Intended Use

The Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse), and Medline Level 4 Surgical Gown (Aurora) is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse), and Medline Level 4 Surgical Gown (Aurora) meets the Level 4 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Device Description

The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) is a Class II medical device regulated under FDA product code FYA, General & Plastic Surgery Panel, and regulation 21 CFR 878.4040. The device description of the Medline Level 4 Surgical Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993.

The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) is a single-use, sterile, disposable medical device that is provided in a variety of sizes and styles. The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) constructed from nonwoven SMS (spubound, meltblown, spunbond), a poly-fabric reinforcement on the chest front panel and on both sleeves. Each gown is designed with a hook and loop closure at the neck, ties at the waist, knitted cuffs at the end of the sleeves as well as a transfer tab for ease in donning. The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) has been tested according to ANSI/AAMI PB70:2012 and meets AAMI Level 4 barrier level protection for a surgical gown.

AI/ML Overview

The provided text is a 510(k) Premarket Notification from Medline Industries, Inc. for their Level 4 Surgical Gowns. This document describes the non-clinical testing performed to demonstrate substantial equivalence to a predicate device, rather than the acceptance criteria and study proving a device meets acceptance criteria in a clinical setting with human interaction or AI.

Therefore, I cannot extract the requested information regarding:

  • A table of acceptance criteria and reported device performance in a clinical or AI context.
  • Sample size used for the test set and data provenance (since it's non-clinical, likely material samples, not patient data).
  • Number of experts used to establish ground truth or their qualifications.
  • Adjudication method.
  • MRMC comparative effectiveness study.
  • Standalone (algorithm only) performance.
  • Type of ground truth (as it relates to clinical diagnosis or outcomes).
  • Sample size for training set.
  • How ground truth for the training set was established.

However, I can provide information based on the non-clinical testing detailed in the document, framed as demonstrating "acceptance criteria" for material performance:


Based on the provided Medline Industries, Inc. 510(k) Premarket Notification for the Medline Level 4 Surgical Gowns (Sirus, Eclipse, Aurora), the "acceptance criteria" and the "study that proves the device meets the acceptance criteria" refer to non-clinical performance testing of the gown's material properties and barrier effectiveness. This document does not pertain to the clinical performance of a diagnostic device or an AI/human-in-the-loop system.

1. A table of acceptance criteria and the reported device performance (for non-clinical material testing):

The acceptance criteria are generally implied by meeting the specified industry standards and guidance documents. The "reported device performance" is that the device meets these requirements. No specific numerical performance values are explicitly tabulated against acceptance criteria in the provided text, but the document states the gown "meets the Level 4 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities."

Acceptance Criteria CategoryStandard/Test Method (Implied Criteria)Reported Device Performance
Liquid Barrier Performance (Primary)ANSI/AAMI PB70:2012 Level 4Meets Level 4 requirements
Material Strength (Breaking)ASTM D5034-09 (2013)Tested and presumably met
Material Strength (Tearing)ASTM D5587-15Tested and presumably met
Material WeightASTM D3776/D3776M-09aTested and presumably met
Material Strength (Bursting)ASTM D3786/D3786M-13Tested and presumably met
Flammability16 CFR 1610Meets requirements of Flame Resistant CPSC 1610 Class 1
Seal LeaksASTM F1929-15Tested and presumably met (Detecting Seal Leaks by Dye Penetration)
Blood-Borne Pathogen PenetrationASTM F1671Tested and presumably met (Resistance of Materials to Penetration by Blood-Borne Pathogens)
Water ResistanceAATCC 42Tested and presumably met (Water Resistance: Impact Penetration Test)
Biocompatibility (Cytotoxicity)ISO 10993-5Met requirements
Biocompatibility (Irritation)ISO 10993-10Met requirements
Biocompatibility (Sensitization)ISO 10993-10Met requirements
Sterilization ResidualsANSI/AAMI/ISO 10993-7:2008(R)2012Tested and presumably met (Ethylene oxide sterilization residuals)

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the exact number of gown samples or material specimens used for each non-clinical test (e.g., how many gowns were tested for barrier performance or how many material swatches for tensile strength).
  • Data Provenance: The tests are explicitly stated as "Non-Clinical Testing" and appear to be laboratory-based tests of the physical gown materials and design. No country of origin for data is stated, implying the testing was performed by or for the manufacturer according to international standards. The data is retrospective in the sense that it was gathered from completed tests to support the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This is not applicable as the "ground truth" for this device is established by objective engineering and material performance standards (e.g., a material either resists liquid penetration to a certain level as defined by a standard, or it doesn't). It does not involve human expert interpretation of clinical data or images.

4. Adjudication method for the test set:

  • Not applicable. The tests are quantitative and objective measurements against predefined standards, not subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a surgical gown, not an AI-assisted diagnostic or clinical decision support system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an algorithm or AI device. The "standalone" performance refers to the physical gown's barrier and material properties.

7. The type of ground truth used:

  • The "ground truth" for this device is based on objective engineering standards and test methods (e.g., ANSI/AAMI PB70:2012, ASTM standards, ISO standards). It relates to the physical and barrier properties of the surgical gown material, not clinical outcomes or expert consensus on a diagnosis.

8. The sample size for the training set:

  • Not applicable. This device does not involve machine learning or a "training set."

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 24, 2019

Medline Industries, Inc. Pauline Maralit Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60093

Re: K182172

Trade/Device Name: Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse), Medline Level 4 Surgical Gown (Aurora) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: March 26, 2019 Received: March 28, 2019

Dear Pauline Maralit:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182172

Device Name

Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse), Medline Level 4 Surgical Gown (Aurora),

Indications for Use (Describe)

The Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse), and Medline Level 4 Surgical Gown (Aurora) is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse), and Medline Level 4 Surgical Gown (Aurora) meets the Level 4 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, block letters, and a white abstract star-like shape is behind the text. The star shape has four points, with the top and bottom points being longer than the side points.

Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093

Medline Level 4 Surgical Gown 510(k) Premarket Submission All information on this page is confidential

510(k) SUMMARY - K182172

Submitter / 510(k) Sponsor

Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093 Registration Number: 1417592

Submission Contact

Pauline Maralit Regulatory Affairs Specialist Phone: 847-949-2283 Email: pmaralit(@medline.com

Summary Preparation Date August 9, 2018

Type of 510(k) Submission Traditional

Device Name / Classification

Regulation Name:Surgical Apparel
Regulatory Panel:General & Plastic Surgery
Regulation #:21 CFR 878.4040
Common Name:Surgical Gown
Classification Name:Surgical Gown
Proprietary Name:Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse),Medline Level 4 Surgical Gown (Aurora).
Product Code:FYA
Classification Panel:II

Predicate Device

PREDICATE DEVICE:

  • Kimberly Clark (Halyard) ULTRA Fluid Barrier Surgical Gown 트
    PREDICATE DEVICE K NUMBER:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Medline logo. The logo features the word "MEDLINE" in white, bold, sans-serif font, stacked on top of a white abstract star-like shape. The background is a solid dark blue. A gray bar is on the right side of the logo.

Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093

  • K080795 (Kimberly-Clark Corp./Halyard) ■

Device Description

The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) is a Class II medical device regulated under FDA product code FYA, General & Plastic Surgery Panel, and regulation 21 CFR 878.4040. The device description of the Medline Level 4 Surgical Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993.

The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) is a single-use, sterile, disposable medical device that is provided in a variety of sizes and styles. The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) constructed from nonwoven SMS (spubound, meltblown, spunbond), a poly-fabric reinforcement on the chest front panel and on both sleeves. Each gown is designed with a hook and loop closure at the neck, ties at the waist, knitted cuffs at the end of the sleeves as well as a transfer tab for ease in donning. The Medline Level 4 Surgical Gown (Sirus, Eclipse, and Aurora) has been tested according to ANSI/AAMI PB70:2012 and meets AAMI Level 4 barrier level protection for a surgical gown.

Item NumberDevice DescriptionSleeve StyleSize
DYNJP2707Aurora Surgical Gown, Poly Reinforced,Set-inLarge
DYNJP2708Aurora Surgical Gown, Poly Reinforced,Set-inX-Large
DYNJP2709Aurora Surgical Gown, Poly Reinforced,Set-inXX-Large
DYNJP2724Aurora Surgical Gown, Poly Reinforced,Set-inLarge, X-Long
DYNJP2725Aurora Surgical Gown, Poly Reinforced,Set-inX-Large, X-Long
DYNJP2726Aurora Surgical Gown, Poly Reinforced,Set-inXX-Large, X-Long
DYNJP2807Aurora Surgical Gown, Poly Reinforced,RaglanLarge
DYNJP2808Aurora Surgical Gown, Poly Reinforced,RaglanX-Large
DYNJP2809Aurora Surgical Gown, Poly Reinforced,RaglanXX-Large
DYNJP220Eclipse Surgical Gown, Poly ReinforcedSet-inLarge
DYNJP2202Eclipse Surgical Gown, Poly ReinforcedSet-inX-Large
DYNJP2203Eclipse Surgical Gown, Poly ReinforcedSet-inXX-Large
DYNJP2204Eclipse Surgical Gown, Poly ReinforcedSet-inLarge, X-Long
DYNJP2205Eclipse Surgical Gown, Poly ReinforcedSet-inX-Large, X-Long
DYNJP2206Eclipse Surgical Gown, Poly ReinforcedSet-inXX-Large, X-Long
DYNJP2201SSirus Surgical Gown, Poly ReinforcedSet-inLarge
DYNJP2202SSirus Surgical Gown, Poly ReinforcedSet-inX-Large
DYNJP2203SSirus Surgical Gown, Poly ReinforcedSet-inXX-Large

Table 1: Device Models

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, with a stylized white starburst symbol placed behind and slightly above the text. The starburst has four points, with the top and bottom points being longer than the side points. There is a gray bar on the right side of the image.

Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093

Medline Level 4 Surgical Gown 510(k) Premarket Submission All information on this page is confidential

DYNJP2204SSirus Surgical Gown, Poly ReinforcedSet-inLarge, X-Long
DYNJP2205SSirus Surgical Gown, Poly ReinforcedSet-inX-Large, X-Long
DYNJP2206SSirus Surgical Gown, Poly ReinforcedSet-inXX-Large, X-Long
DYNJP2219SLSirus Surgical Gown, Poly ReinforcedSet-inXXX-Large, X-Long
DYNJP2601Sirus Surgical Gown, Poly ReinforcedRaglanLarge
DYNJP2602Sirus Surgical Gown, Poly ReinforcedRaglanX-Large
DYNJP2603Sirus Surgical Gown, Poly ReinforcedRaglanXX-Large

Indications for Use

The Medline Level 4 Surgical Gown (Sirus). Medline Level 4 Surgical Gown (Eclipse), and Medline Level 4 Surgical Gown (Aurora) is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Level 4 Surgical Gown (Sirus), Medline Level 4 Surgical Gown (Eclipse), and Medline Level 4 Surgical Gown (Aurora) meets the Level 4 requirements of ANSI/AAMIPB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.

Technological Characteristics

The final proposed finished device was tested in accordance with the Guidance on Premarket Notification [510(k)] Submission for Surgical Gowns and Surgical Drapes- August 1993, Guidance for Industry and FDA Staff - Premarket Notification Requirements Concerning Gowns Intended for Use in Healthcare Settings issued by the FDA- December 9, 2015, and ANSI/AAMIPB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities. Testing was performed on the body and sleeve (equivalent design/technology), sleeve seam, and tie attachment to generate the test data shown in the comparison Table 2. Additional performance testing was conducted to further characterize the device.

Element of ComparisonSubject DevicePredicate DeviceComparisonAnalysis
Product NameMedline Level 4 Surgical Gown(Sirus), Medline Level 4 SurgicalGown (Eclipse), Medline Level 4Surgical Gown (Aurora)ULTRA Fluid Barrier SurgicalGownDifferent
510(k) ReferenceK182172K080795Different
Product OwnerMedline Industries, Inc.Kimberly-Clark Corp. (Halyard)Different
Product CodeFYAFYASame
Intended UseThe Medline Level 4 SurgicalGown is a sterile, single useThe Kimberly-Clark* ULTRASurgical Gown and ULTRA FilmSimilar
surgical apparel intended to beReinforced Surgical Gown are
worn by healthcare professionals tosterile, single use surgical gowns
help protect both the patient andintended to protect surgical patients
the healthcare worker from theand operating room personnel from
transfer of microorganisms, bodythe transfer of
fluids, and particulate matter. Themicroorganisms, body fluids, and
Medline Level 4 Surgical Gownparticulate material. The ULTRA
meets the Level 4 requirements ofSurgical Gown meets Level 3 of the
ANSI/AAMIPB70:2012 LiquidAAMI Liquid Barrier
barrier performance andclassifications, and the ULTRA
classification of protective apparelFilm-Reinforced Surgical Gown
and drapes intended for use inmeets Level 4 of the AAMI Liquid
healthcare facilities.Barrier classifications.
Regulation Number
21 CFR 878.404021 CFR 878.4040Same
Gown ColorBlueBlue
Same
Design ConfigurationsNeck Closure: Hook and LoopHoop & Neck Closures
Belt TiesTie Waist Closure
Knit CuffsCuffsSimilar
Reinforced MaterialFilm Reinforcement
Raglan or Set-in/Standard SleevesRaglan Sleeves
Material CompositionNonwoven SMS polypropyleneNonwoven SMS polypropylene
Anti-Static, RepellencyAnti-Static, Repellency
Poly-Reinforcement FilmFilm ReinforcementSame
White Knitted CuffsCuffs
Prescription vs. OTCOTCOTCSame
Contact DurationsSurface, Intact, < 24 hoursSurface, Intact, < 24 hoursSame
Sterile vs. Non-Sterile
SterileSterileSame
Disposable vs. Non-Disposable
DisposableDisposableSame
Single Use vs. ReusableSingleSingleSame
Performance Specifications:Level 4 barrier protectionLevel 4 barrier protectionSame
BiocompatibilityMet requirements per:Complied with 10993 requirements
ISO 10993-5 Cytoxticityfor surface devices with limited
ISO 10993-10 Irritationcontact with breeched orSame
ISO 10993-10 Sensitizationcompromised surfaces
FlammabilityMeets requirements of FlameResistant CPSC 1610 Class 1NFPA Test Method 702-1980 forClass 1Same
Sterilization MethodEO SterilizationEO SterilizationSame

TABLE 2: COMPARISON OF PROPOSED AND PREDICATE DEVICES

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Medline. The logo is set against a blue background. The word "MEDLINE" is written in bold, sans-serif font. Above the word is a white, four-pointed star-like shape, with each point extending outward from a central point.

Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093

Medline Level 4 Surgical Gown 510(k) Premarket Submission All information on this page is confidential

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. The word "MEDLINE" is written in white, block letters. Above the word is a white, stylized starburst design, with the bottom point of the starburst extending downwards, bisecting the word "MEDLINE".

Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093

The proposed Medline Level 4 Surgical Gown (Sirus, Eclipse, Aurora) is substantially equivalent to the predicate, ULTRA Fluid Barrier Surgical Gown, with regards to claims, design, technology, and intended use.

  • 트 Claim: ANSI/AAMI PB70 Level 4 rating
  • Design: SMS, Single-Use, Disposable
  • " Technology: Reinforcement material along chest and sleeve (critical zones)
  • Intended Use: Surgical apparel intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
ISO 10993-5 CytotoxicityISO MEM Elution Using L-929 Mouse Fibroblast Cells
ISO 10993-10 IrritationISO Intracutaneous Irritation Test
ISO 10993-10 SensitizationISO Guinea Pig Maximization Sensitization Test
ASTM F1929-15Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
ASTM F1671Resistance of Materials Used in Protective Clothing to Penetration byBlood-Borne Pathogens
ASTM D5034-09 (2013)Breaking Strength and Elongation of Textile Fabrics (Grab Test)
ASTM D5587-15Tearing Strength of Fabrics by Trapezoid Procedure
ASTM D3776/D3776M-09aBasis Weight-Mass Per Unit Area (Weight) of Fabric
ASTM D3786/D3786M-13Bursting Strength of Textile Fabrics-Diaphragm Bursting Strength TesterMethod
16 CFR 1610Flammability of Clothing Textiles
ASTM F1929Resistance to Dye Penetration Test
ANSI/AAMI/ISO 10993-7:2008(R)2012Biological evaluation of medical devices –Part 7: Ethylene oxidesterilization residuals
AATCC 42Water Resistance: Impact Penetration Test

Summary of Non-Clinical Testing

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in bold, sans-serif font, and a white star-like shape is behind the text. The star has four points, with two pointing up and down, and two pointing left and right.

Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093

Medline Level 4 Surgical Gown 510(k) Premarket Submission All information on this page is confidential

ANSI/AAMI PB70:2012Liquid Barrier performance and classification of protective apparel and drapes intended for use in health care facilities.
-------------------------------------------------------------------------------------------------------------------------------------------------

Summary of Clinical Testing

Not applicable.

Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the Medline Level 4 Surgical Gowns (Sirus, Eclipse and Aurora) are as safe, as effective, and performs as well as or better than the legally marketed predicate device, Kimberly-Clark (Halyard) ULTRA Fluid Barrier Surgical Gown.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.