K Number
K182232

Validate with FDA (Live)

Date Cleared
2018-09-11

(25 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Octane Mechanical Thrombectomy System is intended for the removal/aspiration of embolic material (thrombus/ debris) from vessels of the arterial and deep venous system, and to infuse/deliver diagnostic or therapeutic agents.

Device Description

The Octane Mechanical Thrombectomy System consists of two components: the Octane mechanical thrombectomy catheter and the Octane hand pump. The Octane mechanical thrombectomy catheter has a working length of 115cm, is compatible with introducer sheaths ≥ 8F and it is intended for use in vessels ≥ 3 mm in diameter. The catheter has a rapid exchange design and includes sliding, co-axial inner and outer lumens. The rapid exchange guidewire lumen is compatible with guidewires ≤ 0.018" in diameter and the 6F inner catheter lumen is compatible with guidewires ≤ 0.038" in diameter. The catheter inner lumen tip has three side wall slot ports and the catheter's outer lumen has a beveled radiopaque tip. To alleviate plugging of the catheter tip caused by thrombus, the actuator on the proximal end of the catheter slides the catheter's outer lumen back and forth over the inner lumen distal tip. The actuator on the proximal end of the catheter is also used to control the directionality of the torque-able catheter. Directionality is further achieved through the gradual curve in the shaft materials at the distal end. The Octane hand pump provides suction to the inner lumen of the Octane mechanical thrombectomy catheter and is manually operated using a spring-loaded handle. It consists of a pump handle fitted with a 30ml piston syringe, tubing, a series of one-way valves, and a 500ml collection bag. The hand pump tubing has a standard luer fitting for connection to the Octane mechanical thrombectomy catheter hub. The Octane Mechanical Thrombectomy System has been sterilized with ethylene oxide.

AI/ML Overview

The provided text describes the Octane Mechanical Thrombectomy System (K182232) and compares it to a predicate device, the Octane Aspiration System (K173266). The submission claims substantial equivalence based on the device's design, intended use, materials, and operating mechanisms.

However, the document does not include a table of acceptance criteria or details of a study that proves the device meets specific acceptance criteria in the manner usually associated with clinical performance or algorithm-based devices. Instead, it focuses on bench testing to demonstrate substantial equivalence to a predicate device following a labeling change.

Therefore, for aspects related to specific acceptance criteria, statistical performance, and ground truth establishment, the information is not present in the provided text.

Here's a breakdown of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

Not explicitly provided in the document.

The document states: "The results of the verification test met the specified acceptance criteria and did not raise different questions of safety and effectiveness." However, the specific acceptance criteria themselves (e.g., specific metrics like effectiveness rates, aspiration volume, etc.) and the reported device performance against those criteria are not detailed in the text. The evaluation method described is "bench tests" to evaluate technological differences and "concomitant device compatibility tests" to support labeling modifications.

2. Sample Size Used for the Test Set and Data Provenance

Not applicable/Not provided.

This document describes bench testing, not a clinical trial with a "test set" in the context of patient data. The "sample size" would refer to the number of devices or test conditions, which is not specified. The concept of "data provenance" (country of origin, retrospective/prospective) is not relevant for bench testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable/Not provided.

As this was bench testing for substantial equivalence, the concept of establishing ground truth by medical experts for a "test set" (e.g., imaging studies) is not relevant. The "ground truth" would be engineering specifications and measurements, overseen by engineers and quality assurance personnel.

4. Adjudication Method for the Test Set

Not applicable/Not provided.

Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies or studies involving expert consensus on medical data. This information is not relevant for the bench testing described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No.

The document does not mention an MRMC study. The study described is bench testing for substantial equivalence to a predicate device, not a comparative effectiveness study involving human readers and AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable.

This device is a mechanical thrombectomy system, not an AI algorithm. Therefore, the concept of "standalone performance" for an algorithm is not relevant.

7. The Type of Ground Truth Used

Engineering Specifications and Bench Test Measurements.

For bench testing of a medical device, the "ground truth" refers to established engineering specifications, mechanical properties, and performance benchmarks measured in a controlled environment. The document states "The results of the verification test met the specified acceptance criteria," indicating that the device's performance was measured against predetermined engineering and design standards.

8. The Sample Size for the Training Set

Not applicable/Not provided.

This refers to bench testing for device performance and substantial equivalence, not the training of an AI algorithm. Therefore, a "training set" in this context is not applicable.

9. How the Ground Truth for the Training Set Was Established

Not applicable/Not provided.

As there is no mention of a "training set" for an algorithm, this question is not relevant. The "ground truth" for the device's performance relies on engineering design specifications and test methods.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

September 11, 2018

Vascular Solutions, Inc. Iroquois Ledbeter Regulatory Product Specialist 6401 Sycamore Ct N Maple Grove, Minnesota 55369

Re: K182232

Trade/Device Name: Octane Mechanical Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE Dated: August 16, 2018 Received: August 17, 2018

Dear Iroquois Ledbeter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Gregory O'Connell 2018.09.11 16:02:05 -04'00' For

  • Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K182232

Device Name

Octane Mechanical Thrombectomy System

Indications for Use (Describe)

The Octane Mechanical Thrombectomy System is intended for the removal/aspiration of embolic material (thrombus/ debris) from vessels of the arterial and deep venous system, and to infuse/deliver diagnostic or therapeutic agents.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Vascular Solutions. The logo features a tan square with a black pattern of squares above it that fades out as it rises. Below the square is the text "Vascular" in a large, bold, serif font, with the word "SOLUTIONS" in a smaller, sans-serif font underneath.

510(k) Summary

[As required by 21 CFR 807.92]

K182232 510(k) Number:

SUBMITTER AND DEVICE

Submitter:Vascular Solutions LLC6464 Sycamore Court NorthMinneapolis, MN 55369 USA
Establishment Registration:2134812
Phone:763-656-4300
Fax:763-656-4253
Contact Person:Iroquois Ledbeter,Regulatory Product Specialist
Date Prepared:August 16, 2018
Name of Device:Octane MechanicalThrombectomy System
Common or Usual Name:MechanicalThrombectomy System
Classification Name:DXE – Catheter, Embolectomy,Cardiovascular (21 CFR 870.5150)
Regulatory Class:II
Product Code:DXE

PREDICATE DEVICE

The legally marketed device to which substantial equivalence is claimed is:

Vascular Solutions Inc., Octane Aspiration System, K173266 cleared December 1, 2017.

DEVICE DESCRIPTION

The Octane Mechanical Thrombectomy System consists of two components: the Octane mechanical thrombectomy catheter and the Octane hand pump. The Octane mechanical thrombectomy catheter has a working length of 115cm, is compatible with introducer sheaths ≥ 8F and it is intended for use in vessels ≥ 3 mm in diameter. The catheter has a rapid exchange design and includes sliding, co-axial inner and outer lumens. The rapid exchange guidewire lumen is compatible with guidewires ≤ 0.018" in diameter and the 6F inner catheter lumen is compatible with guidewires ≤ 0.038" in diameter. The catheter inner lumen tip has three side

{4}------------------------------------------------

wall slot ports and the catheter's outer lumen has a beveled radiopaque tip. To alleviate plugging of the catheter tip caused by thrombus, the actuator on the proximal end of the catheter slides the catheter's outer lumen back and forth over the inner lumen distal tip. The actuator on the proximal end of the catheter is also used to control the directionality of the torque-able catheter. Directionality is further achieved through the gradual curve in the shaft materials at the distal end. The Octane hand pump provides suction to the inner lumen of the Octane mechanical thrombectomy catheter and is manually operated using a spring-loaded handle. It consists of a pump handle fitted with a 30ml piston syringe, tubing, a series of one-way valves, and a 500ml collection bag. The hand pump tubing has a standard luer fitting for connection to the Octane mechanical thrombectomy catheter hub. The Octane Mechanical Thrombectomy System has been sterilized with ethylene oxide.

INDICATIONS FOR USE

The Octane Mechanical Thrombectomy System is intended for the removal/aspiration of embolic material (thrombus/debris) from vessels of the arterial and deep venous system, and to infuse/deliver diagnostic or therapeutic agents.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Octane Mechanical Thrombectomy System is identical in design and intended use to the predicate device, the Octane Aspiration System. Both are aspiration catheter systems intended for the removal of embolic material (thrombus/debris) from vessels of the arterial and deep venous system and to infuse/deliver diagnostic or therapeutic agents. The subject and predicate device are identical in size and working length. The materials used in the subject device are identical to the materials in the predicate device. The operating mechanisms of the Octane Mechanical Thrombectomy System are identical to those of the predicate. The only difference between the subject and predicate device is labeling. The Octane Mechanical Thrombectomy System instructions for use includes an optional, alternate, technique for the deployment and removal of the Octane Mechanical Thrombectomy Catheter using the central extraction lumen of the device. This alternate technique allows for the catheter to be delivered and removed over guidewires ≤0.038" which provides the physician a benefit for ease of use.

PERFORMANCE DATA

There is no change or intended use or fundamental scientific technology between the subject and predicate devices. The following performance data were provided in support of the substantial equivalence determination.

Performance Testing - Bench

The technological differences between the subject and predicate device have been evaluated through bench tests to provide evidence that the subject device is substantially equivalent to the predicate device. The modifications to the device labeling are supported through concomitant device compatibility tests.

{5}------------------------------------------------

The results of the verification test met the specified acceptance criteria and did not raise different questions of safety and effectiveness.

CONCLUSION

The subject Octane Mechanical Thrombectomy System is equivalent to the Octane Aspiration System predicate devices based on comparisons of device functionality, technological characteristics, and indications for use. The results of design verification tests do not raise new or different questions of safety and effectiveness; therefore, the Octane Mechanical Thrombectomy System is substantially equivalent to the predicate device.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).