Intended Use

The intended use of the EPIQ 5, EPIQ 7, Affiniti 30, Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.

Device Description

The proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems are general purpose, software controlled, diagnostic ultrasound systems. Their function is to acquire ultrasound data and to display the data in various modes of operation. The devices consist of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics and optional peripherals (ECG, printers).

The removable transducers are connected to the system using a standard technology, multi-pin connectors. Other than the introductions of the two new transducers, the device description, accessories and components are unchanged, reference table 1.

AI/ML Overview

Here's the breakdown of the acceptance criteria and supporting studies for the Philips EPIQ and Affiniti Diagnostic Ultrasound Systems based on the provided text:

Important Note: The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study for the device's original clearance. Therefore, information regarding specific clinical performance metrics (like sensitivity, specificity, accuracy) from a de novo study is not present. Instead, the document emphasizes compliance with standards and equivalence to a previously cleared device.


1. Table of Acceptance Criteria and Reported Device Performance

The document describes the device meeting safety and performance standards rather than specific clinical performance metrics like sensitivity/specificity for a novel algorithm.

Acceptance Criteria (Safety/Performance Standards)Reported Device Performance (Compliance)
IEC 60601-2-37 Ed 2.1 (Acoustic Output Display Requirements)Complies
IEC 62359 Ed 2.0 (Thermal and Mechanical Indices)Complies
FDA ultrasound guidance document (Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, issued Sept 9, 2008)Complies
System Acoustic Output Limits: Ispta.3 < 720 mw/cm2Reported as: < 720 mw/cm2
System Acoustic Output Limits: MI < 1.9Reported as: < 1.9
System Acoustic Output Limits: TI < 6.0Reported as: < 6.0
ANSI/AAMI 60601-1 (Medical electrical equipment general safety)Complies (2005, Amendment 1, 2012)
IEC 60601-1-2 (EMC Compliance)Complies (2014)
IEC 60601-2-37 (Ultrasonic medical diagnostic and monitoring equipment)Complies (2015)
IEC 62359 (Test methods for thermal and mechanical indices)Complies (2017)
ISO 10993 (Biological evaluation of medical devices)Complies (2009)
Meet intended use and commercial claimsAll pre-determined acceptance criteria were met (based on non-clinical validation testing)
System level requirements (Non-Clinical verification testing)All pre-determined acceptance criteria were met (based on non-clinical verification testing)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a "test set" in the context of clinical performance data for a novel algorithm being evaluated against a ground truth. The testing mentioned is primarily non-clinical, focusing on compliance with safety and performance standards, and demonstrating substantial equivalence to a predicate device. Therefore, a sample size for a separate test set of patient data is not provided. The data provenance is also not applicable in this context as no clinical study for a new performance claim was conducted.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided as the submission is for substantial equivalence and mainly relies on non-clinical testing and comparison to predicate devices, rather than a de novo clinical study with expert-established ground truth.


4. Adjudication Method for the Test Set

This information is not provided as no clinical study with a test set requiring adjudication is described.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

An MRMC comparative effectiveness study is not mentioned in this 510(k) summary. The submission focuses on the safety and effectiveness of the ultrasound system and its transducers, not on an AI-assisted diagnostic workflow affecting human reader performance.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

The document describes an ultrasound system and transducers, not a standalone algorithm. Therefore, "standalone" performance in the context of an algorithm is not applicable or discussed.


7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical performance data, the "ground truth" implicitly refers to the specified values and requirements within the cited technical standards (e.g., acoustic output limits, electrical safety standards). No clinical ground truth (like pathology or outcomes data) is referenced for performance evaluation.


8. The Sample Size for the Training Set

This information is not applicable as the submission describes an ultrasound system, not an AI/algorithm that would require a training set.


9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set for an AI/algorithm mentioned in this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 9, 2018

Philips Ultrasound, Inc. % Mr. Mark Job Official Correspondent Regulatory Technology Services, LLC 1394 25th Street, NW BUFFALO, MN 55313

Re: K182857

Trade/Device Name: EPIQ Diagnostic Ultrasound System, EPIQ 5 Diagnostic Ultrasound System, EPIQ 7 Diagnostic Ultrasound System, Affiniti 30 Diagnostic Ultrasound System, Affiniti 50 Diagnostic Ultrasound System, Affiniti 70 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: October 30, 2018 Received: November 2, 2018

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Hole 2. Nils

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K182857

Device Name

EPIQ 5 Diagnostic Ultrasound System, EPIQ 7 Diagnostic Ultrasound System, EPIQ Diagnostic Ultrasound System, Affiniti 30 Diagnostic Ultrasound System, Affiniti 50 Diagnostic Ultrasound System, Affiniti 70 Diagnostic Ultrasound System

Indications for Use (Describe)

The intended use of the EPIQ 5, EPIQ 7, Affiniti 30, Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.

The clinical environments where the EPIQ and Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.

When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.

The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgment and best clinical procedure.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) No: _K182857

System: EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificPWDCWDColorCombinedOther
(Track I only)(Tracks I & III)BMDoppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,16,17,18, 19
Intra-operative (Cardiac)PPPPPNote: 1,2,3Note: 5,8,9,10, 12,13
Intra-operative (Vascular)PPPPNote: 1,2,3Note: 5,8,9,10, 12,13
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17,18
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17,19
Neonatal CephalicPPPPNote: 1,2,3Note: 5,8,9,10,12,13,17
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10, 12,13,17
Trans-rectalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,17
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13,15,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10, ,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10, ,12,13,17
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,15,17
Other: UrologyPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,17
CardiacCardiac AdultPPPPPNote:1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricPPPPPNote:1,2,3,4Note: 10, ,12,13,14
Trans-esophageal (Cardiac)PPPPPNote:1,2,3,4Note: 10, ,12,13,14
Other (Fetal Echo)PPPPPNote:1,2,3,4Note: 5,6,8,10,12,13,14
PeripheralPeripheral vesselPPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
VesselCerebral vascularPPPPPNote: 1,2,3Note: 5,6,8,9,10, 11,12,13,17

N= new indication P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous submission: K181485

{4}------------------------------------------------

510(k) No: K182857

System: Transducer: EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

3D9-3v

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler**
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7,8,10,12,13,15
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNP
Other: UrologyPPPP
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication; P= previously cleared (; E= previously cleared by earlier letter to file

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{5}------------------------------------------------

510(k) No: _K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: C5-1

Note 6: Imaging for guidance of biopsy

Note 8: Panoramic Imaging

Note 9: Color Power Angio (CPA)

Previous submission: K181485

Note 7: Infertility monitoring of follicle development

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificColorCombined*Other
(Track I only)(Tracks I & III)BMPWDCWDDoppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,16,17,18
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13,17,18
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,7,8, 9,10,11,12,13, 17
Other: UrologyPPPPNote: 1,2,3Note: 5,6,8, 9,10,,12,13, 17
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,10,12,13
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,17
VesselCerebral vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography

Note 16: ElastPQ (for Liver)

Note 19: ElastQ Imaging Linear

Note 18: ElastQ Imaging

Note 17: Percu Nav

{6}------------------------------------------------

510(k) No: K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: C8-E

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorCombined*Other
(Track I only)(Tracks I & III)Doppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
Small Organ (breast, thyroid,
testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD: B+Color+CWD: B+Amplitude+CWDNote 14: TDI

Note 15: Elastography

Note 17: Percu Nav

Note 16: ElastPQ (for Liver)

Note 18: ElastQ Imaging Note 19: ElastQ Imaging Linear

Note 2: Combined modes include: B+M+Color
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWD
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+
Note 5: SonoCT
Note 6: Imaging for guidance of biopsy
Note 7: Infertility monitoring of follicle development

Note 8: Panoramic Imaging

Note 9: Color Power Angio (CPA) Previous Submission: K181485

{7}------------------------------------------------

510(k) No: K182857

System: EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: C9-2

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OBPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13, 17
Fetal Imaging& OtherIntra-operative (Cardiac)
Fetal Imaging& OtherIntra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Fetal Imaging& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10, ,12,13,17
Small Organ (breast, thyroid,testicle)
Fetal Imaging& OtherNeonatal Cephalic
Fetal Imaging& OtherAdult Cephalic
Fetal Imaging& OtherTrans-rectal
Fetal Imaging& OtherTrans-vaginal
Fetal Imaging& OtherTrans-urethral
Fetal Imaging& OtherTrans-esoph. (non-Cardiac)
Fetal Imaging& OtherMusculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13
Musculo-skel. (Superficial)
Fetal Imaging& OtherIntra-luminal
Fetal Imaging& OtherOther: GYNPPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17
Other: UrologyPPPPNote: 1,2,3Note: 5,6,8,9,10, ,12,13,17
CardiacCardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,10,12,13
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13, 17
VesselCerebral Vascular

N= new indication P= previously cleared

Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous submission: K181485

{8}------------------------------------------------

510(k) No: K182857

EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: C10-3v

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note:5,6,7,8,9,10,12,13,17
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7,8,9,10,11,12,13, 15, 17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,15,17
Other: UrologyPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,15, 17
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,6,8,10,12,13
PeripheralPeripheral vessel
VesselCerebral Vascular

N= new indication P= previously cleared

Additional Comments: *Color Doppler includes Color Amplitude Doppler Note 10: Harmonic Imaging Note 11: Contrast Imaging Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+M Note 2: Combined modes include: B+M+Color Note 12: 3D/4D Imaging Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWD Note 13: XRES Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWD Note 14: TDI Note 5: SonoCT Note 15: Elastography Note 6: Imaging for guidance of biopsy Note 16: ElastPQ (for Liver) Note 7: Infertility monitoring of follicle development Note 17: Percu Nav Note 8: Panoramic Imaging Note 18: ElastQ Imaging Note 19: ElastQ Imaging Linear Note 9: Color Power Angio (CPA) Previous Submission: K181485

{9}------------------------------------------------

510(k) No:

Intended Use:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note:5,6,8,9,10,12,13,17
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote: 1,2,3Note: 5,6,7, 8,9,10,1112,13,17
Trans-vaginalPPPPNote: 1,2,3Note: 5,6,7, 8,9,10,12,13,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,2,3Note: 5,6,7,8,9,10,12,13,17
Other: UrologyPPPPNote: 1,2,3Note:5,6,7,8,9,10,11,12,13,17
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,2,3Note: 5,8,9,10,12,13
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17
VesselCerebral Vascular
N= new indicationP= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{10}------------------------------------------------

510(k) No: K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: D2cwc

Note 9: Color Power Angio (CPA)

Previous Submission: K181485

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel.(Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultP
CardiacCardiac PediatricP
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging

{11}------------------------------------------------

510(k) No: _K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: D2tcd

Note 9: Color Power Angio (CPA)

Previous Submission: K181485

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificColorCombined*Other
(Track I only)(Tracks I & III)BMPWDCWDDoppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral VascularP
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging

{12}------------------------------------------------

510(k) No: K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: D5cwc

Note 9: Color Power Angio (CPA) Previous Submission: K181485

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselP
Cerebral VascularP
N= new indication P= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude Doppler
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNav
Note 9: ElastQ ImagingNote 18: FlastQ Imaging

{13}------------------------------------------------

510(k) No: _K182857

System: EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: L12-3

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)PPPP
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10, 12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13, 17
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13

P= previously cleared N= new indication

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{14}------------------------------------------------

510(k) No: _K182857

System: EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: L12-5 50

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
AbdominalPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,10, 12,13, 17
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
VesselCerebral VascularPPPPNote: 1.2.3Note: 5.6.8.9.10. 12.13

N= new indication P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{15}------------------------------------------------

510(k) No: _K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: L15-7io

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)PPPPNote: 1,2,3Note: 5,8,9,10,,12,13
Intra-operative (Vascular)PPPPNote: 1,2,3Note: 5,8,9,10,,12,13
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,8,9,10,,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note: 5,8,10,,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,8,9,10,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,8,9,10,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,8,9,10,,12,13
Cerebral VascularPPPPNote: 1,2,3Note: 5,8,9,10,,12,13
N= new indication P= previously cleared
Additional Comments:*Color Doppler includes Color Amplitude Doppler
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 10: Harmonic Imaging
Note 2: Combined modes include: B+M+ColorNote 11: Contrast Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 12: 3D/4D Imaging
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDI
Note 6: Imaging for guidance of biopsyNote 15: Elastography
Note 16: ElastPQ (for Liver)

Note 17: Percu Nav

Note 18: ElastQ Imaging

Note 19: ElastQ Imaging Linear

Note 9: Color Power Angio (CPA) Previous Submission: K181485

Note 8: Panoramic Imaging

Note 7: Infertility monitoring of follicle development

{16}------------------------------------------------

510(k) No: _K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: eL18-4

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13
Intra-operative (Cardiac)PPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,17,19
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)PPPPNote: 1,2,3Note: 5,6,8,10,12,13, 17
Neonatal CephalicPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13,15,17,19
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Intra-luminalPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral VascularPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
N= new indication P= previously cleared

N= new indication
Additional Commer

Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{17}------------------------------------------------

E10(λ) No: K182857

510(k) No: _K182857

System: EEPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: L18-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
AbdominalPPPPNote: 1,2,3Note:5,6,8,9,10,,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note:5,6,8,9,10,12,13,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
Cerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
N= new indication P= previously clearedAdditional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: 3D/4D Imaging
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{18}------------------------------------------------

510(k) No: _K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: S5-1

Note 9: Color Power Angio (CPA)

Previous Submission: K181485

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 6,8,9,10,12,13, 17
AbdominalPPPPNote: 1,2,3Note: 6,8,9,10, 12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 6,8,9,10, 12,13, 17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10, ,12,13, 17
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14, 17
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10, 12,13,14, 17
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPPNote: 1,2,3,4Note: 10,13,14
PeripheralVesselPeripheral vesselPPPPNote: 1,2,3Note: 6,8,9,10, 12,13, 17
Cerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging

{19}------------------------------------------------

510(k) No: K182857

System: EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: S7-3t

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,
testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{20}------------------------------------------------

510(k) No: K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: 58-3

Note 8: Panoramic Imaging

Note 9: Color Power Angio (CPA)

Previous Submission: K181485

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificColorCombined*Other
(Track I only)(Tracks I & III)BMPWDCWDDoppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 7,8,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 7,8,10,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 8,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav

Note 18: ElastQ Imaging

{21}------------------------------------------------

510(k) No: ___________________________________________________________________________________________________________________________________________________________________ EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: S8-3t

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4Note: 10,,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude Doppler
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+M
Note 2: Combined modes include: B+M+Color
Note 10: Harmonic Imaging
Note 11: Contrast Imaging
Note 12: 3D/4D Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{22}------------------------------------------------

510(k) No: _K182857

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: S12-4

Note 8: Panoramic Imaging

Note 9: Color Power Angio (CPA)

Previous Submission: K181485

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 8,10,,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 8,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: ,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: ,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
P= previously clearedN= new indication
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav

Note 18: ElastQ Imaging

{23}------------------------------------------------

K182857 510(k) No:

EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: V6-2

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
AbdominalPPPPPNote: 1,2,3Note: 5,6,8,9,10,,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPPNote: 1,2,3Note: 5,9,10,,13
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indicationP= previously clearedAdditional Comments:*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic ImagingNote 11: Contrast ImagingNote 12: 3D/4D ImagingNote 13: XRESNote 14: TDINote 15: ElastographyNote 16: ElastPQ (for Liver)Note 17: PercuNavNote 18: ElastQ ImagingNote 19: ElastQ Imaging Linear
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 2: Combined modes include: B+M+ColorNote 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 5: SonoCTNote 6: Imaging for guidance of biopsyNote 7: Infertility monitoring of follicle developmentNote 8: Panoramic ImagingNote 9: Color Power Angio (CPA)Previous Submission: K181485

{24}------------------------------------------------

510(k) No: K182857

System: EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: VL13-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPPNote: 1,2,3Note: 5,6,8,10,12,13
AbdominalPPPPPNote: 1,2,3Note: 5,6,8,10,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote: 1,2,3Note: 5,6,8,10,12,13
Small Organ (breast, thyroid,testicle)PPPPPNote: 1,2,3Note: 5,6,8,10,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPPNote: 1,2,3Note: 5,6,8,10,12,13
Musculo-skel. (Superficial)PPPPPNote: 1,2,3Note: 5,6,8,10,12,13
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPPNote: 1,2,3Note: 5,6,8,10,12,13
Cerebral VascularPPPPPNote: 1,2,3Note: 5,6,8,10,12,13

N= new indication; P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{25}------------------------------------------------

K182857 510(k) No:

EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: XL14-3

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)PPPPNote: 1, 2, 3Note: 5, 6, 8, 9, 10, 11,12, 13, 17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,11,12,13
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{26}------------------------------------------------

K182857 510(k) No:

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer:

X5-1

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 8,9,10,,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 8,9,10, 12,13
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10, 12,13
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10, 12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear

Previous Submission: K181485

{27}------------------------------------------------

K182857 510(k) No:

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: X6-1

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPPNote: 1,2,3Note: 6,8,9,10,12,13
AbdominalPPPPPNote: 1,2,3Note: 6,8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote: 1,2,3Note: 6,8,9,10,12,13,17
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPPNote: 1,2,3Note: 6,8,9,10,12,13
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPPNote: 1,2,3Note:6,8,9,10,11,12,13,17
VesselCerebral VascularPPPPPNote: 1,2,3Note: 6,8,9,10,12,13,17

N= new indication Additional Comments:

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{28}------------------------------------------------

X7-2

510(k) No: K182857

System: Transducer: EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Intended Use:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 6,7,8,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 6,7,8,10,12,13
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPPNote: 1,2,3,4Note: 10,13,14
PeripheralVesselPeripheral vesselNote: 1,2,3
Cerebral Vascular

N= new indication; P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{29}------------------------------------------------

510(k) No:

EPIQ, EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

K182857

Transducer: X7-2t

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
Cardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{30}------------------------------------------------

K182857 510(k) No:

EPIQ and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: X8-2t

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificColorCombined*Other
(Track I only)(Tracks I & III)BMPWDCWDDoppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular

N= new indication P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{31}------------------------------------------------

K182857 510(k) No:

EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: X8-2ti

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
Cardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication P= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude Doppler
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+M
Note 2: Combined modes include: B+M+Color
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWD
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWD
Note 5: SonoCT
Note 6: Imaging for guidance of biopsy
Note 7: Infertility monitoring of follicle development
Note 8: Panoramic Imaging
Note 9: Color Power Angio (CPA)
Note 10: Harmonic Imaging
Note 11: Contrast Imaging
Note 12: 3D/4D Imaging
Note 13: XRES
Note 14: TDI
Note 15: Elastography
Note 16: ElastPQ (for Liver)
Note 17: PercuNav
Note 18: ElastQ Imaging
Note 19: ElastQ Imaging Linear

Previous Submission: K181485

{32}------------------------------------------------

K182857 510(k) No:

EPIQ, EPIQ 5, and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: ട9-2

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificColorCombined*Other
(Track I only)(Tracks I & III)BMPWDCWDDoppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note: 7,8,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 7,8,10,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,2,3Note: 8,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPPNote: 1,2,3,4Note: 10,12,13,14
PeripheralPeripheral vessel
VesselCerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 13: XRES
Note 5: SonoCTNote 14: TDINote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear

Previous Submission: K181485

{33}------------------------------------------------

K182857 510(k) No:

EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems System:

Transducer: V9-2

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBNNNNNote: 1,2,3Note: 5,6,8,9,10,12,13
AbdominalNNNNNote: 1,2,3Note: 5,6,8,9,10,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)NNNNNote: 1,2,3Note: 5,9,10,13
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear

{34}------------------------------------------------

(k) No:K182857

510(k) N
System:

EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems

Transducer: mC7-2

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
AbdominalNNNNNNote: 1,2,3Note:5,6,8,9,10,11,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNNote: 1,2,3Note: 5,6,8,9,10,12,13, 17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indicationP= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 7: Infertility monitoring of follicle developmentNote 16: ElastPQ (for Liver)
Note 8: Panoramic ImagingNote 17: PercuNavNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear

{35}------------------------------------------------

510(k) No: K182857

System: Affiniti 70/50/30 Diagnostic Ultrasound Systems

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,3Note: 5,6,8,9,10,11,12,13,15,16,17
Intra-operative (Cardiac)PPPPNote: 1,3Note: 5,8,10,12,13
Intra-operative (Vascular)PPPPNote: 1,3Note: 5,8,10,12,13
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Small Organ (breast, thyroid,testicle)PPPPNote: 1,3Note:5,6,8,9,10,11,12,13,15,17
Neonatal CephalicPPPPNote: 1,3Note: 5,8,9,10,12,13
Adult CephalicPPPPNote: 1,3Note: 8,10,12,13
Trans-rectalPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Trans-vaginalPPPPNote: 1,3Note:5,6,7,8,9,10,12,13,15,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Intra-luminal
Other: GYNPPPPNote: 1,3Note: 5,6,7,8,9,10,11,12,13,15,17
Other: UrologyPPPPNote: 1,3Note: 5,6,8,9,10,11,12,13,15,17
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)PPPPPNote: 1,2,3,4Note: 5,6,8,10,12,13,14
PeripheralVesselPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Cerebral VascularPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{36}------------------------------------------------

510(k) No: K182857
Affiniti 70 Diagnostic Ultrasound SystemsSystem:
Transducer:CS-1
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Intended Use:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificColorCombined*Other
(Track I only)(Tracks I & III)BMPWDCWDDoppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,3Note: 5,6,8,9,10,11,12,13,15,16,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,3Note: 5,6,7,8, 9,10,12,13,17
Other: UrologyPPPPNote: 1,3Note: 5,6,8, 9,10,12,13,17
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,3Note: 5,6,10,12,13
PeripheralPeripheral Vessel
VesselCerebral Vascular

N= new indication, P= previously cleared Additional Comments:

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{37}------------------------------------------------

510(k) No: K182857

Affiniti 70/50/30 Diagnostic Ultrasound Systems System:

Transducer: C6-2

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,11,12,13,15
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,3Note: 5,6,7,8, 9,10,12,13
Other: UrologyPPPPNote: 1,3Note: 5,6,8, 9,10,12,13
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,3Note: 5,6,10,12,13
PeripheralVesselPeripheral Vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{38}------------------------------------------------

510(k) No: K182857
System: Affiniti 70/50/30 Diagnostic Ultrasound Systems
Transducer: C8-5
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 5,6,8,9,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,3Note: 5,6,8,9,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,12,13
Cerebral VascularPPPPNote: 1,3Note: 5,6,8,9,12,13
N= new indication, P= previously cleared
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{39}------------------------------------------------

510(k) No: K182857
System: Affiniti 70 Diagnostic Ultrasound Systems
Transducer: C9-2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,11,12,13.17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,3Note: 5.6,10,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{40}------------------------------------------------

510(k) No:K182857
System:Affiniti 70 Diagnostic Ultrasound Systems
Transducer:C10-3v
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application
Mode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,3Note: 5,6,7, 8,10,12,13, 15
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,3Note:5,6,7,8,9,10,11,12,13,15
Other: UrologyPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13,15
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,3Note: 5,6,8,10,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication. P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{41}------------------------------------------------

510(k) No: K182857
Affiniti 70/50/30 Diagnostic Ultrasound SystemsSystem:
Transducer:3D9-3v
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificB M PWD CWD ColorDoppler* Combined* Other
(Track I only)(Tracks I & III)(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,3Note: 5,6,7, 8,10,12,13
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
Other: UrologyPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,3Note: 5,6,8,10,12,13
PeripheralPeripheral vessel
VesselCerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{42}------------------------------------------------

510(k) No:K182857
System:Affiniti 70/50/30 Diagnostic Ultrasound Systems
Transducer:C9-4v
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote: 1,3Note: 5,6,7, 8,10,12,13, 15
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYNPPPPNote: 1,3Note:5,6,7,8,9,10,11,12,13,15
Other: UrologyPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13,15
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,3Note: 5,6,8,10,12,13
PeripheralPeripheral vessel
VesselCerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{43}------------------------------------------------

510(k) No: K182857
System: Affiniti 70/50 Diagnostic Ultrasound Systems
Transducer: C10-4ec
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote: 1,3Note: 5,6,7, 8,9,10,12,13,17
Trans-vaginalPPPPNote: 1,3Note: 5,6,7, 8,9,10,12,13,17
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: UrologyPPPPNote: 1,3Note:5,6,7,8,9,10,11,12,13,17
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPNote: 1,3Note: 5,6,8,10,12,13
PeripheralVesselPeripheral Vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{44}------------------------------------------------

510(k) No: R182657System: Affiniti 70/50 Diagnostic Ultrasound SystemsTransducer: BP10-5ec
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
Ophthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: UrologyPPPPNote: 5,6,8,9,10,12,13
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

510(k) No: K182857

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{45}------------------------------------------------

510(k) No:K182857
System:Affiniti 70/50/30 Diagnostic Ultrasound Systems
Transducer:D2cwc
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultP
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{46}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
Peripheral vessel
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{47}------------------------------------------------

510(k) No: K182857
System: Affiniti 70/50 Diagnostic Ultrasound Systems
Transducer: D5cwc
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselP
Cerebral VascularP

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{48}------------------------------------------------

510(k) No: K182857

System: Affiniti 70/50 Diagnostic Ultrasound Systems

Transducer: L12-3

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
PediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Small Organ (breast, thyroid, testicle)PPPPNote: 1,3Note: 5,6,8,9,10,11,12,13
Neonatal CephalicPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Musculo-skel. (Superficial)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Cerebral VascularPPPPNote: 1.3Note: 5.6.8.9.10.12.13

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{49}------------------------------------------------

K182857 510(k) No:

System: Affiniti 70/50/30 Diagnostic Ultrasound Systems

Transducer: L12-4

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,3Note: 5,6,8,9,10,11,12,13
Neonatal CephalicPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Musculo-skel. (Superficial)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Cerebral VascularPPPPNote: 1,3Note: 5,6,8,9,10,12,13

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{50}------------------------------------------------

510(k) No: K182857
Affiniti 70/50/30 Diagnostic Ultrasound SystemsSystem:
L12-5 50Transducer:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Intended Use:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Small Organ (breast, thyroid,testicle)PPPPNote: 1,3Note:5,6,8,9,10,11,12,13,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17
Cerebral VascularPPPPNote: 1,3Note: 5,6,8,9,10,12,13,17

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{51}------------------------------------------------

510(k) No:K182857
System:Affiniti 70/50 Diagnostic Ultrasound Systems
Transducer:L15-7io
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,12,13
Intra-operative (Cardiac)PPPPNote: 1,3Note: 5,8,9,10,12,13
Intra-operative (Vascular)PPPPNote: 1,3Note: 5,8,9,10,12,13
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 5,6,7,8,9,10,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Musculo-skel. (Superficial)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Cerebral VascularPPPPNote: 1,3Note: 5,6,8,9,10,12,13

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{52}------------------------------------------------

510(k) No: K182857

System: Affiniti 70 Diagnostic Ultrasound Systems

Transducer: eL18-4

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,2,3Note:5,6,7,8,9,10,12,13,17
AbdominalPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Small Organ (breast, thyroid,testicle)PPPPNote: 1,2,3Note:5,6,8,9,10,11,12,13,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Musculo-skel. (Superficial)PPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17
VesselCerebral VascularPPPPNote: 1,2,3Note: 5,6,8,9,10,12,13,17

N= new indication, P= previously cleared

Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{53}------------------------------------------------

510(k) No: _K182857

System:

Affiniti 70/50 Diagnostic Ultrasound Systems

Transducer: L18-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,3Note: 5,6,8,10,12,13
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Musculo-skel. (Superficial)PPPPNote: 1,3Note: 5,6,8,9,10,12,13
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,10,12,13
VesselCerebral VascularPPPPNote: 1,3Note: 5,6,8,9,10,12,13

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{54}------------------------------------------------

Clinical ApplicationSpecific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPNote: 1,3Note: 5,6, 8,9,10,12,13,17
AbdominalPPP
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,3Note: 8,9,10,12,13,17
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{55}------------------------------------------------

510(k) No: K182857

System:

Affiniti 70/50/30 Diagnostic Ultrasound Systems

Transducer: S4-2

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,3Note: 5,6, 8,9,10,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,3Note: 8,9,10,12,13
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{56}------------------------------------------------

510(k) No: R182657System: Affiniti 70/50 Diagnostic Ultrasound SystemsTransducer: S7-3t
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Laparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

510(k) No: K182857

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{57}------------------------------------------------

510(k) No: _K182857

System: Affiniti 70/50/30 Diagnostic Ultrasound Systems

Transducer: 58-3

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OBPPPPNote: 1,3Note: 7,8,9,10,12,13
AbdominalPPPPNote: 1,3Note: 8,9,10,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 8,10,12,13
Small Organ (breast, thyroid,testicle)PPPPNote: 1,3Note: 8,10,12,13
Neonatal CephalicPPPPNote: 1,3Note: 8,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPPNote: 1,3Note: 10,12,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{58}------------------------------------------------

510(k) No: K182857

System: Affiniti 70 Diagnostic Ultrasound Systems

Transducer: S8-3t

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,
testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
Cardiac PediatricPPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)
PeripheralPeripheral vessel
VesselCerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{59}------------------------------------------------

510(k) No: K182857
Affiniti 70/50 Diagnostic Ultrasound SystemsSystem:
Transducer:S12-4
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Intended Use:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,3Note: 8,9,10,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,3Note: 8,9,10,12,13
Small Organ (breast, thyroid,testicle)
Neonatal CephalicPPPPNote: 1,3Note: 8,9,10,12,13
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac PediatricPPPPNote: 1,2,3,4Note: 10,12,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{60}------------------------------------------------

510(k) No: _ K182857

System:

Affiniti 70/50/30 Diagnostic Ultrasound Systems

Transducer: V6-2

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
General(Track I only)Specific(Tracks I & III)
Ophthalmic
Fetal / OBPPPPPNote: 1,3Note: 5,6,8,9,10,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)PPPPPNote: 1,3Note: 5,9,10,13
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{61}------------------------------------------------

510(k) No: K182857
System:Affiniti 70/50 Diagnostic Ultrasound Systems
Transducer:X7-2t
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Intended Use:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,12,13,14
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote: 1,2,3,4Note: 10,12,13,14
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{62}------------------------------------------------

510(k) No: K182857
Affiniti 70/50 Diagnostic Ultrasound SystemsSystem:
Transducer:VL13-5
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorCombined*Other
(Track I only)(Tracks I & III)Doppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal / OB
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (breast, thyroid,PPPPNote: 1,3Note: 5,6,8,9,10,12,13
testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel.
(Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralPeripheral vesselPPPPNote: 1,3Note: 5,6,8,9,10,12,13
VesselCerebral VascularPPPPNote: 1,3Note: 5,6,8,9,10,12,13

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{63}------------------------------------------------

510(k) No: K182857

Affiniti 70 Diagnostic Ultrasound Systems System:

Transducer: X5-1

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalPPPPNote: 1,2,3Note: 8,9,10,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote: 1,2,3Note: 8,9,10,13
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult CephalicPPPPNote: 1,2,3Note: 8,9,10,13
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel.(Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac AdultPPPPPNote: 1,2,3,4Note: 10,11,13,14
CardiacCardiac PediatricPPPPPNote: 1,2,3,4Note: 10,13,14
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel
Cerebral Vascular

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{64}------------------------------------------------

510(k) No: K182857
System: Affiniti 70/50 Diagnostic Ultrasound Systems
Transducer: V9-2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal / OBNNNNNote: 1,3Note: 5,6,8,9,10,11,12,13
Abdominal
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid, testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel.(Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)NNNNNote: 1,3Note: 5,9,10,11,13
PeripheralVesselPeripheral vessel
Cerebral Vascular
N= new indication, P= previously cleared
Additional Comments:
*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{65}------------------------------------------------

510(k) No: K182857
System:Affiniti 70/50 Diagnostic Ultrasound Systems
Transducer: mC7-2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal / OB
AbdominalNNNNNote: 1,3Note: 5,6, 8,9,11,12,13
Intra-operative (Cardiac)
Intra-operative (Vascular)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (breast, thyroid,testicle)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel.(Conventional)
Musculo-skel. (Superficial)
Intra-luminal
Other: GYN
Other: Urology
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (Fetal Echo)
PeripheralVesselPeripheral vessel

N= new indication, P= previously cleared

*Color Doppler includes Color Amplitude DopplerNote 10: Harmonic Imaging
Note 1: Combined modes include: B+PWD; B+Color; B+Amplitude; B+MNote 11: Contrast Imaging
Note 2: Combined modes include: B+M+ColorNote 12: 3D/4D Imaging
Note 3: Combined modes include: B+Color+PWD; B+Amplitude+PWDNote 13: XRES
Note 4: Combined modes include: B+CWD; B+Color+CWD; B+Amplitude+CWDNote 14: TDI
Note 5: SonoCTNote 15: Elastography
Note 6: Imaging for guidance of biopsyNote 16: ElastPQ (for Liver)
Note 7: Infertility monitoring of follicle developmentNote 17: PercuNav
Note 8: Panoramic ImagingNote 18: ElastQ Imaging
Note 9: Color Power Angio (CPA)Note 19: ElastQ Imaging Linear
Previous Submission: K181485

{66}------------------------------------------------

K182857

510(k) Summary of Safety and Effectiveness

Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems

This summary of safety and effectiveness information is submitted in accordance with 21CFR §807.92

    1. Submitter's name, address, telephone number, contact person. Philips Ultrasound, Inc. 22100 Bothell Everett Hwy Bothell, WA 98021-8431
Contact person:Kym Rupp, Senior Regulatory Affairs Specialist
Email:kym.rupp@philips.com
Tel:425-487-7127
Fax:425-487-8666
Secondary Contact:Colin Jacob, Senior Regulatory Affairs Specialist
Email:colin.jacob@philips.com
Date prepared:September 17th, 2018
    1. Name of the device. including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common name:Diagnostic ultrasound system and transducers
Proprietary name:EPIQ Diagnostic Ultrasound System
EPIQ 7 Diagnostic Ultrasound System
EPIQ 5 Diagnostic Ultrasound System
Affiniti 70 Diagnostic Ultrasound System
Affiniti 50 Diagnostic Ultrasound System
Affiniti 30 Diagnostic Ultrasound System

These devices are classified as follows:

Classification Description21 CFR SectionProduct Code
Ultrasonic Pulsed Doppler Imaging System892.1550IYN
Ultrasonic Pulsed Echo Imaging System892.1560IYO
Diagnostic Ultrasound Transducer892.1570ITX

As stated in 21 CFR, parts 892.1550, 892.1560, and 892.1570, each of these generic types of devices has been classified as Class II.

3. Substantially Equivalent Devices

{67}------------------------------------------------

Philips Ultrasound believes the proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems are substantially equivalent to the following currently marketed devices:

Predicate Device510(k)
Philips EPIQ Diagnostic Ultrasound System and Philips AffinitiDiagnostic Ultrasound SystemK181485
Reference Device510(k)
TOMTEC-ARENA TTA2K150122

4. Device Description

The proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems are general purpose, software controlled, diagnostic ultrasound systems. Their function is to acquire ultrasound data and to display the data in various modes of operation. The devices consist of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics and optional peripherals (ECG, printers).

The removable transducers are connected to the system using a standard technology, multi-pin connectors. Other than the introductions of the two new transducers, the device description, accessories and components are unchanged, reference table 1.

5. Indications for Use

The intended use of the EPIQ, EPIQ 5, EPIQ7, Affiniti 30, Affiniti 50, and Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use:

Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.

The clinical environments where the Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.

When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.

The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated in the user

{68}------------------------------------------------

information reduces your responsibility for sound clinical judgement and best clinical procedure.

6. Technological Comparison to Predicate Devices

The proposed Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound System and the currently marketed predicate device, Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound System (K181485), employ the same fundamental scientific technology. The primary differences are the addition of the mC7-2 and V9-2 transducers to the Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound System.

A comparison of the proposed Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound Systems to the currently marketed and predicate device, Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound System is provided in Table 1 below:

TABLE 1

Comparison of the proposed Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound Systems to the currently marketed and predicate device, Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound Svstem.

StandardFeatureEPIQK181485 (Predicate Device)EPIQ 4.0(Proposed Device)AffinitiK181485(Predicate Device)Affiniti 4.0(ProposedDevice)
Indicationsfor UseFetal/Obstetric,Abdominal,Intraoperative (Vascular),Intraoperative (Cardiac),Pediatric, Small Organ (Breast,Thyroid, Testicle),Cephalic (Neonatal),Cephalic (Adult),Transrectal, TransvaginalMusculoskeletal (Conventional),Musculoskeletal (Superficial)Gynecological, Other: Urology,Cardiac Adult,Cardiac Pediatric,Transesophageal (Cardiac),Cardiac other (Fetal),Peripheral Vessel, CerebralVascular,Same as EPIQK181485Same as EPIQK181485Same as EPIQK181485
TransducersC5-1, C8-5, C9-2, C10-3v, C10-4ec, D2cwc, D5cwc, eL18-4,L12-3, L12-5 50, L15-7io, L18-5,S5-1, S7-3t, S8-3, S8-3t, S12-4,V6-2, X7-2t, X8-2t, D2tcd, X5-1,X6-1, VL13-5, X7-2, X8-2ti,XL14-3, 3D9-3vSame as EPIQK181485 withaddition of the V9-2 and cM7-2transducers.3D9-3v, C5-1, C6-2,C8-5, C9-2, C9-4V,C10-3v, C10-4ec,D2cwc, D2tcd,D5cwc, BP10-5ec,eL18-4, L12-3, L12-4, L12-5 50, L15-7io,L18-5, S4-2, S5-1,S7-3t, S8-3, S12-4,V6-2, VL13-5, X7-2tSame as AffinitiK181485 withaddition of theV9-2 and cM7-2transducers.
TransducerTypesPhased arrayLinear arrayCurved arrayPencil ProbesMulti-plane TransesophagealMotorized arrays3D Matrix ArraySame as EPIQK181485, both theV9-2 and mC7-2are curved arraytransducersPhased arrayLinear arrayCurved arrayPencil ProbesMulti-plane TransesophagealMotorizedarraysSame as EPIQK181485, boththe V9-2 andmC7-2 are curvedarray transducers
TransducerFrequency1.0 - 18.0 MHzSameSameSame
AcousticOutputDisplay &FDA LimitsIEC 62359ISPTA max=720 mw/cm2MI max =1.9MI displayTI displaySameSameSame
Imaging ModeB mode, M mode, PWD, CWD,Color Doppler, Combinedmodes,SameSameSame
ElastPQSameSameSame
Shear waveElastographyElastQ ImagingSameNANA
ElastQ Imaging LinearSame
# TransmitChannels128SameSameSame
# ReceiveChannels256SameSameSame
510(k) TrackTrack 3SameSameSame
ProductSafetyCertificationIEC 60601-1C22.2 No. 60601-1ANSI/AAMI ES60601-1IEC 60601-2-37SameSameSame
EMCComplianceIEC 60601-1-2SameSameSame
PatientContactMaterialsSame- See section 1.7.3 for an analysis of patient contacting material
AccessoriesVarious hardcopy, recordingand printing devices: videothermal printer, reportprinter, DVD, USBPercuNav InstrumentsBiopsy guidesFootswitchSameSameSame
RenderingEnhancementsTrueVue / GlassVueaRevealMaxVueQuickVueTrueVueMultiVueSame with theaddition of FlexVueand NeedleVisualizationSameSame with theaddition ofFlexVue andNeedleVisualization
QLAB Apps3DQ Advanced3DQGI 3DQ MVNA2DQ aCMQCMQ StressIMT MVIROI SQHeartmodelAutoStrain(referenceK151022)ROI IMTMVI GI 3DQVPQ SQA2DQ aCMQMVNSame

{69}------------------------------------------------

Philips Ultrasound, Inc.

Ultrasound Systems

{70}------------------------------------------------

7. Safety Considerations

As a Track 3 ultrasound device. the subject Philips EPIQ Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound Systems are designed to comply with the acoustic output display requirements of IEC 60601-2-37 Ed 2.1 (Particular requirements for the basic safety and essential performance of ultrasonic medical and monitoring equipment) and IEC 62359, Ed 2.0 (Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields).

The subject devices, Philips EPIO Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems, comply with the referenced standard as well as the FDA ultrasound guidance document, Guidance for Industry and FDA Staff - Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, issued on September 9, 2008.

The system acoustic output limits are identical to the previous version and are:

  • Ispta.3 < 720 MW/cm2 ■
  • MI < 1.9
  • I TI < 6.0

The system and transducers are compliant to:

  • ANSI/AAMI 60601-1: Medical electrical equipment. General requirements I for basic safety and essential performance, 2005. Amendment 1, 2012
  • . IEC 60601-1-2 Medical Electrical Equipment - Part 1-2, General Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility, 2014
  • I IEC 60601-2-37: Medical electrical equipment. Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2015
  • I IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017

8. Nonclinical Performance Data

Philips Ultrasound performed the following testing to ensure the safety and effectiveness of the proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems:

I ANSI/AAMI 60601-1: Medical electrical equipment. General requirements for basic

safety and essential performance, 2005, Amendment 1, 2012

  • IEC 60601-1-2 Medical Electrical Equipment - Part 1-2, General Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility, 2014
  • IEC 60601-2-37: Medical electrical equipment. Particular requirements for the

{71}------------------------------------------------

basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2015

  • IEC 62359. Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017
  • ISO 10993: Biological evaluation of medical devices, 2009.

Non-Clinical verification testing has been performed to cover system level requirements and the risk control measures can be located in Appendix M. Non-Clinical validation testing covered the intended use and commercial claims as well as usability testing with representative intended users.

All testing was used to support substantial equivalence of the subject device and demonstrate that the proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems:

  • comply with the aforementioned international and FDA-recognized consensus ■ standards and FDA ultrasound guidance document, and
  • meet the acceptance criteria and are adequate for their intended use. I

Therefore, the proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems are substantially equivalent to the predicate Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems in terms of safety and effectiveness.

Quality assurance measures applied to the system design and development include, but were not limited to:

  • Risk Analysis
  • Product Specifications
  • l Design Reviews
  • Verification and Validation

9. Clinical Data

The proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems did not require clinical studies since substantial equivalence to the currently marketed predicate devices has been established.

10. Conclusion

For these tests, all pre-determined acceptance criteria were met. The results of these tests show that the proposed Philips EPIO Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems with the new V9-2 and mC7-2 transducers met their intended use. The predicate device and the proposed Philips EPIO Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems:

  • indicated for the diagnostic ultrasonic imaging and fluid flow analysis.
  • have the same gray-scale and Doppler capabilities.

{72}------------------------------------------------

  • I use essentially the same technologies for imaging, Doppler functions and signal processing.
  • I have acoustic output levels within the Track 3 FDA limits.
  • I manufactured under equivalent quality systems.
  • l manufactured of materials with equivalent bio safety.
  • I designed and manufactured to the same electrical and physical safety standards.

The differences between the predicate and subject device do not raise new questions of safety and/or effectiveness and support a determination of substantial equivalence.

514 Performance Standards

There are no Sec. 514 performance standards for the proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems.

Prescription Status

The proposed Philips EPIO Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems are prescription devices. The prescription device statement appears in the labeling.

Sterilization Sites

Not applicable. No components are supplied sterile.

Track

The proposed Philips EPIQ Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems are Track 3 systems.

N/A