children of Concession
And W
Carolina Continued Concession Company of Concession,1NAI A
State Production Come of Concession Comers of Concessional Construction of Construction of Construction of Construction of Children1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------2
114
FeatureProposed DevicePM710-M/PM710-LPrimary Predicate DeviceOmron Avail Wireless DualChannel TENS PM601K172079Reference DeviceQuellK152954Reference DeviceModel R-T1 TENSStimulatorK180956Analysis of TechnologicalDifferences
-Normal Condition(uA)<10uA<10uA<10uA11.4uASame as Avail
-Single FaultCondition (uA)<50uA<50uA<100uA9.6uASame as Avail
Average DC currentthrough electrodeswhen device is onbut no pulse arebeing applied (uA)00<1uA0No difference
Number of outputModes1 TENS mode9 TENS modes1 TENS mode18 TENS modesSame as Quell
Number of outputchannels1ch1ch1chAlternating 2 channelsSame as Avail and Quell
Synchronous orAlternatingNoneNoneN/ABy electrical circuit andsoftwareSame as Avail
Method of ChannelIsolationRegulated CurrentRegulated CurrentRegulated currentRegulated currentNo difference
Regulated Currentor RegulatedVoltageMicroprocessorMicroprocessorYesYesNo difference
Software/Firmware/MicroprocessorControl?NoNoYesYesSame as Avail
Automatic OverloadTrip?YesYesYesYesNo difference
Automatic No-LoadTrip?YesYesYesYesNo difference
Automatic shut Off?Yes, Power On/Off buttonon the deviceYes, Power On/Off button onthe device and in the Appsoftware4 button pressesYesSimilar to Avail but with noApp associated with FocusTENS
User over ridecontrol?

A5-10

{12}------------------------------------------------

1C11
1
C
WA TO WARDS OF THEAAA V & AAA TVAV
------------------------------------------------------------------------------------------------------------------------
And Annual Property of Children
-€
1
11€
Caroline1
No. of Conce
when a need cours nown some and submit1
1
400 WThe Real Property of Children
wa
WALL CALLERY
1Call of Children
12
1
CCaroline of Children
FeatureProposed DevicePM710-M/PM710-LPrimary Predicate DeviceOmron Avail Wireless DualChannel TENS PM601K172079Reference DeviceQuellK152954Reference DeviceModel R-T1 TENSStimulatorK180956Analysis of TechnologicalDifferences
Indication displayOn/Off Status?Yes. LED indicator onmain unitYes on App and LEDindicator on main unitYesYesSimilar to Avail but with noApp associated with FocusTENS
Low Battery?Yes. LED indicator onmain unitYes on AppYesYesSimilar to Avail but with noApp associated with FocusTENS
Voltage/CurrentLevel?Yes. LED indicator onmain unitYes on AppNoYesSimilar to Avail but with noApp associated with FocusTENS
Timer Range(Minutes)30 minutes5-60minutes and30-180minutes60 minutes30 minutesSame as R-T1 and withinrange of Avail
Compliance withVoluntarystandards?ES 60601-1, IEC60601-1-2, IEC60601-2-10, IEC60601-1-11ES 60601-1, IEC60601-1-2,IEC60601-2-10, IEC 60601-1-11IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC62304, IEC 60601-1-11IEC 60601-1, IEC60601-1-2,IEC60601-2-10, IEC 60601-1-11Same as Avail
Compliance with 21CFR898?N/A (no patient cable)N/A (no patient cable)YesYesSame as Avail
WeightDevice: Approx. 1.9 oz (55g)Knee pad: Approx. 0.7 oz(20 g)Knee band M: Approx. 1.0oz (27 g)Knee band L: Approx. 1.0oz (29 g)Charger: Approx. 3.5 oz(100 g)Device: Approx. 42 g (Bothunits have same weight)Pad-L: Approx. 25 gPad-M: Approx. 20 gCharger: Approx. 100 g62g (2.2 oz)0.243 (lbs., oz.)Similar. Difference does notaffect safety and effectivenessof use.
FeatureProposed DevicePM710-M/PM710-LPrimary Predicate DeviceOmron Avail Wireless DualChannel TENS PM601K172079Reference DeviceQuellK152954Reference DeviceModel R-T1 TENSStimulatorK180956Analysis of TechnologicalDifferences
Dimensions (W x Hx D)Device: Approx. 2.3 (W) ×2.8" (H) x 0.6" (D) (60 ×72 × 16 mm)Charger: Approx. 3.5" (W)× 3.1" (H) x 0.9" (D) (90x 80 x 23.5 mm)Knee pad: Approx. 5.1"(W) x 2.3" (H) x 0.6" (D)(130×60×16 mm)Knee Band M: Approx.15.1" (W) x 2.5" (H) (385× 64 mm)Knee Band L: Approx.17.7" (W) x 2.5" (H) (450× 64 mm)Device: Approx. 60 (W)× 72(H)x 15.5 (D)mm(Both units have sameweight)Charger: Approx. 158 (W)x90 (H)x 20.5 (D)mmPad-L: Approx. 219 (W) ×83.5 (H) × 7.5 (D)mmPad-M: Approx. 180 (W) ×79.5 (H) × 7.5 (D)mm11mm (0.4") x 74mm (2.9")x 98mm (3.9")4.82x2.78x1.08 (in.)Similar. Difference does notaffect safety and effectivenessof use.
Operatingconditions10 to 40 °C30 to 80%RH700 to 1060 hPa (non-condensing)10 to 40 °C30 to 80%RH700 to 1060 hPa (non-condensing)-5° C to 40° C15% to 93%70 kPa to 106 kPa5°C40°C15%RH93%RH700 hPa to 1060 hPaSame as Avail
Charging conditions5 to 35 °C (non-condensing)5 to 35 °C (non-condensing)-5° C to 40° C15% to 93%70 kPa to 106 kPaN/A(no rechargeable battery)Same as Avail
Storage conditions0 to 40 °C30 to 80% RH (non-condensing)0 to 40 °C30 to 80% RH (non-condensing)-25° C to 70° C10% to 93%70 kPa to 106 kPa-10°C55°C10%RH90%RH700 hPa to 1060 hPaSame as Avail
Transportingconditions-20 to 60 °C10 to 90% RH (non-condensing)-20 to 60 °C10 to 90% RH (non-condensing)-25° C to 70° C10% to 93%70 kPa to 106 kPa-10°C55°C10%RH90%RH700 hPa to 1060 hPaSame as Avail
Electrode styleHV-KNPAD-ZReusableHV-WPAD-M or HV-WPAD-LReusableQuell electrodesReusableReusableSimilar to Avail and Quell
FeatureProposed DevicePM710-M/PM710-LPrimary Predicate DeviceOmron Avail Wireless DualChannel TENS PM601K172079Reference DeviceQuellK152954Reference DeviceModel R-T1 TENSSimulatorK180956Analysis of TechnologicalDifferences
Patient ContactAccessoryYesYesYesYesNo difference

{13}------------------------------------------------

(Cont.)

Regulatory Information

Substantially Equivalent

A5-12

{14}------------------------------------------------

0(k) SUMMARY (CONT.

Table 1: Substantial Equivalence Table – Regulatory Information (cont.

{15}------------------------------------------------

1
A W A 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
11000 V 1
PM7
or
0I
n
ariso
omp:
C
U
up
U
CS
. .
(
bl
ParameterMode of Program NameKneeSteadyAcupuncture LikeKneadTapLower BackShoulderJointLegArmMicrocurrentStandardAlternative
Focus TENSPM710-M/-LAvail K172079Quell K152954
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
MaximumOutputVoltage [V]@500Ω45.025.938.427.938.338.238.425.636.738.20.049.249.4
@2kΩ68.645.250.837.250.350.450.435.749.550.40.1
@10kΩ78.559.955.240.754.955.855.839.254.655.20.5
MaximumOutputCurrent [mA]@500Ω90.051.876.855.876.676.476.851.273.476.40.198.498.8
@2kΩ34.322.625.418.625.225.225.217.924.825.20.1
@10kΩ7.96.05.54.15.55.65.63.95.55.50.1
Duration of primary phase[µsec]609696969696969696962500000100100
Pulse Duration [µsec]609696969696969696962500000100100
Frequency [Hz]1-25099251.651-20.132-1081-1942.43-1082-51.652-51.650.260-10080
For multiphasic waveforms only:Symmetrical phasesN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
Phase DurationN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
Net Charge(µC per pulse)(@500Ω) [uC]0000000000000

{16}------------------------------------------------

Villa
A Was a last " and the last to the-
1(
-
œ
I
1
PM71
1
1141
0
1
ISON
rist
mpar
C1
A
tput
u
11 10
(J
ble
-
-
5.404.977.375.367.357.337.374.927.057.33125.009.849.88
0.970.160.030.090.110.170.100.120.120.120.000.500.45
2.700.980.030.280.300.790.280.530.360.380.031.971.58
7.59E-035.71E-042.53E-051.73E-042.54E-046.77E-042.41E-043.04E-042.99E-043.24E-041.40E-083.46E-032.79E-03
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A4.0
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A2.1
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A0.1
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A0.2
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A0.1
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A0.4
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A

{17}------------------------------------------------

#11
21(
TEN!1711
4E1l
1l
Inr1 0 10
omparison
N)
tputह1
t1
able 3:WP
ParameterR-T1 TENS Stimulator K180956
Mode of Program NameShoulder(P1)Shoulder(P2)Shoulder(P3)Back(P1)Back(P2)Back(P3)Arm(P1)Arm(P2)Arm(P3)
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
Maximum Output Voltage [V]@500Ω29.4@500Ω29.6@500Ω29.0@500Ω30.0@500Ω29.2@500Ω28.8@500Ω29.2@500Ω28.4@500Ω29.4
@2ΚΩ104.5@2ΚΩ108.0@2ΚΩ109.0@2ΚΩ111.5@2ΚΩ107.0@2ΚΩ111.5@2ΚΩ106.5@2ΚΩ109.5@2ΚΩ111.5
@10ΚΩ111.5@10ΚΩ113.0@10ΚΩ112.5@10ΚΩ114.5@10ΚΩ112.0@10ΚΩ115.5@10ΚΩ108.0@10ΚΩ113.0@10ΚΩ115
Maximum Output Current [mA]@500Ω58.8@500Ω59.2@500Ω58.0@500Ω60.0@500Ω58.4@500Ω57.6@500Ω58.4@500Ω56.8@500Ω58.8
@2ΚΩ52.3@2ΚΩ54.0@2ΚΩ54.5@2ΚΩ55.8@2ΚΩ53.5@2ΚΩ55.8@2ΚΩ53.3@2ΚΩ54.8@2ΚΩ55.75
@10ΚΩ11.2@10ΚΩ11.3@10ΚΩ11.3@10ΚΩ11.5@10ΚΩ11.2@10ΚΩ11.6@10ΚΩ10.8@10ΚΩ11.3@10ΚΩ11.5
Duration of primary phase [usec]250253205256253332252.4197201
Pulse Duration [usec]250253205256253332252.4197201
Frequency [Hz]2-64-82-12510-806-1010022-100100
For multiphasic waveforms only:Symmetrical phasesN/AN/AN/AN/AN/AN/AN/AN/AN/A
Phase DurationN/AN/AN/AN/AN/AN/AN/AN/AN/A
Net Charge(µC per pulse) (@500Ω) [uC]000000000
Maximum Phase Charge (@500Ω) [μC]14.7014.9811.8915.3614.7819.1214.7411.1911.82

{18}------------------------------------------------

Andrew of the propertyWAe)
----------Propertymanufa-
COLLATO21111
1TT 11I0
61111 11. 11.41110
o manage and min of anyบ ทาง ที่ 1 ในปี ค.ศ. 141 ปี ด ท
AC
11------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1
C11€
1Cœ1160 K11P21œ1
ParameterR-T1 TENS StimulatorK180956
Maximum Current Density (@500Ω)[mA/cm2] r.m.s.0.200.240.820.760.260.930.120.700.74
Maximum Average Current (averageabsolute value), mA0.180.242.972.460.303.820.062.242.36
Maximum Average Power Density(@500Ω) [W/cm²]3.24E-044.43E-045.39E-034.61E-035.39E-046.88E-031.08E-043.97E-034.34E-03
Burst Mode(a) Pulses perburstN/AN/AN/AN/AN/AN/AN/AN/AN/A
(b) Bursts persecondN/AN/AN/AN/AN/AN/AN/AN/AN/A
(c) Burst durationN/AN/AN/AN/AN/AN/AN/AN/AN/A
(d) Duty cycle:Line(b)xLine(c)N/AN/AN/AN/AN/AN/AN/AN/AN/A
ON Time (seconds)N/AN/AN/AN/AN/AN/AN/AN/AN/A
OFF Time (seconds)N/AN/AN/AN/AN/AN/AN/AN/AN/A
Additional FeaturesN/AN/AN/AN/AN/AN/AN/AN/AN/A

{19}------------------------------------------------

ParameterR-T1 TENS Stimulator K180956
Mode of Program NameHip(P1)Hip(P2)Hip(P3)Leg(P1)Leg(P2)Leg(P3)Joint(P1)Joint(P2)Joint(P3)KneadRubTap
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
MaximumOutputVoltage[V]@500Ω29.229.42929.629.428.629.629.629.63029.630
@2kΩ111.5112.5112.5112.5111108111109.5112106109.5111.5
@10kΩ114115.5112.5114113112.5112.5115.5114114114116.5
MaximumOutputCurrent[mA]@500Ω58.458.85859.258.857.259.259.259.26059.260
@2kΩ55.7556.2556.2556.2555.55455.554.75565354.7555.75
@10kΩ11.411.5511.2511.411.311.2511.2511.5511.411.411.411.65
Duration of primary phase [usec]152.8255203.5250.7155253201.4156.9155.2220219225.5
Pulse Duration [usec]152.8255203.5250.7155253201.4156.9155.2220219225.5
Frequency [Hz]1006-501004-501006-102-1001008083-13225-79147-291
For multiphasic waveforms only:Symmetrical phasesN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
Phase DurationN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
Net Charge(µC per pulse) (@500Ω) [µC]000000000000
ParameterR-T1 TENS StimulatorK180956
Maximum Phase Charge(@500Ω) [μC]8.9214.9911.8014.849.1114.4711.929.299.1913.2012.9613.53
Maximum Current Density(@500Ω) [mA/cm2] r.m.s.0.640.590.730.590.650.250.740.660.580.900.691.36
Maximum Average Currentaverage absolute value), mA1.781.502.361.481.820.292.381.861.473.482.057.87
Maximum Average PowerDensity (@500Ω) [W/cm²]3.26E-032.76E-034.28E-032.75E-033.35E-035.17E-044.41E-033.44E-032.72E-036.53E-033.79E-031.48E-02
BurstMode(a) Pulses perburstN/AN/A25N/A25N/AN/A25N/AN/AN/AN/A
(b) Bursts persecondN/AN/A2.00N/A2.06N/AN/A2.00N/AN/AN/AN/A
(c) BurstdurationN/AN/A0.24N/A0.24N/AN/A0.24N/AN/AN/AN/A
(d) Duty cycle:Line(b)xLine(c)N/AN/A0.48N/A0.49N/AN/A0.48N/AN/AN/AN/A
ON Time (seconds)N/AN/A0.24N/A0.24N/AN/A0.241N/AN/AN/AN/A
OFF Time (seconds)N/AN/A0.26N/A0.245N/AN/A0.258N/AN/AN/AN/A
Additional FeaturesN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A

arison for R-T1 TENS #2 out Comps Tabla 1. Qui

{20}------------------------------------------------

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).