Intended Use

Zyppah® is intended for use by adult patients (18 years or older) as an aid to reduce snoring.

Device Description

The Zyppah Anti-Snoring Appliance is a single piece tray system for intraoral use. The 'boil-and-bite' device is customized to each patient. It features an elastic strap that spans the device that is designed to keep the tongue in place during sleep, instead of blocking the airway and contributing to snoring. The product is non-sterile and provided in a sealed box with directions for use.

AI/ML Overview

This 510(k) summary (K182312) describes the Zyppah® Anti-Snoring Appliance. It's important to note that the document does not describe a study that proves the device meets specific acceptance criteria in terms of clinical performance. Instead, it relies on substantial equivalence to predicate devices and non-clinical bench testing.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a formal table of acceptance criteria for clinical performance that the device was tested against. The acceptance is based on demonstrating substantial equivalence to predicate devices, primarily through comparison of technological characteristics, materials, and non-clinical testing.

The key "performance" reported is related to structural integrity:

Acceptance Criteria (Implied)Reported Device Performance
Tensile strength of strap14% increase in force withstood compared to predicate

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not applicable for clinical performance. The only "test set" mentioned relates to non-clinical bench testing of the strap. No specific sample size for this bench test is provided in the summary.
  • Data Provenance: The device performance data (tensile strength) is from non-clinical bench testing. The country of origin and whether it was retrospective or prospective is not specified, but it would be prospective for this type of bench test.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. The ground truth for the tensile strength test is objective physical measurement, not expert assessment.

4. Adjudication Method for the Test Set

Not applicable, as no expert adjudication was involved for the non-clinical tensile strength testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The submission explicitly states: "No clinical testing was performed in association with this submission."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Again, no clinical testing (standalone or otherwise) was performed. This device is a physical appliance, not an algorithm.

7. The Type of Ground Truth Used

  • For the non-clinical tensile strength test, the "ground truth" is the physical measurement of the force required to break the strap.
  • The overall "ground truth" for marketing clearance is the substantial equivalence to predicate devices that have prior FDA clearance and a history of safety and effectiveness.

8. The Sample Size for the Training Set

Not applicable. This device is a physical anti-snoring appliance, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

January 24, 2019

Always More Marketing % Colette Cozean Regulatory Consultant EyeDeas Company 21581 Midcrest Drive Lake Forest, California 92630

Re: K182312

Trade/Device Name: Zyppah Anti-Snoring Device Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: December 14, 2018 Received: December 26, 2018

Dear Colette Cozean:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mary S. Runner Runner Runner-S3 Digitally signed by Mary S. Date: 2019.01.24 14:05:59 -23

for

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182312

Device Name Zyppah(R) Anti-Snoring Device

Indications for Use (Describe)

Zyppah(R) is intended for use by adult patients (18 years or older) as an aid to reduce snoring.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K182312

Applicant:Always More Marketing, Inc.24981 Palmilla DriveCalabasas, CA 91302, USAPhone: 818-414-4817
Contact Person:Colette Cozean, Ph.D.21581 Midcrest Dr.Lake Forest, CA 92603 USA949-855-2885ccozean@cox.net
Date Prepared:January 21, 2019
Proprietary Name:Zyppah® Anti-Snoring Appliance
Common Name:Intraoral Device for Snoring
Classification Name:Device, anti-snoring(Class II, 21 CFR 872.5570, Product Code LRK)
Predicate Device:Dr. Greenburg's Hybrid Zyppah (K111680) – Prescription Predicate
Reference Predicates:ZQuiet (K180124) – Over the Counter Reference PredicateSnoreRX (K170825) - Over the Counter Reference Predicate

8. Description of the Device

The Zyppah Anti-Snoring Appliance is a single piece tray system for intraoral use. The 'boil-and-bite' device is customized to each patient. It features an elastic strap that spans the device that is designed to keep the tongue in place during sleep, instead of blocking the airway and contributing to snoring. The predicate The product is non-sterile and provided in a sealed box with directions for use.

9. Indications for Use

{4}------------------------------------------------

Zyppah® is intended for use by adult patients (18 years or older) as an aid to reduce snoring.

10. Technological Characteristics

The appliance is equivalent in design and functionality to the prescription Zyppah device (K111680). Both devices work by depressing the tongue during sleep to open the airway and reduce the incidence of snoring, in addition to advancing the mandible. Both are customized to the individual patient by a standard 'boil and bite' method. Both use the same materials and manufacturing process. Both devices are provided non-sterile. The sole difference between the subject and predicate device is the use of double-hole retention for the silicone strap, rather than the single-hole retention in the predicate device. This adds additional strength to the binding to prevent any breakage or loosening of the strap.

The appliance has similar labeling as the OTC reference predicate devices, SnoreRx and ZQuiet. It utilizes a validated questionnaire to assess the risk of sleep apnea, and contains appropriate warnings and labeling for the device to be used without a prescription.

SubjectPrescriptionPredicateOTC ReferencePredicateOTC ReferencePredicate
NameZyppah® Anti-Snoring ApplianceDr. Greenburg'sHybridSnoreRxZQuiet
ClassificationNameDevice, anti-snoringDevice, anti-snoringDevice, anti-snoringDevice, anti-snoring
510(k)NumberK182312K111680K170825K180124
ClassIIIIIIII
Product CodeLRKLRKLRKLRK
21 CFR872.5570872.5570872.5570872.5570
Technology &Mechanism ofActionRepositioning thetongue,mandibularadvancementRepositioning thetongue,mandibularadvancementMandibularadvancementMandibularadvancement
Indications foruseZyppah® isintended for useby adult patients(18 years orolder) as an aid toreduce snoring.This appliance isindicated forpersons 18 yearsor older, who wishto reduce theincidence ofsnoring and/ormild to moderatesleep apnea."SnoreRx isintended for useon adultpatients 18years of age orolder as an aidfor thereduction ofsnoring""ZQuiet isintended as anaid in thereduction ofsnoring foradults at least18 years old."
PrescriptionStatusOTCPrescriptionOTCOTC

{5}------------------------------------------------

ProvidedSterileNoNoNoNo
MaterialsHard outer shellwiththermoplasticfilling materialand elastic bandHard outer shellwith thermoplasticfilling material andelastic bandThermoplasticElastomer withColorant

10.1 Technology - Mechanism of Action

The subject device and prescription predicate device (K111680) are identical in mechanism of action. They both work by utilizing an intraoral mouthpiece to reposition the tongue in conjunction with a slight advancement of the mandible.

The mechanism of action for the other two OTC predicate devices, the SnoreRx and ZQuiet, is solely advancing the mandible. With any intraoral device that alters the mandible position, mandibular changes can lead to pain or discomfort in the temporomandibular joint (TMJ). The subject device has no new risks in regards to mandibular advancement.

Utilizing the same mechanism of action as the prescription predicate the subject device raises no new concerns regarding effectiveness or safety as compared to predicate devices.

10.2 Technology - Indication for Use

The subject device uses language for the indication for use that is identical in substance to previouslyapproved OTC devices for the reduction of snoring. The subject device is "intended for use by adult patients (18 years or older) as an aid to reduce snoring."

This is almost identical to the indications for use for the two OTC predicate devices, the SnoreRx and ZQuiet, whose indications for use are "SnoreRx is intended for use on adult patients 18 years of age or older as an aid for the reduction of snoring" and "ZQuiet is intended as an aid in the reduction of snoring for adults at least 18 years old," respectively.

As the indications for use are identical, there is no new concerns regarding effectiveness or safety as compared to predicate devices.

10.4 Technology - Prescription Status

The subject device is intended to be sold over-the-counter. Product labeling has therefore eliminated prescribing information. Two of the predicate devices, SnoreRx and ZQuiet, have clearance for over-thecounter use.

Warnings and contraindications are substantially equivalent to those of over-the-counter predicate devices. The labeling specifically states that the device does not treat obstructive sleep apnea, and includes the validated STOP-BANG questionnaire is included in the labeling to help mitigate the risk of undiagnosed OSA among potential Zyppah patients.

{6}------------------------------------------------

The similar predicate prescription device (Dr. Greenburg's Hybrid, K111680) has been on the market for more than seven years. Due to its postmarket record of substantially equivalent performance to other devices of this type and substantially equivalent labeling and instructions for use to that of the over-the-counter predicate devices, there are no new concerns regarding effectiveness or safety as compared to predicate devices.

10.5 Technology - Materials

To support this application, the sponsor has included Biocompatibility testing results demonstrating the subject device is biocompatible under ISO 10993, including cytotoxicity, sensitization, and irritation testing. The device was shown to meet the requirements of the ISO 10993 guidelines for each of these tests. Other biocompatibility tests recommended for permanent mucosal contact were not conducted because all materials used in the device are certified USP Class VI and used in many other currently marketed medical devices.

Because the same materials and manufacturing processes are used as the currently marketed predicate device, there are no new concerns regarding effectiveness or safety as compared to the predicate device.

11. Non-Clinical Testing

The tensile strength of the proposed OTC device was compared to the single-hole design of the cleared predicate in bench testing. It was found that using a two-hole the silicone strap in place increased the force the strap could withstand by 14% over that of a single-hole design used in the predicate device.

The sponsor conducted a risk analysis on the device in accordance with ISO 14971:2007, taking into account the issues raised in the FDA Guidance Document "Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea – Guidance for Industry and FDA." All identified risks have been addressed through device design or with communication with the user through the instructions for use.

12. Biocompatibility

The materials and manufacturing process are similar to the currently marketed materials in the prescription predicate device.

Testing results that have been include cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10) and irritation (ISO 10993-10). As all materials used in the device are certified USP Class VI and used in many other currently marketed medical devices, no additional biocompatibility testing was conducted.

Class VI certification is less stringent than 10993

13. Clinical Testing

{7}------------------------------------------------

No clinical testing was performed in association with this submission.

14. Conclusions

The results of the comparison of design, materials, intended use, labeling, and technological characteristics demonstrate the subject device is substantially equivalent in safety and efficacy the legally marketed predicate devices. Therefore, the sponsor concludes the proposed Zyppah® device is substantially equivalent to the identified predicate devices.

§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.

(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”