K Number
K182242

Validate with FDA (Live)

Date Cleared
2018-12-26

(128 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VO200 - NeurOs Cerebral Oximetry is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain tissue beneath the sensor. The prospective clinical value of data from the VO200 - NeurOs Cerebral Oximetry has not been demonstrated in disease states. The VO200 - NeurOs Cerebral Oximetry should not be used as the sole basis for diagnosis or therapy.

Device Description

The Mesepre VO 200 - NeurOs Cerebral Oximetry System is based on the same optical operating principles of multi-distance diffuse reflectance spectroscopy as the predicate device. It measures oxygenated and de-oxygenated hemoglobin content underneath the sensor probe and allows clinicians to monitor the blood oxygenation level of cerebral tissue. The VO 200 - NeurOs Cerebral Oximetry System consists of a reusable sensor, a single use disposable adhesive, and a display software. The sensor uses two wavelengths within the range from 660nm to 905nm to measure light absorption of tissue, with one wavelength is more sensitive to deoxygenated hemoglobin, the other is more sensitive to oxygenated hemoglobin. Regional cerebral tissue oxygen saturation is calculated through the ratio of tissue absorption to the two wavelengths and validated aqainst blood sample testing through the blood gas analyzer. Sensors are sized to provide desired depth and area coverage for the measurement. The sensor consists of LEDs, photo detectors, pre-amplifier, analogue to digital convertor, and a CUP for controlling the light emission and detection sequence, auto power adjustment for LED, gain adjustment for photo detectors to ensure optimized signal to noise ratio, and data processing and calculations. It also has an isolator between the monitor and the sensor to ensure the electric safety of the sensor. The enclosure of the sensor is made of medical grade ABS. The single use sensor adhesive is made of a soft foam with medical grade adhesives. The Mespere Display Software can be loaded to a tablet, laptop, or stationary PC. The sensor is then connected through a standard USB connector. It displays the measurement through the computer user interface at a refresh rate of approximately one second. Screenshots of the measurement can be saved to a user specified file and can be exported through USB ports or the internet.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the device meets the acceptance criteria for the VO200 - NeurOs Cerebral Oximetry System, based on the provided text:

Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Specification)Reported Device PerformanceNotes
StO2 Measurement Range0 - 100%0 - 100%
StO2 Validated Measurement Range40 - 80%40 - 80%Validation was performed on this specific measurement range.
Trending Accuracy of StO2 (40%-80%)± 1.5%± 1.5%Accuracy is based on a validation study comparing to venous jugular bulb and radial arterial blood samples and a Co-Oximeter on healthy volunteers.
Absolute Accuracy of StO2 (40%-80%)± 4.9%± 4.9%Accuracy is based on a validation study comparing to venous jugular bulb and radial arterial blood samples and a Co-Oximeter on healthy volunteers.
Resolution of StO21%1%

Study Details

  1. Sample Size used for the test set and the data provenance:

    • Sample Size: 12 healthy subjects. A total of 284 paired venous and arterial blood sample readings and 284 VO200 - NeurOs Cerebral Oximetry system readings were used for comparison and statistical analysis.
    • Data Provenance: Prospective clinical study conducted at the University of California San Francisco (UCSF) Hypoxia Labs, San Francisco, California, USA. The subjects were healthy, non-smoking individuals aged 21-49 years old.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    The document does not explicitly state the number or qualifications of experts involved in establishing the ground truth measurements. The ground truth was established through invasive blood samples analyzed by a Co-Oximeter, which implies highly trained medical and laboratory professionals were involved in the process, but their specific roles or qualifications are not detailed.

  3. Adjudication method for the test set:
    The document does not mention an explicit adjudication method (e.g., 2+1, 3+1). The ground truth was established by simultaneously drawing invasive blood samples (from jugular bulb and radial artery) and immediately analyzing them with a Co-Oximeter. This suggests a direct measurement approach rather than an expert consensus/adjudication process on interpretations.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No. This was not an MRMC comparative effectiveness study involving human readers and AI assistance. The study focused on validating the accuracy of the device against direct physiological measurements.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Yes, a standalone performance study was done for the device. The clinical study aimed to validate the accuracy of the VO200 - NeurOs Cerebral Oximetry system's measurements directly against invasive blood samples, without human interpretation of the device's output to determine the primary outcome.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The ground truth was established through invasive physiological measurements (venous jugular bulb and radial arterial blood samples) directly analyzed by a Co-Oximeter.

  7. The sample size for the training set:
    The document does not provide information about a separate training set. The descriptions focus on the clinical validation study (test set). For medical devices, particularly those based on established physiological principles like oximetry, the "training" (calibration) might be integrated into the device's design and manufacturing process rather than through a distinct, large-scale clinical training dataset as would be common for AI/ML algorithms.

  8. How the ground truth for the training set was established:
    As no separate training set is explicitly mentioned, the method for establishing ground truth for a training set is not provided. The accuracy and trending accuracy (which could be considered performance, not training data) values were derived from the validation study described above, where ground truth was established by Co-Oximeter analysis of invasive blood samples.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 26, 2018

Mespere LifeSciences Inc. Rebecca Li Manager of Quality Assurance and Regulatory Affairs 180 Frobisher Drive, Suite 1C Waterloo, N2V 2A2 Ca

Re: K182242

Trade/Device Name: VO200 - NeurOs Cerebral Oximetry System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: QEM Dated: August 10, 2018 Received: August 20, 2018

Dear Rebecca Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Krause =S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -05' -05' -05' -05' -05' -05' -05' -05' -05' -05' -05' -05' 00'

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182242

Device Name

VO200 - NeurOs Cerebral Oximetry System

Indications for Use (Describe)

The VO200 - NeurOs Cerebral Oximetry is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain tissue beneath the sensor. The prospective clinical value of data from the VO200 - NeurOs Cerebral Oximetry has not been demonstrated in disease states. The VO200 - NeurOs Cerebral Oximetry should not be used as the sole basis for diagnosis or therapy.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Appendix 5-1

Image /page/3/Picture/1 description: The image shows the logo for "Mespere Life Sciences". The logo features a large, stylized "M" in gray, with the rest of the company name spelled out in a similar font. Below "Mespere", the words "LIFE SCIENCES" are written in a smaller, blue font. To the left of the "M", there is a vertical blue bar.

180 Frobisher Drive, Suite 1C
Waterloo, ON,

T: 519-884-7575
F: 519-884-8810

K182242

www.mespere.com

510(k) SUMMARY

Submitted by:Mespere LifeSciences Inc.180 Frobisher Drive, Suite 1CWaterloo, Ontario, CanadaN2V 2A2Phone: 519-884-7575Fax: 519-884-8810
Company Contact:Rebecca S. Li, Manager of Quality Assurance and Regulatory Affairs
Date Summaryprepared:8/30/2018
Trade Name:VO200 - NeurOs Cerebral Oximetry System
Common Name:NeurOs Cerebral Oximetry
Classificationname:Oximeter, Tissue Saturation (21CFR 870.2700)
Classification:Class II Medical Device
Product Code:QEM
Predicate Device:K082327INVOS 5100C Cerebral/Somatic Oximeter SystemProduct Code: QEMRegulation Number: 870.2700This predicate has not been subject to a design related recall

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Mespere Life Sciences. Next to the logo is the text "VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo is a stylized "M" with the words "LIFE SCIENCES" below it.

Intended Use/Indications For Use:

The VO200 – NeurOs Cerebral Oximetry is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain tissue beneath the sensor. The prospective clinical value of data from the VO200 - NeurOs Cerebral Oximetry has not been demonstrated in disease states. The VO200 - NeurOs Cerebral Oximetry should not be used as the sole basis for diagnosis or therapy.

Device Description:

The Mesepre VO 200 - NeurOs Cerebral Oximetry System is based on the same optical operating principles of multi-distance diffuse reflectance spectroscopy as the predicate device. It measures oxygenated and de-oxygenated hemoglobin content underneath the sensor probe and allows clinicians to monitor the blood oxygenation level of cerebral tissue.

The VO 200 - NeurOs Cerebral Oximetry System consists of a reusable sensor, a single use disposable adhesive, and a display software. The sensor uses two wavelengths within the range from 660nm to 905nm to measure light absorption of tissue, with one wavelength is more sensitive to deoxygenated hemoglobin, the other is more sensitive to oxygenated hemoglobin. Regional cerebral tissue oxygen saturation is calculated through the ratio of tissue absorption to the two wavelengths and validated aqainst blood sample testing through the blood gas analyzer. Sensors are sized to provide desired depth and area coverage for the measurement.

The sensor consists of LEDs, photo detectors, pre-amplifier, analogue to digital convertor, and a CUP for controlling the light emission and detection sequence, auto power adjustment for LED, gain adjustment for photo detectors to ensure optimized signal to noise ratio, and data processing and calculations. It also has an isolator between the monitor and the sensor to ensure the electric safety of the sensor. The enclosure of the sensor is made of medical grade ABS.

The single use sensor adhesive is made of a soft foam with medical grade adhesives.

The Mespere Display Software can be loaded to a tablet, laptop, or stationary PC. The sensor is then connected through a standard USB connector. It displays the measurement through the computer user interface at a refresh rate of approximately one second. Screenshots of the measurement can be saved to a user specified file and can be exported through USB ports or the internet.

Technology Characteristics:

The VO200 - NeurOs Cerebral Oximetry System has the same technology characteristics as the predicate.

Similarities:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Mespere Life Sciences. The text next to the logo reads "VO200 - NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo is a stylized "M" with the words "LIFE SCIENCES" written below it.

  • Both are based upon multi-distance, diffuse reflectance, near infrared spectroscopy . for analyzing oxygenated and deoxygenated hemoglobin content of blood within cerebral tissue.
  • Both use near infrared light emitters (LED) and photo detectors. .
  • Both uses two near infrared wavelengths ●

Differences:

  • . NeurOs sensor probe are reusable, only the medical grade sensor adhesives are one time disposables. The predicate sensor probes are single-use disposables.
  • . For the predicate sensor, analogue to digital conversion and pre-Amp is made through boxes connected to the probe through cables.

Comparison Table:

CharacteristicsMespere VO200 - NeurOsCerebral Oximetry System(Subject Device)INVOS 5100CCerebral/Somatic OximeterSystem(Predicate)
Intended Use /Indications For UseNeurOs Cerebral Oximetry Systemis intended for use as an adjunctmonitor of regional hemoglobinoxygen saturation of blood in thebrain tissue beneath the sensor.INVOS 5100C is intended for use asan adjunct monitor of regionalhemoglobin oxygen saturation ofblood in the brain or in other tissuebeneath the sensor.
Target PopulationFor adult patients.For adult, pediatric, and neonatepatients.
Anatomical SitesCerebralCerebral
Where used(hospital, home,ambulance, etc.)Hospital, clinic and other healthcare facilitiesHospital, clinic and other health carefacilities
Technologynear infrared, multi-distance diffusereflectance spectroscopy, lightemitters and photo detectorsnear infrared, multi-distance diffusereflectance spectroscopy, lightemitters and photo detectors
SterilityNon-sterileNon-Sterile
CalibrationNo needNo Need
Features/Method ofUseNon-invasiveNon-Invasive
MonitoredparametersRegional Oxygen SaturationRegional Oxygen Saturation
Components- Sensor- Sensor cables- display software- adhesive patch- Sensor- Sensor cables- Preamplifiers- Monitor
DisplayReal-time, continuousReal-time, continuous
RisksNo compromise to patient careNo compromise to patient care
OpticalYes- LEDs and photodetectorsYes- LED and photodetector

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Mespere Life Sciences. The text next to the logo reads "VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo is a stylized "M" in blue and gray.

CharacteristicsMespere VO200 - NeurOsCerebral Oximetry System(Subject Device)INVOS 5100CCerebral/Somatic OximeterSystem(Predicate)
Embedded SoftwareYesYes
Integrated software providesanalytics for measurements to bedisplayedIntegrated software providesanalytics for measurements to bedisplayed
DisplayDisplay StO2 value (right and left)and trend data graphsDisplay rSO2 values (right and left)and trend data graphs
Probe AttachmentAdhesive to skinAdhesive to skin
AlarmNoYes
Device TypeNeeds to be connected to acomputerStandalone Device

Complied Standards:

The design of the Mespere VO200 - NeurOs Cerebral Oximetry system conforms to the following voluntary standards:

  • EN ISO 13485:2016 ●
  • EN ISO 14971:2012 ●
  • EN 60601-1:2006/A1:2013 ●
  • EN 60601-1-2:2007/AC:2010 ●
  • EN 60601-1-6:2010 ●
  • EN 62366:2008 ●
  • EN ISO 14155:2011 ●
  • EN 62304:2006/AC:2008
  • EN 1041:2008 ●
  • EN 980:2008
  • EN 15223-1:2012 ●
  • EN ISO 10993-1:2009 ●
  • EN ISO 10993-5:2009 ●
  • EN ISO 10993-10:2013 .

Non-Clinical Tests:

The VO200 - NeurOs Cerebral Oximetry System was subjected to extensive safety, performance, and verification testing. Extensive benchtop tests verified the performance of the device to meet the design requirements in an ideal non-clinical environment. All tests were performed to ensure that the device complies with industry and safety standards.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Mespere Life Sciences, followed by the text "VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo consists of the word "Mespere" in a combination of blue and gray, with the words "LIFE SCIENCES" in smaller font below. The text indicates that the VO200 is a cerebral oximetry system that has received traditional 510(k) notification.

Clinical Tests:

Clinical tests were performed to verify and validate the clinical performance of the VO200 -NeurOs Cerebral Oximetry system and support the determination of substantial equivalence. The following clinical data was submitted:

Clinical Validation: The primary objective of this study was to validate the accuracy of the Mespere VO200 – NeurOs Cerebral Oximetry system for measuring the regional cerebral tissue blood oxygenation saturation (StO2), by comparing it to invasive blood samples drawn from the jugular bulb and radial artery, and compared by a co-oximeter. The subjects recruited for this study were healthy volunteers.

Methods: With IRB approval and written Informed Consent Forms, healthy volunteers were recruited. An ultrasound-quided small diameter internal jugular vein catheter was inserted and placed at the position of the subject's superior jugular bulb. A radial arterial line was inserted into a radial artery on the left arm. Two standard pulse oximeters were attached to the patient's fingers to help guide the hypoxia state plateau levels. The VO200 -NeurOs Cerebral Oximetry sensor was placed on the forehead. Blood gas analysis to determine oxyhemoglobin saturation was performed using a Co-Oximeter. Each subject had control data taken at the beginning of each experiment, with two control blood samples drawn while breathing room air. Hypoxia was induced to different levels of oxyhemoglobin saturation (between 70-100% SpO2) by having subjects breathe mixtures of nitrogen, room air, and carbon dioxide. Once a steady state level of hypoxia is achieved, a 1.0ml sample of blood are obtained simultaneously from the jugular line and radial arterial line. A second blood pair of samples, at the same steady-state saturation, is taken 30 seconds later. The VO200 - NeurOs Cerebral Oximetry samples were recorded simultaneously to the blood samples. The blood samples were immediately analyzed by the Co-Oximeter. Up to 27 paired samples were obtained on the plateaus across this span for each subject.

Results: The results showed the absolute accuracy of ± 4.9% and trending accuracy of ±1.5% (in range of 40%-80%), which is equivalent or better than the reported absolute accuracy of 5.0% and trending accuracy of ± 4.0 % of the predicate device.

No adverse events were reported.

ParameterSpecificationNotes
StO2 Measurement Range0 - 100%
StO2 Validated Measurement Range40 - 80%1,2,3
Trending Accuracy of StO2(40%-80%)$\pm$ 1.5%1,2,3
Absolute Accuracy of StO2$\pm$ 4.9%1,2,3

Performance specifications

Appendix 5-1

{8}------------------------------------------------

O200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification

(40%-80%)
Resolution of StO21%

NOTES:

  • 1 Validation was performed on a measurement range of 40% - 80%
  • 2 Accuracy is based on a validation study comparison with venous jugular bulb and radial arterial blood samples and a Co-Oximeter on healthy volunteers. In the StO2 range of 40-80%, the trending accuracy was ± 1.5% and the absolute accuracy was ± 4.9% in the study STP-9200001 at the University of California San Francisco, UCSF (San Francisco, California, USA).

3 Summary of validation study STP-9200001 at the University of California San Francisco, UCSF (San Francisco, California, USA)

Number of Study Sites:

Primary Clinical Site:

University of San Francisco California (UCSF)

Hypoxia Labs

513 Parnassus Ave.,

San Francisco, CA, 94143-0542

Subjects/data points:

The subjects recruited for this study consisted of healthy, non-smoking individuals of age 21-49 years' old who underwent hypoxia testing at the University of California at San Francisco (UCSF) located in San Francisco, California. No subjects was anemic (Hemoglobin ≤ gm·dl-1).

Twelve (12) healthy subjects were recruited for the study. A total of 284 paired venous and arterial blood sample readings, and 284 VO200 - NeurOs Cerebral Oximetry system readings were used to perform the comparison and statistical analysis.

Patient Demographics:

SubjectGenderAgeSkin ToneFitzpatrick SkinTone
Subject #1Male26OliveIV
Subject #2Female25Medium/lightoliveIII
Subject #3Male26Medium/lightIII

Appendix 5-1

{9}------------------------------------------------

VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification

olive
Subject #4Female33FairII
Subject #5Male24BrownV
Subject #6Female32Medium/lightoliveIII
Subject #7Male23BrownV
Subject #8Female23FairII
Subject #9Male24BrownV
Subject #10Male30FairII
Subject #11Male31FairII
Subject #12Male26BrownV

Agreement Analysis Results:

The calculated statistical results from the study are as follows:

Trending Accuracy± 1.5%
Absolute Accuracy± 4.9%

Bland Altman Plot of Absolute Measurement

Image /page/9/Figure/6 description: This scatter plot shows the difference between Mespere readings and blood gas readings in percentage on the y-axis versus blood gas readings in percentage on the x-axis. The plot includes data points representing the difference, along with three horizontal lines indicating the lower limit of agreement (LoA) at approximately -10, the upper LoA at around 10, and the mean difference near 0. The blood gas readings range from 40% to 85%, while the difference between readings varies from -15% to 15%.

Linear Regression Analysis of Absolute Measurement with 95% Confidence Interval of Slope and

Appendix 5-1

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image displays the logo for Mespere Life Sciences, followed by the text "VO200 - NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo features a stylized "M" in blue and gray, with the words "LIFE SCIENCES" in smaller font below. The text describes the VO200 NeurOs Cerebral Oximetry System and indicates that it is a Traditional 510(k) Notification.

Intercept using Bootstrap Method

Slope = 0.7346

Intercept = 15.161

Image /page/10/Figure/4 description: The image is a scatter plot comparing Mespere readings to blood gas readings. The x-axis represents blood gas readings in percentage, ranging from 35 to 85. The y-axis represents Mespere readings in percentage, also ranging from 35 to 85. The plot includes data points representing Mespere readings, a line representing y=x, and a linear regression line for the Mespere readings.

Bland Altman of Absolute Trending Measurement

Image /page/10/Figure/6 description: This image is a scatter plot comparing Mespere Trend Readings to Blood Gas Trend Readings. The x-axis represents Blood Gas Trend Readings (%), ranging from approximately -15 to 25. The y-axis represents the difference between Mespere Trend Readings and Blood Gas Trend Readings (%), ranging from -9 to 5. The plot includes a cluster of data points, labeled as 'Difference', along with horizontal lines indicating 'Lower LoA' at approximately -3, 'Upper LoA' at approximately 3, and 'Mean' at approximately 0.

Linear Regression Analysis of Absolute Trending Measurement with 95% Confidence Interval of Slope and Intercept using Bootstrap Method

{11}------------------------------------------------

VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification

Image /page/11/Figure/1 description: This image is a scatter plot comparing Mespere Trend Readings (%) to Blood Gas Trend Readings (%). The x-axis represents Blood Gas Trend Readings (%), ranging from -15 to 30, while the y-axis represents Mespere Trend Readings (%), also ranging from -15 to 30. The plot includes a scatter of blue points representing Mespere Readings, an orange line representing y=x, and a dotted blue line representing a linear trendline with the equation y = 0.9497x + 0.0004.

Conclusions:

The clinical data demonstrates that the VO200 – NeurOs Cerebral Oximetry system is substantially equivalent in performance to the predicate device when measuring regional cerebral tissue blood oxygenation saturation.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).