K Number
K183094

Validate with FDA (Live)

Date Cleared
2019-04-26

(170 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CyclaPlex Correction System is intended for the following surgical indications: to assist in the correction of Hallux Valgus deformity by providing reduction of the 1st intermetatarsal angle.

Device Description

The CyclaPlex Correction Systemis comprised of the Implant, Accessories and Instrument Set. The Implant device pulls the 1ª metatars al inward to wards 2ª metatars al and reduces the internetatarsal angle to medically accepted values. The implant is comprised of two metallic bone screws with internal polymeric housing connected by UHMW polyethylene multiple sutures cord. The cord is looped at Anchor 2 and secured to Anchor 1 by multiple surgical knots. Tw implants (paired) are implanted to reduce local loads on the bone and reduce loads on the device as well.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the CyclaPlex Correction System, a medical device intended to assist in the correction of Hallux Valgus deformity.

Here's an analysis of the provided information regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state numerical acceptance criteria in a table format. Instead, it uses qualitative statements to demonstrate equivalence to the predicate device. The performance is consistently stated as "at least as good as the predicate device" or "same as the predicate device."

Acceptance Criteria (Implied)Reported Device Performance
Mechanical static resistance to biomechanical forcesCan withstand biomechanical forces at least as well as the predicate device
Mechanical fatigue resistance to biomechanical forcesCan withstand biomechanical forces at least as well as the predicate device
Device performance in preclinical animal modelsAt least as good as predicate device in minipig animals
Surgical instrument safety and usabilityProvides safe use at least as good as the predicate device
Suitability of implant structure, surgery procedure, and instruments for implantationDemonstrated suitable for intended implantation in cadaver tests
Biological compatibility of materials post-manufacturing and cleaningMeets ISO 10993-1 requirements
Biological safety assessment (further)Demonstrated through animal testing
Device implantation procedure and surgical instrument usabilityDemonstrated through animal testing
Packaging validation conformanceMet requirements
Sterilization validation conformanceMet requirements
Cleaning validation conformanceMet requirements
MR compatibility (Magnetic field interactions, MR artifacts, MR heating)Demonstrated to be "MR conditional" (per ASTM F2119-07, F2052-15, F2182-11a, F2213-06)
Raises no new safety or effectiveness questions compared to predicateA detailed analysis and testing demonstrated this

2. Sample size used for the test set and the data provenance:

  • Mechanical Testing: Not explicitly stated. The document mentions "The mechanical static and fatigue testing were performed."
  • Preclinical Animal Tests: Not explicitly stated. The document mentions "Preclinical animal tests showed that CyclaPlex device performance in minipig animals..."
  • Cadaver Tests: Not explicitly stated. The document mentions "Cadaver tests further demonstrated..."
  • Biological Evaluation: Not explicitly stated for tests demonstrating ISO 10993-1 compliance. "Animal testing further extended the biological safety assessment."
  • MR Testing: Not explicitly stated.

Data Provenance: The studies are described as "Preclinical animal tests" and "Cadaver tests," suggesting they were conducted specifically for this device submission. The location of these studies (country of origin) is not specified, but the applicant's address is Israel. This would be considered prospective data for the device's evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This type of information is generally not applicable to the non-clinical (mechanical, animal, cadaver, biological, MR) testing performed for this device. The "ground truth" here is objective physical or biological performance measurements, not expert interpretation of images or patient outcomes.

For the Surgical instrument human factor analysis, the document states "Analysis had demonstrated that instruments provide safe use at least as good as the predicate device." It does not specify the number or qualifications of experts involved in this human factors analysis.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. Adjudication methods like 2+1 or 3+1 typically refer to independent review of clinical data (e.g., imaging studies) by multiple experts to establish a consensus ground truth. The studies described are non-clinical, objective tests.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. An MRMC study is a clinical study design used to evaluate the diagnostic performance of a system (often AI-assisted) by multiple human readers across multiple cases. This submission describes a mechanical, biological, and animal testing evaluation of a surgical implant, not an AI-based diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

No. This device is a physical surgical implant, not an algorithm or AI. Therefore, the concept of standalone algorithm performance is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the various tests appears to be based on:

  • Mechanical Testing: Established engineering standards and measurements of force, displacement, and material properties. The comparison is against the predicate device's measured performance.
  • Biological Evaluation: Adherence to international standards (ISO 10993-1) for biocompatibility, and observed biological responses in animal models.
  • Animal & Cadaver Studies: Direct observation of device performance, surgical procedure compatibility, and lack of adverse events or structural failure under simulated surgical conditions.
  • MR Testing: Compliance with specific ASTM standards.

8. The sample size for the training set:

Not applicable. This device is a physical product and not an AI/machine learning algorithm, so there is no "training set."

9. How the ground truth for the training set was established:

Not applicable. As there is no training set for a physical implant, there is no ground truth to be established for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Cycla Orthopedics Ltd % Elaine Duncan President Paladin Medical PO Box 560 Stilwater, Minnesota 55082-0560 April 26, 2019

Re: K183094

Trade/Device Name: CyclaPlex Correction System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: March 22, 2019 Received: March 26, 2019

Dear Elaine Duncan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Daniel S. Ramsey -S 2019.04.26 12:05:54 -04'00'

FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183094

Device Name

CyclaPlex Correction System

Indications for Use (Describe)

The CyclaPlex Correction System is intended for the following surgical indications: to assist in the correction of Hallux Valgus deformity by providing reduction of the 1st intermetatarsal angle.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY- TRADITIONAL-

SUBMITTED on behalf of:

Cycla Orthopedics Ltd. 14 HameyasdimSt. Bat Shlomo, 30992, Israel. Phone 972-4-6443544 Fax 972 3-7256211 Email: info@cycla-or.com

By:Elaine Duncan, M.S.M.E., RAC
President, Paladin Medical, Inc.
PO Box 560
Stillwater, MN 55082
Telephone:715-549-6035
Fax:715-549-5380
CONTACT PERSON:Elaine Duncan
DATE PREPARED:October 30, 2018
TRADE NAME:CyclaPlex Correction System
COMMON NAMEButton/Anchor/suture
CLASSIFICATION NAME:Single/multiple component metallic bone fixation applianceand accessories:
CLASSCLASS II
REGULATION/PROCODE21 CFR 888.3030/ PRO CODE: HTN

INDICATIONS FOR USE:

The CyclaPlex Correction Systemis intended for the following surgical indications: to assist in the correction of Hallux Valgus deformity by providing reduction of the 1st intermetatarsal angle.

DESCRIPTION of the DEVICE:

The CyclaPlex Correction Systemis comprised of the Implant, Accessories and Instrument Set. The Implant device pulls the 1ª metatars al inward to wards 2ª metatars al and reduces the internetatarsal angle to medically accepted values. The implant is comprised of two metallic bone screws with internal polymeric housing connected by UHMW polyethylene multiple sutures cord. The cord is looped at Anchor 2 and secured to Anchor 1 by multiple surgical knots. Tw implants (paired) are implanted to reduce local loads on the bone and reduce loads on the device as well.

SUBSTANTIALLY EQUIVALENT TO:

The CyclaPlex Correction Systemis substantially equivalent to Instratek-Michelangelo Bunion System (MAB). Premarket notification: K 091763. MAB is comprised of two buttons/plates connected by UHMW polyethylene cord. The CyclaPlex Correction Systemis the same as the predicate device in areas of classification, anatomical site, where used, principle of operation, intended use, contraindications and material: both utilize UHM WPE suture and two metallic end caps, both intended for use in the same anatomical location. CyclaPlexhas slight differences in strumentation and sterilization. Additional differences include the end cap geometry, a more modular design with a polymeric insert. Also, the CyclaPlex requires two implants whereas the predicate allows various configurations (one or more).

{4}------------------------------------------------

510(k) Summary-Continued

A detailed analysis and testing have demonstrated that the CyclaPlexCorrection Systemhas the same raises no new safety or effectiveness questions.

SUMMARY of TESTING to DEMONSTRATE SAFETY and EQUIVALENCE:

The mechanical static and fatigue testing were performed to demonstrate that the proposed device can withstand the biomechanical forces, at least as well as the predicate device. Preclinical animal tests showed that CyclaPlex device performance in minipig animals is at least as good as predicate device Surgical instrument human factor analysis was performed to evaluate the safety and usability of surgical instruments in comparison to predicatedevice, Analysis had demonstrated that instruments provide safe use at least as good as the predicate device.

Cadaver tests further demonstrated that the implant structure, surgery procedure and instruments are suitable to perform implantation as intended. Biological evaluation demonstrated that proposed device materials post manufacturing and cleaning meet ISO 10993-1 requirements. Animal testing further extended the biological safety as sessment as well as demonstrated the device implantation procedure and surgical instrument us ability. Packaging, sterilization and cleaning validation were also conducted for implant and met requirements. MR were performed according ASTM relevant standards (F2119-07, F2052-15, F2182-11a, F2213-06) and included Magnetic field interactions, MR artifacts and MR heating, demonstrating that CyclaPleximplant device is "MR conditional."

CONCLUSION:

Based upon the comparative analysis and testing the CyclaPlex Correction Systemis substantially equivalent to the currently marketed predicate device.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.