K Number
K182719

Validate with FDA (Live)

Date Cleared
2019-06-19

(264 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quidel Triage® TOX Drug Screen, 94600 is a fluorescence immunoassay to be used with the Quidel Triage® MeterPro for the qualitative determination of the presence of drugs and/or metabolites in human urine of up to 9 drug assays at or above the threshold concentrations. The threshold concentrations are provided below:

AbbreviationAnalyteCalibratorCutoff
AMPAmphetaminesd-Amphetamine500 ng/mL
mAMPMethamphetaminesd-Methamphetamine500 ng/mL
BARBarbituratesButalbital200 ng/mL
BZOBenzodiazepinesTemazepam200 ng/mL
COCCocaineBenzoylecgonine150 ng/mL
EDDPMethadone MetaboliteEDDP100 ng/mL
OPIOpiatesMorphine300 ng/mL
THCCannabinoids11-nor-9-carboxy-Δ9-THC50 ng/mL
TCATricyclic AntidepressantsDesipramine1000 ng/mL

This test provides only a preliminary test result. Clinical consideration and professional judgment must be applied to any drug test result, particularly in evaluating a preliminary positive result. A more specific alternate chemical must be used to obtain a confirmed analytical result. Gas Chromatography / Mass Spectroscopy (GC/MS), Liquid Chromatography / Mass Spectroscopy / Mass Spectroscopy (LC-MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods.

Quidel Triage® MeterPro:

The Quidel Triage® MeterPro is a portable fluorescence instrument used to measure the results of tests manufactured by Quidel Cardiovascular Inc. The Quidel Triage® MeterPro can be used in a laboratory or in a point-of-care setting.

Device Description

Quidel Triage® TOX Drug Screen, 94600:

The Quidel Triage® TOX Drug Screen, 94600 is a single use test device and is used in conjunction with the Quidel Triage® MeterPro. The device contains murine monoclonal antibody conjugates and drug conjugates labeled with a fluorescent dye or immobilized on the solid phase and stabilizers. The testing device is inserted into and read by the Quidel Triage® MeterPro. Threshold concentrations are used to separate a negative result from a presumptive positive result.

Quidel Triage® MeterPro:

The Quidel Triage MeterPro is a portable fluorescence instrument used to measure the results of tests manufactured by Quidel Cardiovascular Inc. The Quidel Triage MeterPro can be used in a laboratory or in a point-of-care setting.

The Quidel Triage MeterPro uses a laser as a light source. Light from the laser hits a test device that has been inserted in the meter. This causes the fluorescent dye in the test device to give off energy. The more energy the fluorescent dye gives off, the stronger the signal.

AI/ML Overview

The provided document describes the Quidel Triage® TOX Drug Screen, 94600 and the Quidel Triage® MeterPro for the qualitative determination of the presence of drugs and/or metabolites in human urine. The acceptance criteria and the study results are detailed in the "Performance Characteristics" section (1.13) and "Comparison studies" section (1.13.2.a).

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria for each analyte are based on the percentage of positive and negative results at concentrations around the cutoff, as shown in the "Precision/Reproducibility" data in section 1.13.1.a and "Assay cut-off" in section 1.13.1.f. The primary performance metric from the method comparison study (1.13.2.a) is the agreement between the Quidel Triage TOX Drug Screen, 94600 and GC/MS or LC-MS/MS values, particularly for specimens near the threshold and outside the threshold.

Below is a summary table, combining data from the precision/reproducibility testing (1.13.1.a) and the method comparison study (1.13.2.a). For the precision/reproducibility, a high percentage of correct results (e.g., all negative below -75% of cutoff, all positive above +50% of cutoff) would be considered acceptance. For method comparison, "agreement" is implicitly the acceptance criteria, with discordant results requiring explanation.

Analyte (Cutoff)Performance Criteria (e.g., Precision/Reproducibility: Expected % agreement)Reported Device Performance (Precision/Reproducibility)Reported Device Performance (Method Comparison: Agreement with GC/MS/LC-MS/MS, excluding discordant resolved cases)
AMP (500 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 126 ng/mL (-75%): 720/720 Neg (100%) - 760 ng/mL (+50%): 720/720 Pos (100%) - 522 ng/mL (Cutoff): 50 Neg/669 Pos (89.5% Pos)Out of 220 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 11/11 Neg (100%) - Near Threshold Positive (within 50% above threshold): 2 Neg/8 Pos (80% Pos) - Positive (>150% of threshold): 2 Neg/98 Pos (98% Pos) (Discordant results acknowledged and explained related to isomeric cross-reactivity.)
mAMP (500 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 250 ng/mL (-50%): 720/720 Neg (100%) - 742 ng/mL (+50%): 720/720 Pos (100%) - 529 ng/mL (Cutoff): 281 Neg/431 Pos (60.0% Pos)Out of 218 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 5 Neg/5 Pos (50% Pos) - Near Threshold Positive (within 50% above threshold): 5 Neg/6 Pos (54.5% Pos) - Positive (>150% of threshold): 7 Neg/91 Pos (92.9% Pos) (Discordant results acknowledged, with some attributed to isomeric differences in cross-reactivity.)
BAR (200 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 53 ng/mL (-75%): 736/736 Neg (100%) - 306 ng/mL (+50%): 719/719 Pos (100%) - 192 ng/mL (Cutoff): 111 Neg/605 Pos (84.5% Pos)Out of 218 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 3 Neg/8 Pos (72.7% Pos) - Near Threshold Positive (within 50% above threshold): 0 Neg/11 Pos (100% Pos) - Positive (>150% of threshold): 0 Neg/97 Pos (100% Pos) (Discordant results acknowledged and explained.)
BZO (200 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 58 ng/mL (-75%): 704/704 Neg (100%) - 306 ng/mL (+50%): 720/720 Pos (100%) - 219 ng/mL (Cutoff): 318 Neg/402 Pos (55.9% Pos)Out of 221 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 1 Neg/10 Pos (90.9% Pos) - Near Threshold Positive (within 50% above threshold): 1 Neg/10 Pos (90.9% Pos) - Positive (>150% of threshold): 0 Neg/99 Pos (100% Pos) (Discordant results acknowledged and explained.)
COC (150 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 76 ng/mL (-50%): 719/719 Neg (100%) - 218 ng/mL (+50%): 736/736 Pos (100%) - 157 ng/mL (Cutoff): 26 Neg/694 Pos (96.4% Pos)Out of 220 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 5 Neg/6 Pos (54.5% Pos) - Near Threshold Positive (within 50% above threshold): 0 Neg/11 Pos (100% Pos) - Positive (>150% of threshold): 0 Neg/99 Pos (100% Pos) (Discordant results acknowledged and explained.)
EDDP (100 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 29 ng/mL (-75%): 720/720 Neg (100%) - 143 ng/mL (+50%): 716/716 Pos (100%) - 111 ng/mL (Cutoff): 126 Neg/594 Pos (82.5% Pos)Out of 220 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 9 Neg/1 Pos (10.0% Pos) - Near Threshold Positive (within 50% above threshold): 2 Neg/10 Pos (83.3% Pos) - Positive (>150% of threshold): 0 Neg/98 Pos (100% Pos) (Discordant results acknowledged and explained.)
OPI (300 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 165 ng/mL (-50%): 722/722 Neg (100%) - 480 ng/mL (+50%): 720/720 Pos (100%) - 344 ng/mL (Cutoff): 197 Neg/507 Pos (72.0% Pos)Out of 220 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 2 Neg/8 Pos (80% Pos) - Near Threshold Positive (within 50% above threshold): 0 Neg/11 Pos (100% Pos) - Positive (>150% of threshold): 1 Neg/98 Pos (99.0% Pos) (Some discordant results explained by secondary reference testing.)
THC (50 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 12 ng/mL (-75%): 716/716 Neg (100%) - 78 ng/mL (+50%): 4 Neg/700 Pos (99.4% Pos) - 54 ng/mL (Cutoff): 163 Neg/559 Pos (77.4% Pos)Out of 221 samples: - Negative (<50% of threshold): 99/99 Neg (100%) - Near Threshold Negative (within 50% below threshold): 10 Neg/1 Pos (9.1% Pos) - Near Threshold Positive (within 50% above threshold): 3 Neg/8 Pos (72.7% Pos) - Positive (>150% of threshold): 0 Neg/99 Pos (100% Pos) (Discordant results acknowledged and explained.)
TCA (1000 ng/mL)At -75% and below: 100% Neg; At +50% and above: 100% Pos- Neg Control (0%): 720/720 Neg (100%) - 498 ng/mL (-50%): 719/719 Neg (100%) - 1577 ng/mL (+50%): 1 Neg/719 Pos (99.9% Pos) - 996 ng/mL (Cutoff): 218 Neg/518 Pos (70.4% Pos)Out of 220 samples: - Negative (<50% of threshold): 98/98 Neg (100%) - Near Threshold Negative (within 50% below threshold): 1 Neg/10 Pos (90.9% Pos) - Near Threshold Positive (within 50% above threshold): 0 Neg/11 Pos (100% Pos) - Positive (>150% of threshold): 1 Neg/95 Pos (99% Pos) (Some discordant results explained by data entry errors or pH interference.)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Precision/Reproducibility Test Set:
    • For each of the 9 analytes, 720-736 samples were tested for each concentration point (0 to +100% of cutoff, plus negative control). This means for each analyte, n=720-736 replicates were performed at each concentration.
    • Data provenance: Not explicitly stated, but the submission is for the FDA in the US. The study involved "three (3) study sites" which implies data collection within a controlled, prospective study setting, most likely in the US, given the FDA submission. The samples were "blinded and randomized" prior to testing.
  • Method Comparison Test Set:
    • The total number of samples for each analyte in the method comparison study with GC/MS or LC-MS/MS varies slightly per analyte but is around 220 samples (e.g., for AMP, 99+11+2+98+0+0+8 = 218 specimens + 2 discordant cases).
    • Data provenance: The samples were described as "unaltered urine specimens," implying they were clinical samples. The document does not explicitly state the country of origin or whether the data was retrospective or prospective. However, the use of "patient specimens" strongly suggests prospective collection for the purpose of the study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • For the Precision/Reproducibility study, the ground truth was established by the prepared concentrations of the analytes (e.g., negative control, 25%, 50%, 75%, cutoff, 125%, 150%, 175%, and 200% of cutoff). This does not involve human experts establishing ground truth for individual samples, but rather analytical controls.
  • For the Method Comparison study, the ground truth was established by a "reference method," specifically GC/MS or LC-MS/MS values. These are analytical methods considered the gold standard for drug confirmation and quantification. This does not involve human experts establishing ground truth for each case, but rather the laboratory performing these reference methods. No specific number or qualifications of experts operating these reference instruments are mentioned, as the results of these instruments are considered the objective ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • For the Precision/Reproducibility study, no adjudication method is explicitly described as the results are based on measured values against known concentrations.
  • For the Method Comparison study, particularly for discordant results between the device and the reference method, an adjudication process did occur. In some cases, explanations are provided for the discordance (e.g., isomeric cross-reactivity for AMP and mAMP, the presence of multiple metabolites for BZO and OPI, pH interference for TCA, or data entry errors). For OPI, one discordant case (Specimen ID 572644) was sent to a "second reference testing laboratory" for reconfirmation, which then aligned with the device's negative result. This indicates a form of secondary confirmation/adjudication by an additional reference lab for specific discordant cases, rather than a typical 2+1 or 3+1 expert consensus model for image/clinical interpretation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This document describes an in vitro diagnostic (IVD) device that performs qualitative drug screening using a fluorescence immunoassay and a meter (Quidel Triage® MeterPro). It is not an AI-assisted diagnostic device, nor does it involve human readers interpreting images or clinical data. Therefore, an MRMC comparative effectiveness study involving human readers with and without AI assistance is not applicable to this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The Quidel Triage® TOX Drug Screen, 94600 with the Quidel Triage® MeterPro functions as a standalone device in terms of producing a "POS" or "NEG" result for each drug assay. The meter reads the test device and provides a direct qualitative result. While the meter requires a human operator to insert the device and, if desired, print or transmit results, the interpretation of the fluorescence signal into a positive or negative drug screen result is entirely performed by the instrument's programming, which is equivalent to an "algorithm only" performance. The performance data presented in the precision/reproducibility and method comparison studies directly reflect this standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the analytical performance studies (precision/reproducibility, analytical specificity, cutoff characterization), the ground truth was established by known prepared concentrations of drugs/metabolites or spiking experiments.
  • For the method comparison study, the ground truth was established by Gas Chromatography / Mass Spectroscopy (GC/MS) or Liquid Chromatography / Mass Spectroscopy / Mass Spectroscopy (LC-MS/MS) values. These are analytical "gold standard" methods.

8. The sample size for the training set

  • The document describes a 510(k) submission for a new IVD device and presents validation studies. It does not explicitly mention a "training set" in the context of machine learning or AI models. The studies are designed to demonstrate the analytical performance and substantial equivalence of the device to existing predicate devices. Therefore, a specific training set sample size is not applicable/not provided in this type of submission. The performance data is from a test/validation set.

9. How the ground truth for the training set was established

  • As a "training set" is not explicitly referenced in the context of AI/machine learning development, the method of establishing its ground truth is not applicable/not provided. The ground truth for the validation studies (test sets) was established via known concentrations of analytes and reference methods (GC/MS, LC-MS/MS) as detailed above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 19, 2019

Quidel Cardiovascular Inc. Rachael Williamson Senior Manager, Regulatory Affairs 9975 Summers Ridge Road San Diego, CA 92121

Re: K182719

Trade/Device Name: Quidel Triage® TOX Drug Screen, 94600 Quidel Triage® MeterPro Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ, LAF, DIS, JXM, JXO, DJR, DJG, LDJ, LFG, KHO Dated: June 13, 2019 Received: June 14, 2019

Dear Rachael Williamson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182719

Device Name

Quidel Triage® TOX Drug Screen, 94600 Quidel Triage® MeterPro

Indications for Use (Describe)

Quidel Triage® TOX Drug Screen, 94600:

The Quidel Triage® TOX Drug Screen, 94600 is a fluorescence immunoassay to be used with the Quidel Triage® MeterPro for the qualitative determination of the presence of drugs and/or metabolites in human urine of up to 9 drug assays at or above the threshold concentrations. The threshold concentrations are provided below:

AbbreviationAnalyteCalibratorCutoff
AMPAmphetaminesd-Amphetamine500 ng/mL
mAMPMethamphetaminesd-Methamphetamine500 ng/mL
BARBarbituratesButalbital200 ng/mL
BZOBenzodiazepinesTemazepam200 ng/mL
COCCocaineBenzoylecgonine150 ng/mL
EDDPMethadone MetaboliteEDDP100 ng/mL
OPIOpiatesMorphine300 ng/mL
THCCannabinoids11-nor-9-carboxy-Δ9-THC50 ng/mL
TCATricyclic AntidepressantsDesipramine1000 ng/mL

This test provides only a preliminary test result. Clinical consideration and professional judgment must be applied to any drug test result, particularly in evaluating a preliminary positive result. A more specific alternate chemical must be used to obtain a confirmed analytical result. Gas Chromatography / Mass Spectroscopy (GC/MS), Liquid Chromatography / Mass Spectroscopy / Mass Spectroscopy (LC-MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods.

Quidel Triage® MeterPro:

The Quidel Triage® MeterPro is a portable fluorescence instrument used to measure the results of tests manufactured by Quidel Cardiovascular Inc. The Quidel Triage® MeterPro can be used in a laboratory or in a point-of-care setting.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square shape divided into smaller triangles, each filled with a different color of the rainbow, creating a vibrant, mosaic-like effect. Below the colorful square is the word "QUIDEL" in a simple, sans-serif font, with the registered trademark symbol next to it.

K182719

1. 510(K) SUMMARY

1.1. Date Prepared:

June 18, 2019

1.2. Purpose for Submission:

New device

1.3. Measurand:

Amphetamine, Methamphetamine, Barbiturates, Benzodiazepines, Cocaine, EDDP (2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine), Opiates, Cannabinoids, Tricyclic Antidepressants

1.4. Type of Test:

Qualitative, lateral flow immunofluorescence

1.5. Applicant:

Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, California 92121 Telephone: 858-302-0334 Fax: 858-805-8622

Rachael S. Williamson (Submission Contact)

1.6. Proprietary and Established Names:

Quidel Triage® TOX Drug Screen, 94600 Quidel Triage® MeterPro

Regulatory Information: 1.7.

Quidel Triage® TOX Drug Screen, 94600:

ProductCodeClassificationRegulatory SectionPanel
DKZClass II21 CFR 862.3100,Amphetamine test systemToxicology (91)

{5}------------------------------------------------

Image /page/5/Picture/9 description: The image features the logo for QuidelOrtho Corporation. The logo consists of a square shape divided into two sections. The larger section is a gradient of colors, transitioning from red to yellow to green to blue. The smaller section is a darker shade of blue. Below the square is the word "QUIDEL" in a simple, sans-serif font.

ProductCodeClassificationRegulatory SectionPanel
LAFClass II21 CFR 862.3610,Methamphetamine testsystemToxicology (91)
DISClass II21 CFR 862.3150,Barbiturate test systemToxicology (91)
JXMClass II21 CFR 862.3170,Benzodiazepine test systemToxicology (91)
JXOClass II21 CFR 862.3250, Cocaineand cocaine metabolite testsystemToxicology (91)
DJRClass II21 CFR 862.3260,Methadone test systemToxicology (91)
DJGClass II21 CFR 862.3650, Opiatetest systemToxicology (91)
LDJClass II21 CFR 862.3870,Cannabinoid test systemToxicology (91)
LFGClass II21 CFR 862.3910, Tricyclicantidepressant drugs testsystemToxicology (91)

Quidel Triage® MeterPro:

Product CodeClassificationRegulatory SectionPanel
KHOClass I21 CFR 862.2560, Fluorometer forclinical useClinicalChemistry

Intended Use: 1.8.

Quidel Triage® TOX Drug Screen, 94600:

The Quidel Triage® TOX Drug Screen, 94600 is a fluorescence immunoassay to be used with the Quidel Triage® MeterPro for the qualitative determination of the presence of drugs and/or

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Quidel. The logo features a square with rounded corners that is divided into two sections. The top section is a gradient of colors, including yellow, orange, red, purple, and blue. The bottom section is a solid blue color. Below the square is the word "QUIDEL" in a sans-serif font.

metabolites in human urine of up to 9 drug assays at or above the threshold concentrations. The threshold concentrations are provided below:

AbbreviationAnalyteCalibratorCutoff
AMPAmphetaminesd-Amphetamine500 ng/mL
mAMPMethamphetaminesd-Methamphetamine500 ng/mL
BARBarbituratesButalbital200 ng/mL
BZOBenzodiazepinesTemazepam200 ng/mL
COCCocaineBenzoylecgonine150 ng/mL
EDDPMethadone MetaboliteEDDP100 ng/mL
OPIOpiatesMorphine300 ng/mL
THCCannabinoids11-nor-9-carboxy- $\Delta$ -THC50 ng/mL
TCATricyclic AntidepressantsDesipramine1000 ng/mL

This test provides only a preliminary test result. Clinical consideration and professional judgment must be applied to any drug test result, particularly in evaluating a preliminary positive result. A more specific alternate chemical must be used to obtain a confirmed analytical result. Gas Chromatography / Mass Spectroscopy (GC/MS), Liquid Chromatography / Mass Spectroscopy / Mass Spectroscopy (LC-MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods.

Quidel Triage® MeterPro:

The Quidel Triage® MeterPro is a portable fluorescence instrument used to measure the results of tests manufactured by Quidel Cardiovascular Inc. The Quidel Triage® MeterPro can be used in a laboratory or in a point-of-care setting.

1.9. Device Description:

Quidel Triage® TOX Drug Screen, 94600:

The Quidel Triage® TOX Drug Screen, 94600 is a single use test device and is used in conjunction with the Quidel Triage® MeterPro. The device contains murine monoclonal antibody conjugates and drug conjugates labeled with a fluorescent dye or immobilized on the solid phase and stabilizers. The testing device is inserted into and read by the Quidel Triage® MeterPro. Threshold concentrations are used to separate a negative result from a presumptive positive result.

Quidel Triage® MeterPro:

The Quidel Triage MeterPro is a portable fluorescence instrument used to measure the results of tests manufactured by Quidel Cardiovascular Inc. The Quidel Triage MeterPro can be used in a laboratory or in a point-of-care setting.

The Quidel Triage MeterPro uses a laser as a light source. Light from the laser hits a test device that has been inserted in the meter. This causes the fluorescent dye in the test device to give off energy. The more energy the fluorescent dye gives off, the stronger the signal.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image features the logo for QuidelOrtho Corporation. The logo consists of a square shape with a smaller square cut out of the lower right corner. The larger square is filled with a gradient of colors, transitioning from green and yellow on the left to red and orange at the top, and blue and purple on the right. Below the square is the word "QUIDEL" in a bold, sans-serif font.

1.10. Substantial Equivalence Information:

    1. Predicate Device Name:
      GenPrime Drugs of Abuse (DOA) Reader System Immunalysis Barbiturates Urine Enzyme Immunoassay DRI Benzodiazepine Assay Immunalysis EDDP Specific Urine Enzyme Immunoassay Biosite Incorporated Triage TOX Drug Screen Biosite Incorporated Triage Meter
    1. Predicate 510(k) Number:
      K130082, GenPrime Drugs of Abuse (DOA) Reader System K161714, Immunalysis Barbiturates Urine Enzyme Immunoassay K173963, DRI Benzodiazepine Assay K151395, Immunalysis EDDP Specific Urine Enzyme Immunoassay K043242, Biosite Incorporated Triage® TOX Drug Screen K973547, Biosite Incorporated Triage® Meter
    1. Comparison with Predicate:

Quidel Triage® TOX Drug Screen, 94600:

Assays: Amphetamines (AMP), Methamphetamines (mAMP), Cocaine (COC), Opiates (OPI), and Cannabinoids (THC)

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)GenPrime Drugs of Abuse(DOA) Reader System(K130082)
Intended UseFor the qualitativedetermination of drugs ofabuse in human urine.Same
Assay TypeCompetitive assay, whereconcentration of drug isinversely related to the signaldetected by the instrument.Same

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with a gradient of colors, including red, orange, yellow, green, blue, and purple. The word "QUIDEL" is written in a sans-serif font below the square.

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)GenPrime Drugs of Abuse(DOA) Reader System(K130082)
System ProcedureSample is added to a singleuse test device which is thenread by the instrument. Theinstrument is designed to readmultiple assays at the sametime.Same
Specimen TypeHuman urineSame
Single-use Test DeviceYesSame
Analyte Cutoffs(ng/mL)AMP = 500mAMP = 500COC = 150OPI = 300THC = 50AMP = 500 (OS Cup; SK Cup)MET = 500 (OS Cup; SK Cup)COC = 150 (OS Cup)MOP = 300 (SK Cup)THC = 50 (OS Cup; SK Cup)
Differences
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)GenPrime Drugs of Abuse(DOA) Reader System(K130082)
Analyte Cutoffs(ng/mL)BAR = 200BZO = 200EDDP = 100TCA = 1000MTD, OXY, PCP are not panelassays and have no associatedanalyte cutoffs.BAR = 300 (OS Cup)MTD = 300 (SK Cup)MOP = 2000 (SK Cup)PCP = 25 (SK Cup)BZO, EDDP, and TCA are notpanel assays and have noassociated cutoff
Test Device FormatCassetteCup
Storage2-8°C2-30°C
Detection methodMeasures fluorescence ofdiscreet measurement zonesfor each assay.Measures density of visiblelines against background onsingle-use test device.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square shape with rounded corners, divided into four quadrants. The upper-left quadrant displays a gradient of colors resembling a rainbow, while the other three quadrants are various shades of blue. Below the square is the word "QUIDEL" in a simple, sans-serif font.

Differences
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)GenPrime Drugs of Abuse(DOA) Reader System(K130082)
Test Time and TimingMethodOperator adds sample to testdevice and operatestheinstrument.Operator manually times testdevelopment for 5 minutes andthen operates the instrument.
Measurement MethodScans the single-use testdevice to measure signals.Scans the single-use test deviceto detect a signal.
Time to ResultsResult interpretation occurs inapproximately 15 minutes.Results interpretation mustoccur between 5 and 60minutes following specimenapplication
OutputOutputs are "POS" if the resultis at or above the thresholdconcentration or "NEG" if theresult is below the thresholdconcentration. The operatorhas the option to print theresults. If connected, theMeterPro can transmit resultsto the laboratory or hospitalinformation system.Outputs "presumptivepositive", “negative”, and"invalid" test results on agraphic user interfacedisplayed on a computer screenand automatically stores resultsalong with test information.Operator has ability to printand/or export results.

The following cutoff values are being incorporated into the Quidel Triage® TOX Drug Screen, 94600 Test Device. The cutoff values have been modified to accommodate changes in the calibrator used to manufacture the Test Device for barbiturates and benzodiazepines. In addition, the proposed predicates are identified for the EDDP and TCA assays. The tables below provide the similarities and differences with the selected predicates.

Assay: Barbiturates (BAR)

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)ImmunalysisBarbituratesUrine Enzyme Immunoassay(K161714)
Intended UseFor the qualitativedetermination of drugs ofabuse in human urine.Same

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a stylized square with a gradient of colors, including red, orange, yellow, green, blue, and purple. Below the square is the word "QUIDEL" in a sans-serif font.

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)Immunalysis BarbituratesUrine Enzyme Immunoassay(K161714)
Specimen TypeHuman urineSame
Storage2-8°CSame
Measured AnalyteBARSame
Analyte Cutoff(ng/mL)BAR = 200Same
Differences
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)Immunalysis BarbituratesUrine Enzyme Immunoassay(K161714)
Intended UseNot intended for semi-quantitative determination ofthe presence of Barbiturates inhuman urine.For the semi-quantitativedetermination of the presenceof Barbiturates in humanurine.
Assay TypeCompetitiveassay, whereconcentration of drug isinversely related to the signaldetected by the instrument.Enzyme immunoassay
Antibody TypeMouse monoclonal antibodiesRecombinant and monoclonalantibodies

Assay: Benzodiazepines (BZO)

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)DRI Benzodiazepine Assay(K173963)
Intended UseFor the qualitativedetermination of drugs ofabuse in human urine.Same
Specimen TypeHuman urineSame
Storage2-8°CSame

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape divided into two sections. The left section is a gradient of colors, including red, orange, yellow, green, blue, and purple. The right section is a solid blue color. Below the square is the word "QUIDEL" in a sans-serif font.

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)DRI Benzodiazepine Assay(K173963)
Analyte Cutoff(ng/mL)BZO = 200Same
Differences
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)DRI Benzodiazepine Assay(K173963)
Intended UseNot intended for semi-quantitative determination ofthe presence ofbenzodiazepines and theirmetabolites in human urine.For the semi-quantitativedetermination of the presenceof benzodiazepines and theirmetabolites in human urine.
Assay TypeCompetitive assay, whereconcentration of drug isinversely related to the signaldetected by the instrument.Homogenous enzymeimmunoassay
Antibody TypeMouse monoclonal antibodiesPolyclonal sheep antibody

Assay: 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)Immunalysis EDDP SpecificUrine Enzyme Immunoassay(K151395)
Intended UseFor the qualitativedetermination of drugs ofabuse in human urine.Same
Specimen TypeHuman urineSame
Storage2-8°CSame
Assay calibratedEDDPSame
Measured AnalyteEDDPSame
Analyte Cutoff(ng/mL)EDDP = 100Same

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square shape divided into two parts. The upper left part is a gradient of colors, including red, orange, yellow, green, blue, and purple. The lower right part is a solid blue color. Below the square is the word "QUIDEL" in a sans-serif font.

Differences
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)Immunalysis EDDP SpecificUrine Enzyme Immunoassay(K151395)
Intended UseNot intended for semi-quantitative determination ofthe presence of EDDP inhuman urine.For the semi-quantitativedetermination of the presenceof EDDP in human urine.
Assay TypeCompetitive assay, whereconcentration of drug isinversely related to the signaldetected by the instrument.Homogenous enzymeimmunoassay
Antibody TypeMouse monoclonal antibodiesRecombinant fab antibodies
Analyte Cutoffs(ng/mL)Not intended to have cutoffsavailable at 300 ng/mL and1000 ng/mLEDDP = 300 and 1000

Assay: Tricyclic Antidepressants (TCA)

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)Triage® TOX Drug Screen(K043242)
Intended UseFor the qualitativedetermination of drugs ofabuse in human urine.Same
Assay TypeCompetitive assay, whereconcentration of drug isinversely related to the signaldetected by the instrument.Same
Specimen TypeHuman urineSame
Storage2-8°CSame
Detection methodMeasures fluorescence ofdiscreet measurement zonesfor each assay.Same
Assay calibratedDesipramineSame

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image features the logo of Quidel. The logo consists of a square divided into two parts. The left part is a gradient of colors, while the right part is a dark blue color. Below the square is the word "QUIDEL" in a sans-serif font.

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)Triage® TOX Drug Screen(K043242)
Measured AnalyteTCASame
Analyte Cutoff(ng/mL)TCA = 1000Same
Differences
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® TOX DrugScreen, 94600 (Proposed)Triage® TOX Drug Screen(K043242)
AnalytesAmphetamines,Methamphetamines,Barbiturates,Benzodiazepines, Cocaine,Methadone Metabolite,Opiates, Cannabinoids (THC),and Tricyclic AntidepressantsAcetaminophen,Amphetamines,Methamphetamines,Barbiturates, Benzodiazepines,Cocaine, Opiates,Phencyclidine, THC, andTricyclic Antidepressants

Quidel Triage® MeterPro:

Last, the proposed predicate for the Quidel Triage MeterPro is its predecessor the Triage Meter cleared under K973547.

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® MeterPro (Proposed)Triage Meter (K973547)
Device ClassISame
Power supply100-240 VAC, self-switching, or with 4 AA batteriesSame
Max. Voltage of ext. power supply20 VSame
Electrostatic protection to serial port8 / 15 kVSame
Number of serial ports2 (by use of adapter)Same
PrinterIntegratedSame

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape with a smaller square cut out of the bottom right corner. The larger square is filled with a gradient of colors, including red, orange, yellow, green, blue, and purple. Below the square is the word "QUIDEL" in a sans-serif font.

Similarities
ItemProposed DevicePredicate Device
FeaturesQuidel Triage® MeterPro(Proposed)Triage Meter (K973547)
Sample IDManual input or externalhand-held barcode scannerSame
BarcodeIntegrated barcode on eachtest device with lot specificinformationSame
Time to resultApproximately 15 – 20minutesSame
Differences
ItemProposed DevicePredicate Device
FeaturesQuidel Triage®MeterPro(Proposed)Predicate Device
Regulation21 CFR 862.256021 CFR 862.2560 (KHO)21 CFR 862.5680 (DDR)21 CFR 862.1215 (JHX)21 CFR 862.1215 (MMI)
Product CodeKHOKHO, DDR, JHX, MMI
Sample TypeWhole blood, plasma or urineWhole blood or plasma
Top meter housingNew top meter housing moldto provide a larger liquidcrystal display (LCD) and twoelastomer keypadsSmall LCD with one singlemembrane keyboard

1.11. Standard/Guidance Document Referenced:

None referenced.

1.12. Test Principle:

The Quidel Triage TOX Drug Screen, 94600 is a test device utilizing the standard Triage technology. It is a competitive fluorescence immunoassay which contains all the reagents necessary for the qualitative detection, relative to an assigned threshold value, of the major urinary metabolites for the following substances in human urine: amphetamine (AMP), methamphetamine (mAMP), barbiturates (BAR), benzodiazepines (BZO), cocaine (COC), methadone/methadone metabolite (EDDP), opiates (OPI), tetrahydrocannabinol (THC), and tricyclic antidepressants (TCA).

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with a smaller square cut out of the upper right corner. The larger square is filled with a gradient of colors, while the smaller square is blue. Below the square is the word "QUIDEL" in a sans-serif font.

The Test Device contains:

  • Murine monoclonal antibodies against 9 targeted drugs or metabolites
  • Fluorescently labeled antibodies ●
  • Fluorescently labeled metabolites
  • Solid phase
  • Stabilizers

The test procedure involves the addition of a urine specimen to the sample port on the Test Device. After addition of the specimen, the urine passes through a filter. The specimen reacts with fluorescent antibody conjugates or with fluorescent drug conjugates and flows through the Test Device by capillary action. The presence of drug or drug metabolite in the urine specimen prevents binding of the fluorescent conjugates to the solid phase on the detection zone. Excess urine washes the unbound fluorescent conjugates from the detection lane into a waste reservoir.

The Test Device is inserted into the Quidel Triage MeterPro. The Quidel Triage MeterPro is programmed to perform the analysis after the specimen has reacted with the reagents in the Test Device. The analysis is based on the amount of fluorescence the Quidel Triage MeterPro detects within a measurement zone on the Test Device. The positive results are displayed on the Quidel Triage MeterPro screen in about 15 minutes. All results are stored in the Quidel Triage MeterPro memory to display or print when needed. If connected, the Quidel Triage MeterPro can transmit results to the laboratory or hospital information system.

1.13. Performance Characteristics:

    1. Analytical Performance
    • a. Precision/Reproducibility:

Each analyte for the precision study was tested at the following concentrations, as percentages of the cutoffs: A negative control, drug free, 25%, 50%, 75%, cutoff, 125%, 150%, 175%, and 200%. The panels were blinded and randomized prior to testing. Testing was performed at three (3) study sites. Three (3) operators conducted the testing at each study site. Each operator was assigned one test device lot and five (5) Triage MeterPro instruments to conduct the testing. Each operator tested ten (10) samples each day of testing. The ten (10) samples were run in duplicate two (2) times per day for twenty (20) days at each clinical site. Each device was read on one (1) Triage MeterPro. There were approximately seven hundred twenty (720) results per sample. The test results were interpreted as positive (POS) or negative (NEG) for each individual assay based on the assay specific cut-off concentration values. The results of the testing are summarized as follow for the test device.

SampleConcentration(ng/mL)% of Cutoffn# Neg# Pos
AMP (500 ng/mL)

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into two sections. The top left section is a gradient of colors, including yellow, orange, red, green, and blue. The bottom right section is a solid dark blue color. Below the square is the word "QUIDEL" in a sans-serif font.

SampleConcentration(ng/mL)% of Cutoffn# Neg# Pos
Negative Control07207200
007207200
126-757207200
281-507167124
395-2572069426
522Cutoff71950669
650+257222720
760+507200720
884+757040704
991+1007360736
mAMP (500 ng/mL)
Negative Control07207200
007367360
130-757207200
250-507207200
366-2571669719
529Cutoff720281431
652+2571912707
742+507222720
872+757200720
961+1007040704
BAR (200 ng/mL)
Negative Control07207200
007047040
53-757367360
108-507207191
156-2572068931
192Cutoff716111605
233+257203717
306+507190719
355+757220722
406+1007200720
BZO (200 ng/mL)
Negative Control07207200
007207200
58-757047040
107-507367351

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square made up of smaller squares in a rainbow of colors. Below the square is the name "QUIDEL" in a simple, sans-serif font.

SampleConcentration(ng/mL)% of Cutoffn# Neg# Pos
166-2572062694
219Cutoff720318402
259+257169707
306+507200720
378+757190719
399+1007220722
COC (150 ng/mL)
Negative Control07207200
007167160
41-757207200
76-507197190
119-25722519203
157Cutoff72026694
185+257040704
218+507360736
267+757200720
300+1007200720
EDDP (100 ng/mL)
Negative Control07207200
007227220
29-757207200
52-507047022
85-2573664591
111Cutoff720126594
136+257205715
143+507160716
174+757200720
204+1007190719
OPI (300 ng/mL)
Negative Control07207200
007207200
87-757197190
165-507227220
231-257207155
344Cutoff704197507
426+257365731
480+507200720
SampleConcentration(ng/mL)% of Cutoffn# Neg# Pos
548+757200720
589+1007160716
THC (50 ng/mL)
Negative Control07207200
007207200
12-757167160
26-507207173
39-2571967643
54Cutoff722163559
65+257201719
78+507044700
91+757361735
103+1007200720
TCA (1,000 ng/mL)
Negative Control07207200
007197190
236-757227220
498-507207191
741-2570461886
996Cutoff736218518
1,395+257205715
1,577+507201719
1,716+757160716
2,195+1007200720

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into two sections. The top left section is a gradient of colors, including yellow, orange, red, green, and blue. The bottom right section is a solid dark blue color. Below the square is the word "QUIDEL" in a sans-serif font.

  • b. Linearity/Assay Reportable Range:
    Not applicable. These devices are intended for qualitative use only.

  • c. Traceability, Stability, Expected Values

Cold Storage Stability

A shelf-life stability study of the test was performed and the results showed that devices are stable for 3 months when stored in cold storage (2℃ to 8℃). Real time stability studies are ongoing and shelf life will be extended based upon data meeting the acceptance criteria.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square shape with rounded corners, divided into four quadrants. The top-left quadrant displays a gradient of colors, transitioning from yellow to red. The bottom-right quadrant is blue. Below the square is the word "QUIDEL" in a sans-serif font.

Room Temperature Stability

A shelf-life stability study of the test was performed and the results showed that devices are stable for 14 days when stored at room temperature (20℃ to 24℃). Room temperature stability studies are ongoing and shelf life will be extended based upon data meeting the acceptance criteria.

Patient Sample Handling Stability

A patient sample handling stability study of the test was performed and patient sample results were found to be stable when used within 36 hours of sample collection at room temperature and within four (4) days when stored refrigerated. No more than a single freeze/thaw cycle is recommended.

  • d. Detection Limit:
    See assay cutoff characterization data in Section 15.13.1.f. below for assay performance around the claimed cutoff concentrations.

  • e. Analytical Specificity:
    To test cross-reactivity, drug metabolites and other compounds that may be present in human urine samples were tested using nine (9) lots of Quidel Triage TOX Drug Screen, 94600 Test Devices. The following is a summary of the cross-reactivity study. The individual assays are calibrated against the compounds marked with an asterisk (*).

AMP(Cutoff = 500 ng/mL)ResultsPositiveat(ng/mL)% Cross-Reactivity
3,4-Methylenedioxyamphetamine(MDA)1,85027.0
3,4-Methylenedioxyethylamphetamine(MDEA)>200,0000.0
3,4-Methylenedioxymethamphetamine(MDMA)>200,0000.0
d, l -1-(3,4-Methylenedioxyphenyl)-2-Butanamine (BDB)1,50033.3
d, l -Amphetamine1,00050.0
d, l -Phenylpropanolamine>200,0000.0
d-Ephedrine>200,0000.0
d-Amphetamine*500100.0

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape with rounded corners, filled with a gradient of colors ranging from red, orange, yellow, green, blue, and purple. A smaller white square is positioned inside the larger square, with a blue square extending from the bottom right corner of the white square to the edge of the larger square. Below the colorful square is the word "QUIDEL" in a sans-serif font.

AMP(Cutoff = 500 ng/mL)ResultsPositiveat(ng/mL)% Cross-Reactivity
d-Pseudoephedrine>200,0000.0
l-Amphetamine3,00016.7
l-Ephedrine>200,0000.0
Phentermine>200,0000.0
p-Chloroamphetamine (PCA)2,00025.0
p-Hydroxyamphetamine3,50014.3
p-Methoxyamphetamine (PMA)1,75028.6
Tyramine95,0000.5
β-phenylethylamine30,0001.7
mAMP(Cutoff = 500 ng/mL)ResultsPositive at(ng/mL)%Cross-Reactivity
3,4-Methylenedioxyamphetamine (MDA)>200,0000.0
3,4-Methylenedioxyethylamphetamine (MDEA)2,30021.7
3,4-Methylenedioxymethamphetamine (MDMA)75066.7
d, l -1-(3,4-Methylenedioxyphenyl)-2-Butanamine (BDB)25,0002.0
d, l -Methyl-1(3,4-Methylenedioxyphenyl)2-Butanamine (MBDB)500100.0
d-Methamphetamine*500100.0
d-Amphetamine>200,0000.0
d-Ephedrine>150,0000.0
Ethylamphetamine7,0007.1
Fenfluramine5,00010.0
l -Amphetamine>200,0000.0
l -Ephedrine>200,0000.0
l -Methamphetamine>200,0000.0
Isometheptene50,0001.0
Mephentermine25,0002.0
p-Hydroxymethamphetamine1,00050.0
p-Methoxyamphetamine (PMA)>200,0000.0
p-Methoxymethamphetamine (PMMA)2,20022.7
Propylamphetamine>200,0000.0

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square with rounded corners, filled with a gradient of colors ranging from red and orange at the top left to blue and purple at the bottom right. Inside the square is a smaller, white square, creating a window-like effect. Below the square is the word "QUIDEL" in a modern, sans-serif font, with the letters slightly spaced apart.

BAR(Cutoff = 200 ng/mL)ResultsPositiveat(ng/mL)%Cross-Reactivity
Allobarbital30066.7
Alphenal40050.0
Amobarbital25080.0
Aprobarbital30066.7
Barbital30066.7
Butabarbital200100.0
Butalbital*200100.0
Butethal100200.0
Cyclopentobarbital200100.0
Hexobarbital90,0000.2
Mephobarbital3,0006.7
Phenallylmal40050.0
Pentobarbital50040.0
Phenobarbital23087.0
Secobarbital70028.6
Thiopental80,0000.3%
BZO(Cutoff = 200 ng/mL)ResultsPositive at(ng/mL)%Cross-Reactivity
Alprazolam100200.0
Alprazolam, -OH150133.3
Bromazepam75026.7
Chlordiazepoxide8,0002.5
Clobazam75026.7
Clonazepam65030.8
Clonazepam, 7-amino26,0000.8
Clorazepate1,20016.7
Delorazepam35057.1
Demoxepam10,0002.0
Desalkylflurazepam200100.0

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image features the logo for QuidelOrtho Corporation. The logo consists of a stylized square divided into two sections. The left section displays a gradient of colors, transitioning from yellow to red to blue. The right section is a solid blue color. Below the square is the word "QUIDEL" in a simple, sans-serif font.

BZO(Cutoff = 200 ng/mL)ResultsPositiveat(ng/mL)%Cross-Reactivity
Diazepam125160.0
Estazolam40050.0
Flunitrazepam200100.0
Flunitrazepam, 7-amino6,0003.3
Flurazepam80250.0
Halazepam25080.0
Lorazepam200100.0
Lorazepam glucuronide30066.7
Lormetazepam100200.0
Medazepam9,0002.2
Midazolam200100.0
Nitrazepam2,6007.7
Nitrazepam, 7-amino>150,0000.0
Norchlordiazepoxide7,0002.9
Nordiazepam1,10018.2
Oxazepam2,5008.0
Oxazepam glucuronide1,25016.0
Prazepam35057.1
Temazepam*200100.0
Temazepam glucuronide30066.7
Triazolam100200.0
COC(Cutoff = 150 ng/mL)ResultsPositive at(ng/mL)%Cross-Reactivity
Benzoylecgonine*150100.0
Cocaethylene>200,0000.0
Cocaine50,0000.3
Ecgonine>200,0000.0
Ecgonine methyl ester>200,0000.0
m-Hydroxybenzoylecgonine40037.5
Norcocaine>200,0000.0

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape divided into four smaller squares, each filled with a different color gradient, creating a rainbow effect. Below the square is the word "QUIDEL" in a simple, sans-serif font.

EDDP(Cutoff = 100 ng/mL)ResultsPositive at(ng/mL)%Cross-Reactivity
EDDP*100100.0
EMDP40,0000.3
l-iso-methadone>100,0000.0
l-methadone160,0000.1
d-methadone>200,0000.0
d/l-methadone>200,0000.0
l-β-Acetylmethadol (LAAM)>200,0000.0
OPI(Cutoff = 300 ng/mL)ResultsPositiveat(ng/mL)%Cross-Reactivity
6-Acetylcodeine200150.0
6-Acetylmorphine200150.0
Buprenorphine>40,0000.0
Codeine300100.0
Diacetylmorphine200150.0
Dihydrocodeine120250.0
Ethylmorphine300100.0
Hydrocodone70042.9
Hydromorphone1,10027.3
Levorphanol25,0001.2
Morphine*300100.0
Morphine-3-glucuronide300100.0
Nalorphine3,00010.0
Naloxone>230,0000.0
Naltrexone>200,0000.0
Norbuprenorphine>200,0000.0
Norcodeine>200,0000.0
Normorphine>300,0000.0
Oxycodone50,0000.6
Oxymorphone100,0000.3
Thebaine35,0000.9

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a stylized square with a smaller square cut out from the bottom right corner. The larger square is filled with a gradient of colors, transitioning from yellow and green on the left to blue and red on the right. Below the square is the word "QUIDEL" in a simple, sans-serif font.

THC(Cutoff = 50 ng/mL)ResultsPositive at(ng/mL)%Cross-Reactivity
(+/-) 11-hydroxy- Δ9-ΤΗC1,5003.3
11-nor- Δ8-THC-COOH10050.0
11-nor-9 carboxy-Δ9-THC*50100.0
11-nor-9 carboxy-Δ9-THC-glucuronide17,0000.3
Cannabidiol>200,0000.0
Cannabinol, Δ8-3,0001.7
Cannabinol, Δ9-3,0001.7
Tetrahydrocannabinol3,0001.7
TCA(Cutoff = 1,000 ng/mL)ResultsPositiveat(ng/mL)%Cross-Reactivity
Amitriptyline600166.7
Amitriptyline metabolite300333.3
Chlorpromazine>400,0000.0
Chlorprothixene40,0002.5
Clomipramine10,00010.0
Cyclobenzaprine1,40071.4
Desipramine*1,000100.0
Doxepin1,30076.9
Imipramine600166.7
Maprotiline240,0000.4
Nordoxepin1,50066.7
Nortriptyline900111.1
Perphenazine175,0000.6
Phenothiazine280,0000.4
Promazine35,0002.9
Promethazine>200,0000.0
Protriptyline2,50040.0%
Thiothixene>100,0000.0%
Trimeprazine83,5001.2%
Trimipramine3,80026.3%

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square shape with rounded corners, divided into four quadrants. The top-left quadrant displays a gradient of colors, transitioning from yellow to red. The bottom-right quadrant is blue. Below the square is the word "QUIDEL" in a sans-serif font.

Cross-reactivity

Potential interference from pharmaceutical compounds was tested by spiking the listed compounds at a concentration of 100 µg/mL into drug-free urine containing the target drug concentrations at 50% below and 50% above the threshold cutoff level. The following compounds, arranged in alphabetical order, were found not to cross react when tested at concentrations up to at least 100 µg/mL. For the pharmaceutical compounds where interference was observed, the highest concentration that did not cause interference is indicated along with the assay in which it interfered.

5-(4-Hydroxyphenyl)-5-phenylhydantoinAcetaminophenDoxepin (0.65 µg/mL; TCA)Dronabinol (1 µg/mL; THC)O-desmethylvenlafaxined,l-Octopamine
AcetophenetidinAcetopromazine(maleate salt)DroperidolDuloxetineOfloxacinOlanzapine
AmantadineAmoxicillinAmpicillinEfavirenzL-EpinephrineFenfluramine (2 µg/mL; AMP and mAMP)Oxalic AcidOxaprozinPantoprazole
Aspirin (acetylsalicylic acid)AtenololAtorvastatinFenproporexFlunitrazepam (0.2 µg/mL; BZO)FluoxetinePapaverinePentazocinePericyazine
BenzocaineBenzphetamineBenzydamineGamma-HydroxybutyrateGlutethimide (10 µg/mL; BAR)HaloperidolPhenelzinePhenethylamine (2-Phenylethylamine) (6 µg/mL; AMP)Phenmetrazine (37.5 µg/mL; TCA)
BuprenorphineBenztropine MesylateBupropionIbuprofenKetamineKetorolac TromethaminePhenterminePhenylephrinePhenylpropanolamine
ButyrophenoneCarbamazepineChlorpheniramineLabetalolLevofloxacinLevorphanol (12.5 µg/mL; OPI)Promethazined/l -Propranolold-Pseudoephedrine (50 µg/mL; THC)
CimetidineCitalopramMeperidineMeprobamateQuetiapineQuinacrine (50 µg/mL; THC)

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into four quadrants. The top-left quadrant displays a gradient of colors, transitioning from yellow to red to purple. The bottom-right quadrant is a solid blue color. Below the square is the word "QUIDEL" in a sans-serif font.

ClobenzorexMesoridazine BesylateQuinine
Clomipramine(5 $ µg $ /mL; TCA)MethaqualoneRanitidine
ClonidineMethoxyphenamineRifampin
Cotinine [I-Cotinine]MethylphenidateRitodrine
CyproheptadineNalbuphineSelegiline
Dexamphetamine(0.4 $ µg $ /mL; AMP)NalmefeneSertraline
DextromethorphanNaloxoneThioridazine
Dextrorphan TartrateNaltrexoneTramadol
DimethylamineNaproxen[(S)-6-Methoxy-α-methyl-2Naphthaleneacetic acid]Tranylcypromine
DiphenhydramineN-desmethylvenlafaxineTrimethobenzamide
DopamineNiacinamideTyramine (25 $ µg $ /mL; AMP)
Dothiepin (0.7 $ µg $ /mL; TCA)NicotineVerapamil
Doxylamine SuccinateNorpseudoephedrine(25 $ µg $ /mL; AMP)Zolpidem

Exogenous Interference

Potential interference from exogenous compounds was tested by spiking the listed compounds into treated drug-free urine containing the target drug concentrations at 50% below and 50% above the threshold cutoff level. The following exogenous compounds were found not to interfere with test results when tested up to the concentrations identified in the table below.

Interfering SubstanceConcentration
Acetaminophen1 mg/mL
Acetone5 mg/mL
Acetylsalicylic Acid1 mg/mL
Ascorbic Acid15 mg/mL
Caffeine0.125 mg/mL
Ethanol5 mg/mL
Fluoxetine0.5 mg/mL
Hippuric Acid10 µg/mL
Ibuprofen0.75 mg/mL
Ketamine25 mg/mL
Oxalic Acid7 mg/mL
Riboflavin75 µg/mL

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into four smaller squares, each filled with a different color gradient. The top left square transitions from yellow to red, the top right from red to green, the bottom left from green to blue, and the bottom right from blue to purple. Below the square is the word "QUIDEL" in a simple, sans-serif font, with a trademark symbol next to it.

Interfering SubstanceConcentration
Scopolamine62.5 $\mu$ g/mL

Endogenous Interference

Potential interference from endogenous compounds was tested by spiking the listed compounds into drug-free urine containing the target drug concentrations at 50% below and 50% above the threshold cutoff level. The following endogenous compounds were found not to interfere with test results when tested up to the concentrations identified in the table below.

Interfering SubstanceConcentration
Bilirubin2.5 µg/mL
Creatinine2.5 mg/mL
Dextrose20 mg/mL
Gamma Globulin5 mg/mL
Hemoglobin1.2 mg/mL
Human Serum Albumin5 mg/mL
Sodium Chloride30 mg/mL
Urea30 mg/mL

Specific gravity and pH

The effect of specific gravity for each analyte was evaluated by testing positive and negative samples at specific gravities ranging from 1.003 to 1.030 g/mL. No interference was observed for all specific gravities tested.

The effect of pH for each analyte was evaluated by testing positive and negative samples over a range of pH levels of 4.0 to 9.0. The Quidel Triage TOX Drug Screen, 94600 has been validated with specimens that have a pH range of 4.0 -9.0, and no excessive interference was observed for each of the assays. In the case of the TCA assay, there was evidence that increasing urine pH levels at the top of the validated range could have an impact on positive control sample results as the assay cutoff of 1,000 ng/mL was approached. Specimens tested outside of the validated pH range may yield inaccurate results.

Operating Temperature

The effect of operating temperature for each analyte was evaluated by testing positive and negative samples at temperatures ranging from 16℃ to 30℃ (61°F to 86°F). Even though reasonable performance was found for the whole temperature range, non-conformance beyond the set specifications was observed for the cocaine, methamphetamine and opiate assays at the extremes of the temperature range. The Quidel Triage TOX Drug Screen, 94600 has been validated for an operating temperature range from 18℃ to 28℃ (64°F to 82°F).

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into two sections. The larger section is a gradient of colors, transitioning from yellow to green to blue to red. The smaller section is a solid dark blue. Below the square is the word "QUIDEL" in a simple, sans-serif font.

Operating Humidity

The effect of operating humidity for each analyte was evaluated by testing positive and negative samples at relative humidities (RH) ranging from ≤ 10 %RH to 85 %RH. The Quidel Triage TOX Drug Screen, 94600 was validated for an operating relative humidity (RH) range from 10 %RH to 85 %RH and met the specifications set for all assays.

  • f. Assay cut-off:
    Each analyte for the cutoff characterization study was tested at the following concentrations: 0, 25%, 50%, 75%, 100%, +125%, +150%, +175% and +200% of the specific cutoff for each drug assay. Testing was performed using 3 lots of Quidel Triage TOX Drug Screen, 94600 Test Devices and was performed by 3 operators. The results are summarized as follows for each analyte.
Amphetamine (AMP) 500 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
NegativeControl0%315/0
00%315/0
12525%315/0
25050%315/0
37575%313/2
500100%49/266
625125%1/314
750150%0/315
875175%0/315
1000200%0/315
Methamphetamine (mAMP) 500 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
NegativeControl0%315/0
00%315/0
12525%315/0
25050%315/0
37575%312/3
500100%127/188

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into two sections. The larger section is a gradient of colors, including yellow, orange, red, purple, and blue. The smaller section is a solid blue color. Below the square is the word "QUIDEL" in a sans-serif font.

Methamphetamine (mAMP) 500 ng/mL
625125%11/304
750150%2/313
875175%0/315
1000200%0/315
Barbiturates (BAR) 200 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
Negative Control0%315/0
00%315/0
5025%315/0
10050%315/0
15075%309/6
200100%78/237
250125%1/314
300150%1/314
350175%0/315
400200%0/315
Benzodiazepines (BZO) 200 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
NegativeControl0%315/0
00%315/0
5025%315/0
10050%315/0
15075%296/19
200100%242/73
250125%3/312
300150%0/315
350175%0/315
400200%0/315

{30}------------------------------------------------

Image /page/30/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape with rounded corners, divided into two sections. The top and left sections of the square are filled with a gradient of colors, including red, orange, yellow, green, blue, and purple. The bottom right section of the square is filled with a solid blue color. Below the square is the word "QUIDEL" in a sans-serif font.

Cocaine (COC) 150 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
NegativeControl0%315/0
00%315/0
37.525%315/0
7550%315/0
112.575%279/36
150100%10/305
187.5125%1/314
225150%0/315
262.5175%0/315
300200%0/315
Methadone Metabolite (EDDP) 100 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
NegativeControl0%315/0
00%315/0
2525%315/0
5050%315/0
7575%315/0
100100%115/200
125125%11/304
150150%0/315
175175%0/315
200200%0/315
Opiates (OPI) 300 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
NegativeControl0%315/0
00%315/0
7525%315/0
15050%315/0

{31}------------------------------------------------

Image /page/31/Picture/0 description: The image features the logo for QuidelOrtho Corporation. The logo consists of a square divided into two sections. The larger section is a gradient of colors, transitioning from yellow to red to blue. The smaller section is a solid blue square. Below the square is the word "QUIDEL" in a sans-serif font.

Opiates (OPI) 300 ng/mL
22575%315/0
300100%132/183
375125%1/314
450150%0/315
525175%0/315
600200%0/315
Cannabinoids (THC) 50 ng/mL
Concentration(ng/mL)% of cutoffResults#Neg/#Pos
NegativeControl0%315/0
00%315/0
12.525%315/0
2550%315/0
37.575%295/20
50100%20/295
62.5125%0/315
75150%0/315
87.5175%0/315
100200%0/315
Tricyclic Antidepressants (TCA) 1000 ng/mL
Concentration (ng/mL)% of cutoffResults #Neg/#Pos
Negative Control0%315/0
00%315/0
25025%315/0
50050%315/0
75075%300/15
1000100%161/154
1250125%5/310
1500150%1/314
1750175%0/315
2000200%0/315

{32}------------------------------------------------

Image /page/32/Picture/0 description: The image features the logo for QuidelOrtho Corporation. The logo consists of a square shape with rounded corners, filled with a gradient of colors ranging from red, orange, yellow, green, blue, and purple. Inside the square is a smaller white square, with a portion of its lower right corner removed. Below the colorful square is the word "QUIDEL" in a simple, sans-serif font.

    1. Comparison studies:
    • Method comparison with reference method: a.

A method comparison study was conducted using unaltered urine specimens and comparing results from the Quidel Triage TOX Drug Screen, 94600 to a reference method. Results are summarized below.

Quidel Triage TOX Drug Screen, 94600 Test Results vs GC/MS or LC-MS/MS Values

AMP(Cutoff = 500 ng/mL)
Reference Method Result by GC/MS or LC-MS/MS Value
DrugQuidel Triage TOX Drug Screen, 94600Negative (<50% of threshold)Near Threshold Negative (within 50% below threshold)Near Threshold Positive (within 50% above threshold)Positive (>150% of threshold)
AMPNegative99112a2b
Positive00898

3 Patient specimens with Specimen IDs 569740 and 575202 were found to be negative despite a near threshold positive result

b Patient specimens with Specimen IDs 570964 and 572667 were found to be negative despite a high positive result

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS 94600Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
AMP500569740aFalse NegativeAmphetamine697
AMP500575202aFalse NegativeAmphetamine745
AMP500570964bFalse NegativeAmphetamine1003
AMP500572667bFalse NegativeAmphetamine1359

Discordant sample resolution

b The Quidel Triage TOX Drug Screen, 94600 has 100.0% cross-reactivity to the d-amphetamine isomer and 16.7% cross-reactivity to the l-amphetamine isomer, resulting in negative screening results even with amphetamine levels above the cutoff concentration of 500 ng/mL. A specimen with high levels of the Iamphetamine isomer may be associated with prescription use of medications containing amphetamines or compounds that metabolize to amphetamines. A summary of the results for the patient specimens containing both isomers is presented in the table below. After adjudication, Specimen ID 570964 would be classified as "near threshold" negative and would not be classified as a discordant result. Specimen ID 572667 would be classified as a "near threshold" positive patient specimen (within 10.8% of the assay threshold).

{33}------------------------------------------------

Image /page/33/Picture/0 description: The image features the logo for QuidelOrtho Corporation. The logo consists of a square divided into four quadrants, each filled with a different color. The colors transition from red and orange in the top left, to green and yellow in the top right, to a lighter blue in the bottom right, and a darker blue in the bottom left. Below the square is the word "QUIDEL" in a modern, sans-serif font, with a curved line extending from the bottom of the "Q" to underline the rest of the word.

QuidelTriageTOXInitialValueInitialQuidelTriageIsomericComposition (%)IsomericAbundance(ng/mL)Adjudicated
SpecimenIDDrugScreen,94600AMPAssayCutoff(ng/mL)deter-minedbyLC-MS/MS(ng/mL)TOXDrugScreen,94600AMPAssayResult% d-Amphetamine% /-Amphetamined-Amphetamine/-AmphetamineEffectiveAmphetamine(ng/mL)*Quidel TriageTOX DrugScreen, 94600AMP AssayResult
570964b5001003FalseNegative35.364.7354649463Negative
572667b5001359FalseNegative28.971.1393966554False Negative
mAMP(Cutoff = 500 ng/mL)Reference Method Result by GC/MS or LC-MS/MS Value
DrugQuidel Triage TOX Drug Screen, 94600Negative (<50% of threshold)Near Threshold Negative (within 50% below threshold)Near Threshold Positive (within 50% above threshold)Positive (>150% of threshold)
mAMPNegative9955d7e
Positive05c691

← Patient specimens with Specimen IDs 578510, 579777, 579705, 579727, and 579806 were found to be positive despite a near threshold negative result

d Patient specimens with Specimen IDs 586313, 586723, 579791, 586280 and 586293 were found to be negative despite a near threshold positive result

° Patient specimens with Specimen IDs 586276, 586264, 586282, 586275, 586300, and 579757 were found to be negative despite a high positive result

Discordant sample resolution

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result(POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
mAMP500578510cFalse PositiveMethamphetamine325
mAMP500579777cFalse PositiveMethamphetamine402
mAMP500579705cFalse PositiveMethamphetamine445
mAMP500579727cFalse PositiveMethamphetamine490
mAMP500579806cFalse PositiveMethamphetamine495
mAMP500586313dFalse NegativeMethamphetamine535
mAMP500586273dFalse NegativeMethamphetamine555
mAMP500579791dFalse NegativeMethamphetamine588

{34}------------------------------------------------

Image /page/34/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square divided into four quadrants, each filled with a different color gradient. The top-left quadrant transitions from red to yellow, the top-right quadrant is green, the bottom-left quadrant is blue, and the bottom-right quadrant is a darker shade of blue. Below the square is the word "QUIDEL" in a sans-serif font.

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result(POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
mAMP500586280dFalse NegativeMethamphetamine633
mAMP500586293dFalse NegativeMethamphetamine673
mAMP500586276eFalse NegativeMethamphetamine831
mAMP500586269eFalse NegativeMethamphetamine937
mAMP500586264eFalse NegativeMethamphetamine940
mAMP500586282eFalse NegativeMethamphetamine984
mAMP500586275eFalse NegativeMethamphetamine990
mAMP500586300eFalse NegativeMethamphetamine1028
mAMP500579757eFalse NegativeMethamphetamine1568

e The specimens contained both the d-methamphetamine isomers. The initial LC-MS/MS reference method was unable to distinguish between the two enantiomeric forms. The Quidel Triage TOX Drug Screen, 94600 has 100.0% cross-reactivity to the d-methamphetamine isomer and 0.0% cross-reactivity to the I-methamphetamine isomer, resulting in negative screening results even with methamphetamine levels above the cutoff concentration of 500 ng/mL. A specimen with high levels of the l-methamphetamine isomer may be associated with prescription use of medications that contain methamphetamine or compounds that metabolize into methamphetamines. A summary of the results for the patient specimens containing both isomers is presented in the table below. After adjudication, Specimen IDs 586282, 586275 and 586300 were found to be negative despite a near threshold positive result (all within 8.6% of the assay threshold).

ReferenceLaboratory 1LC-MS/MSConfirmatoryValue(ng/mL)Quidel TriageTOX DrugScreen, 94600ThresholdConcentration(ng/mL)Quidel TriageTOX DrugScreen, 94600ResultReference Laboratory 2ResultsAdjudicated QuidelTriage TOX DrugScreen, 94600Result
SpecimenID% d-IsomerConcentrationd-Isomer(ng/mL)
586276°831500False Negative52.3434.6Negative
586269°937500False Negative52.8494.7Negative
586264°940500False Negative52.9497.3Negative
586282°984500False Negative52.2513.6False Negative
586275°990500False Negative52.3517.8False Negative
586300°1028500False Negative52.8542.8False Negative
579757°1568500False Negative19.9312.0Negative

{35}------------------------------------------------

Image /page/35/Picture/0 description: The image features the logo for QuidelOrtho Corporation. The logo consists of a square shape with a gradient of colors, including red, orange, yellow, green, blue, and purple. The square is partially filled, with a section missing from the lower right corner. Below the square is the word "QUIDEL" in a sans-serif font.

BAR(Cutoff = 200 ng/mL)
DrugQuidel TriageTOX DrugScreen, 94600Negative(<50% ofthreshold)Near ThresholdNegative(within 50%below threshold)Near ThresholdPositive(within 50%above threshold)Positive(>150% ofthreshold)
BARNegative99300
Positive08f1197

' Patient specimens with Specimen IDs 582858, 575082, 575079, 586932, 575085, 575084, and 575095 were found to be positive despite a near threshold negative result

Discordant sample resolution

AssayCutoffValue(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
BAR200582858fFalse PositivePhenobarbital120
BAR200575082fFalse PositivePhenobarbital172
BAR200575081fFalse PositiveButalbital150
BAR200575079fFalse PositiveButalbital151
BAR200586932fFalse PositiveAmobarbital164
BAR200575085fFalse PositiveButalbital165
BAR200575084fFalse PositivePhenobarbital210
BAR200575095fFalse PositiveButalbital197

BZO

(Cutoff = 200 ng/mL)
Reference Method Result by GC/MS or LC-MS/MS Value
DrugQuidel TriageTOX DrugScreen, 94600Negative(<50% ofthreshold)Near ThresholdNegative(within 50%below threshold)Near ThresholdPositive(within 50%above threshold)Positive(>150% ofthreshold)
BZONegative9911h0
Positive010g1099

8 Patient specimens with Specimen IDs 586239, 586234, 586235, 582865, 570784, 569935, 585873, 570860, 570941, and 578527 were found to be positive despite a near threshold negative result

ʰ Patient specimen with Specimen ID 578563 was found to be negative despite a near threshold positive result

{36}------------------------------------------------

Image /page/36/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a stylized square divided into two sections. The larger section is a gradient of colors, transitioning from red and orange at the top left to green and yellow in the middle, and blue and purple at the top right. The smaller section is a solid blue color. Below the square is the word "QUIDEL" in a modern, sans-serif font.

Discordant sample resolution

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
BZO200586239gFalse PositiveAlprazolam-OH79
BZO200586234gFalse PositiveAlprazolam-OH80
BZO200586235gFalse PositiveAlprazolam-OH86
7-aminoclonazepam293
BZO200582865gFalse PositiveLorazepam53
Oxazepam379
7-aminoclonazepam1134
BZO200570784gFalse PositiveNordiazepam63
Oxazepam124
Temazepam90
BZO200569935gFalse PositiveNordiazepam68
Oxazepam244
Temazepam93
BZO200585873gFalse PositiveAlprazolam-OH105
7-aminoclonazepam2500
BZO200570860gFalse PositiveNordiazepam83
Oxazepam218
Temazepam126
BZO200570941gFalse PositiveAlprazolam-OH126
BZO200578527gFalse PositiveAlprazolam-OH129
7-aminoclonazepam146
BZO200578563hFalse NegativeLorazepam214
COC(Cutoff = 150 ng/mL)
Reference Method Result by GC/MS or LC-MS/MS Value
DrugQuidel TriageTOX DrugScreen, 94600Negative(<50% ofthreshold)Near ThresholdNegative(within 50%below threshold)Near ThresholdPositive(within 50%above threshold)Positive(>150% ofthreshold)
COCNegative99500
Positive06i1199

¹ Patient specimens with Specimen IDs 569915, 579796, 579821, 569790, 570955 and 577368 were found to be positive despite a near threshold negative result

{37}------------------------------------------------

Image /page/37/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a stylized square divided into four quadrants, each filled with a different color gradient. The top-left quadrant transitions from red to orange, the top-right quadrant from orange to yellow, the bottom-left quadrant from green to blue, and the bottom-right quadrant is a solid blue. Below the square is the word "QUIDEL" in a sans-serif font.

Discordant sample resolution

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
COC150569915COC False PositiveBenzoylecgonine113
COC150579796False PositiveBenzoylecgonine126
COC150579821False PositiveBenzoylecgonine128
COC150569790False PositiveBenzoylecgonine129
COC150570955False PositiveBenzoylecgonine144
COC150577368False PositiveBenzoylecgonine144
EDDP(Cutoff = 100 ng/mL)
Reference Method Result by GC/MS or LC-MS/MS Value
DrugQuidel TriageTOX DrugScreen, 94600Negative(<50% ofthreshold)Near ThresholdNegative(within 50%below threshold)Near ThresholdPositive(within 50%above threshold)Positive(>150% ofthreshold)
EDDPNegative9992i0
Positive01k1098

أ Patient specimen with Specimen ID 572444 was found to be positive despite a near threshold negative result

  • Patient specimens with Specimen IDs 572432 and 572426 were found to be negative a near threshold positive result

Discordant sample resolution

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
EDDP100572444jFalse PositiveEDDP76
EDDP100572432kFalse NegativeEDDP100
EDDP100572426kFalse NegativeEDDP110
OPI(Cutoff = 300 ng/mL)
Reference Method Result by GC/MS or LC-MS/MS Value
DrugQuidel TriageTOX DrugScreen, 94600Negative(<50% ofthreshold)Near ThresholdNegative(within 50%below threshold)Near ThresholdPositive(within 50%above threshold)Positive(>150% ofthreshold)
OPINegative99201m

{38}------------------------------------------------

Image /page/38/Picture/0 description: The image features the logo for Quidel. The logo consists of a square divided into four quadrants, each filled with a different color gradient. The top-left quadrant transitions from red to orange to yellow, the top-right quadrant transitions from yellow to green, the bottom-left quadrant transitions from green to blue, and the bottom-right quadrant transitions from blue to purple. Below the square is the word "QUIDEL" in a simple, sans-serif font.

Positive08'1198
Patient specimens with Specimen IDs 572508, 570415, 586327, 582860, 572488, 586330, 582932 and

586359 were found to be positive despite a near threshold negative result

™ Patient specimen with Specimen ID 572644 was found to be negative despite a high positive result

Discordant sample resolution

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
OPI300572508'False PositiveHydrocodone353
Oxycodone24859
OPI300570415'False PositiveOxymorphone217
OPI300586327'False PositiveMorphine228
OPI300582860'False PositiveHydrocodone469
OPI300572488'False PositiveMorphine239
OPI300586330'False PositiveMorphine260
OPI300582932'False PositiveHydrocodone488
Hydromorphone67
OPI300586329'False PositiveMorphine285
Morphine2097
OPI300572644mFalse NegativeOxycodone2674
Oxymorphone137

™ The initial reference testing laboratory result for Specimen ID 572644 was positive for opiates while the Quidel Triage TOX Drug Screen, 94600 OPI assay result was negative. Specimen ID 572644 was sent to a second reference testing laboratory and was confirmed to be negative for opiates. The second result is concordant with the Quidel Triage TOX Drug Screen, 94600 OPI assay result.

THC(Cutoff = 50 ng/mL)
DrugQuidel TriageTOX DrugScreen, 94600Negative(<50% ofthreshold)Near ThresholdNegative(within 50%below threshold)Near ThresholdPositive(within 50%above threshold)Positive(>150% ofthreshold)
THCNegative99100
Positive01ⁿ899

ª Patient specimen with Specimen ID 570021 was found to be positive despite a near threshold negative result

0 Patient specimens with Specimen IDs 570366, 575089, and 569400 were found to be negative despite a near threshold positive result

{39}------------------------------------------------

Image /page/39/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a stylized square divided into four quadrants. The top-left quadrant is a gradient of colors, including yellow, orange, red, and purple. The bottom-right quadrant is blue. Below the square is the word "QUIDEL" in a sans-serif font.

Discordant sample resolution

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
THC50570021nFalse Positive11-nor-9-Carboxy-delta 9-THC31
THC50570366oFalse Negative11-nor-9-Carboxy-delta 9-THC51
THC50575089oFalse Negative11-nor-9-Carboxy-delta 9-THC56
THC50569400oFalse Negative11-nor-9-Carboxy-delta 9-THC63
TCA(Cutoff = 1,000 ng/mL)
DrugQuidel TriageTOX DrugScreen, 94600Reference Method Result by GC/MS or LC-MS/MS Value
Negative(<50% ofthreshold)Near ThresholdNegative(within 50%below threshold)Near ThresholdPositive(within 50%above threshold)Positive(>150% ofthreshold)
TCANegative98101r
Positive1p10q1195

م Patient specimen with Specimen ID 586712 was found to be positive despite a high negative result 9 Patient specimens with Specimen IDs 586353, 585878, 582936, 585866, 586232, 585830, 585879, 582925, and 585874 were found to be positive despite a near threshold negative result

「 Patient specimen with Specimen ID 586694 was found to be negative despite a high positive result

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
TCA1,000586712pFalse PositiveDesipramine0
TCA1,000586353qFalse PositiveDesipramine637
TCA1,000585878qFalse PositiveDesipramine639
TCA1,000582936qFalse PositiveAmitriptyline385
TCA1,000582936qFalse PositiveNortriptyline160
TCA1,000572581qFalse PositiveAmitriptyline149
TCA1,000572581qFalse PositiveNortriptyline512
TCA1,000585866qFalse PositiveDesipramine680
TCA1,000586232qFalse PositiveDesipramine727
TCA1,000585830qFalse PositiveDesipramine730

Discordant sample resolution

{40}------------------------------------------------

Image /page/40/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into four quadrants, each filled with a different color gradient. The top-left quadrant is a blend of red, orange, and yellow, while the top-right quadrant transitions from yellow to green. The bottom-left quadrant displays a gradient from green to blue, and the bottom-right quadrant shows a blend of blue, indigo, and violet. Below the square is the word "QUIDEL" in a sans-serif font.

AssayCutoff Value(ng/mL)SpecimenIDQuidel Triage TOX DS94600 Result (POS/NEG)Drug/MetaboliteDetectedGC/MS or LC/MS/MSvalue (ng/mL)
TCA1,000585879qFalse PositiveDesipramine758
TCA1,000582925qFalse PositiveAmitriptylineNortriptyline162599
TCA1,000585874qFalse PositiveDesipramineDoxepin608238
TCA1,000586694rFalse NegativeDesipramine1646

P A data entry error occurred with Specimen ID 586712. This patient specimen was positive for TCA when tested on the Quidel Triage TOX Drug Screen, 94600 and reconfirmed as a near cut-off negative at a secondary reference testing laboratory with value of 864 ng/mL desipramine (within 10.8% of the assay threshold).

Specimen ID 586694 was confirmed to have a pH value of 9.4. A pH value at this level is above the upper limit of the expected range for normal human urine8 and above the upper limit evaluated for performance on the Quidel Triage TOX Drug Screen, 94600. The Quidel Triage TOX Drug Screen, 94600 has been validated with specimens that have a pH range of 4.0 – 9.0, and no excessive interference was observed for each of the assays. In the case of the TCA assay, there was evidence that increasing urine pH levels at the top of the validated range could have an impact on positive control sample results as the assay cutoff of 1,000 ng/mL was approached. Specimens tested outside of the validated pH range may yield inaccurate results.

  • b. Method comparison with reference method:
    Not applicable. These devices are for use with human urine samples only.

    1. Clinical studies:
    • a. Clinical sensitivity:

Not applicable.

  • b. Clinical specificity:
    Not applicable.

  • c. Other clinical supportive data (when a. and b. are not applicable):
    Not applicable.

    1. Clinical cut-off:
      Not applicable.
    1. Expected values/Reference range:

{41}------------------------------------------------

Image /page/41/Picture/0 description: The image features the logo for Quidel. The logo consists of a square shape with rounded corners, divided into two sections. The larger section displays a gradient of colors, including red, orange, yellow, green, blue, and purple. The smaller section is a solid blue color. Below the square is the word "QUIDEL" in a sans-serif font.

Not applicable.

1.14. Conclusion:

These studies demonstrated the substantial equivalence of the Quidel Triage TOX Drug Screen, 94600 to the GenPrime Drugs of Abuse (DOA) Reader System (K130082), Immunalysis Barbiturates Urine Enzyme Immunoassay (K161714), DRI Benzodiazepine Assay (K173963), Immunalysis EDDP Specific Urine Enzyme Immunoassay (K151395), Triage® TOX Drug Screen (K043242), and Triage Meter (K973547). Such studies are a critical element in establishing the fundamental safety and effectiveness of the product and its appropriateness for commercial distribution.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).