K Number
K183156

Validate with FDA (Live)

Device Name
V-Laser
Manufacturer
Date Cleared
2019-04-24

(160 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The V-Laser laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

Device Description

The V-Laser is a Nd:YAG laser operating at wavelengths of 1,064 nm and 532 nm. The V-Laser consists of the main body, optical fiber cable, user-undetachable laser handpiece tip, footswitch, and handpiece cable cradle. The laser output is delivered to the optical fiber terminated by the handpiece. The fluence (energy density), frequency and pulse are controlled from the LCD display/Touch Pad located on the front of the main unit. The LCD display is used to obtain feedback from the system, such as the number of pulses delivered or spot size selected. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the V-Laser system. It describes the device, its intended use, and compares it to a predicate device to demonstrate substantial equivalence. Crucially, this document does not contain information about acceptance criteria, device performance metrics, or a study proving the device meets acceptance criteria in the context of clinical outcomes or diagnostic accuracy.

The "Clinical Test Summary" section explicitly states: "No clinical studies were considered necessary and performed." This means the device's performance was not evaluated through clinical trials for human use, and therefore, there are no reported device performance metrics against acceptance criteria related to clinical efficacy or safety from such studies.

Instead, the submission focuses on bench tests to verify that the device meets design specifications and complies with non-clinical standards related to electrical safety, electromagnetic compatibility, laser safety, and software validation. It also includes biocompatibility testing for materials that may come into contact with the body.

Here's an attempt to answer your questions based only on the provided document, acknowledging the significant limitations:


1. A table of acceptance criteria and the reported device performance

Based on the provided document, there are no acceptance criteria or reported device performance for clinical efficacy or diagnostic accuracy because "No clinical studies were considered necessary and performed."

The "acceptance criteria" discussed are primarily compliance with international and national standards for medical electrical equipment, electromagnetic compatibility, laser safety, and biocompatibility. The "reported device performance" is that the device "met all design specifications" and "complies with the following standards."

Acceptance Criteria (Non-Clinical Standards)Reported Device Performance
Electrical Safety & Essential Performance: Compliance with AAMI ES60601-1 (Edition 3.1, 2012)"Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with... AAMI ES60601-1"
Electromagnetic Compatibility: Compliance with IEC 60601-1-2 (Edition 3, 2007)"Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with... IEC 60601-1-2"
Particular Requirements for Laser Equipment: Compliance with IEC 60601-2-22 (Edition 3.1, 2012)"Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with... IEC 60601-2-22"
Safety of Laser Products: Compliance with IEC 60825-1 (Edition 3, 2014)"Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with... IEC 60825-1"
Software Validation: Designed and developed according to a software development process, verified and validated (for a MODERATE level of concern software), and in accordance with FDA guidance "The content of premarket submissions for software contained in medical devices, on May 11, 2005.""The V-Laser contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated. The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005."
Biocompatibility: Compliance with ISO 10993-1 and FDA Blue Book Memo #G95-1, specifically passing: Cytotoxicity (ISO 10993-5:2009), Intracutaneous (intradermal) reactivity (ISO 10993-10:2010), and Skin sensitization (ISO 10993-10:2010) tests for materials in contact with the body."According to the recommendations of ISO 10993-1 and FDA Blue Book Memo #G95-1, the following tests were performed: -Cytotoxicity test according to ISO 10993-5:2009 -Intracutaneous (intradermal) reactivity test according to ISO 10993-10:2010 -Skin sensitization test according to ISO 10993-10:2010" (Implied "passed" as part of demonstrating substantial equivalence).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable and therefore not provided, as "No clinical studies were considered necessary and performed." The "test set" for non-clinical bench testing would typically refer to the specific laser units, components, or materials tested, not a patient cohort. The document does not specify the number of units or samples subjected to each bench test.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and therefore not provided, as "No clinical studies were considered necessary and performed." Ground truth established by medical experts is relevant for studies involving human diagnosis or treatment outcomes, which were not conducted for this submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and therefore not provided, as "No clinical studies were considered necessary and performed." Adjudication methods are used in clinical trials or diagnostic accuracy studies involving human interpretation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was done. The device is a laser system for surgical and aesthetic applications, not an AI-powered diagnostic or assistive tool. "No clinical studies were considered necessary and performed."

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable to the V-Laser system, as it is a physical laser device, not a standalone AI algorithm. "No clinical studies were considered necessary and performed."

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

This information is not applicable as "No clinical studies were considered necessary and performed." For the non-clinical tests, the "ground truth" would be the specifications and requirements outlined in the various engineering and biocompatibility standards.

8. The sample size for the training set

This information is not applicable. Training sets are relevant for machine learning algorithms. The V-Laser is a hardware device. "No clinical studies were considered necessary and performed."

9. How the ground truth for the training set was established

This information is not applicable. "No clinical studies were considered necessary and performed."


Summary of the Study that Proves the Device Meets the Acceptance Criteria (Non-Clinical):

The "study" in this context refers to a series of non-clinical bench tests and validations. The document states:
"Verification and validation activities were conducted to establish the performance and safety characteristics of the V-Laser. The results of these activities show that there are no any new safety and effectiveness issues. Therefore, the V-Laser is considered substantially equivalent to the predicate."

These activities included:

  • Electrical Safety, Electromagnetic Compatibility, and Performance Bench Tests: These tests verified that the device met all design specifications and complied with AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards.
  • Software Validation: The software, identified as having a MODERATE level of concern, was designed, developed, verified, and validated according to a software development process and FDA guidance.
  • Biocompatibility Testing: Tests were performed according to ISO 10993-1 and FDA Blue Book Memo #G95-1, covering cytotoxicity, intracutaneous reactivity, and skin sensitization for materials in contact with the body.

The "proof" that the device meets these non-clinical "acceptance criteria" is the statement that the tests were conducted and the device "complies" with the listed standards and its design specifications. No detailed results, specific measurements, or pass/fail thresholds beyond the general statement of compliance are provided in this summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name is "U.S. Food & Drug Administration".

WONTECH Co., Ltd. Erin Park Staff of Regulatory Affair 64 Techno 8-Ro. Yuseong-Gu Daejeon, 305-500 KR

April 24, 2019

Re: K183156

Trade/Device Name: V-Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 9, 2018 Received: March 6, 2019

Dear Erin Park:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Digitally signed by Neil Neil R R Ogden -S Date: 2019.04.24 Oaden -S 14:51:18 -04'00' For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K183156

Image /page/2/Picture/1 description: The image shows the logo for WONTECH. The logo features a stylized letter "W" in orange and yellow gradient, followed by the word "WON" in red and "TECH" in gray. The text is in a sans-serif font.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 305-500 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491

005_Indications for Use Statement (FDA Form 3881)

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K183156

Device Name V-Lasaer

Indications for Use (Describe)

The V-Laser laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K183156

Image /page/4/Picture/1 description: The image shows the logo for WONTECH. The logo consists of an orange and yellow circular graphic on the left, followed by the text "WONTECH" in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491

510(k) Summary

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

October 30, 2018

2. Submitter's Information [21 CFR 807.92(a)(1)]

  • Name of Sponsor: ● WON TECH Co., Ltd. Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, -305-500
  • Contact Name: Erin Park / Staff of Regulatory Affair
  • Name of Manufacturer: Same as sponsor -Address: Same as sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade NameV-Laser
Device Classification NamePowered Laser Surgical Instrument
Regulation Number21CFR878.4810
Classification Product CodeGEX
Device ClassClass II
510k Review PanelGeneral & Plastic Surgery

{5}------------------------------------------------

K183156

Image /page/5/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized letter "W" in orange and yellow, followed by the text "WONTECH" in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491

4. Identification of Predicate Device [21 CFR 807.92(a)(3)[

The identified predicate device within this submission are shown as follow:

Primary Predicate

  • K153671 ● 510(k) Number:
  • Applicant: Cutera, Inc.
  • Classification Name: ● Powered Laser Surgical Instrument
  • Trade Name: Family of CoolGlide Aesthetic Lasers ●

5. Description of the Device [21 CFR 807.92(a)(4)]

The V-Laser is a Nd:YAG laser operating at wavelengths of 1,064 nm and 532 nm. The V-Laser consists of the main body, optical fiber cable, user-undetachable laser handpiece tip, footswitch, and handpiece cable cradle. The laser output is delivered to the optical fiber terminated by the handpiece. The fluence (energy density), frequency and pulse are controlled from the LCD display/Touch Pad located on the front of the main unit. The LCD display is used to obtain feedback from the system, such as the number of pulses delivered or spot size selected. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.

6. Indications for Use [21 CFR 807.92(a)(5)]

The V-Laser laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for WONTECH. The logo features a stylized letter "W" in orange and yellow gradient, followed by the text "WON" in red and "TECH" in gray. The text is in a bold, sans-serif font.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491

7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)]

There are no significant differences in the technological characteristics of this device compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the V-Laser and the predicate device:

Proposed DevicePredicate Device
K NumberK183156K153671
ManufacturerWON TECH Co., Ltd.Cutera, Inc.
ModelV-LaserFamily of CoolGlide Aesthetic Lasers
Product CodeGEXGEX
Intended Use/Indications for UseThe V-Laser laser system is intended for usein surgical and aesthetic applications in themedical specialties of dermatology andgeneral and plastic surgery.Excel V is Intended use is for surgical andaesthetic applications in the medicalspecialties of dermatology and general andplastic surgery.
Laser TypeNd:YAG LaserNd:YAG Laser
Wavelength1064 nm532 nm1064 nm532 nm
Spot Size1064 nm: 2 to 20 mm532 nm: 2 to 20 mmGenesis Mode: 8 mm1064 nm: 3 to 18 mm532 nm: 2 to 12 mmGenesis Mode: 8 mm
Output1064 nm: Max. 50 J532 nm: Max. 10 J1064 nm: Max. 100 J532 nm: Max. 12 J
Fluence1064 nm: 2 to 300 J/cm2532 nm: 1.8 to 42 J/cm2Genesis Mode: 4 to 7 J/cm21064 nm: 2 to 300 J/cm2532 nm: 1.8 to 42 J/cm2Genesis Mode: 4 to 7 J/cm2
Pulse Duration1064 nm: Max. 60 ms532 nm: Max. 40 msGenesis Mode: Max. 0.3 ms1064 nm: Max. 60 ms532 nm: Max. 40 msGenesis Mode: Max. 0.3 ms
Repetition RateMax. 10 HzMax. 10 Hz
Laser MediaFlashlamp-pumped solid state rodFlashlamp-pumped solid state rod
Aiming Beam635 nm630 to 680 nm
Dynamic Cooling Device (DCD)
CryogenHFC 134aNot applicable
DCD Spray Duration10 to 100 ms
DCD Delay DurationMax. 100 ms
DCD Post SprayDuration0 to 20 ms

The key differences between V-Laser and the predicate are (1) Spot Size and (2) Output, and which do not raise any new safety and effectiveness issues. The V-Laser and predicate device, K153671 have similar applicable fluence ranges, which are given in joules/cm2.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized "W" in orange and yellow, followed by the text "WON" in red and "TECH" in gray. The "W" is made up of curved lines that create a sense of movement.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491

Verification and validation activities were conducted to establish the performance and safety characteristics of the V-Laser. The results of these activities show that there are no any new safety and effectiveness issues. Therefore, the V-Laser is considered substantially equivalent to the predicate.

Non-Clinical Test Summary [21 CFR 807.92(b)(1)]

  1. Electrical Safety, Electromagnetic Compatibility and Performance

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

Standards No.StandardsOrganizationStandard TitleVersionPublicationYear
ES60601-1AAMIMedical Electrical Equipment - Part 1: General Requirements for BasicSafety and Essential PerformanceEdition 3.12012
60601-1-2IECMedical Electrical Equipment - Part 1-2: General Requirements forSafety - Collateral Standard: Electromagnetic Compatibility -Requirements and TestsEdition 32007
60601-2-22IECMedical Electrical Equipment - Part 2-22: Particular Requirements forBasic Safety and Essential Performance of Surgical, Cosmetic,Therapeutic and Diagnostic Laser EquipmentEdition 3.12012
60825-1IECSafety of Laser Products - Part 1: Equipment Classification, andRequirementsEdition 32014
    1. Software Validation
      The V-Laser contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.

    1. Biocompatibility
      According to the recommendations of ISO 10993-1 and FDA Blue Book Memo #G95-1, the following tests were performed:
  • -Cytotoxicity test according to ISO 10993-5:2009

  • -Intracutaneous (intradermal) reactivity test according to ISO 10993-10:2010

  • -Skin sensitization test according to ISO 10993-10:2010

Clinical Test Summary [21 CFR 807.92(b)(2)]

No clinical studies were considered necessary and performed.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized letter "W" in orange and yellow gradient, followed by the text "WONTECH" in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491

8. Conclusion [21 CFR 807.92(b)(3)]

Based on the information provided in this premarket notification WON TECH Co., Ltd., concludes that the V-Laser is substantially equivalent in safety and effectiveness to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.