K Number
K182839

Validate with FDA (Live)

Date Cleared
2019-02-28

(142 days)

Product Code
Regulation Number
870.3600
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended uses of Lifetech Cardio Model 8301 Temporary Pacemaker are

  1. to provide temporary single chamber synchronous or asynchronous anti-brady pacing therapy, and
  2. to test electrical performance of an implanted lead system, including pacing threshold, sensing sensitivity and impedance.
    Lifetech Cardio Model 8301 Temporary Pacemaker is only intended to be operated
  3. in a clinical setting, and
  4. by trained professionals.
    Indications for anti-brady pacing therapy may be based on symptomatic bradycardia conditions that:
  5. results from an acute and reversible cause and will likely not require permanent pacing, and
  6. causes symptoms and/or severe hemodynamic impairment and when permanent cardiac pacing is not immediately indicated or available.
    Examples of specific indications for temporary pacing may include:
    a) Sick sinus syndrome
    b) Sinus bradycardia
    c) Atrial and/or ventricular arrhythmias
    d) Complete atrioventricular block
    e) Asystole
    f) Bradycardia accompanied by congestive heart failure
    g) Patient support, management and evaluation before the implantable pacemaker
    h) Support during the replacement of implantable pacemaker
    i) Cardiac complications occurring during intervention or surgery
    j) Support after the cardiac surgery
    k) Acute myocardial infarction complicated with cardiac conduction block
Device Description

Lifetech Cardio Model 8301 Temporary Pacemaker (hereinafter called "8301 TPM") is a handheld device powered by two common size AA 1.5V Alkaline (LR6) batteries, which is intended for temporary single chamber anti-brady pacing therapy and implanted system analysis. 8301 TPM offers a complete set of pacing and sensing controls and supports either SSI (synchronous) or SOO (asynchronous) pacing modes. Its analysis features enable pacing threshold, sensing sensitivity and impedance measurement.
Patient Cable is an accessory provided along with 8301 TPM and is designed to connect a pacing lead system to 8301 TPM.

AI/ML Overview

This document is a 510(k) premarket notification for the "Lifetech Cardio Model 8301 Temporary Pacemaker." It describes the device's intended use, technological characteristics, and performance compared to a predicate device, the "Medtronic Model 5318 Temporary Pacemaker / Implant Tool."

Here's an analysis of the provided information to answer your questions regarding acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of specific numerical acceptance criteria (e.g., "pacing rate accuracy within X%") alongside reported performance values. Instead, it makes a general statement of compliance with standards and successful verification and validation.

However, based on the Summary of Testing section and the Conclusion, we can infer the acceptance criteria categories and the high-level assessment of performance:

Acceptance Criteria CategoryReported Device Performance
Electrical SafetyComplies with IEC 60601-1 (general medical electrical equipment safety) and IEC 60601-2-31 (specific safety requirements for external pacemakers).
Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2 (general requirements for basic safety and essential performance – electromagnetic compatibility).
Software Functionality & SafetySoftware verification and validation testing conducted, and documentation provided as recommended by FDA guidance ("Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"). The software was considered "major" level of concern, implying stringent testing was applied to ensure that "a failure or latent flaw in the software could [not] directly result in serious injury or death to the patient or operator."
Physical CharacteristicsBench testing confirmed adherence to established requirements. (Specific characteristics and numerical criteria are not detailed in this summary.)
Electrical Parameters & FeaturesBench testing confirmed adherence to established requirements for pacing parameters and analysis features (pacing threshold, sensing sensitivity, impedance measurement). (Specific parameters and numerical criteria are not detailed in this summary.)
Cleaning Method ValidationBench testing confirmed the effectiveness and safety of the cleaning method. (Specific criteria not detailed.)
Hardware ReliabilityBench testing confirmed adherence to established requirements. (Specific criteria not detailed.)
Environmental RequirementsBench testing confirmed adherence to established requirements (e.g., operating temperature, humidity, storage conditions). (Specific criteria not detailed.)
Overall Performance & Intended UseThe device "performs as intended when used in accordance with its labeling" and is "substantially equivalent to Medtronic Model 5318 Temporary Pacemaker / Implant Tool." This implies the device meets the performance expectations for temporary single-chamber anti-brady pacing therapy and testing electrical performance of an implanted lead system.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not provide details on specific sample sizes for particular tests (e.g., number of devices tested for electrical safety, number of software test cases). It broadly states that "Design verification and validation was performed."

  • Sample Size: Not explicitly stated for any specific test. It implies adequate samples were used to demonstrate compliance with the cited standards and internal criteria.
  • Data Provenance: The testing was conducted by Shenzhen Lifetech Cardio Medical Electronics Co., Ltd. (based in China) as part of their design verification and validation process. The data would therefore originate from their internal testing activities.
  • Retrospective or Prospective: These would be prospective tests conducted as part of the device's development and regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable to the type of device and testing described. The "Lifetech Cardio Model 8301 Temporary Pacemaker" is a hardware medical device with embedded software. Its performance is evaluated against engineering specifications, international standards (IEC), and its ability to replicate the functions of a predicate device.

Ground truth, in the context of expert consensus, pathology, or outcomes data, is typically relevant for diagnostic AI/ML devices or devices whose output requires subjective interpretation. For this pacemaker, ground truth is established by objective measurements against established engineering principles and medical device standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 are used to resolve discrepancies in expert interpretations, primarily in studies involving human readers or subjective assessments (e.g., image analysis, clinical diagnoses). The testing of this pacemaker involves objective, measurable electrical and mechanical performance parameters, and compliance with standards, not subjective expert judgment requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. MRMC studies are relevant for diagnostic devices, especially those incorporating AI, where the device's output influences human readers' performance (e.g., radiologists interpreting images). The Lifetech Cardio Model 8301 Temporary Pacemaker is a therapeutic and diagnostic tool for direct patient intervention (pacing) and objective measurements (threshold, sensitivity, impedance), not an AI-powered diagnostic aide that assists human readers in making interpretations.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is a "standalone" medical device in the sense that its core functions (pacing, sensing, measurement) are performed by the device and its embedded software. The software verification and validation testing described in the document is essentially evaluating the "algorithm only" performance within the device's complete system. The device is intended to be operated "by trained professionals," meaning there's always a human in the loop for clinical application and decision-making, but the internal functions and measurements are automated by the device's algorithms and hardware.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is based on:

  • Engineering specifications and design requirements: These define the expected performance metrics (e.g., pacing rate range, output current accuracy, sensing sensitivity limits).
  • International standards: Specifically cited are IEC 60601-1, IEC 60601-2-31, and IEC 60601-1-2. Compliance with these standards indicates adherence to generally accepted safety and performance criteria for external pacemakers.
  • Equivalence to a predicate device: The fundamental claim is substantial equivalence to the Medtronic Model 5318. This implies that the performance (pacing parameters, analysis features, safety, etc.) of the Lifetech device should be comparable to or meet the established performance of the predicate.
  • FDA guidance: For software, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" served as the basis for software verification and validation.

8. The sample size for the training set

Not applicable. This device is not an AI/ML device that requires a "training set" in the conventional machine learning sense. The software it contains is traditional embedded software, developed using standard software engineering principles and verified against specifications, not trained on a dataset.

9. How the ground truth for the training set was established

Not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

February 28, 2019

Shenzhen Lifetech Cardio Medical Electronics Co., Ltd. Stephy Pan Quality and Regulatory Affairs Manager Rm 201, Bldg No.2, Huahan Sci-tech Industrial Park, No.16 Jinniu West Rd Shenzhen, CN 518000 Guangdong

Re: K182839

Trade/Device Name: Lifetech Cardio Model 8301 Temporary Pacemaker Regulation Number: 21 CFR 870.3600 Regulation Name: External Pacemaker Pulse Generator Regulatory Class: Class III Product Code: DTE Dated: January 9, 2019 Received: January 17, 2019

Dear Stephy Pan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

If

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182839

Device Name

Lifetech Cardio Model 8301 Temporary Pacemaker

Indications for Use (Describe)

The intended uses of Lifetech Cardio Model 8301 Temporary Pacemaker are

  1. to provide temporary single chamber synchronous or asynchronous anti-brady pacing therapy, and

  2. to test electrical performance of an implanted lead system, including pacing threshold, sensing sensitivity and impedance.

Lifetech Cardio Model 8301 Temporary Pacemaker is only intended to be operated

  1. in a clinical setting, and

  2. by trained professionals.

Indications for anti-brady pacing therapy may be based on symptomatic bradycardia conditions that:

  1. results from an acute and reversible cause and will likely not require permanent pacing, and

  2. causes symptoms and/or severe hemodynamic impairment and when permanent cardiac pacing is not immediately indicated or available.

Examples of specific indications for temporary pacing may include:

a) Sick sinus syndrome

b) Sinus bradycardia

c) Atrial and/or ventricular arrhythmias

d) Complete atrioventricular block

e) Asystole

f) Bradycardia accompanied by congestive heart failure

g) Patient support, management and evaluation before the implantable pacemaker

h) Support during the replacement of implantable pacemaker

i) Cardiac complications occurring during intervention or surgery

i) Support after the cardiac surgery

k) Acute myocardial infarction complicated with cardiac conduction block

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared:September 25th, 2018
510(k) Owner/Submitter:Shenzhen Lifetech Cardio Medical Electronics Co.,Ltd.
Address:Shenzhen Lifetech Cardio Medical Electronics Co.,Ltd.A07-A08, Room 201, Building No. 2, Huahan Sci-techIndustrial Park, No. 16 Jinniu West Road, PingshanDistrict, Shenzhen, 518000, P. R. China
Contact:Stephy PanQuality and Regulatory Affairs Manager
Telephone:+86 755 86026855-8833+86 13632729557
Fax:+86 755 8602651
E-mail:panyuying@lifetechmed.com
Proprietary Name:Lifetech Cardio Model 8301 Temporary Pacemaker
Common Name:External Pacemaker Pulse Generator
Device Classification:Class II (special controls), 21 CFR 870.3600External Pacemaker Pulse Generator
Product Code:DTE
Regulation Medical SpecialtyCardiovascular

A. General Information

B. Predicate Device

Trade NameMedtronic Model 5318 Temporary Pacemaker /Implant Tool
510(k) NumberK971474
Product CodeDTE
ManufacturerMedtronic, Inc.

C. Device Description

Lifetech Cardio Model 8301 Temporary Pacemaker (hereinafter called "8301 TPM") is a handheld device powered by two common size AA 1.5V Alkaline (LR6) batteries, which is intended for temporary single chamber anti-brady pacing therapy and implanted system analysis. 8301 TPM offers a complete set of pacing and sensing controls and supports either SSI (synchronous) or SOO (asynchronous) pacing modes. Its analysis features enable pacing threshold, sensing sensitivity and impedance

{4}------------------------------------------------

measurement.

Patient Cable is an accessory provided along with 8301 TPM and is designed to connect a pacing lead system to 8301 TPM.

D. Intended Use/ Indications for Use

The intended uses of Lifetech Cardio Model 8301 Temporary Pacemaker are

  1. to provide temporary single chamber synchronous or asynchronous anti-brady pacing therapy, and

  2. to test electrical performance of an implanted lead system, including pacing threshold, sensing sensitivity and impedance.

Lifetech Cardio Model 8301 Temporary Pacemaker is only intended to be operated

    1. in a clinical setting, and
    1. by trained professionals.

Indications for anti-brady pacing therapy may be based on symptomatic bradycardia conditions that:

  1. results from an acute and reversible cause and will likely not require permanent pacing, and

  2. causes symptoms and/or severe hemodynamic impairment and when permanent cardiac pacing is not immediately indicated or available.

Examples of specific indications for temporary pacing may include:

  • a) Sick sinus syndrome
  • b) Sinus bradycardia
  • c) Atrial and/or ventricular arrhythmias
  • d) Complete atrioventricular block
  • e) Asystole
  • f) Bradycardia accompanied by congestive heart failure
  • g) Patient support, management and evaluation before the implantation of implantable pacemaker
  • h) Support during the replacement of implantable pacemaker
  • i) Cardiac complications occurring during intervention or surgery
  • j) Support after the cardiac surgery
  • k) Acute myocardial infarction complicated with cardiac conduction block

E. Technological Characteristics

Intended use, technological characteristic and performance are substantially equivalent to the predicate devices referenced.

{5}------------------------------------------------

Proposed product: Lifetech Cardio Model 8301 Temporary Pacemaker When compared to the predicate device, Lifetech Cardio Model 8301 Temporary

Pacemaker presented in this submission has the same:

Intended Use

  • Intended Use ●
  • . Anatomical Sites
  • Target Population .
  • . Indications for Use
  • Contraindications

Technological Characteristic

  • Accessory
  • Physical Characteristics
  • Principles of Operation

Performance

  • Pacing Parameters ●
  • Analysis Features .
  • Safety Features .
  • Standards ●

F. Summary of Testing

Design verification and validation was performed to demonstrate that Lifetech Cardio Model 8301 Temporary Pacemaker and its accessory meet established performance criteria to support substantial equivalence to the referenced predicate device.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Lifetech Cardio Model 8301 Temporary Pacemaker with associated cables as necessary for test. The system complies with the IEC 60601-1 and IEC 60601-2-31 standards for safety and the IEC 60601-1-2 of the standard for EMC.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

Bench Testing

  • Physical Characteristics
  • Electrical Parameters and Features ●
  • Cleaning Method Validation ●
  • Hardware Reliability

{6}------------------------------------------------

Proposed product: Lifetech Cardio Model 8301 Temporary Pacemaker

K182839 Version: A/0

  • Environmental Requirements

G. Conclusion

Design verification and validation results demonstrate that Lifetech Cardio Model 8301 Temporary Pacemaker performs as intended when used in accordance with its labeling and is considered substantially equivalent to Medtronic Model 5318 Temporary Pacemaker / Implant Tool .

§ 870.3600 External pacemaker pulse generator.

(a)
Identification. An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment.
(2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source (
i.e., battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following:
(i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling;
(ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use;
(iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices (
e.g. data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use;
(ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals;
(iii) The labeling must list all pacing modes available in the device;
(iv) Labeling must include a detailed description of any special capabilities (
e.g., overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.