K Number
K183037

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-12-04

(33 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.

The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for:

  • contouring
  • image manipulation
  • simulation
  • image fusion
  • plan optimization
  • QA and plan review
Device Description

Monaco is a radiation treatment planning system that first received FDA clearance in 2007 (K071938). The modified system received clearance in 2009, when Volumetric Modulated Arc Therapy (VMAT) planning capability was added (K091179), again when Dynamic Conformal Arc planning was added (K110730), and electron planning, support for stereotactic cones, and SUV calculation were added (K132971). Finally, specialty image creation was added in 2015 (K15123). A 510(k) was filed in 2017 for the addition of carbon ion planning. The 510(k) was withdrawn because there was no hardware cleared for the US market capable of delivering carbon ion plans. Monaco's carbon ion planning functionality remains licensed off and inaccessible to US users.

The Monaco system accepts patient imaging data and "source" dosimetry data from a linear accelerator. The system then permits the user to display and define (contour) the target volume to be treated and critical structures which must not receive above a certain level of radiation on these diagnostic images.

Based on the prescribed dose, the user, a Dosimetrist or Medical Physicist, can create multiple treatment scenarios involving the number, position(s) and energy of radiation beams and the use of a beam modifier (MLC, block, etc.) between the source of radiation and the patient to shape the beam. The Monaco system then produces a display of radiation dose distribution within the patient, indicating doses to the target volume and surrounding structures. The "best" plan satisfying the clinican prescription is then selected, one that maximizes dose to the target volume while minimizing dose to surrounding healthy volumes.

AI/ML Overview

The Monaco RTP System is a radiation treatment planning system. The provided text indicates that no clinical trials were performed. Instead, validation testing involved simulated clinical workflows using actual patient data, and algorithm testing verified the accuracy of the new dose calculation algorithm.

1. Table of acceptance criteria and reported device performance:

Acceptance Criteria (Pre-defined pass/fail criteria)Reported Device Performance (Summary of Non-Clinical Testing)
Equivalent to predicate (K151233)Demonstrated equivalent accuracy to the predicate device.
Conformity to the same pass/fail criteria as the predicate (K151233)Monaco 5.40 was deemed substantially equivalent in safety and effectiveness.

2. Sample size used for the test set and data provenance:

  • Test Set Sample Size: The document mentions "actual patient data" was used for simulated clinical workflows and algorithm testing, but it does not specify the sample size of this patient data.
  • Data Provenance: The document does not specify the country of origin. It indicates the data was "actual patient data," but does not state whether it was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and their qualifications:

  • The document states that "qualified clinicians" review and approve treatment plans and that the process "can be performed such that no human subjects are exposed to risk." This implies that the validation testing was based on expert consensus or established clinical standards rather than live patient outcomes. However, the number of experts and their specific qualifications are not specified.

4. Adjudication method for the test set:

  • The document mentions that treatment plans are "reviewed and approved by qualified clinicians," and that a "flaw in the treatment plan [could] escape the notice of the qualified professionals." This suggests a review process by experts. However, a formal adjudication method (e.g., 2+1, 3+1) is not explicitly described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical trials were not performed as part of this product."

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, a standalone component was done for algorithm testing. The document states: "Algorithm testing verified the accuracy of the new dose calculation algorithm in Monaco 5.40 using the same test methods as the predicate version of Monaco." This implies evaluation of the algorithm's performance independent of a human user.

7. The type of ground truth used:

  • The ground truth appears to be based on established clinical standards and expert review/consensus rather than direct pathology or outcomes data from clinical trials. The document mentions "pre-defined pass/fail criteria" and validation against "simulated clinical workflows using actual patient data." The acceptance criteria focused on demonstrating "equivalent accuracy" to the predicate device, which would have its own established standards.

8. The sample size for the training set:

  • The document does not specify a training set sample size. The focus of the provided text is on validation and verification testing of the already developed Monaco RTP system, which includes a new dose calculation algorithm. It does not describe the development or training phase of the algorithm itself.

9. How the ground truth for the training set was established:

  • Since a training set is not explicitly mentioned and the document focuses on the validation of an existing system's new algorithm, the method for establishing ground truth for a hypothetical training set is not described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 4, 2018

Elekta. Inc % Ms. Kathryn Stinson Senior Regulatory Affairs Engineer 13723 Riverport Drive, Suite 100 MARYLAND HEIGHTS MO 63043

Re: K183037

Trade/Device Name: Monaco RTP System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ Dated: October 30, 2018 Received: November 1, 2018

Dear Ms. Stinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and

Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183037

Device Name Monaco RTP System

Indications for Use (Describe)

The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.

The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for:

· contouring

  • · image manipulation
  • · simulation
  • · image fusion
  • plan optimization
  • · QA and plan review
Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

October 30, 2018

K183037

Monaco RTP System Premarket Notification (510(k)) Summary of Safety and Effectiveness

INTRODUCTION

This document summarizes the safety and effectiveness information contained within the Monaco RTP System 510(k). The Summary of Safety and Effectiveness contains no confidential or trade secret information and is intended for full public disclosure and distribution.

PREMARKET NOTIFICATION INFORMATION

  • Product Information: 1. Product Trade Name Monaco a. Release Version Number Release 5.40 ڪ 2. Classification Information: a. Classification Name Medical charged-particle radiation therapy system Common/Usual Name Radiation Treatment Planning System ﻗ Product Classification Class II C. d. Product Code MUJ 21 CFR 892.5050 Reference e. f. Review Panel Radiology 3. Establishment Information: a. Submitter Elekta, Inc. Submitter Address 13723 Riverport Dr., Suite 100 b. Maryland Heights, MO 63043 C. Establishment Number 1937649 Kathryn Stinson, Principal RA Engineer d. Contact Contact Phone 314-993-0003 e.
    314-812-4496

  • f. Contact Fax

{4}------------------------------------------------

PREDICATE DEVICE INFORMATION

The Monaco RTP System is substantially equivalent to the following devices that the Food and Drug Administration (FDA) has cleared for distribution and are currently being actively marketed in the United States. Monaco is substantially equivalent to these products in intended use and safety and effectiveness.

    1. Monaco RTP System Elekta, Inc. K151233
    1. ViewRay Treatment Planning and Delivery ViewRav Inc. K102915

MONACO INTENDED USE/INDICATIONS FOR USE

The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.

The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for:

  • · contouring
  • · image manipulation
  • · simulation
  • · image fusion
  • · plan optimization
  • · QA and plan review

DESCRIPTION OF THE PRODUCT

Monaco is a radiation treatment planning system that first received FDA clearance in 2007 (K071938). The modified system received clearance in 2009, when Volumetric Modulated Arc Therapy (VMAT) planning capability was added (K091179), aqain when Dynamic Conformal Arc planning was added (K110730), and electron planning, support for stereotactic cones, and SUV calculation were added (K132971). Finally, specialty image creation was added in 2015 (K15123). A 510(k) was filed in 2017 for the addition of carbon ion planning. The 510(k) was withdrawn because there was no hardware cleared for the US market capable of delivering carbon ion plans. Monaco's carbon ion planning functionality remains licensed off and inaccessible to US users.

Since the 2015 510(k) clearance, IMPAC Medical Systems, Inc. merged with Elekta, Inc., both wholly owned subsidiaries of Elekta AB. There were no changes in location, personnel, or quality system as a result of the merger. However, the name of the manufacturer for Monaco as changed from IMPAC Medical Systems, Inc, to Elekta, Inc.

{5}------------------------------------------------

The Monaco system accepts patient imaging data and "source" dosimetry data from a linear accelerator. The system then permits the user to display and define (contour) the target volume to be treated and critical structures which must not receive above a certain level of radiation on these diagnostic images.

Based on the prescribed dose, the user, a Dosimetrist or Medical Physicist, can create multiple treatment scenarios involving the number, position(s) and energy of radiation beams and the use of a beam modifier (MLC, block, etc.) between the source of radiation and the patient to shape the beam. The Monaco system then produces a display of radiation dose distribution within the patient, indicating doses to the target volume and surrounding structures. The "best" plan satisfying the clinican prescription is then selected, one that maximizes dose to the target volume while minimizing dose to surrounding healthy volumes.

LEVEL OF CONCERN

Item 4b of Table 1 in the FDA Guidance document entitled, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices asks, "Does the Software Device control the delivery of potentially harmful energy that could result in death or serious injury, such as radiation treatment systems…." Monaco does not directly control the linear accelerator that delivers the radiation. Once completed, plans are reviewed and approved by qualified clinicians and may be subject to quality assurance practices before treatment actually takes place. There is no automatic link between the Monaco software and the linear accelerator. However, should a flaw in the treatment plan escape the notice of the qualified professionals using the Monaco system, serious injury or death could result. Therefore, we believe Monaco to be of major level of concern.

SUMMARY OF CLINICAL TESTING

Clinical trials were not performed as part of this product. Clinical testing on patients is not advantageous in demonstrating substantial equivalence or safety and effectiveness of the device since testing can be performed such that no human subjects are exposed to risk. Validation testing involved simulated clinical workflows using actual patient data, such as patient images, described in detail in section 20. Algorithm testing verified the accuracy of the new dose calculation algorithm in Monaco 5.40 using the same test methods as the predicate version of Monaco. Pre-defined passifail criteria were also equivalent to that of the predicate, K151233. Equivalent accuracy was demonstrated, and the product was deemed substantially equivalent and fit for clinical use.

SUMMARY OF NON-CLINICAL TESTING

Verification tests were written and executed to ensure that the system is working as designed. Over 600 test procedures were executed, including tests to verify requirements for new product functionality, tests to ensure that risk mitigations function as intended, and regression tests to ensure continued safety and effectiveness of existing functionality. Verification testing was performed according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, IEC 62304 Software Life Cycle standard, and ISO 14971 Risk Management Standard, as was the predicate version, K151233. Quality System procedures goverening the testing process, including pre-defined pass/fail criteria, were equivalent to procedures used in the testing of the previous Monaco version cleared under K151233. Conformity to the same passifail criteria as the predicated that Monaco 5.40 was substantially equivalent in safety and effectiveness. Monaco 5.40 was deemed safe and effective for its intended use.

{6}------------------------------------------------

Monaco for ElektaUnity MR-LinacMonaco(K151233)ViewRay TreatmentPlanning & DeliverySystem(K102915)
Intended Use and Indicationsfor Use
ContouringYesYesYes
Dose CalculationYesYesYes
Plan OptimizationYesYesYes
Image Manipulation & FusionYesYesYes
CT SimulationYesYesYes
QA/Plan ReviewYesYesYes
Integrated Planning and DeliverySoftwareThe Elekta MR-Linacsystem uses Monaco forplanning and the MOSAIQOIS for delivery. Theproducts were designed towork together and tested inan integrated environmentto ensure seamlessinteroperability.NoThe ViewRay system isbranded as a single devicethat handles both planning anddelivery.
Technological Characteristics
Dose Calculation AlgorithmsMonte Carlo (electron &photon), Collapsed Cone(photon), Pencil Beam(optimization only),GPUMCD for MR-linacMonte Carlo (electron& photon), CollapsedCone (photon),Pencil Beam(optimization only)Monte Carlo
Calculates dose for MR-Linac(including magnetic field, coils &cryostat)Yes (GPUMCD algorithm)NoYes
Calculation and display ofstandardized uptake valueYesYesUnknown
Local Biological MeasureOptimizationYesYesNo
Support for various treatment aidsYesYesYes
Support for Dynamic DeliveryMethodsYes (conventional linacsonly, not for MR-linacenvironment)YesNo
Operating SystemWindowsWindowsWindows
DICOM RT SupportYesYesYes
Modalities Supported: Full RTPWorkflowPhoton, ElectronPhoton, ElectronPhoton only

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.