K Number
K182265

Validate with FDA (Live)

Date Cleared
2018-11-20

(91 days)

Product Code
Regulation Number
882.5550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Codman Certas Plus Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

Device Description

This submission includes two additional configurations of the Codman Certas Plus Programmable Valve: Certas Plus Inline Small and Certas Plus Right Angle. The Codman Certas Plus Programmable Valves are sterile, single use, implantable devices designed for shunting cerebrospinal fluid (CSF) for the treatment of hydrocephalus.

The Codman Certas Plus Programmable Valve is a pressure-regulating valve utilizing the ruby ball-in-cone principle with a pressure inducing spring design. Intraventricular pressure is maintained by the ball and cone valve seat design. As the differential pressure across the shunt increases, the ball further displaces from the cone, through which CSF flows, thereby increasing flow and re-establishing the selected pressure. The ball is manufactured of synthetic ruby, as is the matching cone. Together these components provide a precise fit for regulating the flow of CSF through the valve.

The valve is available with 8 different performance settings for constant intraventricular pressure and drainage of CSF. Seven (7) of the settings provide a change in operating pressure, with a range of 25 to 215 mmH2O. The eighth (8) setting provides a minimum opening pressure of '400' mmH20, thus allowing a physician to turn the valve "virtually off" without the need to surgically remove the valve to limit flow. The pressure of the valve is set preoperatively and can be noninvasively changed postimplantation by using the Codman Certas Tool Kit.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device, specifically the Codman Certas Plus Programmable Valve. This type of submission is for demonstrating substantial equivalence to a predicate device, not for proving the device meets acceptance criteria through an AI-based study or a multi-reader, multi-case (MRMC) comparative effectiveness study. The document primarily focuses on bench testing (Verification Testing) and simulated use (Validation Testing) to establish that the modified versions of the device perform equivalently to the predicate.

Therefore, many of the questions related to AI-specific study design (like ground truth establishment, expert consensus, MRMC studies, training/test set sizes, and specific performance metrics like AUC, sensitivity, specificity for diagnostic AI) are not applicable to this document.

However, I can extract information related to the physical device's performance, acceptance criteria for those tests, and the general approach to validating the device's functionality.

Here's a breakdown of the requested information based on the provided document:


Acceptance Criteria and Device Performance for Codman Certas Plus Programmable Valve

The device is a non-AI, physical medical device (a CSF shunt) and therefore the typical "acceptance criteria" and "study" described in an AI context (e.g., AUC, sensitivity, specificity, MRMC studies) are not present. Instead, the document describes design verification and validation activities against established standards and internal criteria to demonstrate substantial equivalence to a predicate device.

1. A table of acceptance criteria and the reported device performance

The document states that "All samples in design verification testing met predefined acceptance criteria" and "test results demonstrated that the acceptance criteria were met." However, the specific numerical acceptance criteria for each test are not explicitly detailed in this summary. The results are reported as "PASS."

Test CategoryTest Method SummaryReported Device Performance/Result
Verification Testing
Shunt Safety and Performance TestingTesting per ISO 7197:2009 standard, including: • 4.2. Radiopacity • 4.3. Biocompatibility • 4.4. Resistance to leakage • 4.5. Control of the implanted shunt • 4.6. Pressure-flow characteristics • 4.7. Identification of shunts in vivo • 4.8. Ability to withstand overpressure • 4.9. Dynamic breaking strength • 4.10. Behavior under MR imaging conditions • 4.11. Bursting pressure • 5.1.1. Reflux performance • 8.2.d. Method for puncture and indication of how often puncturing is possiblePASS
Shelf Life StudyTesting per ISO 7197:2009 standard for shunt safety and performance following accelerated aging. Clauses related to mechanical performance of the valve were tested to specifically stress the valve's silicone housings.PASS
SterilizationTesting per ISO 17665-1: 2006 and ISO 17665-2: 2009 for sterilization, moist heat. • Confirm assembly process does not create a higher risk of bioburden. • Test ability of Certas Plus Inline valve sterilization cycle to sterilize the subject devices. • Natural Product Resistance Test • Microbial ChallengePASS
Transit TestingISO 11067-1: 2009 and ISO11607-2: 2006: Packaging for Terminally Sterilized Medical Devices. Confirm that unit box and double blister prevents damage to the product in normal conditions of transit, handling, and storage in accordance with ISTA 3A (Vibration, drop, and environmental conditioning).PASS; Cosmetic defects observed on unit box with no effect on packaging or valve performance.
Valve Fixation TestingConfirm that silicone strength of Certas Plus Inline Small valve suture flange is no worse than the predicate device. Test compatibility of bone screw fixation with Certas Plus Right Angle valve per ISO 7197:2009 method of Dynamic Breaking Strength.PASS
Valve Flexibility TestingMeasure force to flex subject devices to known curvature compared to predicate device.PASS; Certas Plus Inline Small valve is more flexible and Certas Plus Right Angle valve is equivalent to predicate device.
Adjustment and Indication TestingVerify ability to use Codman Certas toolkits to adjust and indicate a valve.PASS
Validation Testing
System Safety Study (Simulated Post-Implantation Use)Clinicians evaluated the maximum acceptable rate for false positives and maximum success rate for completing a full procedure. Testing conducted on a simulated head model with variable valve orientation.PASS

2. Sample sizes used for the test set and the data provenance

The document does not specify the exact sample sizes (number of units tested) for each verification and validation test. It generally refers to "all samples" meeting criteria. This is typical for 510(k) summaries where the detailed test protocols and sample sizes are provided in the full submission, but not always summarized in this public document.

Data provenance: The testing was conducted by Integra LifeSciences Production Corp. (manufacturer/applicant). The document doesn't explicitly state the country of origin of the data in terms of patient source, as this is bench and simulation testing, not clinical data from patients. The testing is retrospective in the sense that it's performed on manufactured devices, not derived from real-time patient observations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. For a physical device like a shunt, "ground truth" generally refers to the known physical properties and performance specifications, established through engineering and quality control benchmarks, and adherence to international standards (e.g., ISO). For the "System Safety Study" (validation), "clinicians" were involved, but their number and specific qualifications are not detailed. Their role was to evaluate the device in simulated use.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This concept is relevant for studies involving human interpretation of data, typically in diagnostic imaging or AI performance assessment. For physical device performance, reconciliation/adjudication is typically handled by senior engineers or quality assurance personnel reviewing failed tests, not by a formal multi-reader adjudication process.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device, not an AI diagnostic or assistance system. The "System Safety Study" involved clinicians but was a usability/performance study in a simulated environment, not an MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm. The device functions mechanically.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is based on:

  • Adherence to international standards (e.g., ISO 7197:2009, ISO 17665-1: 2006, ISO 17665-2: 2009, ISO 11067-1: 2009, ISO11607-2: 2006, ISTA 3A).
  • Predefined engineering specifications and performance characteristics (e.g., pressure-flow characteristics, strength, sterilization effectiveness).
  • Comparison to the performance of the legally marketed predicate device (K152152) and reference device (K053107) to demonstrate substantial equivalence.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 20, 2018

Integra LifeSciences Corporation Nancy MacDonald Regulatory Affairs Manager 11 Cabot Boulevard Mansfield. Massachusetts 02048

Re: K182265

Trade/Device Name: Codman Certas Plus Programmable Valve (Certas Plus Inline Small and Certas Plus Right Angle) Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: August 21, 2018 Received: August 22, 2018

Dear Nancy MacDonald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182265

Device Name

Codman Certas Plus Programmable Valve (Certas Plus Inline Small and Certas Plus Right Angle)

Indications for Use (Describe)

The Codman Certas Plus Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

Type of Use (Select one or both, as applicable)

For activities identified in 6 NYCRR Part 360, Series 11
For other Tier I composting activities

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. Submitter Integra LifeSciences Production Corp. 11 Cabot Blvd Mansfield, MA 02048

On behalf of: Medos International SARL Chemin-Blanc 38 CH 2400 LeLocle, Switzerland

Primary Contact: Nancy MacDonald Phone: (781) 971-5664

Secondary Contact: Jocelyn Raposo Phone: (781) 971-5688

Date of Preparation: November 1, 2018

II. Device(s)

510(k) NumberK182265
Device Proprietary NameCodman Certas Plus Programmable Valve (Certas PlusInline Small and Certas Plus Right Angle)
Common NameHydrocephalus Shunt
Classification NameCentral Nervous System Fluid Shunt and Components(21 CFR 882.5550)
Regulatory ClassificationII
Product CodeJXG

III. Predicate

The predicate and reference devices for this submission are:

Device(s)

510(k)NumberDate ClearedTitle
PredicateK152152Oct. 27, 2015Codman Certas PlusProgrammable Valve
ReferenceK053107Jan. 19, 2006Codman Hakim ProgrammableValve with/without SiphonGuard

IV. Device Description(s)

This submission includes two additional configurations of the Codman Certas Plus Programmable Valve: Certas Plus Inline Small and Certas Plus Right Angle. The Codman Certas Plus Programmable Valves are sterile,

{4}------------------------------------------------

single use, implantable devices designed for shunting cerebrospinal fluid (CSF) for the treatment of hydrocephalus.

The Codman Certas Plus Programmable Valve is a pressure-regulating valve utilizing the ruby ball-in-cone principle with a pressure inducing spring design. Intraventricular pressure is maintained by the ball and cone valve seat design. As the differential pressure across the shunt increases, the ball further displaces from the cone, through which CSF flows, thereby increasing flow and re-establishing the selected pressure. The ball is manufactured of synthetic ruby, as is the matching cone. Together these components provide a precise fit for regulating the flow of CSF through the valve.

The valve is available with 8 different performance settings for constant intraventricular pressure and drainage of CSF. Seven (7) of the settings provide a change in operating pressure, with a range of 25 to 215 mmH2O. The eighth (8) setting provides a minimum opening pressure of '400' mmH20, thus allowing a physician to turn the valve "virtually off" without the need to surgically remove the valve to limit flow. The pressure of the valve is set preoperatively and can be noninvasively changed postimplantation by using the Codman Certas Tool Kit.

The Indications for Use statement of the proposed device remains identical

Equivalence ComparisonIndications for Use
Codman Certas Plus Programmable Valve(Predicate K152152)The Codman Certas Plus ProgrammableValve is an implantable device thatprovides constant intraventricularpressure and drainage of CSF for themanagement of hydrocephalus.
Codman Hakim Programmable Valve(Reference K053107)The Codman Hakim Programmable Valves(CHPV) are implantable devices that provideconstant intraventricular pressure anddrainage of CSF for the management ofhydrocephalus.
Codman Certas Plus Inline Small andCertas Plus Right Angle ProgrammableValves(Subject of This Submission - K182265)The Codman Certas Plus ProgrammableValve is an implantable device thatprovides constant intraventricular pressureand drainage of CSF for the managementof hydrocephalus.

V. Indications for Use

to the predicate device.

VI. Comparison

to Predicate and Reference Devices

The Codman Certas Plus Inline Small and Right Angle Programmable Valves are substantially equivalent to the predicate Codman Certas Plus Programmable Valve, which has been cleared under K152152 on October 27, 2015. The Certas Plus Inline Small and Right Angle Valves have the

{5}------------------------------------------------

same intended use, design principles, operational principles and materials as the predicate Certas Plus Programmable Valve. Compared to the predicate device, the Codman Certas Plus Inline Small and Right Angle valves include the modifications listed in Table 1.

The Certas Plus Right Angle Valve is also similar in design to the reference device, the Codman Hakim Programmable Valve, Right Angle (K053107).

Table 1. Proposed Modifications for the Codman Certas Plus ProgrammableValve
DeviceModificationRationale
Certas PlusInline Small• Smaller silicone housing andreservoir• Curved silicone housing• Angled proximal connector• Angled distal connector(applicable to valves withoutunitized catheters and valveswithout SiphonGuard)• Modified silicone backing toaccommodate siliconehousing• Needle Guard materialchanged to Polyethersulfone(PES). PES is currently usedfor other Certas Plus parts• Modification to the innerblister packagingMinor modifications are beingintroduced to respond to marketdemand without changing thefundamental scientific technologyor the intended use.
Certas PlusRight Angle• Right angle silicone housing• Right angle proximalconnector• Angled distal connector(applicable to valves withoutunitized catheters and valveswithout SiphonGuard)• Modified silicone backing toaccommodate siliconehousing• Needle Guard materialchanged to Polyethersulfone(PES). PES is currently usedfor other Certas Pluscomponents.Minor modifications are beingintroduced to respond to marketdemand without changing thefundamental scientific technologyor the intended use.A right-angle configuration iscurrently available with thelegally marketed reference device,the Codman HakimProgrammable Valve (K053107).
LabelingRevised the labeling to includethe Certas Plus Inline Small andCertas Plus Right AngleValves.Added details to MRI SafetyLabeling updates to reflect theproposed valve configurationswith no changes to the indicationsor intended use.Additional details requested from
Table 1. Proposed Modifications for the Codman Certas Plus ProgrammableValve
DeviceModificationRationale
Information') to clarify how thevalve was MR tested.

{6}------------------------------------------------

VII. Performance Data

The following performance data is provided in support of the substantial equivalence determination.

Verification Testing

The following table summarizes the design verification activities. All samples in design verification testing met predefined acceptance criteria, and the proposed devices passed design verification test activities. These test results, in conjunction with the similarities to the predicate device, demonstrate that the Certas Plus Programmable Inline Small and Right Angle Valves are substantially equivalent to the predicate device.

Table 2: Design Verification Activities – Codman Certas Plus Inline Small andRight Angle Programmable Valves
TestTest Method SummaryResults
Shunt SafetyandPerformanceTestingTesting per ISO 7197:2009 standard for shunt safetyand performance including the following clauses:• 4.2. Radiopacity• 4.3. Biocompatibility• 4.4. Resistance to leakage• 4.5. Control of the implanted shunt• 4.6. Pressure-flow characteristics• 4.7. Identification of shunts in vivo• 4.8. Ability to withstand overpressure• 4.9. Dynamic breaking strength• 4.10. Behavior under MR imagingconditions• 4.11. Bursting pressure• 5.1.1. Reflux performance• 8.2.d. Method for puncture and indication ofhow often puncturing is possiblePASS
Shelf LifeStudyTesting per ISO 7197:2009 standard for shunt safetyand performance following accelerated aging. Clausesrelated to mechanical performance of the valve weretested to specifically stress the valve's siliconehousings.PASS
SterilizationTesting per ISO 17665-1: 2006 and ISO 17665-2:2009 for sterilization of healthcare products, moistheat.• Confirm assembly process does not create ahigher risk of bioburden.• Test ability of Certas Plus Inline valvesterilization cycle to sterilize the subjectdevices.○ Natural Product Resistance Test○ Microbial ChallengePASS

{7}------------------------------------------------

TransitTestingISO 11067-1: 2009 and ISO11607-2:2006: Packaging for Terminally Sterilized MedicalDevicesConfirm that unit box and double blister preventsdamage to the product in normal conditions oftransit, handling, and storage in accordance withISTA 3A. Vibration, drop, and environmental conditioningo (Packaged-Products for Parcel DeliverySystem Shipments 70kg (150 lbs) or Less(standard, small, flat or elongated).PASS;Cosmeticdefects observedon unit box withno effect onpackaging orvalveperformance.
ValveFixationTestingConfirm that silicone strength of Certas PlusInline Small valve suture flange is no worsethan the predicate device. Test compatibility of bone screw fixation withCertas Plus Right Angle valve per ISO7197:2009 method of Dynamic BreakingStrength.PASS
ValveFlexibilityTestingMeasure force to flex subject devices to knowncurvature compared to predicate device.PASS;Certas PlusInline Smallvalve is moreflexible andCertas PlusRight Anglevalve isequivalent topredicatedevice.
AdjustmentandIndicationTestingVerify ability to use Codman CERTAS toolkits toadjust and indicate a valve.PASS

Validation Testing Simulated Post-Implantation Use

Codman conducted validation testing with clinicians which included indication and adjustment post-implantation using simulated skin thicknesses.

Table 3: Design Validation Activities – Codman Certas Plus Inline Small andRight Angle Programmable Valves
TestTest Method SummaryResults
SystemSafety StudyClinicians evaluated the maximum acceptable rate forfalse positives and maximum success rate for completing afull procedure. Testing conducted on a simulated headmodel with variable valve orientation.PASS

Test results demonstrated that the acceptance criteria were met; therefore, the Codman Certas Plus Programmable Inline Small and Right Angle Valves conform to the expected device performance and intended use. Results of the verification and validation testing have demonstrated that the proposed valves are substantially equivalent to the predicate Codman

{8}------------------------------------------------

Certas Plus Programmable Valve, and that the modifications do not impact the safety or effectiveness of the proposed valves.

Magnetic Resonance Imaging

The metallic components of the proposed Certas Plus Inline Small Valve and Certas Plus Right Angle Valve are the same as the current Certas Plus Programmable Valve. Since there were no changes to the metallic components within the valve assembly, there are no additional risks associated with the MRI safety of the proposed valves.

Biocompatibility Testing

The materials of the proposed Certas Plus Inline Small Valve and Certas Plus Right Angle Valve are the same as the current Certas Plus Programmable Valve and are sterilized using the same sterilization method as the predicate devices in the Certas Plus Programmable Valve product family.

Animal Studies

No animal studies were required as appropriate verification and validation of the design modifications were achieved based on the similarities of the proposed devices to the predicate device, and from result of bench testing.

Clinical Studies

No clinical studies were required as appropriate verification and validation of the design modifications were achieved based on the similarities of the proposed devices to the predicate device, and from result of bench testing.

Conclusion

Based upon the intended use, design, materials, function, comparison to currently marketed devices, and testing performed, it is concluded that the Codman Certas Plus Inline Small and Right Angle Valves are substantially equivalent to the predicate Codman Certas Programmable Valve (K152152) and, therefore, do not raise any new questions of safety and effectiveness.

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).