K Number
K182329

Validate with FDA (Live)

Date Cleared
2018-11-08

(72 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SureCure Orthodontic Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition. The aligner system repositions teeth by way of continuous gentle force.

Device Description

Digital Orthodontic Care's SureCure Orthodontic Aligner System consists of a series of customized clear plastic removable aligners that are fabricated from a clear, thin thermoformed polyurethane. The aligners are designed to gradually move the patient's teeth incrementally, repositioning them from their original misalignment to a more aligned state.

Digital Orthodontic Care manufactures the customized aligners based on either standard impressions or intraoral scans sent to the company by the prescribing dentist or orthodontist. Digital Orthodontic Care manufactures models from the impressions and those models are scanned using standard validated software. The digital files are used to produce the aligner series with the thermoplastic polyurethane.

The thermoplastic material used for fabrication of the aligners is commonly used in many dental appliances, including the predicate aligners.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the SureCure Orthodontic Aligner System. However, it does not contain information about acceptance criteria or specific study results that prove the device meets acceptance criteria in the format requested.

The document primarily focuses on establishing substantial equivalence to a predicate device (ClearCorrect System, K113618) based on material, intended use, mode of action, and manufacturing methods.

Here's what can be extracted based on the provided text, and what information is missing:

1. Table of Acceptance Criteria and Reported Device Performance:

  • Acceptance Criteria: Not explicitly stated in the document. The substantial equivalence argument relies on similarity to the predicate device.
  • Reported Device Performance: No specific performance metrics or clinical outcomes are reported in the context of acceptance criteria. The document states that the material passed biocompatibility testing and the device passed shelf-life testing, which are general safety and quality assurance measures rather than performance criteria for efficacy in tooth movement.
Acceptance CriteriaReported Device Performance
Not specifiedNo performance metrics provided for efficacy
Biocompatibility (ISO 10993)Passed
Shelf-life5 years (after real-time storage)

2. Sample size used for the test set and the data provenance:

  • The document states that "Neither animal nor human testing are required for this product because it is composed of the same materials, is designed similarly, and is manufactured by a similar method as the predicate device."
  • Therefore, there is no specific test set or associated data provenance for evaluating the clinical performance of the device itself.
  • Biocompatibility and shelf-life testing would have used material samples, but the sample sizes are not provided, nor is the provenance of raw material.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • As there was no clinical human or animal testing outlined for the device's efficacy, there was no test set for which ground truth was established by experts.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Since no clinical test set for performance evaluation was conducted, there was no adjudication method described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • The device described is an orthodontic aligner system, not a diagnostic AI device that involves human "readers" or AI assistance for interpretation.
  • Therefore, no MRMC comparative effectiveness study was done in the context of AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This question is not applicable as the device is a physical orthodontic aligner system, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc.):

  • No specific ground truth for clinical efficacy was used as no clinical trials were conducted for the new device.
  • The ground truth for biocompatibility and shelf-life would be based on established international standards (ISO 10993) and real-time degradation measurements, respectively.

8. The sample size for the training set:

  • This question is typically relevant for AI/machine learning models.
  • The document mentions that "Digital Orthodontic Care manufactures the customized aligners based on either standard impressions or intraoral scans sent to the company by the prescribing dentist or orthodontist. Digital Orthodontic Care manufactures models from the impressions and those models are scanned using standard validated software."
  • This indicates the use of software in the manufacturing process (e.g., for 3D printing models), but it does not describe a "training set" in the context of an AI algorithm or a study for the aligner's clinical performance.

9. How the ground truth for the training set was established:

  • Not applicable, as no training set for an AI model or clinical performance study for the aligner system is described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Digital Orthodontic Care % Patsy Trisler Regulatory Consultant Trisler Consulting 5600 Wisconsin Avenue, #509 Chevy Chase, Maryland 20815

Re: K182329

Trade/Device Name: SureCure Orthodontic Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: November 2, 2018 Received: November 2, 2018

Dear Patsy Trisler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

November 8, 2018

{1}------------------------------------------------

ence determination does not mean that FDA

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mary S. Mary S. Runner -S3 Runner -S33 Date: 2018.11.08 08:47:46 -05'00'

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182329

Device Name

SureCure Orthodontic Aligner System

Indications for Use (Describe)

The Sure Orthodontic Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition. The aligner system repositions teeth by way of continuous gentle force.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K182329

510(k) Summary

Submitter Name:Digital Orthodontic Care
Submitter Address:348 Bronte Street, Unit 16Milton, ON L9T 5B6 CANADA
Phone Number:647-977-7253
Contact Person:Mr. Jeff SheppardPresident
Date Prepared:November 7, 2018
Device Trade Name:SureCure Orthodontic Aligner System
Common NameAligner, Sequential
Classification NameNumberProduct CodeRegulatory ClassOrthodontic Plastic Bracket21 CFR 872.5470NXC2
Primary PredicateDevice:K113618, ClearCorrect System, ClearCorrect, LLC
Reference Device:K152086, Ortho Analyzer, 3Shape A/S
Statement ofIntended UseThe SureCure Orthodontic Aligner System is indicated for thetreatment of tooth malocclusion in patients with permanentdentition. The aligner system repositions teeth by way ofcontinuous gentle force.
Device Descriptionand Summary ofTechnologicalCharacteristicsDigital Orthodontic Care's SureCure Orthodontic Aligner Systemconsists of a series of customized clear plastic removable alignersthat are fabricated from a clear, thin thermoformed polyurethane.The aligners are designed to gradually move the patient's teethincrementally, repositioning them from their original misalignmentto a more aligned state.
Digital Orthodontic Care manufactures the customized alignersbased on either standard impressions or intraoral scans sent tothe company by the prescribing dentist or orthodontist.Digital Orthodontic Care manufactures models from theimpressions and those models are scanned using standardvalidated software. The digital files are used to produce thealigner series with the thermoplastic polyurethane.
The thermoplastic material used for fabrication of the aligners iscommonly used in many dental appliances, including thepredicate aligners.
Mechanism of Action Based on the clinician's treatment plan, each aligner is used for adefined period of time to exert gentle force to achieve progressive
correction has been achieved.realignment of the teeth. This occurs over time until the final

{4}------------------------------------------------

Device Testing Laboratory Testing

Test data were submitted to validate the processes used for the design and manufacture of the aligners.

Biocompatibility

The thermoplastic polyurethane used for making the aligner series has been tested according to Good Laboratory Practices for its biocompatibility according to ISO 10993. In addition, the final finished aligner device was tested according to ISO 10993, Part 5 (Cytotoxicity - MEM Elution).

The testing has shown that the material is safe and biocompatible for the stated intended use.

Shelf-life Testing

The aligner material has been tested after real-time storage for a shelf life of 5 years.

Animal | Human Testing

Neither animal nor human testing are required for this product because it is composed of the same materials, is designed similarly, and is manufactured by a similar method as the predicate device.

Comparison to A side-by-side comparison of SureCure to the predicate device is Predicate Device: presented in the Substantial Equivalence Comparison table.

Based on the documentation presented in the 510(k), as Substantial Equivalence summarized above and in the following table, it can be concluded Conclusion that Digital Orthodontic Care's SureCure Orthodontic Aligner System is substantially equivalent to the predicate device.

{5}------------------------------------------------

Trade Name:SureCure OrthodonticClearCorrect System
Aligner System
510(k) NumberK182329K113618
ManufacturerDigital Orthodontic CareClearCorrect, LLC
Classification # &Product CodeClass21 CFR 852.5470NXC221 CFR 852.5470NXC2
Intended UseThe SureCure OrthodonticAligner System is indicated forthe treatment of toothmalocclusion in patients withpermanent dentition. Thealigner system repositions teethby way of continuous gentleforce.The ClearCorrect System isindicated for the treatment oftooth malocclusion in patientswith permanent dentition (i.e. allsecond molars). The ClearCorrect System positions teethby way of continuous gentleforce.
Mode of ActionAlignment of teeth byapplication of continuous gentleforce, by sequential use ofpreformed plastic trays.Alignment of teeth byapplication of continuous gentleforce, by sequential use ofpreformed plastic trays.
Method of UseEach preformed plastic tray isworn by the patient asprescribed by the dentalpractitioner, usually a fewweeks prior to using the nextsequential aligner tray.Each preformed plastic tray isworn by the patient asprescribed by the dentalpractitioner, usually a fewweeks prior to using the nextsequential aligner tray.
MaterialThin thermoformedpolyurethaneThin thermoformedpolyurethane
Method ofManufactureUsing software, molds/modelsare 3D printed based on thetreatment plan. Aligners arefabricated on the molds using athermoforming machine.Using software, molds/modelsare 3D printed based on thetreatment plan. Aligners arefabricated on the molds using athermoforming machine.
BiocompatibleYesYes
OTC or RxRxRx
Software UseYesYes
SterileNoNo

Substantial Equivalence Comparison

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.