K Number
K182924

Validate with FDA (Live)

Date Cleared
2019-04-15

(178 days)

Product Code
Regulation Number
878.4810
Age Range
0.52876712328 - 120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diode Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description

The Diode Laser Treatment System consists of the main unit and a hand piece. The system uses a diode laser as an active medium placed in an optical cavity to produce amplified beam at the wavelength of 808 nm. A microprocessor is used to control electronics for the front panel. A self-contained water cooling system is built into the power supply unit. The device provides 36 working modes, which are six modes for men and six modes for women. The men or women mode includes face, armpit, arm, body, bikini, leg mode respectively for different treatment part. The diode laser operates in a pulsed mode with a fixed pulse width and fixed pulse duration of the pulse train for each mode. The number of pulses can be adjusted within the preset range.

AI/ML Overview

This document describes a 510(k) premarket notification for the Weifang KM Electronics Co., Ltd. Diode Laser Treatment System. The notification aims to demonstrate substantial equivalence to legally marketed predicate devices for hair removal and permanent hair reduction.

Based on the provided text, a clinical study proving the device meets specific performance-based acceptance criteria for hair reduction was not described. The document focuses on demonstrating substantial equivalence primarily through technological characteristics and non-clinical performance data (biocompatibility, electrical safety, EMC, and laser safety).

Here's an analysis based on your requested information:

1. A table of acceptance criteria and the reported device performance

The provided text does not explicitly state performance-based acceptance criteria for hair reduction (e.g., a percentage of hair reduction required) or reported device performance in terms of clinical efficacy. Instead, it defines "Permanent hair reduction" as "the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime." This is a definition rather than an acceptance criterion with a specific threshold.

The document does include acceptance criteria and reported performance for non-clinical testing:

Acceptance CriteriaReported Device Performance
Biocompatibility:
ISO 10993-1:2009 for "Surface – Mucosal Membrane" with contact duration "< 24 hours"All evaluation acceptance criteria were met for Cytotoxicity, Irritation, and Sensitization.
Electrical Safety:
IEC 60601-1:2012The system has been tested to comply with this standard.
IEC 60601-2-22:2007The system has been tested to comply with this standard.
Laser Safety:
IEC 60825-1: 2007The system has been tested to comply with this standard.
Electromagnetic Compatibility (EMC):
IEC 60601-1-2:2007The system has been tested to comply with this standard.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

There is no information provided in the document about a clinical test set, its sample size, or data provenance related to hair reduction efficacy. The non-clinical tests mentioned (biocompatibility, electrical safety, EMC, laser safety) do not typically involve a patient sample size or data provenance in the context of clinical efficacy.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

Since no clinical test set for assessing hair reduction efficacy is described, there is no information provided regarding experts used to establish ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Since no clinical test set for assessing hair reduction efficacy is described, there is no information provided regarding an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The device is a Diode Laser Treatment System for hair removal and not an AI-assisted diagnostic or interpretation device that would typically involve a multi-reader multi-case (MRMC) study or human readers. Therefore, no such study was done or reported in this document.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This point is not applicable to a laser treatment system that performs a physical action (hair removal). The "standalone" performance for such a device would refer to its inherent operational capabilities, which are covered by the non-clinical tests (electrical safety, laser safety, etc.) but not in the context of algorithm-only performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical tests, the "ground truth" is defined by the requirements of the specific international standards (ISO, IEC). For example, for biocompatibility, the ground truth is whether the device materials elicit a toxic, irritant, or sensitizing response according to the defined test methods. As no clinical efficacy study is described, no clinical ground truth (like pathology or outcomes data for hair reduction) is mentioned.

8. The sample size for the training set

The document describes a physical medical device (Diode Laser Treatment System), not an AI/ML algorithm that would require a "training set." Therefore, this is not applicable, and no information is provided.

9. How the ground truth for the training set was established

As the device is not an AI/ML algorithm requiring a training set, this is not applicable, and no information is provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Weifang KM Electronics Co., LTD Andy Pang General Manager No. 7999 Health East Street, High-tech District Weifang, 261000 CN

April 15, 2019

Re: K182924

Trade/Device Name: Diode Laser Treatment System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 11, 2019 Received: March 13, 2019

Dear Andy Pang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Neil RDigitally signed by Neil
Ogden -SR Ogden -S
Date: 2019.04.15
09:23:59 -04'00'
ForBinita S. Ashar, M.D., M.B.A., F.A.C.S.
Director
Division of Surgical Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{2}------------------------------------------------

Section 4 Indications for use

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K182924

Device Name

Diode Laser Treatment System

Indications for Use (Describe)

The Diode Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Section 5 510(k) summary

l Submitter

Device submitter: Weifang KM Electronics Co., Ltd.

No.7999 Health East Street, High-tech District, Weifang City, Shandong, China

Contact person: Andy PANG General Manager Phone: +86 136 0646 4486 Fax: +86-536-8885351 Email: 51050359@qq.com Date of preparation: Oct 12th, 2018

II Device

Trade Name of Device: Diode Laser Treatment System Models: KM200D, KM300D, KM600D, KM800D, KM900D Common name: Powered Laser Surgical Instrument Regulation Number: 21 CFR 878.4810 Regulatory Class: II Product code: GEX Review Panel: General & Plastic Surgery

III Predicate Devices

Trade Name of Device: Diode Laser Therapy Machine Common name: Powered Laser Surgical Instrument Regulation Number: 21 CFR 878.4810 Requlatory Class: II Product code: GEX Review Panel: General & Plastic Surgery 510(k) number: K161692

Trade Name of Device: Diode Laser Hair Removal System Common name: Powered Laser Surgical Instrument Regulation Number: 21 CFR 878.4810 Regulatory Class: II Product code: GEX Review Panel: General & Plastic Surgery 510(k) number: K141973

{4}------------------------------------------------

IV Device description

The Diode Laser Treatment System consists of the main unit and a hand piece. The system uses a diode laser as an active medium placed in an optical cavity to produce amplified beam at the wavelength of 808 nm. A microprocessor is used to control electronics for the front panel. A self-contained water cooling system is built into the power supply unit.

The device provides 36 working modes, which are six modes for men and six modes for women. The men or women mode includes face, armpit, arm, body, bikini, leg mode respectively for different treatment part. The diode laser operates in a pulsed mode with a fixed pulse width and fixed pulse duration of the pulse train for each mode. The number of pulses can be adjusted within the preset range.

V Indications for use

The Diode Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

VI Comparison of technological characteristics with the predicate devices

The Diode Laser Treatment System has the same technological characteristics and fundamental design as the predicate device. The subject device and the predicate devices are all designed for hair removal on different parts of the body. The differences between the subject device and predicate devices do not alter suitability of the proposed device for its intended use.

Device featureDiode Laser Treatment System (subject device)Diode Laser Therapy Machine FG 2000-B K161692Diode Laser Hair Removal System K141973
Product codeGEXGEXGEX
Regulation number21 CFR 878.481021 CFR 878.481021 CFR 878.4810
Indications for useThe Diode Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skinThe Diode Laser Therapy Machine is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skinThe Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin
type I-VI), includingtype I-VI), includingtype I-VI), including
tanned skin.tanned skin.tanned skin.
Permanent hairreduction is defined asthe long-term, stablereduction in the numberof hairs regrowing whenmeasured at 6, 9, and12 months after thecompletion of atreatment regime.Permanent hairreduction is defined asthe long-term, stablereduction in the numberof hairs regrowing whenmeasured at 6, 9, and 12months after thecompletion of atreatment regime.Permanent hairreduction is defined asthe long-term, stablereduction in the numberof hairs regrowing whenmeasured at 6, 9, and 12months after thecompletion of atreatment regime.
Operation principleMelanin could absorbthe energy from thelaser, which would resultin temperature rapidincrease, to destroysurrounding hairfollicles, and finallyremove hair.Melanin could absorbthe energy from thelaser, which would resultin temperature rapidincrease, to destroysurrounding hair follicles,and finally remove hair.Melanin could absorbthe energy from thelaser, which would resultin temperature rapidincrease, to destroysurrounding hairfollicles, and finallyremove hair.
Laser typeDiode laserDiode laserDiode laser
Laser classificationClass IVClass IVClass IV
Laser wavelength808nm808nm808nm
Spot Size1.44 cm²1.44 cm²1.44 cm²
Fluence2-120J/cm²2-120J/cm²1-120J/cm²
Frequency1-10Hz1-10Hz1Hz, 2Hz, 3Hz, 10Hz
Pulse Duration10-300ms9-143ms2.9-348ms
Power supply100-240V AC, 50/60HzAC110V/50-60HzAC110V, 60Hz
DimensionKM200D 604238cmKM300D 604235cmKM600D 604235cmKM800D 604240cmKM900D 554230cm426354cm3854120cm
Weight35kg30kg55kg
Patient contactmaterialSapphire in handpieceand handpiece tip(stainless steel)Sapphire in handpieceand handpiece tip(stainless steel)Sapphire in handpieceand handpiece tip(stainless steel)
BiocompatibilityComply withComply withComply with

{5}------------------------------------------------

{6}------------------------------------------------

ElectricalSafetyComplyIEC60601-1,IEC60601-2-22withComplyIEC60601-1,IEC60601-2-22withComplyIEC60601-1,IEC60601-2-22with
EMCComplyIEC60601-1-2,withComplyIEC60601-1-2,withComplyIEC60601-1-2,with
Laser safetyComplyIEC60825-1,IEC60601-2-22withComplyIEC60825-1,IEC60601-2-22withComplyIEC60825-1,IEC60601-2-22with

VII Performance data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

Biocompatibility of the Laser Treatment System was evaluated in accordance with ISO 10993-1:2009 for the body contact category of "Surface – Mucosal Membrane" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended: Cytotoxicity, Irritation and Sensitization. All evaluation acceptance criteria were met

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Laser Treatment System. The system has been tested to comply with the following standards:

  • । IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
  • -IEC 60601-2-22:2007, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment;
  • -IEC 60825-1: 2007, Safety of laser products - Part 1: Equipment classification and requirements.
  • -IEC 60601-1-2:2007, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests.

VIII Conclusion

The Diode Laser Treatment System is substantially equivalent to its predicate devices. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.