CerusEndo Microcatheter

K182487 · Cerus Endovascular, Inc. · DQY · Jul 20, 2019 · Cardiovascular

Device Facts

Record IDK182487
Device NameCerusEndo Microcatheter
ApplicantCerus Endovascular, Inc.
Product CodeDQY · Cardiovascular
Decision DateJul 20, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The CerusEndo Microcatheter is intended for general intravascular use to deliver therapeutic devices to the peripheral, coronary, and neurovasculature.

Device Story

CerusEndo Microcatheter is a percutaneous, intravascular catheter designed for delivery of therapeutic devices to peripheral, coronary, and neurovascular sites. Device features include a proximal Luer hub, hydrophilic coating, radiopaque marker, polymer tapered shaft, stainless steel reinforcement, and Teflon-lined inner lumen. Available in three configurations (150 cm or 160 cm lengths). Physician operates device in hospital interventional neuroradiology suites; access gained via introducer sheath, advanced over guidewire through guide catheter. Distal tip is steam-shapeable at point of use to navigate tortuous anatomy. Device facilitates precise placement of interventional tools; output is the successful delivery of therapeutic devices to target vascular locations. Benefits include improved access to complex vascular anatomy for minimally invasive procedures.

Clinical Evidence

Bench testing only. Testing included tensile strength, shaft flexibility, shape retention, kink resistance, static burst, simulated use, particulate assessment, coating durability, and sterilization validation. Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility (hemolysis, complement activation, thrombogenicity), and genotoxicity (Ames assay, micronucleus). All tests met acceptance criteria.

Technological Characteristics

Percutaneous intravascular catheter. Materials: Polyether outer layer, stainless steel braid/coil, PTFE/polyether liner, platinum/iridium marker, nylon hub, Pebax strain relief. Hydrophilic coating. Dimensions: 150-160 cm length, 0.0205 in min ID. Steam-shapeable distal tip. Sterilization: Ethylene oxide. Single-use. No software/electronics.

Indications for Use

Indicated for patients undergoing vascular procedures requiring delivery of therapeutic devices to peripheral, coronary, or neurovasculature. No known contraindications.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 20, 2019 Cerus Endovascular, Inc. Theresa Brandner Vice President of Regulatory 47757 Fremont Boulevard Fremont, California 94538 Re: K182487 Trade/Device Name: CerusEndo Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: June 11, 2019 Received: June 19, 2019 Dear Theresa Brandner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Xiaolin Zheng, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182487 Device Name CerusEndo Microcatheter ### Indications for Use (Describe) The CerusEndo Microcatheter is intended for general intravascular use to deliver therapeutic devices to the peripheral, coronary, and neurovasculature. Type of Use (Select one or both, as applicable) | <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> | |-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------| | <input checked="" type="checkbox"/> | <input type="checkbox"/> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K182487 This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 8807.92. #### I. SUBMITTER Submitter Name: Cerus Endovascular, Inc. Address: 47757 Fremont Boulevard, Fremont, CA 94538 Phone Number: (510) 651-4000 Fax Number: (510) 405-8356 Contact Person: Theresa Brandner VP of Regulatory Date Prepared: July 17, 2019 #### II. DEVICE Name of Device: CerusEndo Microcatheter Common Name: Catheter, Percutaneous Classification Name: Percutaneous Catheter 21 CFR 870.1250 Regulatory Class: Class II Product Code: DQY #### III. PREDICATE DEVICE K093160: Headway 21 Microcatheter manufactured by MicroVention Inc. The predicate device has not been subject to a design-related recall. #### IV. DEVICE DESCRIPTION The subject device is a microcatheter available in three versions: a 150 cm length microcatheter with a 15 cm length distal segment; a 150 cm length microcatheter with a 34 cm length distal segment; and a 160 cm length microcatheter with a 42 cm length {4}------------------------------------------------ distal segment. The distal segment of the catheter is flexible to facilitate access into tortuous anatomy, and the distal tip of the catheter is formable, allowing the physician to shape it according to the needs of the procedure at the point of use. The CerusEndo Microcatheter has a hydrophilic coating, radiopaque marker, Luer hub on the proximal end, polymer tapered shaft construction, stainless steel reinforced shaft, and Teflon lined inner lumen. The subject device is controlled by user manipulation to access discrete locations within the vascular anatomy. It is intended to deliver other interventional or therapeutic devices through its inner lumen. It is designed to be used in peripheral, coronary, and neurovascular locations. #### V. INDICATIONS FOR USE The CerusEndo Microcatheter is intended for general intravascular use to deliver therapeutic devices to the peripheral, coronary, and neurovasculature. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The subject device has similar design, dimensions, materials, intended use, and technological characteristics to the legally marketed predicate cleared under K093160. Non-clinical testing has been performed to demonstrate that any differences in technological characteristics do not raise new questions of safety and effectiveness. A table comparing the intended use and technical characteristics of the proposed device and the legally marketed predicate is provided in Table 1. | | Subject Device | Predicate Device | | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------| | Manufacturer | CerusEndo Microcatheter<br>(Cerus Endovascular) | Headway Microcatheter<br>(MicroVention) | Comparison | | Clinical Attributes | | | | | Indications for Use | The CerusEndo Microcatheter is<br>intended for general intravascular<br>use to deliver therapeutic devices to<br>the peripheral, coronary, and<br>neurovasculature. | The Headway Microcatheter is<br>intended for general intravascular<br>use, including the peripheral,<br>coronary and neuro vasculature for<br>the infusion of diagnostic agents,<br>such as contrast media, and<br>therapeutic agents, such as<br>occlusion coils. | Same | | Environments of Use | Hospital interventional<br>neuroradiology suites | Hospital interventional<br>neuroradiology suites | Identical | | Patient Population | Patients undergoing vascular<br>procedures | Patients undergoing vascular<br>procedures | Identical | | Contraindications | No known contraindications | No known contraindications | Identical | | Functions | To facilitate introduction of<br>therapeutic devices into the<br>vasculature | To facilitate introduction of<br>diagnostic and therapeutic<br>devices into the vasculature | Same | | | Subject Device | Predicate Device | | | Manufacturer | CerusEndo Microcatheter | Headway Microcatheter | Comparison | | | (Cerus Endovascular) | (MicroVention) | | | Patient Access | Device access is gained using an | Device access is gained using an | Identical | | | introducer sheath inserted into | introducer sheath inserted into | | | | the vasculature and is advanced | the vasculature and is advanced | | | | over a guidewire and through a | over a guidewire and through a | | | | guide catheter | guide catheter | | | Intraoperative | Yes | Yes | Identical | | Use | | | | | Technological Attributes | | | | | General | Percutaneous Catheter | Percutaneous Catheter | Same | | Description | Intravascular Catheter | Diagnostic Intravascular Catheter | | | Device | Proximal Luer hub, hydrophilic | Proximal Luer hub, hydrophilic | Same | | Configuration | coating, radiopaque marker, | coating, radiopaque marker, | | | | polymer tapered shaft | polymer tapered shaft | | | | construction, stainless steel | construction, stainless steel | | | | reinforced shaft, and lined inner | reinforced shaft, and lined inner | | | | lumen | lumen | | | Catheter Body | Polyether outer layer, stainless steel | Polyurethane outer layer, | Similar | | Materials | polyether inner layer, braid/coil, | polyether/polyamide inner layer, | | | | PTFE/polyether liner | stainless steel braid/coil,<br>PTFE/polyolefin liner | | | Marker | Platinum/iridium | Platinum/iridium | Same | | Hub | Nylon | Nylon | Same | | Strain Relief | Pebax | Pebax | Same | | Introducer | Not applicable | Pebax | The CerusEndo | | | | | Microcatheter is not | | | | | provided with an | | | | | introducer | | Shaping Mandrel<br>Proximal ID/OD | Stainless Steel<br>0.0205 in min / 0.036 in | Stainless steel<br>0.021 in min / 0.033 in (2.5 Fr) | Same<br>Similar | | | | | | | Distal ID/OD | 0.0205 in min / 0.031 in | 0.021 in min / 0.026 in (2.0 Fr) | Similar | | | | | | | Effective Length | 150 cm, 160 cm | 150 cm, 156 cm | Similar | | Coating | Hydrophilic coating | Hydrophilic coating | Similar | | Tip Configuration | Straight - Steam Shapeable by | Straight - Steam Shapeable by | Same | | Guidewire | physician prior to use<br>≤ 0.018 in | physician prior to use<br>≤ 0.018 in | Same | | Compatibility | | | | | Accessories | Shaping mandrel | Shaping mandrel, introducer | The CerusEndo | | | | | Microcatheter is | | | | | provided with a | | | | | shaping mandrel but is | | | | | not provided with an | | | | | introducer | | Method of Supply | Sterile, single use | Sterile, single use | Same | | Sterilization | Ethylene oxide,<br>single patient use | Ethylene oxide,<br>single patient use | Same | | Biocompatibility | Biocompatible per ISO 10993-1 | Biocompatible per ISO 10993-1 | Same | Table 1 - Comparison Table of Proposed Device and Predicate Device {5}------------------------------------------------ {6}------------------------------------------------ There are no significant technological differences between the subject device and predicate device. The technological differences between the subject and predicate device are indicated in the table above. #### VII. PERFORMANCE DATA Non-clinical tests were performed to demonstrate safety and substantial equivalence. Bench testing performed is summarized in Table 2. Biocompatibility testing performed is summarized in Table 3. | Test | Test Method Summary | Results | |-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Tensile Strength | This test measures the tensile strength<br>of the catheter bonds using a tensile<br>tester and are pulled to failure. | Device met acceptance<br>criteria | | Shaft Flexibility<br>(stiffness) | This test measures the bending stiffness<br>of the distal and proximal catheter shaft<br>segments. | Device met acceptance<br>criteria | | Shape Retention | This test measures the ability of the<br>catheter tip to form and retain a steam<br>shape using conventional catheter lab<br>shaping techniques. | Device met acceptance<br>criteria | | Kink Resistance | This test measures the distal and<br>proximal catheter shaft resistance to<br>kinking. | Device met acceptance<br>criteria | | Static Burst | This test measures the catheters<br>resistance to burst failure by using a<br>high pressure injector to pressurize the<br>lumen while the distal tip is occluded. | Device met acceptance<br>criteria | | Simulated Use | The device was used in accordance to<br>the Instructions for Use. | Device met acceptance<br>criteria | | Particulate | This test assesses the coating integrity<br>by measuring the quantity and size of<br>particles generated during simulated<br>use of the device in an anatomical<br>model. | Device met acceptance<br>criteria | | Coating Friction and<br>Durability | This test measures the lubricity of the<br>coating and the durability after repeated<br>abrasion cycles. | Device met acceptance<br>criteria | | Sterilization<br>Validation | These tests confirms a minimum 6 log<br>sterility assurance level, assesses<br>ethylene oxide sterilant residual levels,<br>and detects pyrogens | Device met acceptance<br>criteria | | Test | Extract/Test System | Results | | Cytotoxicity (MEM<br>Elution) | The test extract, a positive control, and<br>a negative control were extracted at<br>37°C for 24 hours in MEM solution<br>(5% serum supplemented cell culture<br>medium) and exposed to mouse<br>fibroblast cells. | Non-cytotoxic.<br>The test article is<br>considered non-<br>cytotoxic<br>to cells. | | Sensitization<br>(ISO Kligman<br>Guinea Pig<br>Maximization Test) | Test samples and control blanks were<br>extracted in normal saline and in cotton<br>seed oil at 37°C for 72 hours. | Non-sensitizing.<br>The test article did not<br>elicit a sensitization<br>response. | | Irritation/<br>ISO Intracutaneous<br>Toxicity in<br>Rabbits | Test samples were extracted in normal<br>saline and in cotton seed oil at 37°C for<br>72 hours. | Non-irritant.<br>No evidence of<br>irritation. | | Systemic Toxicity<br>(ISO Systemic<br>Injection) | Test samples and negative control<br>blanks were extracted and were<br>prepared in normal saline and in cotton<br>seed oil at 37°C for 72 hours. | Non-cytotoxic.<br>No weight loss,<br>mortality, or evidence of<br>systemic toxicity from<br>the extract exposure to<br>the mice. | | Systemic Toxicity<br>(Rabbit Material-<br>Mediated<br>Pyrogenicity) | Test samples were extracted and<br>negative control blanks were prepared<br>in normal saline and in cotton seed oil<br>at 37°C for 72 hours. | Non-pyrogenic.<br>All individual rabbits for<br>both the test article and<br>negative control showed<br>a total rise in<br>temperature of < 0.5°C<br>and were determined to<br>be nonpyrogenic. | | Hemocompatibility<br>(ISO In Vitro<br>Hemocompatibility<br>- Direct Contact) | Blood samples from three human<br>donors were pooled and diluted. The<br>test article is added to aliquots of<br>human blood and incubated at 37 °C<br>for a minimum of 3 hours. | Non-hemolytic. There<br>were no differences<br>between the hemolytic<br>index of the test article<br>and the negative control. | | Hemocompatibility<br>(Hemolysis -<br>Indirect) | Test samples were extracted in<br>phosphate buffered saline at 37°C for<br>72 hours. The test article extract was<br>incubated at 37°C for a minimum of 3<br>hours. | Non-hemolytic.<br>There were no<br>significant differences<br>between the test article<br>extract/solid and<br>negative control article<br>results. | | Hemocompatibility<br>(ISO Complement<br>Activation C3 and<br>SC5b-9 Test –<br>Direct Contact) | The test article, predicate, negative<br>control, and positive control (latex)<br>were added to the serum pooled from<br>three human blood samples. All were<br>incubated at 37 °C for 30, 60, and 90<br>minutes. | C3a and SC5b-9<br>complement proteins<br>were considered to be<br>non-activated as<br>compared to the<br>negative control. | | Genotoxicity – Gene<br>Mutation<br>(Ames Assay,<br>S-9 Activation) | Test samples and negative control<br>blanks were extracted in normal saline<br>and PEG 400 at 37°C for 72 hours. | Non-mutagenic.<br>Based on the acceptance<br>criteria under the<br>experimental conditions<br>utilized, the test article<br>extracts were both<br>deemed non-mutagenic<br>in all strains under both<br>non-activated and<br>activated conditions. | | Genotoxicity –<br>Micronucleus | Test samples and negative control<br>blanks were extracted in polar (saline)<br>and non-polar solvents at 37 °C for 72<br>hours. The extracts are placed on<br>Chinese Hamster Ovary (CHO) cells<br>with and without metabolic activations<br>(S-9). The cells are then incubated at<br>37 °C, with 5% CO2, for 15 – 21 hours.<br>The cells are harvested and the<br>chromosomes stained and examined for<br>aberrations. | Not genotoxic.<br>No significant increase<br>of aberrations when<br>compared to the<br>negative controls | | Hemocompatibility<br>(Thrombogenicity) | Devices were placed in a canine carotid<br>vessel. | Non-thrombogenic. No<br>significant thrombus<br>was observed on any of<br>the subject catheters,<br>and the device was<br>determined to not show<br>thrombogenic potential | ## Table 2 - Bench Testing Summary {7}------------------------------------------------ # Table 3 – Biocompatibility Testing Summary {8}------------------------------------------------ ## VIII. CONCLUSION Differences between the subject device and the predicate device do not raise different questions of safety and effectiveness. After a comparison of the subject CerusEndo Microcatheter intended use, technological characteristics, and expected performance to the legally marketed predicate, the Headway 21 Microcatheter (K093160), Cerus Endovascular concludes that the CerusEndo Microcatheter performs as safely and effectively as the predicate device for the same intended use.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%