K Number
K182407

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-07-19

(318 days)

Product Code
Regulation Number
878.4430
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Exceed is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I, II, III and IV in adults aged 22 years or older.

Device Description

The Exceed micro needling device is intended to create many, very tiny, microscopic punctures in the epidermal and dermal layers of the skin using sterile stainless-steel needles. The Exceed micro needling device consists of 7 component parts; Control Unit, Sterile single use Safety needle cartridge, Handpiece, Handpiece holder, Handpiece cover, Footswitch, Power supply.

The control unit switches the device on and off and contains the power source (5.5-12 V, DC (150mA)). The control unit adjusts the frequency of the needle stroke from 100-150Hz using a digital display and a keypad. The control unit receives power via a coaxial connector from a standard 100-240V, 50-60Hz, 1.2A wall socket transformer with an output of 15V. The handpiece contains a motor that moves the needles and a needle protrusion gauge that allows the user to control the depth of the needle protrusion from 0 - 1.9mm. The handpiece contains a scale that allows for the needle protrusion depth to be adjusted by the operator. The scale has a tolerance of ±0.15 mm. The standard safety needle cartridge is a 6-stainless steel micro needle plate of 1.5 mm length. The safety needle cartridge is screwed into the handpiece. The safety needle cartridge is sterile and for single use only. This standard safety needle cartridge is used for Microneedling only. The length of needles is 1.5 mm. In combination with the handpiece needle protrusion can be adjusted between 0.0 mm (min) and a maximum of 1.5 mm.

AI/ML Overview

This document describes the safety and effectiveness study for the Exceed Microneedling device.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied through the study's focus on demonstrating improvement in facial acne scars and safety outcomes. While no explicit percentage or threshold for "improvement" is stated as an acceptance criterion in an absolute sense, the study aims to show statistically significant or clinically meaningful improvement, and acceptable safety profiles compatible with similar aesthetic devices.

Acceptance Criterion (Implied)Reported Device Performance (Effectiveness)Reported Device Performance (Safety)
Improvement in appearance of facial acne scars (Physician-assessed)Mean ASAS score reduction from 2.89 at baseline to 2.27 at final follow-up.- 59% of subjects improved by 1 grade (as agreed by 2 of 3 blinded physicians).- 83% of subjects showed some improvement (improvement > 0 but < 1).N/A
Improvement in appearance of facial acne scars (Subject-assessed - SASIS)88% of subjects reported some improvement (1-99% improvement).- 41.9% reported 1-25% improvement.- 34.9% reported 25-50% improvement.- 9.3% reported 50-75% improvement.- 2.3% reported 75-99% improvement.N/A
Improvement in appearance of facial acne scars (Subject-assessed - SGAIS)77% of subjects reported some improvement.- 9.3% reported 'Much Improved'.- 67.4% reported 'Improved'.N/A
Subject satisfaction with treatment86% of subjects noticed improvement in how their acne scars looked.81.5% were satisfied or extremely satisfied (16.3% extremely, 41.9% satisfied, 23.3% slightly satisfied).88% would recommend the treatment to friends and family.N/A
Acceptable safety profile (Adverse Events)N/ANo serious adverse events reported.No adverse events persisted at final follow-up.26 adverse events in 18 subjects (32%).- 88.5% mild, 11.5% moderate.- Most common: Bruising (6 events), Swollen lymph-nodes (6 events).
Acceptable safety profile (Pain and Discomfort)N/AMean pain score: 2.8 on treatment evening, <1 by day 3, 0.05 by day 7.Mean discomfort score: 3.20 on day 1, <1 by day 4, 0.12 by day 7.
Acceptable safety profile (Redness)N/A49% moderately red, remainder mild or less on treatment evening.94% minimal or absent redness by day 5.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 56 subjects were recruited for the clinical study. 47 subjects (83.9%) completed all treatment and follow-up visits.
  • Data Provenance: The document does not explicitly state the country of origin. It indicates "a single center" for the study. Since the applicant is "MT.DERM GmbH" from Germany, it is plausible the study was conducted there, but not definitively stated. The study was prospective, clinical, and interventional.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: 3 physicians.
  • Qualifications of Experts: The document states they were "3 blinded physicians" but does not provide specific qualifications (e.g., years of experience, specialty beyond 'physician'). They used a validated 5-point photo numeric descriptive grading scale (Acne Scar Assessment Scale - ASAS) to grade acne scarring.

4. Adjudication Method for the Test Set

  • The digital images were "randomized and analyzed independently by 3 physicians."
  • For the "Physician measurement of effectiveness," one of the reported metrics is "Subjects improved 1 grade as agreed upon by 2 of 3 blinded physicians." This indicates a 2+1 majority consensus method for this specific outcome.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This study is for a microneedling device used for aesthetic purposes, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No, a standalone (algorithm only) performance study was not done. This device is a physical microneedling device for aesthetic treatment, not a software algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

  • The ground truth for effectiveness was established through:
    • Expert Consensus: Independent assessment by 3 blinded physicians using the validated Acne Scar Assessment Scale (ASAS) based on digital photographs. A 2-out-of-3 consensus was used for some metrics.
    • Subject-Reported Outcomes: Self-assessed Scar Improvement Scale (SASIS), Subject Global Aesthetic Improvement Scale (SGAIS), and direct satisfaction questions.
  • The ground truth for safety was established through:
    • Physician-recorded adverse events.
    • Subject-reported pain, discomfort, and redness via daily diaries and visual analogue/descriptive scales.

8. The Sample Size for the Training Set

  • There is no mention of a separate "training set" in the context of this microneedling device study, as it is a physical treatment device, not an AI model requiring a training dataset. The 56 subjects represent the clinical study (test) population.

9. How the Ground Truth for the Training Set was Established

  • Not applicable, as there was no explicit "training set" for an AI model. The mentioned clinical study focused on evaluating the device's performance in a prospective manner on its target population.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 19, 2019

Mt. Derm GmbH % Susan D'Arcy Director Ismart Marketing Svcs 129 Green Lanes, Wylde Green Birmingham, B735LT Gb

Re: K182407

Trade/Device Name: Exceed Microneedling device Regulation Number: 21 CFR 878.4430 Regulation Name: Microneedling Device For Aesthetic Use Regulatory Class: Class II Product Code: QAI Dated: July 7, 2019 Received: July 9, 2019

Dear Susan D'Arcy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cynthia Chang, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182407

Device Name Exceed Microneedling device

Indications for Use (Describe)

The Exceed is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I, II, III and IV in adults aged 22 years or older.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).

Date Prepared: August 22nd 2018

Date amended: July 01th 2019

Submitter's Name: MT.DERM GmbH

Submitter's Address: Gustav-Krone-Str. 3. 14167 Berlin, Germany

Submitter's telephone number: +49 30 845 885 171

Contact Person: Andreas Pachten, Director QM and Regulatory Affairs

Device Trade Name: Exceed Microneedling device

1. Device Classification Information:

RegulationNumberDeviceClassificationnameDeviceClassProductCodeGeneric descriptionClassificationPanelType
21 CFR878.4430Microneedlingdevice foraesthetic useClass 2QAIA microneedlingdevice for aestheticuse is a device usingone or more needles tomechanically punctureand injure skin tissuefor aesthetic use. Thisclassification does notinclude devicesintended fortransdermal deliveryof topical productssuch as cosmeticsdrugs, or biologicsGeneral & PlasticSurgeryTraditional510 (k)

2. Device Description

The Exceed micro needling device is intended to create many, very tiny, microscopic punctures in the epidermal and dermal layers of the skin using sterile stainless-steel needles. The Exceed micro needling device consists of 7 component parts; Control Unit, Sterile single use Safety needle cartridge, Handpiece, Handpiece holder, Handpiece cover, Footswitch, Power supply.

The control unit switches the device on and off and contains the power source (5.5-12 V, DC (150mA)). The control unit adjusts the frequency of the needle stroke from 100-150Hz using a digital display and a keypad. The control unit receives power via a coaxial connector from a standard 100-240V, 50-60Hz, 1.2A wall socket transformer with an output of 15V. The handpiece contains a motor that moves the needles and a needle protrusion gauge that allows the user to control the depth of the needle protrusion from 0 - 1.9mm. The handpiece contains a scale that allows for the needle protrusion depth to be adjusted by the operator. The scale has a tolerance of ±0.15 mm. The standard safety needle cartridge is

{4}------------------------------------------------

a 6-stainless steel micro needle plate of 1.5 mm length. The safety needle cartridge is screwed into the handpiece. The safety needle cartridge is sterile and for single use only. This standard safety needle cartridge is used for Microneedling only. The length of needles is 1.5 mm. In combination with the handpiece needle protrusion can be adjusted between 0.0 mm (min) and a maximum of 1.5 mm .

Indications/Intended Use 3.

The Exceed is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I, III and IV in adults aged 22 years or older."

The device is a prescription use device and complies with 21CFR 801.109.

Predicate Device 4.

DEN160029 SkinPen Precision Microneedling system

SkinPen® Precision System DEN160029 is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older.

The Exceed Micro needling device is predicated against the SkinPen Precision microneedling system because both systems are microneedling devices containing one or more needles to mechanically puncture and injure the skin tissue for aesthetic use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "MT.DERM" in white letters against a red background. The letters are sans-serif and appear to be slightly blurred, giving them a soft, glowing effect. The text is centered in the image and takes up most of the space.

5. Technological Characteristics

PropertyExceed MicroneedlingdeviceDEN160029SkinPen PrecisionMicroneedling systemSignificant differences
DeviceManufacturerMT.DERM GmbHGustav-Krone-Str. 3,14167 Berlin,GermanyBellus Medical, LLC, 4505 ExcelParkway, Suite 100, Addison, TX75001Not applicable
Device TradeNameExceed Microneedling deviceSkinPen Precision SystemNot applicable
510(K) NumberK182407DEN160029Not applicable
DeviceClassificationnameMicroneedling device foraesthetic useMicroneedling device foraesthetic useIdentical
Device ProductCodeQAIQAIIdentical
DeviceClassificationClass IIClass IIIdentical
Regulationnumber21 CFR 878.443021 CFR 878.4430Identical
UsePrescription usePrescription useIdentical
Intended Locationof UseFaceFaceIdentical
Intended use andIndicationsThe Exceed is a microneedlingdevice and accessories intendedto be used as a treatment toimprove the appearance offacial acne scars in Fitzpatrickskin types I, II, III and IV inadults aged 22 years or older."SkinPen® Precision System is amicroneedling device andaccessories intended to be used asa treatment to improve theappearance of facial acne scars inadults aged 22 years or older.Substantially Equivalent
PropertyExceed MicroneedlingdeviceDEN160029SkinPen PrecisionMicroneedling systemSignificant differences
Geometry6 needles, squared arrangement14 needles, radial arrangementDissimiliar. Clinical and Non-clinicalperformance testing demonstrates that theExceed microneedling device does notpose any undue or additional risks.
Needle protrusionsetting0 - 1.9 mm0 - 2.5 mmDissimilar in general, but maximumneedle penetration within the face forboth devices is 1.5mm. Clinical and Non-clinical performance testing demonstratesthat the Exceed microneedling devicedoes not pose any undue or additionalrisks.
Maximum needlepenetration(maximum needlelength)1.5mm1.5 mmSubstantially equivalent. Both devicesuse a maximum needle penetration lengthof 1.5mm for the indication of treatingacne scars (Skinpen references acapability of extending from 1.5mm to amaximum of 2.5mm however this has notbeen demonstrated to be safe andeffective for this indication[Manufacturers claim1].
Frequency100-150 Hz (±10%)6300RPM to 7700 RPM1(105 - 128.3 Hz)Dissimilar. Clinical and Non-clinicalperformance testing demonstrates that theExceed microneedling device does notpose any undue or additional risks.

{6}------------------------------------------------

1 Data obtained from marketing material and user guide

6. Substantial Equivalency and Comparison of Technological Similarities & Differences

6.1. Key Similarities.

  • i. The device classification (generic description) and basic technologies are equivalent in that both devices are micro needling systems containing >1 needle that mechanically punctures or injures the skin for aesthetic use.
  • ii. Identical intended use.
  • iii. Intended for prescription use.
  • iv. Location of use (Face)
  • The depth of penetration of the needles (needle length) is equivalent 1.5mm v.

6.2. Differences.

Although the systems share the basic generic description and technologies they do differ in several areas.

  • Geometry and needle count a.
  • b. Needle protrusion setting

{7}------------------------------------------------

c. Frequency

These differences have been addressed by the manufacturer through the applicable safety standards (General controls and mitigation measures) and through clinical and non-clinical performance testing (Special controls).

  • Geometry and needle count the difference in needle count and geometry has been a. addressed in terms of safety through specific non-clinical performance testing demonstrating that the needle geometry of the proposed device is uniform and clinical performance testing to demonstrate substantial equivalence. The difference in geometry and needle count does not raise any new questions in relation to safety or effectiveness compared to the predicate device.
  • Needle protrusion setting the needle protrusion setting of the proposed device is within b. the range of the predicate device. The predicate device used a maximum of 1.5mm in its reported clinical study and Instructions for Use and therefore is comparable to the proposed device that does not exceed 1.5mm. Non- Clinical performance testing demonstrates the needle protrusion setting of the proposed device is reproduceable within the technical specifications of the device. Clinical performance testing demonstrates safety and effectiveness in use.
  • Frequency the frequency range (oscillation) of the Exceed microneedling device is c. greater than the predicate device. In terms of performance testing the upper range of 150 Hz was considered to be worst case scenario and therefore relevant non-clinical performance testing has been carried out at this maximum setting. Results from these range of tests provides evidence that this difference does not raise any new concerns regards safe use of the device in its intended use. Clinical performance testing demonstrates safety and effectiveness in use.

Although there are differences between the proposed device and predicate device DEN160029 SkinPen Precision System where there are differences, General controls of the FD&C Act and "Special controls" including non-clinical and clinical performance testing demonstrate that the Exceed microneedling device raises no new questions in relation to safety and effectiveness compared to the predicate device.

{8}------------------------------------------------

MT.DERM

7. General controls and mitigation measures

To demonstrate safe and effective performance and support substantial equivalence the eystem has undergone a number of non-clinical performance tests in line with rems of general requirements, biocompatibility, electrical safety and software. The following non-clinical performance data is provided in support of the substantial equivalence determination.

7.1. Summary of Risk and mitigation measures

Identified Risk to HealthMitigation MeasuresEvidence
Adverse tissue reactionBiocompatibility evaluationISO 10993 – 5,2009 10993-7:2008, 10993-10:2010 and ISO 10993-11:2017(E) and USP 40<151> Pyrogen Test, 2017 and 10993-12:2012FDA - Blue Book Memorandum #G95-1
LabelingISO 15223-1:2012EN 1041:2008IEC 82079-1:2012IEC62366:2007
Sterilization validationISO11135-1:2007ISO/TS11135-2:2008ISO11737-1:2006
Cross contamination andinfectionReprocessing validationA cleaning validation was performed for reusable components of the device. (ReprocessingMedical Devices in Health Care Settings: Validation Methods and Labeling)
Non-clinical performance testingNon-clinical performance data demonstrates that the device performs as intended underanticipated conditions of use.The following performance characteristics are testedi. Accuracy of needle penetration depth and puncture rate in pig skin with high -speed camera measurement;ii. Safety features built into the device to protect against cross-contamination,including fluid ingress protection due to a safety membrane; andiii. Identification of the maximum safe needle penetration depth for the device inpig skin with highspeed camera measurement.
Shelf life testingPerformance data supports the shelf life of the device by demonstrating continued sterility,package integrity, and device functionality over the intended shelf lifeaccording to ISO 11607-1:2006+A1:2014 and ISO11607-2:2006+A1:2014
LabelingISO 15223-1:2012EN 1041:2008IEC 82079-1:2012IEC62366:2007
Identified Risk to HealthMitigation MeasuresEvidence
Electrical shock orElectromagnetic interferencewith other devicesEMC testing and electricalsafety testingIEC60601-1:2005IEC60601-1-2:2014IEC60601-1-6:2010
LabelingISO 15223-1:2012EN 1041:2008IEC 82079-1:2012IEC62366:2007
Damage to underlying tissueincluding nerves and bloodvessels, scarring and hyperand hypopigmentation.Exceeding safe penetrationdepthMechanical failureNon-clinical performance testingNon-clinical performance data demonstrates that the device performs as intended underanticipated conditions of use.The following performance characteristics are testedi.Accuracy of needle penetration depth and puncture rate in pig skin with high -speed camera measurement;ii.Safety features built into the device to protect against cross-contamination,including fluid ingress protection due to a safety membrane; andiii.Identification of the maximum safe needle penetration depth for the device inpig skin with highspeed camera measurement.
Software malfunctionTechnological characteristicsNon-clinical performance data demonstrates that the device performs as intended underanticipated conditions of use.The following performance characteristics are testedi.Accuracy of needle penetration depth and puncture rate in pig skin with high -speed camera measurement;ii.Safety features built into the device to protect against cross-contamination,including fluid ingress protection due to a safety membrane; andiii.Identification of the maximum safe needle penetration depth for the device inpig skin with highspeed camera measurement.
Shelf life testingISO 11607 1:2006ISO11607 2:2006
LabelingISO 15223-1:2012EN 1041:2008IEC 82079-1:2012IEC62366:2007
Software verification, validationand hazard analysisIEC62304:2006

{9}------------------------------------------------

MT.DERM

K182407

{10}------------------------------------------------

8. Special Controls - Non clinical performance testing

In addition to the general standards and risk mitigation measures identified above, the Exceed Microneedling device has been subjected to the special controls outlined in 21 CFR 878.4430.

9. Clinical performance testing

A clinical study was conducted to support the safe and effective performance of the Exceed Microneedling device for the treatment of acne scars on the face.

The study was conducted at a single center. 56 otherwise healthy volunteers with facial acne scarring were recruited. Subjects underwent 4 micro needling sessions 4 weeks apart. Treatments were conducted by a medical physician.

Subjects were assessed at baseline and 12 weeks after the last treatment.

The subjects face was cleansed using a skin disinfecting product and numbed using a topical anaesthetic prior to the treatment. A sterile gel was applied to the subject's face to prevent abrasion and friction during the treatment. The physician was instructed to start with a minimum needle protrusion of 0.5 mm and increase until pinpoint bleeding was achieved. During the treatment blood was with a sterile saline solution, to prevent encrustation.

Forty-seven subjects (83.9%, (47/56)) completed all treatment and follow up visits.

The mean age of the subjects was 35.8 years (range 18-62 years). The study population was comprised of 18 (32.1%) males and 38 (67.9%) females. Fitzpatrick phototypes ranged from I -IV. No subjects with Fitzpatrick skin type V or VI were included in the study.

A summary of the subject demographics is given in Table 1.

Table 1: Subject demographics at baseline.

All Subjects
N56
Age (years)
Mean (SD)35.8 (9.9)
Minimum18
Median32
Maximum62
N(%)

{11}------------------------------------------------

Sex
Male1832.1
Female3867.9
Fitzpatrick Skin Type
I11.8
II2951.8
III2030.8
IV610.7
Acne Scar Assessment Scale
Very mild (1)59
Mild (2)1425
Moderate (3)2137
Severe (4)1629

Measurement of effectiveness

Physician measurement of effectiveness

At each visit digital photographs (VISIA) were taken of the subject's face. At the end of the study the digital images of the subject's face were randomized and analyzed independently by 3 physicians using the Acne Scar Assessment Scale (ASAS). The ASAS scale is a validated 5-point photo numeric descriptive grading scale used to grade acne scarring (table 2).

Table 2. Acne Scar Assessment Scale
---------------------------------------
GradeTermDescription
0ClearNo depressions are seen in the treatment area. Macular discoloration may beseen.
1Very MildA single depression is easily noticeable with direct lighting (deep). Most or allof the depressions seen are only readily apparent with tangential lighting(shallow).
2MildA few to several, but less than half of all the depressions are easily noticeablewith direct lighting (deep). Most of the depressions seen are only readilyapparent with tangential lighting (shallow).

{12}------------------------------------------------

GradeTermDescription
3ModerateMore than half of the depressions are apparent with direct lighting (deep).
4SevereAll or almost all the lesions can be seen with direct lighting (deep).

Subject measurement of effectiveness

At the end of the study period subjects were asked to grade the effectiveness of the treatment using the following subject reported measures.

    1. Self-assessed Scar Improvement Scale (SASIS)
    1. Subject Global Aesthetic Improvement Scale (SGAIS)
    1. Subject Satisfaction of treatment.

Measurement of Safety

During the study the physician recorded adverse events.

Using a subject diary, subjects graded their pain, discomfort and redness/swelling on the day of the treatment and thereafter daily until day 7. Pain and discomfort were recorded using an 11-point visual analogue scale (0-10) where 0 was equivalent to "no pain" or "no discomfort" to 10, "most intense pain ever" and "most discomfort ever". A descriptive grading scale was used for subjects to evaluate their skin redness with the addition of photography depicting grades of skin redness in different phototypes (table 3).

Table 3 Redness grading scale

GradeDescription
NoneNo erythema or redness. Skin is normal color.
MinorVery faint erythema or redness. Skin has a slight redness to it.
MildBlotchy, visible redness that does not cover the entire face. Skin has a noticeable inconsistentredness to it.
ModerateDull red color to the skin. Skin has a very definite redness to it.
SevereBright or dark red color to the skin. Skin is severely red.

{13}------------------------------------------------

Results – Safety

No serious adverse events were reported. No adverse events persisted at the final follow up. In 18 out of 56 subjects (32%), 26 adverse events were reported during the study (Table 4).

There were 22 anticipated or expected adverse events and 4 unanticipated or unexpected adverse events. Three of the unexpected adverse events were attributable to the device (n=2) and bleeding of treatment area 2 days post intervention (n=1). One unexpected event (fever and rigors) was not attributable to the device.

Of the 26 reported events, 23 of 26 (88.5%) were recorded as mild, and 3 of 26 (11.5%) as moderate. All but one event (Fever and ague, 1 of 26 (3.8%)) was attributable to the test device. No adverse event was directly linked or resulted in study discontinuation.

Bruising (6 events in 5 subjects) and swollen lymph-nodes (6 events in 3 subjects) were the most common adverse events. Table 4 lists all adverse events.

{14}------------------------------------------------

Table 4 – Adverse events

Adverse Event (AE)CategoryNumber ofAdverseeventsNumber ofsubjectsAdverse events in relation to the deviceExpected/UnexpectedFitzpatrickskintype
RelatedNon-Related
Bleeding 2 days posttreatment111 (100%)0UnexpectedFP III
Bruising656 (100%)0Expected4/5 FP II1/5 FP III
Fever and ague1101 (100%)UnexpectedFP III
Flaking of the skinsurface313 (100%)0ExpectedFP II
Headache222 (100%)0Unexpected1/2 FP II1/2 FP III
Herpes simplex111 (100%)0ExpectedFP III
Hyperpigmentation444 (100%)0Expected1/4 FP II2/4 FP III1/4 FP IV
Pustules and rash111 (100%)0ExpectedFP III
Serous fluid leakage111 (100%)0ExpectedFP II
Swollen lymph nodes636 (100%)0Expected1/3 FP II2/3 FP III
All categories261825 (96%)1 (4%)-

{15}------------------------------------------------

Subject reported outcomes - Pain and Discomfort

Subject evaluation of pain on the evening of the treatment showed a mean pain score of 2.8 (mean of 4 treatments, (range 0-9)). Mean pain scores receded to <1 by day 3 and receded further to 0.05 by Day 7. Discomfort was highest on Day 1 with a mean score of 3.20 across the 4 treatments, range 0-10. Discomfort had receded to <1 by day 4 and receded further to 0.12 by Day 7 (figure 1-2).

Image /page/15/Figure/2 description: Figure 1 shows that a subject recorded pain scores over 4 treatments. The title of the figure is "Figure 1. Subject recorded pain scores over 4 treatments1". The figure is likely a graph or table that shows the pain scores for each of the 4 treatments. More information is needed to fully describe the image.

Image /page/15/Figure/3 description: The image is a bar graph showing pain levels after different treatments over the course of 7 days. The y-axis represents pain intensity, ranging from 0 (no pain) to 10 (intense pain), with an indicator at 5 for moderate pain. The x-axis represents the day, from 1 to 7. There are four different treatments shown on the graph: after first treatment, after second treatment, after third treatment, and after fourth treatment.

  • based on a scale 0-10 0 = No pain. 10 = Worst pain ever 1

{16}------------------------------------------------

Image /page/16/Figure/0 description: The image is a box plot that shows the level of discomfort reported by patients after four different treatments over a period of seven days. The y-axis represents the level of discomfort, ranging from 0 (no discomfort) to 10 (intense discomfort). The x-axis represents the day of the treatment. The discomfort levels generally decrease over time for all four treatments, with the most significant decrease occurring within the first few days. By day 6 and 7, the discomfort levels are near zero for all treatments.

Figure 2. Subject recorded discomfort scores over 4 treatments2

    1. based on a scale 0-10 0 = No discomfort. 10 = Worst discomfort ever

{17}------------------------------------------------

Subject reported outcomes - Redness

27 of 55 subjects (49%) reported their skin to be moderately red on the evening of the treatment, the remainder reporting their signs to be mild or less than mild. By day 5, 48 of 51 subjects (94%) reported their skin redness as either minimal or absent (figure 3).

Image /page/17/Figure/2 description: The bar graph illustrates the mean skin redness recorded from the evening of day 1 to day 7. The x-axis represents the days, from day 1 to day 7, and the y-axis represents the number of subjects, ranging from 0 to 50. Each day has five bars representing the levels of skin redness: None, Minimal, Minor, Mild, and Moderate. On day 7, 46 subjects had no skin redness, while 3 subjects had minimal redness.

Figure 3. Subject recorded skin redness recorded from the evening of day 1 to day 7

Results - Effectiveness

Physician measurement of effectiveness

Results of the photograding using the Acne Scar Assessment Scale (ASAS) demonstrated that at baseline the mean population score was 2.89. At final follow up (3 months after the last treatment) the mean grade was 2.27 (table 5).

10/56 (18%) subject could not be evaluated by the 3 blinded physicians at final follow up, 3 months after last treatment. The reasons that 10 subjects could not be evaluated are listed below;

  • 6 subjects did not complete all 4 treatments -
  • 3 subjects did not appear at the final follow-up -
  • -1 subject had no photographs at final follow-up

Table 5. Results of photograding of Acne Scar Assessment Scale from 3 blinded physicians

TimepointnMeanStandardDeviationMinimumMedianMaximum
Baseline562.890.96134
Final follow up 3 monthsafter the last treatment462.271.03134

{18}------------------------------------------------

TimepointnSubjectsimproved1 grade asagreedupon by 2of 3blindedphysiciansSubjectshowed someimprovement(improvementis greaterthan 0 butless than 1)NOimprovementSubjectworsenedMeanchangeStandarddeviationRange
Final followup 3months posttreatment4627 (59%)38(83%)8(17%)00.590.380-1

Table 6. Change from Baseline of photograding of ASAS

Of the 38 subjects that showed some improvement; 20/24(83%) subjects were Fitzpatrick skin type II, 14/18(78%) were Fitzpatrick skin type III and 4/4(100%) were Fitzpatrick skin type IV

Of the 17 subjects that improved by 1 grade; 8/24(33%) subjects were Fitzpatrick skin type II, 5/18 (28%) were Fitzpatrick skin type III and 4/4(100%) were Fitzpatrick skin type IV.

Results - Effectiveness - Subject measurement of effectiveness

Self-assessed Scar Improvement Scale (SASIS)

Subjects assessed their acne scarring compared to baseline using the Self-assessed Scar Improvement Scale (SASIS). Thirty-eight, 38/43 (88%) of subjects reported some improvement in their acne scarring. Five subjects 5/43 (12%) reported no improvement. No subjects reported a worsening of their acne scars (table 7).

{19}------------------------------------------------

Table 7. Self-assessed Scar Improvement Scale (SASIS)

Self-assessed ScarImprovement Scale(SASIS)Compared to the beginning of the study please rate your facial scars now?
Grading- 1Exacerbationof Acne Scars0No change inappearance ofmy acne scars11% - 25%improvementin appearanceof my acnescars225% - 50%improvementin appearanceof my acnescars350% - 75%improvementin appearanceof my acnescars475% - 99%improvementin appearanceof my acnescars
Number of Subjects(n=43)05181541
% of Subjects0.011.641.934.99.32.3

Subject Global Aesthetic Improvement Scale (SGAIS)

Subjects also assessed their acne scarring compared to baseline using the Subject Global Aesthetic Improvement Scale (SGAIS). Thirty-three, 33/43 (77%) of subjects reported some improvement in their acne scarring. Ten subjects 10/43 (23%) reported that the appearance of their acne scarring was essentially the same as the original condition. No subjects reported that the appearance of their acne scars worsened (table 8).

Table 8. Subject Global Aesthetic Improvement Scale (SGAIS)

Subject GlobalAestheticImprovementScaleCompared to the beginning of the study please rate your facial scars now?
Grading1Very MuchImproved:Optimal cosmeticresult.2Much Improved:Markedimprovement inappearance from theinitial condition, butnot completelyoptimal.3Improved:Obviousimprovement inappearance frominitial condition.4No Change: Theappearance isessentially thesame as theoriginalcondition5Worse: Theappearance isworse than theoriginalcondition.
Number ofSubjects (n=43)0429100
% of Subjects0.09.367.423.30.0

{20}------------------------------------------------

Additional questions of subject's satisfaction

    1. Do you notice any improvement in how your acne scars look in the treated area?
Yes [N, (%)]No [N, %]
37 (86%)6 (14%)

2. How would you characterize your satisfaction with the treatment?

GradingExtremelysatisfiedSatisfiedSlightlysatisfiedNeithersatisfiednordissatisfiedSlightlydissatisfiedDissatisfiedExtremelydissatisfied
Number ofSubjects (n=-43)718107010
% of Subjects16.341.923.316.30.02.30.0

How would you characterize your satisfaction with the treatment?

Would you recommend this treatment to your friends and family members? ന്

Yes [N, %]No [N, %]
38 (88%)5 (12%)

{21}------------------------------------------------

10. Statement of Substantial Equivalence:

513(i) of the FD&C Act (21 U.S.C. 360c(i) states that for substantial equivalence a proposed device is required to have the same intended use and similar technological characteristics as the predicate device. Where there are differences in technological characteristics, these can be negated by appropriate clinical or scientific data demonstrating that the proposed device is as safe and effective as the predicate device, and that the proposed device does not raise any different questions of safety and effectiveness than the predicate device for the same intended use.

MT.DERM GmbH has demonstrated that the Exceed Microneedling device has the same generic classification (generic description) and basic technologies as the predicate device DEN160029. Both devices are micro needling systems containing >1 needle that mechanically punctures or injures the skin for aesthetic use. Both devices have used needle depths of up to 1.5mm in clinical investigations.

Where there are differences between the Exceed micro needling device and the predicate (DEN160029) MT.DERM GmbH has conducted clinical and non-clinical performance testing applicable to those general and special controls deemed necessary by the agency for this product classification and has determined that the Exceed microneedling device raises no new questions relating to safety or effectiveness and therefore has demonstrated that the Exceed microneedling device is substantially equivalent to the referenced predicate DEN160029.

§ 878.4430 Microneedling device for aesthetic use.

(a)
Identification. A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.