K Number
K183153

Validate with FDA (Live)

Date Cleared
2019-07-08

(236 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microwave Ablation System is intended for coagulation (ablation) of soft tissue. It is not intended for use in cardiac procedures.

Device Description

The Microwave Ablation System M150E is microwave-based system intended to deliver energy through an antenna inserted into soft tissue for the purpose of coagulating (ablation) of a defined volume of that tissue. The Microwave Ablation System utilizes a 2450MHz generator to deliver power to single microwave ablation antenna or double microwave ablation antennas. The Microwave Ablation Generator provides for user setting of ablation time and ablation power. The proposed device consists of Microwave Ablation Generator, Microwave Ablation Electrode Kits, Temperature Probe and Foot Switch.

AI/ML Overview
  1. Table of Acceptance Criteria and Reported Device Performance:

The document describes the device as a Microwave Ablation System. Rather than specific performance metrics like sensitivity/specificity for a diagnostic device, the acceptance criteria for this therapeutic device are based on demonstrating equivalent technical characteristics, safety, and performance to a predicate device.

Acceptance Criteria / Performance ElementPredicate Device (Emprint™ Ablation System K133821)Proposed Device (Microwave Ablation System M150E)Result / Comparison
Indications for UseIntended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors. Not intended for use in cardiac procedures.Intended for the coagulation (ablation) of soft tissues. Not intended for use in cardiac procedures.Equivalent
Intended Purposecoagulation and ablation of tissuecoagulation and ablation of tissueEquivalent
DesignSingle channel; cooling-water cycle, thermal ablation, TEMP probe, foot switch.Double channels; cooling-water cycle, thermal ablation, TEMP probe, foot switch. (Note: Double channel is a difference, but determined not to affect safety/performance.)Similar/Different
Operating Temperature10°C-30°C5°C-40°CSimilar
Operating Humidity20%~80% non condensation20%~80% non condensationEquivalent
Output Parameters2450MHz±50MHz, 0-100W2450MHz±10MHz, 0-100WEquivalent
Output Impedance50Ω nominal50Ω nominalEquivalent
Voltage Supply100-240VAC 50-60 Hz100-240VAC 50-60 HzEquivalent
Fuse Rating6.3 A, 250V6.3 A, 250VEquivalent
Physicochemical PropertiesNot contain any medications, animal tissues or blood components, especially in contact with human tissue does not contain the above substances.Not contain any medications, animal tissues or blood components, especially in contact with human tissue does not contain the above substances.Equivalent
Working PrincipleMicrowave energy delivered via antenna to coagulate tissue; transforms into heat, raising temperature to lose bioactivity.Same working principle.Equivalent
Antenna Length (mm)CA15L1: 150; CA20L1: 200; CA30L1: 300SS-MWA-1526C: 150; SS-MWA-1531C: 150; SS-MWA-2026C: 200; SS-MWA-2031C: 200; SS-MWA-2526C: 250; SS-MWA-2531C: 250 (Note: Different lengths, but stated not to affect safety/performance in terms of depth of tumor position.)Different
Emission Area Length (mm)CA15L1: 28; CA20L1: 28; CA30L1: 28SS-MWA-1526C: 26; SS-MWA-1531C: 31; SS-MWA-2026C: 26; SS-MWA-2031C: 31; SS-MWA-2526C: 26; SS-MWA-2531C: 31 (Note: Wider range, but stated not to affect safety/performance.)Different
OD (mm)CA15L1: 2.40; CA20L1: 2.40; CA30L1: 2.40SS-MWA-1526C: 2.08; SS-MWA-1531C: 2.08; SS-MWA-2026C: 2.08; SS-MWA-2031C: 2.08; SS-MWA-2526C: 2.08; SS-MWA-2531C: 2.08 (Note: Thinner diameter to reduce trauma, determined not to affect ablation area or safety/performance.)Different
Antenna MaterialNot explicitly stated for predicate in comparison table, but implied to be similar biocompatible material.304 stainless steel, biocompatible per ISO 10993-1.Similar
Disposable/Single-useAntennas are disposable and single-use.Antennas are disposable and single-use.Equivalent
SterilityAccessories sterilized with EO (SAL: 10-6).Accessories sterilized with EO (SAL: 10-6).Equivalent
BiocompatibilityPatient-contacting materials are biocompatible.Patient-contacting materials are biocompatible.Equivalent
Device Temperature MonitoringTemperature monitoring features used to ensure system safety.Temperature monitoring features used to ensure system safety.Equivalent
Device CoolingPumped normal saline is used to cool the Emprint™ Antenna.Pumped normal saline is used to cool the antenna.Equivalent
  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable. The submission states, "Clinical studies were not required to demonstrate the substantial equivalence of the microwave ablation system and the predicated device." The testing primarily involved non-clinical bench and lab testing (e.g., electrical safety, EMC, performance, thermal effects, temperature monitoring, shelf life, package verification, sterilization validation, biocompatibility).
    • Data Provenance: Not applicable, as no clinical test set was used. The non-clinical testing data provenance is implied to be from the manufacturer's internal testing.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable, as no clinical test set was used that required expert ground truth establishment in this type of 510(k) submission.
  3. Adjudication method for the test set:

    • Not applicable, as no clinical test set was used for adjudication.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a therapeutic microwave ablation system, not a diagnostic imaging AI device that would involve human readers or AI assistance for interpretation.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable in the context of diagnostic AI performance. The device is a physically operated medical system. However, software validation was performed, indicating the internal algorithms and software governing the device's operation were tested in a standalone manner (without human intervention in the algorithm's execution during ablation).
  6. The type of ground truth used:

    • For the non-clinical tests, the "ground truth" was established by adherence to recognized national and international standards (e.g., IEC 60601-1, IEC 60601-2-6, IEC 60601-1-2, ISO 11607-1, ISO 11135, ISO 10993), and engineering specifications for system performance, thermal effects, and temperature monitoring. These standards and specifications themselves represent a form of "ground truth" for acceptable device characteristics and safety.
  7. The sample size for the training set:

    • Not applicable, as this device does not involve machine learning or AI models that require a training set in the conventional sense. The "training" here refers to the device's design, development, and adherence to established engineering principles and standards.
  8. How the ground truth for the training set was established:

    • Not applicable, as no machine learning training set was used. The "ground truth" for the device's design and operational parameters was derived from existing medical device standards, predicate device characteristics, and established engineering and physiological principles related to microwave ablation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 8, 2019

Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. Guofang Ma Quality Director No.1 XinXing Yilu Road, Emerging Industrial Cluster Area Zonghan Subdistrict, Cixi City, Zhejiang, China 315300

Re: K183153

Trade/Device Name: Microwave Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: NEY Dated: November 2, 2018 Received: November 14, 2018

Dear Guofang Ma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183153

Device Name Microwave Ablation System

Indications for Use (Describe)

The Microwave Ablation System is intended for coagulation (ablation) of soft tissue. It is not intended for use in cardiac procedures.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section III 510(k) Summary

This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.

There is no prior submission for the device.

3.1 Submitter Information

  • 510(k) Submitter/Holder: Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. No.1 Xinxing Yilu Road, Emerging Industrial Cluster Area, Zonghan Subdistrict, Cixi City, Zhejiang, China

. Contact

Guofang Ma Quality Director Telephone: +86-10-64116115 Fax: +86-10-64113261 Email: QARA@surgnova.com

● Date Prepared: August 02, 2018

3.2 Proposed Device Information

Device Common Name: Microwave ablation system and accessories Device Trade/Proprietary Name: Microwave Ablation System Model: M150E Classification Name: Electrosurgical cutting and coagulation device and accessories Regulation Number: 878.4400 Product Code: NEY Class: Il Panel: General & Plastic Surgery

3.3 Predicate Device

510(k) Number: K133821 Common Name: Microwave ablation system and accessories Device Trade/Proprietary Name: Emprint™ Ablation System Classification Name: Electrosurgical cutting and coagulation device and accessories Regulation Number: 878.4400 Product Code: NEY Class: II Panel: General & Plastic Surgery Manufacturer: Covidien LLC

{4}------------------------------------------------

3.4 Device Description

The Microwave Ablation System M150E is microwave-based system intended to deliver energy through an antenna inserted into soft tissue for the purpose of coagulating (ablating) a defined volume of that tissue. The Microwave Ablation System utilizes a 2450MHz generator to deliver power to single microwave ablation antenna or double microwave ablation antennas. The Microwave Ablation Generator provides for user setting of ablation time and ablation power.

The proposed device consists of Microwave Ablation Generator, Microwave Ablation Electrode Kits, Temperature Probe and Foot Switch. Wherein, The Microwave Ablation Electrode Kits consists of Microwave Ablation Antenna, water pipe and Coax cable, to be used for puncturing the patient's lesion position during operation and also for outputting microwave energy to ablate the tumor tissue through connection with Microwave Ablation Generator. The Temperature Probe is applied to monitor the temperature of the target location and to protect important organs and tissues in the periphery of the lesion from unexpected damage by microwave energy.

It supports both power and temperature modes.

Power mode

Under the power mode, at the beginning, the system output microwave power according manually setting. When the tissue temperature reaches the target temperature, the system adjusts the duty cycle in real time according to the specific algorithm and load power to keep the tissue temperature on the target temperature. Two microwave ablation antennas can be used for microwave ablation.

Temperature mode

Under the temperature mode, at the beginning, the system output microwave power according manually setting. When the tissue temperature reaches the target temperature, the system adjusts the output of microwave power in real time according to the specific algorithm to keep the tissue temperature on the target temperature. Two microwave ablation antennas can be used for microwave ablation.

The Microwave Ablation Generator is non-sterile, while Microwave Ablation Electrode Kits and Temperature Probe are sterile.

The device is software-driven and the software validation is provided in Section of Software.

{5}------------------------------------------------

3.5 Comparison list of the technological characteristics

Based on the comparison of the clinical and ablation antennas of the predicate device—Emprint™ ablation system, there are no significant differences between the material, the key technology, the design characteristics, the applicable population, the site of action, etc. Confirming that both devices are equivalent in terms of clinical performance and safety. The specific results are as follows:

Comparison ElementsPredicate Device(K133821)Proposed DeviceComparison(Equivalent,Similar,Different)
Product NameEmprint™ Ablation SystemMicrowave Ablation System/
Regulation No.21 CFR 878.440021 CFR 878.4400Equivalent
ClassificationIIIIEquivalent
Classification NameElectrosurgical cutting and coagulation device and accessoriesElectrosurgical cutting and coagulation device and accessoriesEquivalent
Product CodeNEYNEYEquivalent
ClinicalIndications for UseThe Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors.The Emprint™ Ablation System is notThe Microwave Ablation System is intended for the coagulation (ablation) of soft tissues.The Microwave Ablation System is not intended for use in cardiac procedures.Equivalent

{6}------------------------------------------------

intended for use in cardiac procedures.
Intended purposecoagulation and ablation of tissuecoagulation and ablation of tissueEquivalent
Comparison StatementThe proposed device has the same indications for use and classification as the predicate device.
TechnicalDesignSingle channelDouble channelsDifferent
cooling-water cycle, thermal ablation,the probe of TEMP, foot switchcooling-water cycle, thermal ablation,the probe of TEMP, foot switchEquivalent
ComparisonStatementCompared with the predicate device, the proposed device has double channel to output energy. The twochannel not only can output energy at the same time but also can work respectively.
Operatingtemperature10°C-30°C5°C-40°CSimilar
Operatinghumidity20%~80% non condensation20%~80% non condensationEquivalent
ComparisonStatementThe proposed device operating environment temperature is wider than that of the equivalent product. Wehave confirmed through environmental tests that this temperature range does not affect the product'ssafety and performance.
OutputParameters2450MHz±50MHz,0-100W2450MHz±10MHz,0-100WEquivalent
Output Impedance50Ω nominal50Ω nominalEquivalent

{7}------------------------------------------------

Premarket Notification 510(k) Submission—K183153
--------------------------------------------------------
Voltage Supply100-240VAC 50-60 Hz100-240VAC 50-60 HzEquivalent
Fuse Rating6.3 A, 250V6.3 A, 250VEquivalent
PhysicochemicalPropertiesNot contain any medications, animaltissues or blood components, especiallyin contact with human tissue does notcontain the above substances.Not contain any medications, animal tissuesor blood components, especially in contactwith human tissue does not contain theabove substances.Equivalent
Working PrincipleDuring the surgery, the microwaveablation antenna is accurately placed inthe tumor target area by imagingtechniques (such as CT, US, etc.). Themicrowave energy generated by themicrowave generator transmits to themicrowave ablation antenna through thecoaxial cable, and then it is radiated outthrough the microwave antenna andabsorbed by water molecules in thetumor tissue. The microwave energytransform into heat, and thetemperature rises rapidly result in tumortissue losing bioactivity.During the surgery, the microwave ablationantenna is accurately placed in the tumortarget area by imaging techniques (such asCT, US, etc.). The microwave energygenerated by the microwave generatortransmits to the microwave ablationantenna through the coaxial cable, and thenit is radiated out through the microwaveantenna and absorbed by water molecules inthe tumor tissue. The microwave energytransform into heat, and the temperaturerises rapidly result in tumor tissue losingbioactivity.Equivalent
Comparison StatementThe proposed device and predicated device have the same working principle and similar technicalparameter.
Disposableaccessories-Antenna Length(mm)CA15L1: 150SS-MWA-1526C: 150SS-MWA-1531C: 150Different
AblationAntennasCA20L1: 200SS-MWA-2026C: 200
SS-MWA-2031C: 200
CA30L1: 300SS-MWA-2526C: 250
SS-MWA-2531C: 250
ComparisonStatementThe ablation antenna length of the proposed device is different from that of the predicate device, butthis difference is only reflected in the depth of the position of the tumor to be ablated and does notaffect the product's safety and performance.
Emission arealength (mm)CA15L1: 28SS-MWA-1526C: 26Different
CA20L1: 28SS-MWA-1531C: 31
SS-MWA-2026C: 26
CA30L1: 28SS-MWA-2031C: 31
SS-MWA-2526C: 26
SS-MWA-2531C: 31
ComparisonStatementThe range of microwave emission area of the proposed device is wider than that of the predicate device,We have proved through performance tests that this difference does not affect the product's safety andperformance.
OD(mm)CA15L1: 2.40SS-MWA-1526C: 2.08Different
SS-MWA-1531C: 2.08
CA20L1: 2.40SS-MWA-2026C: 2.08
CA30L1: 2.40SS-MWA-2031C: 2.08
SS-MWA-2526C: 2.08
SS-MWA-2531C: 2.08
ComparisonStatementBased on not affecting the area of the ablation, the diameter of the microwave antenna is thinner, sothat the trauma caused by the puncture is smaller.
ComparisonStatementThe 304 stainless steel usually has excellent corrosion resistance, good weldability and formability, goodresistance to hydrogen embrittlement, in addition to high ductility and toughness. The material of theproposed device has passed the Bicompatibility test according to ISO 10993-1.
Disposable /Single-use DeviceThe antennas are disposable and are tobe used within a single patientprocedure only.The antennas are disposable and are to beused within a single patient procedure only.Equivalent
SterilityThe accessories are sterilized withEO(SAL: 10-6)The accessories are sterilized with EO(SAL:10-6)Equivalent
BiocompatibilityPatient-contacting materials arebiocompatible.Patient-contacting materials arebiocompatible.Equivalent
DeviceTemperatureMonitoringTemperature monitoring features usedto ensure system safetyTemperature monitoring features used toensure system safetyEquivalent
Device coolingPumped, normal saline is used to coolthe Emprint™ Antenna.Pumped, normal saline is used to cool theantenna.Equivalent
Comparison StatementThe proposed device and predicated device have the similar disposable accessories-Ablation Antennas andsame sterilization method.

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

3.6 Indications for use

The Microwave Ablation System is intended for the coagulation (ablation) of soft tissues. The MW Ablation System is not intended for use in cardiac procedures.

3.7 Testing

Non-Clinical Testing

The Microwave Ablation System and Accessories and the predicate device are substantially equivalent in design concepts, technologies and materials. The Microwave Ablation System and Accessories has been designed and will be manufactured in accordance with both mandatory and voluntary standards, including:

• IEC 60601-1: 2005/A1:2012 Medical Electrical Equipment-Part 1: General requirements for safety.

• IEC 60601-2-6:2012 Medical electrical equipment –Part 2-6: Particular requirements for the basic safety and essential performance of microwave therapy equipment.

• IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

•ISO 11607-1:2016 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.

• ISO 11135: 2014 Sterilization of health care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices

The Software Validation is in compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices".

The shelf life of Microwave Ablation Electrode Kits and Temperature Probe is 2 years.

No.Test Name
1Electrical Safety Test According to IEC 60601-1
2Electromagnetic Compatibility Test According to IEC 60601-1-2

The list of non-clinical test performed on the proposed device.

{11}------------------------------------------------

3Performance Test according to IEC 60601-2-6
4System Performance Test
5Thermal Effects test according to FDA Guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for General Surgery
6Temperature Monitoring test according to FDA Guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for General Surgery
7Shelf Life Test
8Package Verification Test according to ISO 11607-1
9Sterilization validation according to ISO 11135
10Irritation, Sensitization, Cytotoxicity, Pyrogenicity, Acute Systemic Toxicity Test according to ISO 10993

Clinical Testing

Clinical studies were not required to demonstrate the substantial equivalence of the microwave ablation system and the predicated device.

3.8 Determination of substantial equivalence

The proposed device is equivalent with respect to the basic system design and function to that of the predicate device. The proposed device isn't the implants and high-risk device. And it doesn't have new intended purposes, new medical, new target populations, and new users and so on. What's more, it can't use the medicinal substances or animal tissues. So differences between the predicate and proposed device do not raise new questions of safety or effectiveness.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.