K Number
K182658

Validate with FDA (Live)

Device Name
Bioliner
Date Cleared
2019-06-19

(267 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bioliner™ is an aligner system indicated for use in the alignment teeth through orthodontic treatment of misalignment and malocclusion.

Device Description

The Bioliner™ aligner system is used for treatment of misalignment and malocclusion of permanent teeth by use of clear plastic aligners. The Bioliner™ aligner system is a sequential set of removable aligners prescribed by a dentist that are custom made for a specific patient.

Each Bioliner™ sequential aligner introduces subtle changes in the position of the teeth from the previous aligner. Each Bioliner™ is fabricated from 1.0mm thermoplastic. For each patient, the Bioliner™ will be designed by a Five Star Orthodontic Laboratory & Supply, Inc. technician utilizing 3Shape (K152086) orthodontic validated software and tooth movement recommendations from the prescribing dentist. Once fabricated, the Bioliner™ aligners are provided to the dentist, and over a period of time, the dentist provides the next sequential aligner to the patient in order to move the teeth to the desired final position.

The aligners are removable at any time and treatment may be discontinued at any time. The Bioliner™ aligners are a non-sterile, single use device.

AI/ML Overview

The provided text is a 510(k) Pre-Market Notification from the FDA regarding the Bioliner™ orthodontic aligner system. It details the device, its intended use, a comparison to a predicate device, and summaries of testing.

However, the document explicitly states:
"Summary of Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the Bioliner™ to the predicate device."
"Summary of Non-Clinical Testing: Biocompatibility testing was performed for the Bioliner™ aligners in accordance with ISO 10993-1 and its applicable parts. The mechanical properties of the Essix Ace plastic (K062828) has been previously demonstrated by the manufacturer as appropriate for use with aligners, therefore no additional testing was required."

Therefore, based on the provided document, the following information cannot be extracted:

  • A table of acceptance criteria and reported device performance (as no clinical or mechanical performance testing was performed for this 510(k) submission).
  • Sample size used for the test set and data provenance.
  • Number of experts used to establish the ground truth for the test set and their qualifications.
  • Adjudication method for the test set.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size.
  • If a standalone (algorithm only) performance was done.
  • The type of ground truth used.
  • The sample size for the training set.
  • How the ground truth for the training set was established.

The 510(k) submission for the Bioliner™ relied on demonstrating substantial equivalence to a predicate device and existing clearances for its materials and manufacturing software, rather than new clinical or detailed performance studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 19, 2019

Five Star Orthodontic Lab & Supply Andrew Loch Sales Manager 2928 Metro Street, Suite 102 Denton, Texas 76207

Re: K182658

Trade/Device Name: Bioliner™ Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: May 14, 2019 Received: May 17, 2019

Dear Andrew Loch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182658

Device Name Bioliner™

Indications for Use (Describe)

The Bioliner™ is an aligner system indicated for use in the alignment teeth through orthodontic treatment of misalignment and malocclusion.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

ال Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo for "Five Star Orthodontic Laboratory & Supply". The logo has a graphic of an atom-like structure with five stars orbiting the center. The text is in bold, and the words "Laboratory & Supply" are smaller than the words "Five Star Orthodontic".

510(k) Summarv

Submitter:Five Star Orthodontic Lab & Supply, Inc.2928 Metro Street. Suite 102Denton, TX 76207
Contact Person:Andrew LochSales ManagerPhone: (800) 521-2351andrew@fivestarortho.com
Submission Date:June 17, 2019
Trade Name:Bioliner™
Common Name:Aligner, Sequential
Classification Name:Orthodontic Plastic Bracket; 21 CFR 872.5470; NXC
Device Classification:Class II
Predicate Device:K173785, Derby Dental Laboratory Custom Clear Aligner System
Reference Devices:K062828, Dentsply Intl, Inc. Mouthguard and Aligner MaterialsK152086, 3Shape A/S 3Shape Ortho System

Description of Device:

The Bioliner™ aligner system is used for treatment of misalignment and malocclusion of permanent teeth by use of clear plastic aligners. The Bioliner™ aligner system is a sequential set of removable aligners prescribed by a dentist that are custom made for a specific patient.

Each Bioliner™ sequential aligner introduces subtle changes in the position of the teeth from the previous aligner. Each Bioliner™ is fabricated from 1.0mm thermoplastic. For each patient, the Bioliner™ will be designed by a Five Star Orthodontic Laboratory & Supply, Inc. technician utilizing 3Shape (K152086) orthodontic validated software and tooth movement recommendations from the prescribing dentist. Once fabricated, the Bioliner™ aligners are provided to the dentist, and over a period of time, the dentist provides the next sequential aligner to the patient in order to move the teeth to the desired final position.

The aligners are removable at any time and treatment may be discontinued at any time. The Bioliner™ aligners are a non-sterile, single use device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo for "Five Star Orthodontic Laboratory & Supply". The logo features a stylized atom symbol with five stars orbiting the nucleus. The text "ORTHODONTIC" is prominently displayed, with "LABORATORY & SUPPLY" written in a smaller font underneath. The words "FIVE STAR" are located above and to the right of the word "ORTHODONTIC".

Intended Use / Indications:

The Bioliner™ is an aligner system indicated for use in the alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion.

Predicate Device Comparison:

Proposed DevicePredicate Device
Property or CharacteristicFive Star Orthodontic Laboratory& Supply, Inc.Bioliner™Derby Dental LaboratoryCustom Clear Aligner System(K173785)
Device Classification NameOrthodontic Plastic BracketOrthodontic Plastic Bracket
Product CodeNXCNXC
ClassificationIIII
Indications for UseThe Bioliner™ is an alignersystem indicated for use in thealignment of permanent teeththrough orthodontic treatment ofmisalignment and malocclusion.The Custom Clear AlignerSystem is indicated for use in thealignment of permanent teeththrough orthodontic treatment ofmisalignment and malocclusion.
Prescription or OTCPrescriptionPrescription
Mode of ActionAlignment of teeth byapplication of continuous gentleforce, by sequential use ofpreformed plastic trays.Alignment of teeth byapplication of continuous gentleforce, by sequential use ofpreformed plastic trays.
Number of aligner sets tocomplete treatmentPatient specific - determinedduring design of alignersutilizing orthodontic softwarePatient specific - determinedduring design of alignersutilizing orthodontic software
MaterialEssix Ace Plastic (K062828)Thin thermoformed polyurethane
SizesPatient specificPatient specific
Body LocationMouthMouth
Manufacturing ProcessThermoformedThermoformed
Software Used duringmanufacturingUse of 3Shape Ortho System(K152086)Use of 3Shape Ortho System(K152086)
Provided Non-SterileYesYes

The Bioliner™ aligners are manufactured from Essix Ace Plastic (K062828), whereas the predicate device is manufactured from polyurethane. Polyurethane is a more flexible material as compared to the thermoplastic material that the Bioliner™ aligners are manufactured from. The thermoplastic material (Essix Ace Plastic) used for the Bioliner™ aligners has been cleared for orthodontic use in

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Five Star Orthodontic Laboratory & Supply. The logo features a stylized atom symbol with five stars orbiting the center. The text "FIVE STAR ORTHODONTIC LABORATORY & SUPPLY" is displayed to the right of the atom symbol.

K062828 and the material labeling specifically indicates use of the material for aligners.

A biocompatibility risk assessment has been completed for the Bioliner™ aligners taking into account not only the material used for fabrication, but also the manufacturing processes utilized during fabrication. As identified below, all applicable biocompatibility testing has been performed on the Bioliner™ aligners.

Summary of Clinical Testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the Bioliner™ to the predicate device.

Summary of Non-Clinical Testing:

Biocompatibility testing was performed for the Bioliner™ aligners in accordance with ISO 10993-1 and its applicable parts. The mechanical properties of the Essix Ace plastic (K062828) has been previously demonstrated by the manufacturer as appropriate for use with aligners, therefore no additional testing was required. There is sufficient data to demonstrate substantial equivalence to the predicate device.

The Bioliner™ aligners are manufactured using 3Shape orthodontic software (K152086). Software validation by Five Star Orthodontic Laboratory and Supply was not required since 3Shape has received 510(k) clearance and has been validated by the manufacturer.

A manufacturing validation was performed to ensure the dimensional accuracy of the Bioliner™ aligners through the manufacturing process. This validation demonstrated that the finished Bioliner™ aligners match the software output specifications.

Conclusion:

Based upon the information presented in this submission, it is concluded that the Bioliner™ system is substantially equivalent to the predicate device in regard to indications for use, design and technology.

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.