K Number
K182429

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-06-07

(274 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors in patients in healthcare environments. The O3 Regional Oximeter is only to be used with Masimo O3 sensors. The use of any other sensor is not supported or recommended by Masimo and could give erroneous results.

When used with the O3 Adult Sensor, the O3 Regional Oximeter is indicated for measuring absolute and trending regional hemoglobin oxygen saturation of blood (rSO2) in adults ≥ 40 kg.

When used with the O3 Pediatric Sensor, the O3 Regional Oximeter is indicated for measuring absolute and trending regional hemoglobin oxygen saturation of blood (rSO2) in pediatrics ≥ 5 kg and < 40 kg.

When used with the O3 Neonatal Sensor, the O3 Regional Oximeter is indicated for measuring only trending regional hemoglobin oxygen saturation of blood (rSO2) in neonates < 10 kg.

Device Description

The Masimo Regional Oximetry System monitors regional hemoglobin oxygen saturation of blood (rSO2) under the sensors. The O3 System includes the O3 Sensors that acquire physiological signals and the O3 Module that processes those signals. The FDA has previously cleared the O3 System in K160526 (with an O3 Adult Sensor) and K162603 (with an O3 Pediatric Sensor). In this submission, Masimo seeks clearance of its O3 System with an 03 Neonatal Sensor.

Similar to the cleared O3 Adult and Pediatric Sensors, the O3 Neonatal Sensor is a singlepatient use, adhesive sensor and is supplied non-sterile. The O3 Neonatal Sensor attaches to the patient's forehead. The sensor includes four emitters and two detectors. The emitters radiate multiple wavelengths of near infrared light, while the detectors sense the reflected light. The detector outputs are physiological signals and these signals pass through the other end of the sensor that connects to a patient cable, passing these signals to the O3 Module for processing.

The O3 Module is unchanged from K160526. It includes Masimo technology for processing those signals and outputting regional oximetry (tSO2) measurements. Specifically, the O3 Module includes Near Infra Red Spectroscopy (NIRS) technology. When O3 module is connected to an O3 Neonatal Sensor, the O3 Monitor continuously and accurately determines the trending measurement of regional blood oxygen saturation in the tissue (rSO2) in neonates. In turn, the Host/Backboard device displays this measurement. The O3 Module can connect to up to two O3 Sensors, both connected to a patient.

The O3 System does not have an internal battery or an AC power input. The O3 Module, instead, receives power via its connection to a Host/Backboard Device, such as the Root Monitoring System (Root). Root in turn receives power from either AC power or internal rechargeable batteries.

Similar to K160526, the O3 System using an O3 Neonatal Sensor provides the following key measurements:

  • Regional Oxygenation (rSO2): Regional tissue oxygenation level in the deep tissue local to the sensor site, including cerebral tissue
  • Delta Baseline (Abase): Relative difference in rSO2 with respect to baseline rSO2
  • Area Under the Limit (AUL index): Index that quantifies the duration (amount of time the patient stays below rSO2 low alarm limit) and depth (refers to the gap between the patient's rSO2 level and the rSO2 low alarm limit) of patient's stay below the userdefined rSO2 low alarm limit (LAL)
  • . Delta SpO2 (4SpO2): The difference between SpO2 and rSO2. The source of SpO2 is from peripheral SpO2 measurement (using pulse oximeter).
AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Masimo O3 Regional Oximeter System, specifically focusing on the O3 Neonatal Sensor, based on the provided text:

Acceptance Criteria and Device Performance for Masimo O3 Regional Oximeter System (O3 Neonatal Sensor)

1. Table of Acceptance Criteria and Reported Device Performance:

FeatureAcceptance Criteria (Specification)Reported Device Performance (O3 Neonatal Sensor)
rSO2 Display Range0-99%0-99%
rSO2 Display Resolution1%1%
Δbase Display Range0-99%0-99%
Δbase Display Resolution1%1%
Trending ARMS (Neonates < 10 kg)3% for SavO2 of 45%-85%Not explicitly stated as meeting a specific acceptance criterion in the provided text. The text states "Trending ARMS, 3% for SavO2 of 45%-85%" under specifications but then describes how it was calculated in the clinical study, implying this is the desired performance. The study aimed to determine this.
Absolute Accuracy (Convenience Sample)N/A (not a claimed feature for this submission)Less than 6% (favorable trend of rSO2 with SavO2)

Notes on the "Reported Device Performance":

  • For the display ranges and resolutions, the "Reported Device Performance" can be inferred to match the "Acceptance Criteria (Specification)" as these are direct specifications of the device.
  • For Trending ARMS, the document states the specification is "3% for SavO2 of 45%-85%" and explains the methodology to determine this for the Neonatal sensor, but doesn't explicitly state "the study demonstrated the device met the 3% Trending ARMS." However, it's listed as a specification, so the implication is that the outcome of the calculation supports this specification.
  • Absolute accuracy was not a claim but reported as a "convenience sample," indicating it's not a formal acceptance criterion for this submission's claims.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size (Clinical Study for Trending ARMS): The text indicates a combined approach:
    • One clinical study for the O3 Adult Sensor to establish its trending accuracy with respect to blood testing (sample size not specified).
    • A separate study where relative rSO2 data was collected for both the O3 Adult Sensor and O3 Neonatal Sensor on the same subject to calculate the O3 Neonatal Sensor's Trending ARMS relative to the O3 Adult Sensor (sample size for this specific part is not provided).
  • Sample Size (Clinical Study for Absolute Accuracy - convenience sample): 11 hospitalized patients.
  • Data Provenance: The document does not specify the country of origin. The studies appear to be prospective clinical studies conducted by Masimo to support the device's performance claims.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • Not applicable / Not specified. The ground truth for this device's performance is established through direct physiological measurements (blood testing/SavO2), not through expert interpretation of images or other data that would require multiple human experts for ground truth establishment.

4. Adjudication Method for the Test Set:

  • None. As the ground truth is established via quantitative physiological measurements (SavO2 from blood draws), there is no need for expert adjudication methods in this context.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, Effect Size:

  • No. An MRMC comparative effectiveness study was not performed. This device is a diagnostic/monitoring tool that directly measures a physiological parameter, not an AI-assisted diagnostic imaging tool where human reader performance would be a relevant metric for comparison.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done:

  • Yes. The performance of the O3 Regional Oximeter System (including the O3 Neonatal Sensor) is evaluated as a standalone device. Its "Trending ARMS" and "Absolute Accuracy" are measured against physiological ground truth (SavO2 from blood). There is no "human-in-the-loop" component in its core functional performance evaluation described here, as it directly outputs rSO2 measurements.

7. The Type of Ground Truth Used:

  • Physiological Ground Truth:
    • For Trending ARMS: Blood testing (specifically SavO2, which is defined as "0.3 SaO2 + 0.7 SjvO2"). This is considered the gold standard for blood oxygen saturation. The Neonatal Sensor's performance was also compared relative to the O3 Adult Sensor, which was previously validated against blood testing.
    • For the absolute accuracy convenience sample: Invasive SavO2 blood draws.

8. The Sample Size for the Training Set:

  • Not specified. The document describes clinical studies performed for performance validation (test set), but does not provide information about any training sets used for the development or calibration of the device's algorithms. As an oximetry device, its core algorithms are likely based on established biophotonics principles and may not involve a separate "training set" in the same way machine learning algorithms do, but rather extensive calibration and verification data.

9. How the Ground Truth for the Training Set was Established:

  • Not specified. Since details about a specific "training set" are not provided, the method for establishing its ground truth is also not mentioned. It is generally understood that the algorithms for such devices are developed and calibrated using a controlled experimental setup with known ground truth parameters for oxygen saturation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Masimo Corporation Sindura Penubarthi Regulatory Affairs Manager 52 Discovery Irvine, California 92618

Re: K182429

Trade/Device Name: Masimo O3 Regional Oximeter System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: April 30, 2019 Received: May 1, 2019

Dear Sindura Penubarthi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182429

Device Name Masimo O3 Regional Oximeter System

Indications for Use (Describe)

The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood (tSO2) in the cerebral region under the sensors in healthcare environments. The O3 Regional Oximeter is only to be used with Masimo O3 sensors. The use of any other sensor is not supported or recommended by Masimo and could give erroneous results.

When used with the O3 Adult Sensor, the O3 Regional Oximeter is indicated for measuring absolute and trending regional hemoglobin oxygen saturation of blood (rSO2) in adults ≥ 40 kg.

When used with the 03 Pediatric Sensor, the 03 Regional Oximeter is indicated for measuring absolute and trending regional hemoglobin oxygen saturation of blood (rSO2) in pediatrics ≥ 5 kg and < 40 kg.

When used with the 03 Neonatal Sensor, the 03 Regional Oximeter is indicated for measuring only trending regional hemoglobin oxygen saturation of blood (rSO2) in neonates < 10 kg.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a stylized font, with a red checkmark symbol to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

Submitter and Address ofManufacturing Facility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7541FAX: (949) 297-7592
Date:September 4, 2018
Contact:Sindura PenubarthiRegulatory Affairs Manager
Trade Name:Masimo O3 Regional Oximeter System
Common Name:Oximeter, Tissue Saturation
Classification Regulation/Product Code:21 CFR 870.2700, Class II/MUD
Establishment RegistrationNumber:2031172
Reason for PremarketNotification:New Device - Masimo O3 Neonatal Sensor
Predicate Device:K160526 – Masimo O3 Regional Oximeter SystemReference Predicate: K082327- Somanetics Invos 5100Cwith Infant/Neonatal Cerebral OxyAlert NIRSensor
Performance StandardsThere are no performance standards pursuant to Section 514of the Food, Drug and Cosmetic Act for the above device.

Device Description

The Masimo Regional Oximetry System) monitors regional hemoglobin oxygen saturation of blood (rSO2) under the sensors. The O3 System includes the O3 Sensors that acquire physiological signals and the O3 Module that processes those signals. The FDA has previously cleared the O3 System in K160526 (with an O3 Adult Sensor) and K162603 (with an O3 Pediatric Sensor). In this submission, Masimo seeks clearance of its O3 System with an 03 Neonatal Sensor.

Similar to the cleared O3 Adult and Pediatric Sensors, the O3 Neonatal Sensor is a singlepatient use, adhesive sensor and is supplied non-sterile. The O3 Neonatal Sensor attaches to the patient's forehead. The sensor includes four emitters and two detectors. The emitters radiate multiple wavelengths of near infrared light, while the detectors sense the reflected light. The detector outputs are physiological signals and these signals pass through the other

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a stylized red checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" in a smaller font, followed by the address "52 Discovery, Irvine, CA 92618".

end of the sensor that connects to a patient cable, passing these signals to the O3 Module for processing.

The O3 Module is unchanged from K160526. It includes Masimo technology for processing those signals and outputting regional oximetry (tSO2) measurements. Specifically, the O3 Module includes Near Infra Red Spectroscopy (NIRS) technology. When O3 module is connected to an O3 Neonatal Sensor, the O3 Monitor continuously and accurately determines the trending measurement of regional blood oxygen saturation in the tissue (rSO2) in neonates. In turn, the Host/Backboard device displays this measurement. The O3 Module can connect to up to two O3 Sensors, both connected to a patient.

The O3 System does not have an internal battery or an AC power input. The O3 Module, instead, receives power via its connection to a Host/Backboard Device, such as the Root Monitoring System (Root). Root in turn receives power from either AC power or internal rechargeable batteries.

Similar to K160526, the O3 System using an O3 Neonatal Sensor provides the following key measurements:

  • Regional Oxygenation (rSO2): Regional tissue oxygenation level in the deep tissue ● local to the sensor site, including cerebral tissue
  • Delta Baseline (Abase): Relative difference in rSO2 with respect to baseline rSO2
  • Area Under the Limit (AUL index): Index that quantifies the duration (amount of time ● the patient stays below rSO2 low alarm limit) and depth (refers to the gap between the patient's rSO2 level and the rSO2 low alarm limit) of patient's stay below the userdefined rSO2 low alarm limit (LAL)
  • . Delta SpO2 (4SpO2): The difference between SpO2 and rSO2. The source of SpO2 is from peripheral SpO2 measurement (using pulse oximeter).

The O3 System has the following specifications:

O3 System Specifications
FEATURESPECIFICATION
Display
Display RangeRegional Oxygen Saturation (rSO2): 0-99%
Δbase: 0-99%(Δbase is the difference between regional and baseline rSO2)
Display Resolution for MeasurementsrSO2: 1%
Δbase: 1%
SO2 Measurement Accuracy,Neonates < 10 kgTrending ARMS, 3% for SavO2 of 45%-85%
General

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a stylized font, with a red checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION", followed by the address "52 Discovery, Irvine, CA 92618".

O3 System Specifications
FEATURESPECIFICATION
Visual/audible alarmHost/backboard device (Root per K140188, K151644 and K153225) is IEC60601-1-8 compliant
Storage/recordingHost/backboard device (Root per K140188, K151644 and K153225) trend/data storage
Electrical
AC Power orBattery, RechargeableHost/backboard device (Root per K140188, K151644 and K153225) provides power to O3 System
Interface
03 Module ConnectionMOC-9 interface with host/backboard device (Root per K140188, K151644 and K153225)
Mechanical
03 Module: Dimensions/Weight5x1.8x0.6 inch/ 7 oz
03 Neonatal Sensor: Dimensions/Weight2.6x1.1x0.1 inch/0.18 oz.
Environmental
03 Module
• Operating Temperature0°C to +40°C, ambient humidity
• Storage Temperature-40°C to +70°C, ambient humidity
• Operating/ Storage Humidity10% to 95%, non-condensing
• AltitudeUp to 12,000 feet (3700 meters)
03 Adult Sensor, O3 Pediatric, and O3Neonatal Sensor
• Operating Temperature+5°C to +40°C
• Storage Temperature-40°C to +60°C
• Humidity15% to 90% relative humidity

SavO2 = 0.3 SaO2 + 0.7 SjvO2

Intended Use

The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors in patients in healthcare environments. The O3 Regional Oximeter is only to be used with Masimo O3 sensors. The use of any other sensor is not supported or recommended by Masimo and could give erroneous results.

Indications For Use

The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors in patients in healthcare environments. The O3 Regional Oximeter is only to be used with Masimo O3 sensors. The use of any other sensor is not supported or recommended by Masimo and could give erroneous results.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, stylized font. To the right of the logo, the text "MASIMO CORPORATION 52 Discovery Irvine, CA 92618" is displayed in a smaller, sans-serif font.

When used with the O3 Adult Sensor, the O3 Regional Oximeter is indicated for measuring absolute and trending regional hemoglobin oxygen saturation of blood (rSO2) in adults ≥ 40 kg.

When used with the O3 Pediatric Sensor, the O3 Regional Oximeter is indicated for measuring absolute and trending regional hemoglobin oxygen saturation of blood (rSO2) in pediatrics ≥ 5 kg and < 40 kg.

When used with the O3 Neonatal Sensor, the O3 Regional Oximeter is indicated for measuring only trending regional hemoglobin oxygen saturation of blood (rSO2) in neonates < 10 kg.

Technological Characteristics

Principle of Operation (same as in K160526)

The O3 System's operating principle is based on multi-distance diffuse reflectance spectroscopy. The O3 System use light to examine a cross-section tissue microvasculature (a mixed bed of arterioles, capillaries and venules) and analyzes the light returned after having passed through the tissues. The spectroscopic analysis determines concentrations of hemoglobin in its oxygenated and deoxygenated states.

Mechanism of Action for Achieving the Intended Effect (same as in K160526)

The O3 Sensor is noninvasively applied to the patient on one end. The other end of the O3 Sensor connects to the O3 Module. In turn, the O3 Module connects to a Host/Backboard device. The O3 Sensor collects patient physiological signals which are processed by the O3 Module. The processed signals which resulted in rSO2 measurements are displayed on the Host/Backboard device.

{7}------------------------------------------------

Non-clinical Testing

Masimo performed the following tests, as applicable, for the qualification of the subject devices, O3 System under K160526 in accordance with the requirements of the design control regulations and established quality assurance processes to demonstrate substantial equivalence with the predicate device:

  • Electrical safety testing per IEC-60601-1
  • EMC testing per IEC-60601-1-2 ●
  • Alarm testing per IEC-60601-1-8
  • Biocompatibility testing per ISO-10993
  • Software verification per FDA Software Guidance
  • Mechanical and environmental testing .

Masimo performed the following tests, as applicable, for the qualification of the subject devices, O3 Neonatal sensor in accordance with the requirements of the design control regulations in and established quality assurance processes to demonstrate substantial equivalence with the predicate device cleared in K160526

  • . Usability testing per FDA Human Factors and Usability Guidance
    The results demonstrate that all requirements and performance specifications were satisfied, and that the subject device is substantially equivalent to the predicate device.

Clinical Testing

Masimo performed a clinical study to determine the O3 Neonatal Sensor's Trending ARMS. Previously, Masimo performed a clinical study on the O3 Adult Sensor to establish its trending accuracy with respect to blood testing.

In this submission. Masimo collected relative rSO2 data for both the O3 Adult Sensor and O3 Neonatal Sensor on the same subject to calculate the O3 Neonatal Sensor's Trending Arms relative to the O3 Adult Sensor. Masimo then calculated the Trending ARMs for the O3 Neonatal Sensor relative to blood from a combination of the O3 Adult Sensor's Trending ARMs relative to blood testing and 03 Neonatal Sensor's Trending ARMs relative to the 03 Adult Sensor.

Masimo also conducted a study to test the absolute accuracy of the subject sensor on 11 hospitalized patients weighing less than 10 kg with invasive SavO2 blood draws of 46 % to 82 %. The study showed a favorable trend of rSO2 with SavO2 with an accuracy of less than 6 %. Note that absolute accuracy is not a claim that is being pursued in this submission. These data are merely being reported as convenience samples.

{8}------------------------------------------------

Conclusion

The clinical evaluation, non-clinical testing including safety testing, as included in this 510(k) submission, demonstrate that the subject device, O3 Neonatal Sensor, is substantially equivalent to its predicate with respect to safety and effectiveness.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).