K Number
K182530

Validate with FDA (Live)

Date Cleared
2018-11-09

(56 days)

Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BIOEASY Multi-Drug Test Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Marijuana50 ng/mL
Methamphetamine1000 ng/mL
Morphine300 ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL

Configuration of the BIOEASY Multi-Drug Test Cup tests can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

Device Description

The BIOEASY Multi-Drug Test Cup tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamphetamine, Phencyclidine, Methadone, Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices. Each test kit contains a Test Device, a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the successful outcomes of the various performance studies. The reported device performance is excellent, demonstrating near-perfect agreement with the LC/MS ground truth for samples outside the +/-25% cutoff range, and reasonable agreement within the +/-25% cutoff range, which is expected for qualitative tests around the cutoff.

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance Summary
PrecisionConsistent and accurate results across different lots and concentrations.For -100% to -25% of cutoff: 100% negative results. For +25% to +100% of cutoff: 100% positive results. At cutoff: varying split between positive/negative (e.g., Amphetamine Lot 1: 23-/27+), which is expected for qualitative assays at the decision point.
StabilityDevice remains effective for its stated shelf life.Stable at 4-30 °C for 24 months (accelerated stability). Real-time studies ongoing.
InterferenceNo significant interference from common physiological/pathological substances.No interference observed for a wide range of common substances at specified concentrations (summarized in tables on page 9-10).
Specificity (Cross-reactivity)Correct identification of target drugs/metabolites, with acceptable cross-reactivity to similar compounds.Detailed cross-reactivity tables provided for each drug, showing specificity to the target analyte and acceptable cross-reactivity percentages for related compounds (pages 10-13).
Effect of Urine Specific Gravity & pHPerformance unaffected by variations in urine specific gravity and pH.No differences observed for samples at +/-25% cut-off levels with specific gravity from 1.000 to 1.035 and pH 4-9.
Method Comparison (Professional User)High agreement with LC/MS reference method.For each drug, out of 80 samples (40 negative, 40 positive), most samples outside the near cut-off range were correctly identified. Discordant results primarily occurred in the near cut-off positive/negative ranges, where the device correctly identified values slightly above the cutoff as positive and slightly below as negative in many cases, even when LC/MS showed a value across the cutoff. Overall strong agreement.
Lay-user StudyHigh percentage of correct results by untrained users, and ease of understanding instructions.For -100% to -50% of cutoff: 100% correct negative results. For +50% to +100% of cutoff: 100% correct positive results. At -25% and +25% of cutoff: 95% correct results for most drugs, with Nortriptyline at +25% having 90%. All lay users found instructions easy to follow, and Flesch-Kincaid score indicated Grade Level 7 reading.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Studies:

    • For each of the 14 analytes, 8 concentrations were tested (ranging from -100% cut off to +100% cut off).
    • For each concentration, tests were performed two runs per day for 25 days, using 3 different lots of the device.
    • Total samples per lot per concentration: 50 (2 runs/day * 25 days).
    • Total samples per drug per concentration for 3 lots: 150 (50 * 3 lots).
    • Total samples per drug: 1,200 (150 * 8 concentrations).
    • Total samples for all 14 drugs in precision study: 16,800.
    • Data Provenance: The text does not explicitly state the country of origin but implies it was performed "in-house" by the manufacturer (Shenzhen Bioeasy Biotechnology Co., Ltd. in China). These samples were prepared by spiking drug in negative samples. The study appears to be prospective in nature for the purpose of device validation.
  • Interference & Specificity Studies:

    • "Three batches of each device" were used.
    • The number of individual samples tested for each interfering substance or cross-reactant is not explicitly stated, but it involved "drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels."
  • Effect of Urine Specific Gravity and pH:

    • "Three lots of each device" were used.
    • "Urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels." The exact number of samples at each specific gravity/pH level is not detailed.
  • Method Comparison (Professional User) Studies:

    • For each of the 14 analytes: 80 unaltered clinical samples (40 negative and 40 positive).
    • Total samples for 14 drugs: 1120 samples.
    • Data Provenance: "clinical samples" are mentioned, but their country of origin is not specified. The study is retrospective, as these are "unaltered clinical samples" compared to LC/MS results.
  • Lay-user Study:

    • 300 lay persons were involved.
    • The text describes sample numbers for various concentrations. For each drug, a total of 300 samples were prepared the following way: 20 at -100% Cutoff, 20 at -75%, 160 at -50%, 20 at -25%, 20 at +25%, 40 at +50%, 20 at +75%.
    • Total samples per drug: 300.
    • Total samples for all 14 drugs in lay-user study: 4,200.
    • Data Provenance: The text does not explicitly state the country of origin for the urine samples, but the study was conducted at "three intended user sites." These samples were prepared by spiking drugs into drug-free pooled urine specimens. The study itself is prospective in terms of collecting performance data from lay users.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Method Comparison Studies: The ground truth was established by LC/MS (Liquid Chromatography-Mass Spectrometry). This is a highly accurate analytical method, considered the gold standard for drug confirmation in urine. No human experts are explicitly mentioned for establishing the ground truth in these studies, as the LC/MS directly provides it. The "three laboratory assistants" who ran the devices are operators, not ground truth experts.

  • Precision, Interference, Specificity, Effect of Urine Specific Gravity and pH, Lay-user Studies: The ground truth for these studies was established by LC/MS to confirm the drug concentrations in the spiked urine samples.

4. Adjudication Method for the Test Set

  • Precision Studies: Not explicitly stated, however, the results were quantified as positive/negative based on the criteria for each concentration and then tabulated. No formal adjudication process among multiple readers is described for the precision data, as the device's output (presence/absence of line) is objectively recorded.

  • Method Comparison Studies: Three "laboratory assistants" viewed the device results, and their readings were compared against the LC/MS ground truth. There is no mention of an adjudication process (e.g., 2+1, 3+1) among these three assistants. Each viewer's discordant results were individually reported.

  • Lay-user Study: The results from each lay person were recorded (positive/negative) and then compared to the known LC/MS-confirmed concentration of the sample. No adjudication among lay users is mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This device is a rapid, qualitative drug test cup, which provides a direct visual result (lines appearing/not appearing). It does not involve AI assistance, nor does it involve human readers interpreting complex images or data that AI would augment. Therefore, there's no discussion of human readers improving with AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, in the context of the device's output. The "Precision" studies and the "Interference," "Specificity," and "Effect of Urine Specific Gravity and pH" studies represent the standalone performance of the device itself to detect the analytes, as the results are based on the chemical reactions within the cup. The "Method Comparison" study also evaluates the standalone performance of the device's reading mechanism when interpreted by professional users, directly comparing it to the LC/MS reference.

7. The Type of Ground Truth Used

  • The primary ground truth used across all analytical and clinical (in-house) studies was LC/MS (Liquid Chromatography-Mass Spectrometry). For the precision, interference, specificity, and pH/SG studies, the samples were spiked with known concentrations confirmed by LC/MS. For the method comparison study, unaltered clinical samples were directly compared against LC/MS results.

8. The Sample Size for the Training Set

  • This device is a lateral flow immunochromatographic assay, not an AI/machine learning model. Therefore, there is no concept of a "training set" in the traditional sense for an algorithm. The development and optimization of the test's chemical components and cutoff concentrations implicitly involve internal R&D and calibration, but this is distinct from training an AI model on a dataset.

9. How the Ground Truth for the Training Set was Established

  • As stated above, this is not applicable as it's not an AI/machine learning device requiring a training set. The "ground truth" for establishing the device's analytical performance (e.g., cutoff levels, cross-reactivity) would have been determined through empirical testing and comparison against established analytical methods like LC/MS during the device's development and validation.

{0}------------------------------------------------

Build Correspondence

Image /page/0/Picture/2 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 9, 2018

Shenzhen Bioeasy Biotechnology Co., Ltd. % Joe Shia, Director LSI International 504E Diamond Ave., Suite I Gaithersburg, MD 20877

Re: K182530

Trade/Device Name: BIOEASY Multi-Drug Test Cup Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: NFT, NFW, NFY, NGG, NGI, NFV, NGL, PTH, NGM, PTG, QAW, QBF Dated: September 9, 2018 Received: September 14, 2018

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182530

Device Name BIOEASY Multi-Drug Test Cup

Indications for Use (Describe)

BIOEASY Multi-Drug Test Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Marijuana50 ng/mL
Methamphetamine1000 ng/mL
Morphine300 ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL

Configuration of the BIOEASY Multi-Drug Test Cup tests can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K182530 510(k) SUMMARY

  • November 9, 2018 1. Date:
  • Shenzhen Bioeasy Biotechnology Co., Ltd. 2. Submitter: No.2-1, Liuxian 1st Road Baoan District Shenzhen, China 518101
  • Joe Shia 3. Contact person: LSI International Inc. 504 East Diamond Ave. Gaithersburg, MD 20877 Telephone: 240-505-7880 Email: shiajl@yahoo.com
    1. Device Name: BIOEASY Multi-Drug Test Cup
Classification:Class 2
Product CodeClassificationRegulation SectionPanel
NFTAmphetamineII21 CFR § 862.3100, AmphetamineTest SystemToxicology (91)
NFWCannabinoidsII21 CFR § 862.3870, CannabinoidsTest SystemToxicology (91)
NFYCocaineII21 CFR § 862.3250, Cocaine andCocaine Metabolites Test SystemToxicology (91)
NGGMethamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology (91)
NGIMorphineII21 CFR § 862.3640, MorphineTest SystemToxicology (91)
NFVOxazepamII21 CFR § 862.3170,Benzodiazepine Test SystemToxicology (91)
NGLOxycodoneII21 CFR § 862.3650, Opiate TestSystemToxicology (91)
PTHSecobarbitalII21 CFR § 862.3150, BarbiturateTest SystemToxicology (91)
NGLBuprenorphineII21 CFR § 862.3650,Opiate Test SystemToxicology (91)
NGGMethylenedioxy-methamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology (91)
NGMPhencyclidineunclassifiedEnzyme ImmunoassayPhencyclidineToxicology (91)
PTGMethadoneII21 CFR § 862.3620, MethadoneTest SystemToxicology (91)
QAWNortriptylineII21 CFR, 862.3910 TricyclicAntidepressant Drugs Test SystemToxicology (91)
QBFPropoxypheneII21 CFR, 862.3700 PropoxypheneTest SystemToxicology (91)
  1. Predicate Devices: K153050

{5}------------------------------------------------

The CO-INNOVATION BIOTECH Rapid Multi-Drug Test Dip Card and Rapid Multi-Drug Test Cup

    1. Indications for Use
      BIOEASY Multi-Drug Test Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:
Drug (Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Marijuana50 ng/mL
Methamphetamine1000 ng/mL
Morphine300 ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL

Configuration of the BIOEASY Multi-Drug Test Cup tests can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. For in vitro diagnostic use only.

    1. Device Description
      The BIOEASY Multi-Drug Test Cup tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamphetamine, Phencyclidine, Methadone, Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices. Each test kit contains a Test Device, a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch
    1. Substantial Equivalence Information
      A summary comparison of features of the BIOEASY Multi-Drug Test Cup tests and the predicate devices is provided in following table.

{6}------------------------------------------------

ItemDevicePredicate - K153050
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
Calibrator and Cut-OffValuesAmphetamine (AMP): 1,000 ng/mlOxazepam (BZO):300 ng/mlCocaine(COC): 300 ng/mlMarijuana (THC):50 ng/mlMethamphetamine (MET): 1,000 ng/mlMorphine (MOR): 300ng/mlOxycodone(OXY) : 100 ng/mlSecobarbital (BAR): 300 ng/mlBuprenorphine (BUP): 10 ng/mlMethylenedioxy-methamphetamine(MDMA): 500 ng/mlPhencyclidine (PCP): 25 ng/mlMethadone (MTD): 300 ng/mlNortriptyline (TCA): 1000 ng/mlPropoxyphene (PPX): 300 ng/mlSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor over-the-counterSame
ConfigurationsCupSame

Table 1: Features Comparison of BIOEASY Multi-Drug Test Cup tests and the Predicate Devired

9. Test Principle

The BIOEASY Multi-Drug Test Cup tests are rapid tests for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline and Propox yphene in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibodycoated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices

{7}------------------------------------------------

regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

    1. Analytical Performance

a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables.

Amphetamine

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

Secobarbital

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

Buprenorphine

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

Oxazepam

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

Cocaine

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoff+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-

{8}------------------------------------------------

Methylenedioxy-methamphetamine

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoff+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/24+50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-

Methamphetamine

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+29-/21+50+/0-50+/0-50+/0-50+/0-

Morphine

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

Methadone

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+29-/21+50+/0-50+/0-50+/0-50+/0-

Oxycodone

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

Phencyclidine

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

Propoxyphene

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoff+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

Nortriptyline

{9}------------------------------------------------

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+21-/29+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Marijuana

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-

c. Stability

The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 45 °C. Real time stability determinations are ongoing at both 4 °C and 30 °C.

d. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL (albumin was tested at 100 mg/dL) are summarized in the following tables.

Acetaminophenβ-EstradiolOxalic acid
AcetophenetidinErythromycinOxolinic acid
N-AcetylprocainamideEthanolOxymetazoline
Acetylsalicylic acidFenoprofenPapaverine
Albumin (100 mg/dL)FurosemidePenicillin G
AminopyrineGentisic acidPerphenazine
AmoxicillinHemoglobinPhenelzine
AmpicillinHydralazinePrednisone
ApomorphineHydrochlorothiazide(±)-Propranolol
Ascorbic acidHydrocortisonePseudoephedrine
AspartameO-Hydroxyhippuric acidQuinine
Atropine3-HydroxytyramineRanitidine
Benzilic acidIbuprofenSalicylic acid
Benzoic acidIsoproterenolSerotonin (5- Hydroxytyramine)
BilirubinIsoxsuprineSulfamethazine
Chloral hydrateKetamineSulindac
ChloramphenicolKetoprofenTetrahydrocortisone 3-(β-
Dglucuronide)
ChlorothiazideLabetalolTetrahydrocortisone 3-acetate
ChlorpromazineLoperamideTetrahydrozoline
CholesterolMeperidineThiamine
ClonidineMeprobamateThioridazine
CortisoneMethoxyphenamineTriamterene
(-)-CotinineNalidixic acidTrifluoperazine
CreatinineNaloxoneTrimethoprim
DeoxycorticosteroneNaltrexoneDL-Tryptophan
DextromethorphanNaproxenTyramine
DiclofenacNiacinamideDL-Tyrosine

{10}------------------------------------------------

DiflunisalNifedipineUric acid
DigoxinNorethindroneVerapamil
DiphenhydramineNoscapineZomepirac
Ecgonine methyl ester(±)-Octopamine

e. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below.

AMP(Amphetamine)(D - Amphetamine, Cut-off=1000 ng/mL)Result%Cross-Reactivity
D - AmphetaminePositive at 1000 ng/mL100%
L - AmphetaminePositive at 20000 ng/mL5%
DL - AmphetaminePositive at 3000 ng/mL33%
PhenterminePositive at 30000 ng/mL3.3%
HydroxyamphetaminePositive at 8000 ng/mL12.5%
Methylenedioxyamphetamine (MDA)Positive at 20000 ng/mL5%
d-MethamphetaminePositive> 100000 ng/mL<1%
1-MethamphetaminePositive> 100000 ng/mL<1%
EphedrinePositive> 100000 ng/mL<1%
Methylenedioxyethylamphetamine (MDE)Positive> 100000 ng/mL<1%
3,4-methylenedioxy-methamphetamine(MDMA)Positive> 100000 ng/mL<1%
BAR(Barbital)(Secobarbital, Cut-off=300 ng/mL)Result%Cross-Reactivity
SecobarbitalPositive at 300 ng/mL100%
AmobarbitalPositive at 1000 ng/mL30%
AlphenalPositive at 62.5 ng/mL480%
AprobarbitalPositive at 250 ng/mL120%
ButabarbitalPositive at 100 ng/mL300%
ButethalPositive at 500 ng/mL60%
ButalbitalPositive at 5000 ng/mL6%
CyclopentobarbitalPositive at 500 ng/mL60%
PentobarbitalPositive at 200 ng/mL150%
PhenobarbitalPositive at 300 ng/mL100%
BUP(Buprenorphine)(Buprenorphine, Cut-off=10 ng/mL)Result%Cross-Reactivity
BuprenorphinePositive at 10 ng/mL100%
Buprenorphine -3-D-GlucuronidePositive at 10 ng/mL100%
NorbuprenorphinePositive at 50 ng/mL20%
Norbuprenorphine-3-D-GlucuronidePositive at 10 ng/mL10%
MorphinePositive> 100000 ng/mL<0.01%
OxymorphonePositive> 100000 ng/mL<0.01%
HydromorphonePositive> 100000 ng/mL<0.01%
BZO(Benzodiazepines)Result%Cross-Reactivity
-------------------------------------------------

{11}------------------------------------------------

(Oxazepam, Cut-off=300 ng/mL)
OxazepamPositive at_300_ng/mL100%
AlprazolamPositive at_150_ng/mL200%
a-HydroxyalprazolamPositive at_1000_ng/mL30%
BromazepamPositive at_1000_ng/mL30%
ChlordiazepoxidePositive at_63_ng/mL476.2%
ClonazepamPositive at_2500_ng/mL12%
ClobazamPositive at_75_ng/mL400%
Clorazepate dipotassiumPositive at_100_ng/mL300%
DesalkylflurazepamPositive at_500_ng/mL60%
DiazepamPositive at_500_ng/mL60%
EstazolamPositive at_500_ng/mL60%
FlunitrazepamPositive > 50000 ng/mL<0.6%
D,L-LorazepamPositive at_10000_ng/mL3%
MidazolamPositive at_10000_ng/mL3%
NitrazepamPositive at_75_ng/mL400%
NorchlordiazepoxidePositive at_62.5_ng/mL480%
NordiazepamPositive at_125_ng/mL240%
TemazepamPositive at_75_ng/mL400%
TriazolamPositive at_1000_ng/mL33%
COC(Cocaine)(Benzoylecgonine, Cut-off=300 ng/mL)Result%Cross-Reactivity
BenzoylecgoninePositive at 300 ng/mL100%
Cocaine HClPositive at 750 ng/mL40%
CocaethylenePositive at 12500 ng/mL2.4%
EcgoninePositive at 32000 ng/mL0.9%
NorcocainePositive at 100000 ng/mL0.3%
MDMA(Methylenedioxymethamphetamine)(Methylenedioxymethamphetamine,Cut-off=500 ng/mL)Result%Cross-Reactivity
Methylenedioxymethamphetamine (MDMA)Positive at 500 ng/mL100%
3,4-Methylenedioxyamphetamine (MDA)Positive at 5000 ng/mL10%
3,4-Methylenedioxyethylamphetamine (MDEA)Positive at 300 ng/mL166.7%
d-methamphetaminePositive > 50000 ng/mL<1%
d-amphetaminePositive > 50000 ng/mL<1%
l-amphetaminePositive > 50000 ng/mL<1%
l-methamphetaminePositive > 50000 ng/mL<1%
MET(Methamphetamine)(D(+)-Methamphetamine, Cut-off=1000 ng/mL)Result%Cross-Reactivity
D(+)-MethamphetaminePositive at 1000 ng/mL100%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)Positive at 10000 ng/mL10%
D/L-MethamphetaminePositive at 1000 ng/mL100%
p-HydroxymethamphetaminePositive at 10000 ng/mL10%

{12}------------------------------------------------

D-AmphetaminePositive > 100000 ng/mL<1%
L-AmphetaminePositive >100000 ng/mL<1%
ChloroquinePositive at 50000 ng/mL2%
(+/-)-EphedrinePositive at 4000 ng/mL25%
L-MethamphetaminePositive at 10000 ng/mL10%
(+/-)3,4-Methylenedioxyamphetamine (MDA)Positive >100000 ng/mL<1%
β-PhenylethylaminePositive at 7500 ng/mL13.3%
TrimethobenzamidePositive at 20000 ng/mL5%
(+/-)3,4-methylenedioxymethamphetamine(MDMA)Positive at 10000 ng/mL10%
MOP(Morphine)(Morphine, Cut-off=300 ng/mL)Result%Cross-Reactivity
MorphinePositive at_300 ng/mL100%
CodeinePositive at 300 ng/mL100%
EthylmorphinePositive at_310 ng/mL96.8%
HydrocodonePositive at_25000 ng/mL1.2%
HydromorphonePositive 10000 ng/mL3%
LevorphanolPositive > 100000 ng/mL<0.3%
6-AcetylmorphinePositive at_250 ng/mL120%
Morphine-3- β -D-glucuronidePositive at 10000 ng/mL3%
NormorphinePositive at 100000 ng/mL0.3%
OxycodonePositive > 10000 ng/mL<3%
OxymorphonePositive > 10000 ng/mL<3%
ProcainePositive > 10000 ng/mL3%
ThebainePositive > 10000 ng/mL<3%
HeroinPositive at 500 ng/mL60%
MTD(Methadone)(Methadone, Cut-off=300 ng/mL)Result%Cross-Reactivity
MethadonePositive at 300ng/mL100%
DoxylaminePositive at 5000ng/mL6%
LAAMPositive at 10000ng/mL3%
Alpha MethadolPositive at 2000ng/mL15%
EDDPPositive > 100000ng/mL<0.3%
EMDPPositive > 100000ng/mL<0.3%
OXY(Oxycodone)(Oxycodone, Cut-off=300 ng/mL)Result%Cross-Reactivity
OxycodonePositive at 100 ng/mL100%
DihydrocodeinePositive >100000ng/mL<0.1%
CodeinePositive >100000ng/mL<0.1%
HydromorphonePositive >100000ng/mL<0.1%
MorphinePositive >100000ng/mL<0.1%
BuprenorphinePositive >100000ng/mL<0.1%
EthylmorphinePositive >100000ng/mL<0.1%
OxymorphonePositive at 250 ng/mL40%

{13}------------------------------------------------

r y y are a1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -0 00/
HydrocodonePositive at 3125 ng/mLJ. L. / V
PCP(Phencyclidine)(Phencyclidine, Cut-off=25 ng/mL)Result%Cross-Reactivity
PhencyclidinePositive at 25 ng/mL100%
4-HydroxyphencyclidinePositive at 75 ng/mL33.3%
PPX(Propoxyphene)(d-Propoxyphene, Cut-off=300ng/mL)Result%Cross-Reactivity
d-PropoxyphenePositive at 300 ng/mL100%
D-NorpropoxyphenePositive at 333 ng/mL90.1%
TCA(Nortriptyline)(Nortriptyline, Cut-off=1000 ng/mL)Result%Cross-Reactivity
NortriptylinePositive at 1000 ng/mL100%
AmitriptylinePositive at 750 ng/mL133.3%
ClomipraminePositive at 10000 ng/mL10%
DesipraminePositive at 200 ng/mL500%
DoxepinPositive at 1250ng/mL80%
ImipraminePositive at 625 ng/mL160%
MaprotilinePositive at 2000 ng/mL50%
NordoxepinPositive at 1000 ng/mL100%
PromazinePositive at 1500 ng/mL66.7%
PromethazinePositive at 25000 ng/mL4%
TrimipraminePositive at 3000 ng/mL33.3%
CyclobenzaprinePositive at 5000 ng/mL20%
NorclomipraminePositive at 3000 ng/mL33.3%
THC(Cannabinoids)(11-nor-△9-THC-9-COOH, Cut-off = 50 ng/mL)Result% Cross-Reactivity
11-nor-△9-THC-9-COOHPositive at 50 ng/mL100%
11-Hydroxy-△9-TetrahydrocannabinolPositive at 50 ng/mL100%
11-Nor-△8-Tetrahydrocannabinol-9-COOHPositive at 50 ng/mL100%
CannabinolPositive at 20000 ng/mL0.25%
△8-TetrahydrocannabinolPositive at 15000 ng/mL0.33%
△9-TetrahydrocannabinolPositive at 15000 ng/mL0.33%
CannabidiolPositive> 100000 ng/mL<0.05%
11-Nor-△9-THC-carboxy glucuronidePositive at 75 ng/mL66.7%
(-)-11-nor-9-carboxy-△ 9-THCPositive at 50 ng/mL100%

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different devices.

{14}------------------------------------------------

2. Comparison Studies

Method comparison studies for the BIOEASY Multi-Drug Test Cup tests were performed inhouse with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results. The results are presented in the tables below.

Amphetamine

NegativeLow Negative by LC/MS (less than -50%)Near Cutoff Negative by LC/MS (Between -50% and cutoff)Near Cutoff Positive by LC/MS (Between the cutoff and +50%)High Positive by LC/MS (greater than +50%)
Viewer A Positive0021919
Viewer A Negative5161720
Viewer B Positive0042019
Viewer B Negative5161510
Viewer C Positive0031919
Viewer C Negative5161620
Discordant Results
ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer AAMPC439981Positive
Viewer AAMPC449768Positive
Viewer BAMPC439981Positive
Viewer BAMPC449768Positive
Viewer BAMPC412976Positive
Viewer BAMPC387981Positive
Viewer CAMPC412976Positive
Viewer CAMPC493961Positive
Viewer CAMPC387981Positive
Viewer AAMPC3411010Negative
Viewer AAMPC4821010Negative
Viewer BAMPC4821010Negative
Viewer CAMPC3411010Negative
Viewer CAMPC3521120Negative

Secobarbital

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0021722
ANegative6151710
ViewerPositive0021622
BNegative6151720
ViewerPositive0011722

{15}------------------------------------------------

CNegative6151810
Discordant Results
ViewerSample NumberLC/MS ResultBIOEASY Cup Viewer Results
Viewer ABARC477291Positive
Viewer ABARC375265Positive
Viewer BBARC477291Positive
Viewer BBARC375265Positive
Viewer CBARC364269Positive
Viewer ABARC384312Negative
Viewer BBARC384312Negative
Viewer BBARC414315Negative
Viewer CBARC414315Negative

Buprenorphine

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0012018
ANegative5181620
ViewerPositive0002018
BNegative5181720
ViewerPositive0022118
CNegative5181510

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer ABUPC3799.67Positive
Viewer CBUPC3598.94Positive
Viewer CBUPC3799.67Positive
Viewer ABUPC32210.1Negative
Viewer ABUPC42911.2Negative
Viewer BBUPC32210.1Negative
Viewer BBUPC34810.2Negative
Viewer CBUPC34810.2Negative

Oxazepam

LowNear CutoffNear Cutoff
NegativeNegative byNegative byPositive byHigh Positive
LC/MSLC/MSLC/MSby LC/MS
(less than(Between(Between the(greater than
-50%)-50% andcutoff and+50%)
cutoff)+50%)

{16}------------------------------------------------

Viewer0022018
APositive0022018
Negative5151820
Viewer0012118
BPositive0012118
Negative5151910
Viewer0022118
CPositive0022118
Negative5151810

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY Cup Viewer Results
Viewer ABZOC343298Positive
Viewer ABZOC322296Positive
Viewer BBZOC322296Positive
Viewer CBZOC343298Positive
Viewer CBZOC435292Positive
Viewer ABZOC301314Negative
Viewer ABZOC348312Negative
Viewer BBZOC348312Negative
Viewer CBZOC301314Negative

Cocaine

NegativeLow Negative by LC/MS(less than -50%)Near Cutoff Negative by LC/MS(Between -50% and cutoff)Near Cutoff Positive by LC/MS(Between the cutoff and +50%)High Positive by LC/MS(greater than +50%)
ViewerPositive0022119
ANegative6171500
ViewerPositive0022019
BNegative6171510
ViewerPositive0012019
CNegative6171410
Discordant Results

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY Cup Viewer Results
Viewer ACOCC368298Positive
Viewer ACOCC340294Positive
Viewer BCOCC368298Positive
Viewer BCOCC315282Positive
Viewer CCOCC340294Positive
Viewer BCOCC317301Negative
Viewer CCOCC317301Negative

Methylenedioxymethamphetamine

{17}------------------------------------------------

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0011920
ANegative5161810
ViewerPositive0021920
BNegative5161710
ViewerPositive0021920
CNegative5161710

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer AMDMAC310490Positive
Viewer BMDMAC306497.5Positive
Viewer BMDMAC365484.5Positive
Viewer CMDMAC310490Positive
Viewer CMDMAC306497.5Positive
Viewer AMDMAC316505Negative
Viewer BMDMAC316505Negative
Viewer CMDMAC484565Negative

Methamphetamine

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0022018
ANegative5161720
ViewerPositive0012118
BNegative5161810
ViewerPositive0012118
CNegative5161810

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer AMETC498921Positive
Viewer AMETC479984Positive
Viewer BMETC467973Positive
Viewer CMETC479984Positive
Viewer AMETC3431020Negative
Viewer AMETC4731110Negative
Viewer BMETC4731110Negative

{18}------------------------------------------------

Viewer CMETC3431020Negative
-----------------------------------

Morphine

NegativeLow Negative by LC/MS (less than -50%)Near Cutoff Negative by LC/MS (Between -50% and cutoff)Near Cutoff Positive by LC/MS (Between the cutoff and +50%)High Positive by LC/MS (greater than +50%)
ViewerPositive0021821
ANegative5161710
ViewerPositive0021821
BNegative5161710
ViewerPositive0021721
CNegative5161720

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer AMOPC355287Positive
Viewer AMOPC429298Positive
Viewer BMOPC408286Positive
Viewer BMOPC429298Positive
Viewer CMOPC408286Positive
Viewer CMOPC433276Positive
Viewer AMOPC393301Negative
Viewer BMOPC416339Negative
Viewer CMOPC416339Negative
Viewer CMOPC393301Negative

Methadone

NegativeLow Negative by LC/MS (less than -50%)Near Cutoff Negative by LC/MS (Between -50% and cutoff)Near Cutoff Positive by LC/MS (Between the cutoff and +50%)High Positive by LC/MS (greater than +50%)
Viewer APositive0022118
Viewer ANegative5151810
Viewer BPositive0022018
Viewer BNegative5151820
Viewer CPositive0032218
Viewer CNegative5151700

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY Cup Viewer Results
Viewer AMTDC314289Positive
Viewer AMTDC389299Positive

{19}------------------------------------------------

Viewer BMTDC314289Positive
Viewer BMTDC408290Positive
Viewer CMTDC389299Positive
Viewer CMTDC373279Positive
Viewer CMTDC408290Positive
Viewer AMTDC447309Negative
Viewer BMTDC447309Negative
Viewer BMTDC427336Negative

Oxycodone

LowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive041822
Negative61500
ViewerBPositive031822
Negative61500
ViewerCPositive031822
Negative61600

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer AOXYC46797.5Positive
Viewer AOXYC36994.6Positive
Viewer AOXYC33498.4Positive
Viewer AOXYC47095.1Positive
Viewer BOXYC46797.5Positive
Viewer BOXYC31096.1Positive
Viewer BOXYC47095.1Positive
Viewer COXYC31096.1Positive
Viewer COXYC36994.6Positive
Viewer COXYC33498.4Positive

Phencyclidine

NegativeLow Negative by LC/MS(less than -50%)Near Cutoff Negative by LC/MS(Between -50% and cutoff)Near Cutoff Positive by LC/MS(Between the cutoff and +50%)High Positive by LC/MS(greater than +50%)
Viewer APositive0011820
Negative3181820
Viewer BPositive0022020
Negative3181700

{20}------------------------------------------------

Viewer0021920
CPositive0021920
CNegative3181710

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer APCPC44124.8Positive
Viewer BPCPC41124.7Positive
Viewer BPCPC44124.8Positive
Viewer CPCPC46224.7Positive
Viewer CPCPC41124.7Positive
Viewer APCPC47725.3Negative
Viewer APCPC36827.8Negative
Viewer CPCPC47725.3Negative

Propoxyphene

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0011820
ANegative5151920
ViewerPositive0021920
BNegative5151810
ViewerPositive0011920
CNegative5151910

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY Cup Viewer Results
Viewer APPXC472286Positive
Viewer BPPXC472286Positive
Viewer BPPXC320278Positive
Viewer CPPXC320278Positive
Viewer APPXC493318Negative
Viewer APPXC382306Negative
Viewer BPPXC382306Negative
Viewer CPPXC493318Negative

Nortriptyline

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0022019

{21}------------------------------------------------

ANegative5151810
ViewerPositive0021819
BNegative5151830
ViewerPositive0021919
CNegative5151820

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer ATCAC319979Positive
Viewer ATCAC470944Positive
Viewer BTCAC319979Positive
Viewer BTCAC346958Positive
Viewer CTCAC346958Positive
Viewer CTCAC470944Positive
Viewer ATCAC4791190Negative
Viewer BTCAC3981050Negative
Viewer BTCAC3311180Negative
Viewer BTCAC4791190Negative
Viewer CTCAC3981050Negative
Viewer CTCAC3311180Negative

Marijuana

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0021918
ANegative6141830
ViewerPositive0022018
BNegative6141820
ViewerPositive0021918
CNegative6141830

Discordant Results

ViewerSample NumberLC/MS ResultBIOEASY CupViewer Results
Viewer ATHCC40246Positive
Viewer ATHCC36848.8Positive
Viewer BTHCC36346.95Positive
Viewer BTHCC40246Positive
Viewer CTHCC36346.95Positive
Viewer CTHCC36848.8Positive
Viewer ATHCC47855.5Negative
Viewer ATHCC49350.5Negative
Viewer ATHCC49657.5Negative
Viewer BTHCC47855.5Negative

{22}------------------------------------------------

Viewer BTHCC31856Negative
Viewer CTHCC49350.5Negative
Viewer CTHCC31856Negative
Viewer CTHCC49657.5Negative

Lay-user study

A lay user study was performed at three intended user sites with 300 lay persons for the device. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested. Summary results are shown below.

The results summary for AMP:

% of CutoffNumber ofsamplesDrug Concentrationby LC/MS/MS(ng/mL)Lay person ResultsNo. of PositiveLay person ResultsNo. of NegativeThe percentage ofcorrect results (%)
-100% Cutoff200020100
-75% Cutoff20261020100
-50% Cutoff1605070160100
-25% Cutoff2077111995
+25% Cutoff20129019195
+50% Cutoff401560400100
+75% Cutoff201870200100

The results summary for BAR:

% of CutoffNumber of samplesDrug Concentration by LC/MS/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff2075.9020100
-50% Cutoff1601500160100
-25% Cutoff2022011995
+25% Cutoff2036019195
+50% Cutoff40429400100
+75% Cutoff20501200100

The results summary for COC:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2081.5020100
-50% Cutoff1601510160100
-25% Cutoff2022511995
+25% Cutoff2039519195
+50% Cutoff40455400100

{23}------------------------------------------------

00. 7-0/.EACV International Comments of the Comments of Children of the Children Comments of the Children Children of the Children Children of the Children Children of the Children ChildAn1 V V

The results summary for BUP:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff202.57020100
-50% Cutoff1605.140160100
-25% Cutoff206.7611995
+25% Cutoff2012.819195
+50% Cutoff4015.1400100
+75% Cutoff2017.2200100

The results summary for MET:

% of CutoffNumber of samplesDrug Concentration by LC/MS/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff20268020100
-50% Cutoff1605260160100
-25% Cutoff2076911995
+25% Cutoff20127019195
+50% Cutoff401560400100
+75% Cutoff201780200100

The results summary for MTD:

% of CutoffNumber of samplesDrug Concentration by LC/MS/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff2076.8020100
-50% Cutoff1601470160100
-25% Cutoff2022611995
+25% Cutoff2037519195
+50% Cutoff40441400100
+75% Cutoff20504200100

The results summary for MOP:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage of
No. ofPositiveNo. ofNegativecorrect results(%)
-100% Cutoff200020100
-75% Cutoff2079020100
-50% Cutoff1601580160100
-25% Cutoff2024611995
+25% Cutoff20389200100
+50% Cutoff40469400100

{24}------------------------------------------------

+75% Cutoff20530200100
The results summary for OYY.

The results summary for OXY:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2024.5020100
-50% Cutoff16049.30160100
-25% Cutoff2071.111995
+25% Cutoff2011819195
+50% Cutoff40147400100
+75% Cutoff20169200100

The results summary for PCP:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff206.27020100
-50% Cutoff16012.50160100
-25% Cutoff2017.911995
+25% Cutoff2030.819195
+50% Cutoff4036.4400100
+75% Cutoff2042.8200100

The results summary for THC:

% of CutoffNumber of samplesDrug Concentration by LC/MS/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff2013020100
-50% Cutoff16025.30160100
-25% Cutoff204111995
+25% Cutoff206519195
+50% Cutoff4079400100
+75% Cutoff2093200100

The results summary for BZO:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2070.8020100
-50% Cutoff1601480160100
-25% Cutoff2022411995
+25% Cutoff2039019195
+50% Cutoff40452400100
+75% Cutoff20504200100

{25}------------------------------------------------

The results summary for MDMA:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20137020100
-50% Cutoff1602500160100
-25% Cutoff2035111995
+25% Cutoff2060019195
+50% Cutoff40745400100
+75% Cutoff20925200100

The results summary for TCA:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff20273020100
-50% Cutoff1605090160100
-25% Cutoff2080911995
+25% Cutoff20119018290
+50% Cutoff401510400100
+75% Cutoff201680200100

The results summary for PPX:

% of CutoffNumber ofsamplesDrug ConcentrationbyLC/MS/MS(ng/mL)Lay person ResultsThe percentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2077.4020100
-50% Cutoff1601500160100
-25% Cutoff2022711995
+25% Cutoff2035119195
+50% Cutoff40420400100
+75% Cutoff20492200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the device, it's concluded that the BIOEASY Multi-Drug Test Cup tests are substantially equivalent to the predicate.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).