K Number
K183146

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-04-25

(163 days)

Product Code
Regulation Number
876.1075
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cartilage Biopsy Needle is intended to obtain a sample of articular cartilage tissue during arthroscopy.

Device Description

The Ranfac Cartilage Biopsy Needle is a manual, sterile disposable needle intended to obtain a sample of cartilage during arthroscopy. The device is comprised of an outer cannula with handle and an inner stylet. The product is provided with probe guide and probe to assist with the extraction of the cartilage from the needle. The probe guide facilitates the insertion of the probe through the distal end of the needle. The probe is introduced into the probe guide at the distal tip of the needle and advanced forward to move and expel the specimen out through the needle handle. This premarket notification is for both an 8-gauge and 11-gauge needle (product catalog numbers: CBN-84 and CBN-114, respectively).

AI/ML Overview

The provided document is a 510(k) Summary for the Ranfac Cartilage Biopsy Needle. It includes some performance data (bench testing) but does not describe acceptance criteria, a specific study proving it meets those criteria, or details regarding sample sizes for training/test sets, expert ground truth establishment, or MRMC studies. The document indicates "No clinical studies were conducted for this submission."

Here's an analysis of the available information based on your request:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria or detailed performance metrics. It provides qualitative statements about bench testing results:

Acceptance Criteria (Not explicitly stated, inferred from text)Reported Device Performance
Structural integrity (tensile/torque)All samples met or exceeded acceptance criteria.
Suitability for acquisition of cartilage biopsyDemonstrated suitability.
Maintenance of sample integrity (histological analysis)Demonstrated maintenance of sample integrity.

2. Sample size used for the test set and the data provenance

The document does not specify a sample size for the bench testing beyond "All samples." It states that articular cartilage was harvested from "ex vivo bovine knees." This indicates the data provenance is ex-vivo animal tissue (bovine), and the nature of the test is prospective in the sense that the testing was conducted for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document mentions "histological analysis" for evaluating sample integrity but does not specify the number or qualifications of experts involved in this analysis for the bench testing.

4. Adjudication method for the test set

The document does not describe any adjudication method for the bench testing results.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. The device described is a physical biopsy needle, not an AI-powered diagnostic tool. The document explicitly states: "No clinical studies were conducted for this submission."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical biopsy needle. There is no algorithm or AI component.

7. The type of ground truth used

For the bench testing, the ground truth for "sample integrity" was established via histological analysis of the cartilage samples acquired.

8. The sample size for the training set

Not applicable. The device is a physical biopsy needle, not a machine learning model. There is no training set mentioned or implied.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth for a training set to be established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

April 25, 2019

Ranfac Corporation Eric Kreuz Director of Quality Assurance/Regulatory Affairs 30 Doherty Ave Avon, Massachusetts 02322

Re: K183146

Trade/Device Name: Ranfac Cartilage Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW, FCG Dated: November 8, 2018 Received: November 13, 2018

Dear Eric Kreuz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows a digital signature. The signature is from Long H. Chen -S. The date of the signature is 2019.04.25 at 06:21:18, with a time zone of -04'00'. The signature is for an unspecified purpose.

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183146

Device Name Ranfac Cartilage Biopsy Needle

Indications for Use (Describe)

The Cartilage Biopsy Needle is intended to obtain a sample of articular cartilage tissue during arthroscopy.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Page 1 of 4

510(k) Summary

I. SUBMITTER

Sponsor:Ranfac Corp.30 Doherty AveAvon, MA
Contact Person:Eric Kreuz
Date:March 23, 2019Ph: 1-508-588-4400, Ext. 137Fax: 1-508-584-8588

II. DEVICE

Device Trade Name:Ranfac Cartilage Biopsy Needle
Common or Usual Name:Biopsy Needle
Device Classification:Class II
Classification Name:Instrument, Biopsy
Regulation:876.1075
Device Regulation Panel:Gastroenterology/Urology
Device Product Code:KNW & FCG

III. PREDICATE DEVICES

Ranfac Bone Marrow Aspiration Needle (K131157). This device has not been subject to a design-related recall.

A reference predicate for this device is the CareFusion Jamshidi-Type Bone Marrow Biopsy/Aspiration Needle (K171531).

IV. DEVICE DESCRIPTION

The Ranfac Cartilage Biopsy Needle is a manual, sterile disposable needle intended to obtain a sample of cartilage during arthroscopy. The device is comprised of an outer cannula with handle and an inner stylet. The product is provided with probe guide and probe to assist with the extraction of the cartilage from the needle. The probe guide facilitates the insertion of the probe through the distal end of the needle. The probe is introduced into the probe guide at the distal tip of the needle and advanced forward to move and expel the specimen out through the needle handle. This premarket notification is for both an 8-gauge and 11-gauge needle (product catalog numbers: CBN-84 and CBN-114, respectively).

{4}------------------------------------------------

Page 2 of 4

INDICATIONS FOR USE V.

The Cartilage Biopsy Needle is intended to obtain a sample of articular cartilage tissue during arthroscopy.

The Indications for Use statement for the Cartilage Biopsy Needle is not identical to the predicate device which is used for obtaining a sample of bone marrow; however, the differences do not alter the high-level intended use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices are used to harvest a tissue specimen near bone.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VI. THE PREDICATE DEVICE

The Ranfac Cartilage Biopsy Needles ("CBN)" are similar to the predicate Ranfac Bone Marrow Aspiration Needle ("BMA Needle" - K131157) in intended use (i.e., acquisition of tissue), fundamental scientific technology, operating principles, fundamental mechanical design, performance characteristics, packaging, sterility, shelf-life and biocompatibility. The proposed and predicate devices share a similar design (needle cannula and stylet) and are provided sterile for single use, have the same materials, and operate in the Ranfac CBN with the primary predicate Ranfac BMA Needle is provided in Table 5-1. This table details the closely shared indications for use, materials and design and principle of operation therefore establishing substantial equivalence of the proposed device with the presently commercialized predicate product.

Ranfac Cartilage BiopsyNeedle (This Submission)Ranfac Bone Marrow AspirationNeedle (K131157)
RegulationNumber21 CFR §876.1075Same
Intended UseIndication for UseFor harvest of a tissue sample/tissue biopsyThe Ranfac Cartilage Biopsy Needle isintended to obtain a sample of articularcartilage tissue during arthroscopy.SameThe Ranfac Bone Marrow AspirationNeedle is intended for use inaspirating bone marrow
Overall ProductDesignSingle-use, sterile disposable needle toacquire tissue specimen. The device iscomprised of an outer cannula with handleand an inner stylet. The device is providedin two gauge sizes; i.e., 8 gauge (CBN-84)and 11 gauge (CBN-114).Single-use, sterile disposable needle toacquire tissue specimen. The device iscomprised of an outer cannula withhandle and an inner stylet. The device isan 11-gauge needle.
Mechanics ofOperationManual instrumentManual instrument
Patient/TissueContact MaterialsStainless steel and plasticStainless steel and plastic
Gauge x Length8 and 11-Gauge x 4 inches11 Gauge x 4 inches

Table 5-1. Comparison of the Proposed Ranfac Cartilage Biopsy Needle to the Primary Predicate Ranfac Bone Marrow Aspiration Needle

{5}------------------------------------------------

Page 3 of 4

Ranfac Cartilage Biopsy Needle(This Submission)Ranfac Bone Marrow AspirationNeedle (K131157)
CannulaConfiguration304 stainless steel hollow cannula with no sideports304 stainless steel hollow cannulawith side ports
StyletConfigurationSolid stainless-steel wireSame
HandleConfigurationMating plastic handle on cannula and styletSame
Sample ExtractionAccessoriesProbe guide and probe provided to expel tissuespecimen from needle cannulaNone: tissue is aspirated throughneedle using standard hypodermicsyringe.
PackagingTyvek/Mylar PouchSame
SterilizationSupplied Sterile via Ethylene Oxide validated to10-6 Sterility Assurance LevelSame
Shelf-Life5 years5 years
Table 5-1. Comparison of the Proposed Ranfac Cartilage Biopsy Needle to the Primary
Predicate Ranfac Bone Marrow Aspiration Needle

The differences between the proposed and predicate devices is that the Ranfac CBN is intended for acquisition of cartilage; whereas, the Ranfac BMA needle is used to acquire bone marrow, both bone marrow aspirate and cartilage can be characterized as soft tissues and in both cases the needles must penetrate bone for purposes of acquiring the target tissue. The Ranfac BMA Needle is an 11-gauge needle and includes side-holes for aspiration of the bone marrow; whereas, the Ranfac CBN is provided in an 8 and 11-gauge configuration and without side holes. The Ranfac CBN is provided with a probe to facilitate tissue acquisition. These differences are not unique to the Ranfac CBN. The Jamshidi-type Bone Marrow Biopsy/Aspiration Needles (reference K171531) classified under 21 CFR §876.1075 (procode KNW) is provided in both an 8 gauge and 11-gauge configuration (no side ports on needle cannula), and provided with a probe for expelling the tissue sample. Both the Ranfac CBN and the reference Jamshidi-type needle acquire a solid core of tissue (trephine biopsy) using similar methods. The general procedures and guidelines in the proposed Ranfac CBN device's labeling provide instructions for cartilage biopsy sample collection.

In summary, the proposed changes were appropriately assessed and do not raise any new or significant questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate devices presented.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

{6}------------------------------------------------

Page 4 of 4

Biocompatibility Testing:

The biocompatibility evaluation for the Ranfac CBN was conducted in accordance with FDA's Guidance "Use of International Standard ISO-10993-1, "Biological evaluation of medical devices Part 1: Evaluation and Testing Within a Risk Management Process," as published by FDA on June 16, 2016. The battery of testing included the following tests with passing results for all:

  • Cytotoxicity
  • Sensitization
  • Intracutaneous Reactivity ●
  • Acute Systemic toxicity
  • Material-Mediated Pyrogenicity ●

The Ranfac CBN is considered tissue contacting for a duration of less than 24 hours.

Bench Testing:

Structural integrity testing (tensile/torque) was conducted on the Ranfac CBN demonstrating robustness of the design. All samples met or exceeded acceptance criteria. Additionally, testing was performed to demonstrate that the Ranfac CBN was suitably designed for its intended purpose; i.e., acquisition of a cartilage biopsy, while maintaining the integrity of the sample as evaluated via histological analysis in a model in which articular cartilage was harvested from ex vivo bovine knees.

Clinical Studies:

No clinical studies were conducted for this submission.

VIII. CONCLUSION

Performance testing and comparison of characteristics between the subject and predicate devices have demonstrated that the Ranfac CBN is substantially equivalent to the primary and reference predicate devices with regard to intended use (biopsy for harvest of a tissue specimen), operation, function, and technological characteristics.

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.