(81 days)
Head Coil - 12 is a 12 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Head Coil - 32 is a 32 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Carotid Coil - 8 is an 8 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for obtaining diagnostic images of carotid. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Temporomandibular Joint Coil - 4 is a 4 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for diagnostic images of the mandible region. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Infant Coil - 24 is a 24 channel receive-only RF coil designed for use for imaging of the infant's brain, spine, heart, torso and extremities. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Cardiac Coil - 24 is a 24 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The indications for use include imaging of the heart, and mediastinum regions. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Foot & Ankle Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for foot and ankle imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 4, Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24 are receive-only multichannel phased array coils designed for use with UIH 3T MRI systems. The suffix number in coil name stands for the receive channel of the resonant nucleus is hydrogen protons (1H). All coil elements are protected from transmit field by both active and passive decoupling circuits. Each coil is designed to provide optimum image quality for its dedicated human anatomy region as defined in intended use.
The provided text is a 510(k) summary for several MRI coils manufactured by Shanghai United Imaging Healthcare Co., Ltd. This document primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and non-clinical performance data and does not contain detailed information about specific acceptance criteria or an explicit study proving performance against those criteria in the context of diagnostic accuracy.
Here's a breakdown of the requested information based on the provided text, with an acknowledgment of what is not present:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list specific quantitative acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, accuracy) for a given clinical task. Instead, the "performance" demonstrated for these MRI coils relates to their technical imaging capabilities and adherence to safety and image quality standards.
The non-clinical tests performed, and their implied "acceptance criteria" are that the devices perform "as expected" and are "substantially equivalent" to predicate devices. The "reported device performance" is essentially their successful compliance with these standards.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Safety and Electrical Performance: | Compliant: Meets requirements of ES60601-1:2005/(R)2012 and IEC 60601-2-33 Ed. 3.1:2013 |
| Biocompatibility: | Compliant: Meets requirements of ISO 10993-5:2009 and ISO 10993-10:2010 |
| Image Quality (Technical): | Compliant: Meets requirements for SNR and uniformity as per MS 1-2008(R2014), MS 3-2008(R2014), MS 6-2008(R2014), and MS 9-2008(R2014). Generates "diagnostic quality images." |
| Substantial Equivalence: | Achieved: Devices found substantially equivalent to predicate devices based on non-clinical tests and technological characteristics. |
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "Sample clinical images were provided to support the ability of the proposed devices to generate diagnostic quality images." However, it does not specify the sample size of these clinical images, nor does it provide details about their provenance (e.g., country of origin, retrospective or prospective collection).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided in the document. The text states that images, "When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis." However, it does not detail how ground truth was established for any specific diagnostic performance evaluation.
4. Adjudication Method for the Test Set
This information is not provided in the document.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size.
No MRMC comparative effectiveness study is mentioned in the provided text. The submission focuses on demonstrating substantial equivalence of the coils themselves, not the impact of AI assistance on human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done.
The devices described are MRI coils, which are physical hardware components used to acquire images. They are not AI algorithms. Therefore, a standalone (algorithm only) performance study is not applicable and not mentioned.
7. The Type of Ground Truth Used
The document describes the coils as producing images that, "When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis." This implies that the images are to be used in conjunction with a physician's expertise for diagnosis. For the "sample clinical images" provided, the type of ground truth is not specified. It's generally inferred that for diagnostic imaging devices, the gold standard would be expert clinical diagnosis, possibly correlated with pathology or outcomes where available, but this is not explicitly stated for these coils.
8. The Sample Size for the Training Set
The concept of a "training set" is relevant for machine learning algorithms. Since these devices are MRI coils (hardware), there is no training set in the typical AI/ML sense. The non-clinical tests relate to engineering specifications and image quality measurements, not algorithm training.
9. How the Ground Truth for the Training Set was Established
As there is no training set for these hardware devices, this question is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
Shanghai United Imaging Healthcare Co., Ltd. Xin Gao Regulatory Affairs Specialist No. 2258 Chengbei Rd., Jiading Industrial District Shanghai, 201807 Cn
February 8, 2019
Re: K183186
Trade/Device Name: Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil -4. Infant Coil - 24. Cardiac Coil - 24. Foot & Ankle Coil - 24 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: November 15, 2018 Received: November 19, 2018
Dear Xin Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K183186
Device Name
Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24
Indications for Use (Describe)
Head Coil - 12 is a 12 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Head Coil - 32 is a 32 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Carotid Coil - 8 is an 8 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for obtaining diagnostic images of carotid. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Temporomandibular Joint Coil - 4 is a 4 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for diagnostic images of the mandible region. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Infant Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for imaging of the infant's brain, spine, heart, torso and extremities. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Cardiac Coil - 24 is a 24 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The indications for use include imaging of the heart, and mediastinum regions. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Foot & Ankle Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for foot and ankle imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol, which is also in a bold, sans-serif font. The color of the text and the "U" symbol is a dark gray or blue-gray.
510 (k) SUMMARY
1. Date Prepared [21 CFR 807.92(a)(1)] November 15, 2018
2. General Information [21 CFR 807.92(a)(1)]
| Manufacturer: | Shanghai United Imaging Healthcare Co., Ltd2258 Chengbei Rd., Jiading District, Shanghai, 201807 |
|---|---|
| Contact Person: | Xin GAORegulatory Affairs Specialist |
| Tel: +86 (21) 67076888-5386 | |
| Fax: +86-021-67076889 | |
| Email: xin.gao@united-imaging.com |
3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | Head Coil - 12, Head Coil - 32, Carotid Coil - 8,Temporomandibular Joint Coil - 4, Infant Coil - 24,Cardiac Coil - 24, Foot & Ankle Coil - 24 |
|---|---|
| Common Name: | Magnetic Resonance Diagnostic Device |
| Model: | Head Coil - 12, Head Coil - 32, Carotid Coil - 8,Temporomandibular Joint Coil - 4, Infant Coil - 24,Cardiac Coil - 24, Foot & Ankle Coil - 24 |
| Product Code: | MOS |
| Regulation Number: | 21 CFR 892.1000 |
| Device Class: | II |
4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification of predicate devices within this submission are as follow:
| Predicate Device 1 | |
|---|---|
| Manufacturer: | GE Healthcare Coils (USA Instruments, Inc.) |
| Device Name: | 48CH Head Coil |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K163205 |
| Predicate Device 2 | |
| Manufacturer: | Quality Electrodynamics, LLC |
| Device Name: | 16ch Foot/ Ankle SPEEDER |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K181697 |
| Predicate Device 3 | |
| Manufacturer: | INVIVO |
| Device Name: | CARDIOVASCULAR ARRAY COILS |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K061952 |
| Predicate Device 4 | |
| Manufacturer: | Resonance Innovations LLC |
| Device Name: | Orbit and Mandible Array Family |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K133986 |
| Predicate Device 5 | |
| Manufacturer: | Shanghai Chenguang Medical Technologies Co., Ltd |
| Device Name: | Carotid Coil |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K112002 |
| Predicate Device 6 | |
| Manufacturer: | Quality Electrodynamics, LLC |
| Device Name: | Pediatric 16, A 3T Tim Coil |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K140998 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, sans-serif font. To the right of the text is a stylized icon that resembles a shield with a horizontal line through the middle. The logo is simple and modern in design.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the use of bold fonts gives it a strong and confident look.
5. Device Description [21 CFR 807.92(a)(4)]
Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 4, Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24 are receive-only multichannel phased array coils designed for use with UIH 3T MRI systems. The suffix number in coil name stands for the receive channel of the resonant nucleus is hydrogen protons (1H). All coil elements are protected from transmit field by both active and passive decoupling circuits. Each coil is designed to provide optimum image quality for its dedicated human anatomy region as defined in intended use.
6. Intended Use [21 CFR 807.92(a)(5)]
- Head Coil - 12 is a 12 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Head Coil - 32 is a 32 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Carotid Coil - 8 is an 8 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for obtaining diagnostic images of carotid. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Temporomandibular Joint Coil - 4 is a 4 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for diagnostic images of the mandible region. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Infant Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for imaging of the infant's brain, spine, heart, torso and extremities. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Cardiac Coil - 24 is a 24 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The indications for use include imaging of the heart, and mediastinum regions. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, uppercase letters, stacked on top of each other. To the right of the text is a stylized graphic that resembles the letter "U" with a horizontal line through the middle, creating a cross-like shape within the "U".
- Foot & Ankle Coil 24 is a 24 channel receive-only RF coil designed for use ● with UIH 3.0T MRI systems. The coil is indicated for use for foot and ankle imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
7. Technological Characteristic [21 CFR 807.92(a)(6)]
Head Coil - 12. Head Coil - 32. Carotid Coil - 8. Temporomandibular Joint Coil - 4. Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24 are built with wellestablished technology and design principles in the industry.
All coils are formed with linear loop coils resonant at the MR system frequency into a phased array. Each coil element is connected to an independent low noise amplifier with a low noise figure to maintain optimum SNR.
Two means of protection from RF transmit field are used for each coil element. An active decoupling circuit is controlled by a PIN diode driven by external control signal provided by the MR system. A passive decoupling circuit is also used in case the coil is left unplugged in the system bore.
The rigid coil enclosures are designed with shape matching the anatomy. Flexible flaps are also used in some coils to enhance patient comfort and adjustability. Coil materials are test to be flame and impact resistant, and biocompatible.
8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]
Summary of Non-Clinical Tests:
The following testing was conducted on the proposed devices as the predicate devices:
- A ES60601-1:2005/(R)2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- A IEC 60601-2-33 Ed. 3.1:2013, Medical Electrical Equipment - Part 2-33: Particular Requirements For The Basic Safety And Essential Performance Of Magnetic Resonance Equipment For Medical Diagnostic
- A ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- A ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- A MS 1-2008(R2014), Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Images
- A MS 3-2008(R2014), Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- A MS 6-2008(R2014), Determination of Signal-to-Noise Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging
-
MS 9-2008(R2014), Standards Publication Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized graphic element that resembles the letter "U" or a shield, with a vertical white line running through the center, creating a negative space "I" shape. The color scheme is a muted teal or gray-blue.
The test results demonstrated that the devices performs as expected and thus, it is substantially equivalent to the predicate devices to which it has been compared.
Summary of Clinical Tests:
- Sample clinical images were provided to support the ability of the proposed A devices to generate diagnostic quality images in accordance with the MR guidance on premarket notification submissions.
9. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, we conclude that the proposed devices substantially equivalent to the predicate devices. They do not introduce new indications for use, and have the same technological characteristics and do not introduce new potential hazards or safety risks.
N/A