K Number
K182565

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-11-09

(52 days)

Product Code
Regulation Number
882.4250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use in blocking pain by temporarily ablating peripheral nerves.

Device Description

The cryolCE cryoSPHERE cryoablation (CRYOS) probe is a single use device that achieves cryoablation of peripheral nerves by allowing a surgeon to insert the probe through an incision to reach the target tissue, place the probe tip on the target site, and in conjunction with an AtriCure Cryo Module (ACM), temporarily freeze the tissue in contact with the probe tip by circulating a cryogenic agent, nitrous oxide, through the device. The CRYOS probe is offered in two probe length configurations: approximately 11" and 18" long. The probe is malleable, and is capable of being bent by the end user. At the distal end, the CRYOS device features an 8 mm diameter ball tip shaped probe.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the AtriCure cryoICE cryoSPHERE cryoablation probe. It states that the device is substantially equivalent to a previously cleared predicate device based on bench testing. However, the document does not contain information about the device's performance in a study involving human or animal subjects, nor does it detail acceptance criteria related to diagnostic performance (e.g., sensitivity, specificity, accuracy) that would be relevant for an AI/ML-based medical device.

The provided document focuses on demonstrating substantial equivalence for a physical medical device (a cryoablation probe) through non-clinical bench testing. Therefore, it does not provide the information requested in the prompt regarding acceptance criteria and study data for an AI/ML device's diagnostic performance.

Specifically, the document states:

  • Performance Data: "The following bench testing was conducted for design and performance elements deemed appropriate to demonstrate equivalence to the previously cleared CRYO2 device met the predetermined acceptance criteria ensuring substantial equivalence to the previously cleared CRYO2 device. No new safety or performance issues were raising during testing."
  • Non-clinical Bench Testing:
    • Reliability Testing
    • Transit
    • Shelf-life
    • Cryogen Performance/Thermal Insulation
    • Mechanical testing
    • Biocompatibility

Given this, I cannot construct a table of acceptance criteria and reported device performance for an AI/ML diagnostic device, nor can I provide details about sample sizes, expert ground truth establishment, MRMC studies, or standalone algorithm performance, as these concepts are not applicable to the information provided in this 510(k) summary for a physical cryoablation probe.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

November 9, 2018

AtriCure, Inc. Jonathan McElwee Regulatory Affairs Manager 7555 Innovation Way Mason, Ohio 45040

Re: K182565

Trade/Device Name: AtriCure cryoICE cryoSPHERE cryoablation probe Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: Class II Product Code: GXH Dated: September 17, 2018 Received: September 18, 2018

Dear Jonathan McElwee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew C. Krueger -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182565

Device Name cryoICE® cryoSPHERE™ cryoablation probe

Indications for Use (Describe)

AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use in blocking pain by temporarily ablating peripheral nerves.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "AtriCure" in a stylized font. The first three letters, "Atri", are in blue, while the rest of the word, "Cure", is in orange. The "C" in "Cure" has a blue dot in the middle. There is a registered trademark symbol to the right of the word.

510(k) Summary

l. Applicant Information

Manufacturer:AtriCure®, Inc.7555 Innovation WayMason, Ohio 45040P: 513-644-4736F: 513-895-9013
Contact Person:Jonathan McElwee, RACManager, Regulatory Affairs
Date Prepared:9/17/2018
II.Device Information
Proprietary Name:cryoICE™ cryoSPHERE® cryo-ablation probe (CRYOS)
Common Name:Cryosurgical Probe
Classification:Cryogenic surgical deviceRegulatory Class: Class II; per 21 CFR 882.4250Product Code: GXHClassification Panel: Neurodiagnostic and Neurosurgical Devices
Predicate Device:cryoICE cryoablation probe (CRYO2)(K180138, February 15, 2018)
Reference Device:cryoICE cryoablation probe (CRYO2)(K142203, November 25, 2014)

III. Device Description

The cryolCE cryoSPHERE cryoablation (CRYOS) probe is a single use device that achieves cryoablation of peripheral nerves by allowing a surgeon to insert the probe through an incision to reach the target tissue, place the probe tip on the target site, and in conjunction with an AtriCure Cryo Module (ACM), temporarily freeze the tissue in contact with the probe tip by circulating a cryogenic agent, nitrous oxide, through the device. The CRYOS probe is offered in two probe length configurations: approximately 11" and 18" long. The probe is malleable, and is capable of being bent by the end user. At the distal end, the CRYOS device features an 8 mm diameter ball tip shaped probe.

IV. Intended Use/ Indications for Use

AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use blocking pain by temporarily ablating peripheral nerves.

{4}------------------------------------------------

AtriCure®

Comparison of Technological Characteristics (CRYO2 K180138) V.

  • The devices include the same intended use, and; ●
  • No changes were made in operating principle, or specifications of performance, and; .
  • The results of the verification and validation testing: .
    • o Demonstrated equivalency in performance
    • O Did not raise any new issues of safety

Modifications included in the cryoSPHERE probe end-effector were for differing body habitus and surgeon preference for the use of the device specifically intended to blocking pain.

CRYO2 (Predicate)CRYOS (Subject)
Market ProductNamecryoICE cryoablation probecryoICE cryoSPHERE cryoablation probe
510(k) NumberK180138Subject of this Submission
Intended UseAtriCure's cryoICE cryo-ablation probes aresterile, single use devices intended for use in thecryosurgical treatment of cardiac arrhythmias byfreezing target tissues, creating an inflammatoryresponse (cryonecrosis) that blocks the electricalconduction pathway. The probe is also intendedfor use in blocking pain by temporarily ablatingperipheral nerves.AtriCure's cryolCE cryoSPHERE cryoablationprobes are sterile, single use devices intendedfor use blocking pain by temporarily ablatingperipheral nerves.
OperatingPrincipleJoule-Thompson EffectJoule-Thompson Effect
TechnologyThe system consists of cryoprobes that are usedfor freezing target tissue. A console is used tocontrol the supply of gas to the cryoprobe.The system consists of cryoprobes that are usedfor freezing target tissue. A console is used tocontrol the supply of gas to the cryoprobe.
Energy UsedNitrous OxideNitrous Oxide
OperatingTemperature-50°C to 70°C-50°C to 70°C
Human FactorsHand-held device containing cryogen withactivation button on console or footswitch.Hand-held device containing cryogen withactivation button on console or footswitch.
CryotipMaterial: aluminum alloyMaterial: aluminum alloy
Construction: smooth malleable cylindrical linearprobeConstruction: smooth spherical ball welded to amalleable cylindrical aluminum shaft.
Diameter: 4 mmDiameter: ball - 8mm; aluminum shaft - 5mm
Length: 10cmLength: reference the "end-effector" length
End-effectorLength – Shaftand Cryotip28 cm28cm (CRYOS)46cm (CRYOS-L)
BiocompatibilityBiocompatible patient contacting materials.Biocompatible patient contacting materials.
Probe PackagingSterile - Single Use disposable deviceSterile - Single Use disposable device
SterilizationGamma IrradiationGamma Irradiation
Power SourceMains PoweredMains Powered

{5}------------------------------------------------

VI. Performance Data

The following bench testing was conducted for design and performance elements deemed appropriate to demonstrate equivalence to the previously cleared CRYO2 device met the predetermined acceptance criteria ensuring substantial equivalence to the previously cleared CRYO2 device. No new safety or performance issues were raising during testing.

Non-clinical Bench Testing:

  • Reliability Testing ●
  • Transit ●
  • Shelf-life
  • Cryogen Performance/Thermal Insulation ●
  • Mechanical testing .
  • Biocompatibility .

VII. Conclusions

AtriCure has demonstrated that the cryolCE cryoSPHERE cryoablation probe is substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principal, and intended use/ indication for use as the previously cleared device: cryolCE cryoablation probe.

§ 882.4250 Cryogenic surgical device.

(a)
Identification. A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.(b)
Classification. Class II (performance standards).