K Number
K183162

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-12-14

(29 days)

Product Code
Regulation Number
888.3020
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Affixus Hip Fracture Nail System is intented to treat stable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal non-unions, malunions, revision procedures and tumor resections.

Device Description

Not Found

AI/ML Overview

This document is a 510(k) premarket notification from the FDA for a medical device called the "Affixus Hip Fracture Nail System." It is a regulatory clearance letter, not a study report. As such, it does not contain the detailed information about acceptance criteria and study data that you have requested for an AI/ML medical device.

The document discusses the substantial equivalence of the Affixus Hip Fracture Nail System to a previously cleared predicate device (K100238). It focuses on mechanical and material characteristics rather than AI/ML performance.

Therefore, I cannot provide the requested information based on the provided text. The document does not describe:

  1. A table of acceptance criteria and reported device performance for an AI/ML model. The acceptance criteria mentioned are related to equivalence with a predicate device, not performance metrics like sensitivity/specificity for an AI diagnosis.
  2. Sample sizes for a test set or data provenance for an AI/ML model.
  3. Number of experts and their qualifications for establishing ground truth.
  4. Adjudication method for a test set.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study results.
  6. Standalone (algorithm only) performance.
  7. Type of ground truth used (expert consensus, pathology, outcome data).
  8. Sample size for a training set.
  9. How ground truth for a training set was established.

The "Summary of Performance Data (Nonclinical and/or Clinical)" section explicitly states:

  • Non-Clinical Tests: "Study of insertion torque before and after end-of o line cleaning operations"
  • Clinical Tests: "N/A"

This confirms that the clearance was based on non-clinical engineering tests, not clinical performance data or AI/ML model evaluations.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

December 14, 2018

Biomet Inc. Dhaval Saraiya Regulatory Affairs Sr. Specialist 56 East Bell Drive Warsaw, Indiana 46582

Re: K183162

Trade/Device Name: Affixus Hip Fracture Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: November 14, 2018 Received: November 15, 2018

Dear Dhaval Saraiya:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,
Vesa VuniqiDigitally signed by Vesa Vuniqi -S
-SDate: 2018.12.1408:54:37 -05'00'
For:Mark N. Melkerson
Director
Division of Orthopedic Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510{k} Number (if known) K183162

Device Name

Affixus Hip Fracture Nail System

Indications for Use (Describe)

The Affixus Hip Fracture Nail System is intented to treat stable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal non-unions, malunions, revision procedures and tumor resections.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Affixus Hip Fracture Nail System - Nails 510(k) premarket notification.

Sponsor:Biomet Inc.56 East Bell DrivePO Box 587Warsaw, IN 46581Establishment Registration Number: 1825034
Contact Person:Dhaval SaraiyaSenior Regulatory Affairs SpecialistTelephone: (305-269-6386)Fax: (305-269-6400)
Date:14th November, 2018
Subject Device:Trade Name: Affixus Hip Fracture Nail SystemCommon Name: Intramedullary Fixation RodClassification Name:• HSB - Rod, Fixation, Intramedullary and accessories(21 CFR 888.3020)
Predicate Device(s):K100238                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        &5000;

{4}------------------------------------------------

implants are manufactured from titanium alloy.

Intended Use and Indications for Use:The Affixus Hip Fracture Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions, malunions, revision procedures and tumor resections.
Summary of Technological Characteristics:The rationale for substantial equivalence is based on consideration of the following characteristics: Intended Use: Same as the predicate deviceIndications for Use: Same as the predicate deviceMaterials: Same as the predicate deviceDesign Features: Same as the predicate deviceSterilization: Same as the predicate device
Summary of Performance Data (Nonclinical and/or Clinical)

. Non-Clinical Tests:

  • Study of insertion torque before and after end-of o line cleaning operations
  • Clinical Tests:
    • o N/A

Substantial Equivalence Conclusion

The information provided within this premarket notification demonstrates that the Affixus Hip Fracture Nail System - Nails are substantially equivalent to the predicate device.

N/A