K Number
K182170

Validate with FDA (Live)

Date Cleared
2019-02-05

(179 days)

Product Code
Regulation Number
888.3020
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PRECICE® BONE TRANSPORT™ System is indicated for limb-lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions or bone transport of long bones.

Device Description

The PRECICE® Bone Transport System includes the PRECICE® Bone Transport Nail, locking screws, end caps, surgical instruments, and external remote controller (ERC). The PRECICE® Bone Transport nails and end caps are supplied sterile by gamma radiation while the locking screws and instruments are supplied non-sterile and must be sterilized prior to use. The system is designed to achieve limb correction through gradual lengthening or compression of the intercalary bone segment and providing internal fixation for fractures of long bones. The PRECICE® Bone Transport intramedullary nail is implanted using locking screws, end caps, and reusable surgical instruments. The PRECICE® Bone Transport nail contains an enclosed rare earth magnet, distraction rod, and planetary gearing which allows the extension of the distraction rod to be adjusted non-invasively by the External Remote Controller (ERC). The PRECICE® Bone Transport nail is available in various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies and implantation methods. The locking screws are also available in a variety of diameters, lengths, and thread styles. The ERC is available in several compatible models.

AI/ML Overview

The provided text is a 510(k) summary for the PRECICE® BONE TRANSPORT™ System. It details the device, its indications for use, and a comparison to predicate devices to establish substantial equivalence. However, this document does not describe a study involving a comparison of the device's performance against detailed acceptance criteria using metrics such as sensitivity, specificity, or accuracy in a clinical or diagnostic context.

Instead, the document focuses on non-clinical performance verification testing to demonstrate functional substantial equivalence to a predicate device. Therefore, I cannot provide the information requested in the prompt as it pertains to a clinical efficacy study with acceptance criteria and reported device performance in terms of diagnostic or clinical metrics.

Here's a breakdown of why the requested information cannot be fully provided based on the given text:

  1. A table of acceptance criteria and the reported device performance:

    • The document lists various non-clinical tests (Static Compression, Dynamic Compression Bending strength, Torsion, Torque Resistance, Axial Pullout, Biocompatibility, Sterilization, Package Distribution, Distraction Force Testing, Retention Force Testing, Fluid Ingress Testing, Biomechanical Axial Compression Testing).
    • It also lists the applicable standards (e.g., ASTM F1264, ASTM F543, ISO 10993, ISO 11137).
    • However, it does not provide specific numerical acceptance criteria for each test (e.g., "must withstand X N of force") nor does it report the actual numerical device performance results (e.g., "device withstood Y N of force"). It only states, "The results demonstrate that the subject PRECICE® Bone Transport System is substantially equivalent to the predicate device." This is a qualitative statement of equivalence, not a quantitative report against specific acceptance criteria.
    • Crucially, these are engineering/mechanical/material performance criteria, not clinical diagnostic performance criteria.
  2. Sample size used for the test set and the data provenance: Not applicable. The tests are non-clinical, likely involving a certain number of device samples rather than patient data. The document does not specify the number of device samples or their "provenance."

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no human "ground truth" establishment in these non-clinical engineering tests.

  4. Adjudication method: Not applicable for non-clinical engineering tests.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, not mentioned. This type of study is for evaluating human interpretation of diagnostic data, often with AI assistance. The device in question is an intramedullary fixation rod, not a diagnostic imaging AI.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI algorithm.

  7. The type of ground truth used: For the non-clinical tests, the "ground truth" would be established by the physical properties and performance characteristics dictated by the ASTM/ISO standards and internal design specifications. There's no clinical ground truth (like pathology or outcomes data) mentioned.

  8. The sample size for the training set: Not applicable. There is no AI model or training set described.

  9. How the ground truth for the training set was established: Not applicable.

In summary, the provided FDA 510(k) clearance document is for a physical medical device (intramedullary fixation rod) and details non-clinical engineering and material performance tests to establish substantial equivalence, not a clinical study involving AI or diagnostic performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services logo. The FDA logo features the agency's acronym in a blue square, followed by the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is on the left side of the image and features a symbol with three stylized human profiles. The logos are placed side by side.

NuVasive Specialized Orthopedics, Inc. Regualtory Affairs Specialist 101 Enterprise. Suite 100

February 5, 2019

Re: K182170

Aliso Viejo, California 92656

Vidhi Bhatt

Trade/Device Name: PRECICE® BONE TRANSPORT™ System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB, HWC Dated: September 21, 2018 Received: September 24, 2018

Dear Vidhi Bhatt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone

(1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel S. Ramsey -S 2019.02.05 16:31:33 -05'00'

FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182170

Device Name

PRECICE® BONE TRANSPORT™ System

Indications for Use (Describe)

The PRECICE® BONE TRANSPORT™ System is indicated for limb-lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions or bone transport of long bones.

Type of Use (Select one or both, as applicable)
✖ Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized, abstract shape on the left, resembling a curved, open form in purple and silver. To the right of the shape, the word "NUVASIVE" is written in a sans-serif font, with a silver gradient effect. Below "NUVASIVE", the words "SPECIALIZED ORTHOPEDICS, INC." are written in a smaller, simpler font.

PRECICE® Bone Transport System 510(k) Summary - K182170

In accordance with Title 21 of the Code of Federal Regulations. Part 807, and in particular 21 CFR $807.92, the following summary of information is provided:

Submitted by: A.

Vidhi Bhatt Regulatory Affairs Specialist NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo, CA 92656 Telephone: (949) 563-0655

Date Prepared: February 5, 2019

B. Device Name

Trade or Proprietary Name:PRECICE® Bone Transport System
Common or Usual Name:Intramedullary Fixation Rod
Classification Name:Rod, Fixation, Intramedullary and AccessoriesScrew, Fixation, Bone
Device Class:Class II
Classification:21 CFR § 888.3020
Product Code:HSB, HWC

C. Predicate Devices

Predicate DevicePRECICE® System (K173129)
Reference DevicePRECICE® Stryde System (K180503)

D. Device Description

The PRECICE® Bone Transport System includes the PRECICE® Bone Transport Nail, locking screws, end caps, surgical instruments, and external remote controller (ERC). The PRECICE® Bone Transport nails and end caps are supplied sterile by gamma radiation while the locking screws and instruments are supplied non-sterile and must be sterilized prior to use. The system is designed to achieve limb correction through gradual lengthening or compression of the intercalary bone segment and providing internal fixation for fractures of long bones. The PRECICE® Bone Transport intramedullary nail is implanted using locking screws, end caps, and reusable surgical instruments. The PRECICE® Bone Transport nail contains an enclosed rare earth magnet, distraction rod, and planetary gearing which allows the extension of the distraction rod to be adjusted non-invasively by the External Remote Controller (ERC). The PRECICE® Bone Transport nail is available in various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies and implantation methods. The locking screws are also available in a variety of diameters, lengths, and thread styles. The ERC is available in several compatible models.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic to the left of the company name. The text "NUVASIVE" is in a sans-serif font, and below it, in smaller letters, is "SPECIALIZED ORTHOPEDICS, INC."

The purpose of this submission is to implement design changes to the predicate PRECICE® System (K173129) and reference device PRECICE® Stryde System (K180503) to create a product line extension which includes the PRECICE® Bone Transport nail and compatible locking screws and end caps. The subject nail performance meets predicate device performance specifications. The design changes of subject nail inform revised contraindications, warnings. precautions and implantation procedure on the label to address new risks compared to predicate device.

These modifications to the PRECICE® System do not change the indications for use or intended use of the device, nor do they change the fundamental scientific technology of the device.

E. Indication for Use

The PRECICE® Bone Transport System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, or bone transport of long bones.

F. Comparison of Technological Characteristics with the Predicate Device

The subject PRECICE® Bone Transport System is substantially equivalent to the predicate PRECICE® System (K173129) and reference device PRECICE® Stryde System (K180503) previously cleared by the FDA for commercial distribution in the United States.

Substantial equivalence between subject, predicate and reference device is established based on identical indications for use, material composition, technological characteristics, principles of operation and fundamental scientific principles.

Substantial equivalence is also based on device dimension, specifically, the nail diameters, locking screws, device configurations and end caps which are the same as the reference device. The predicate, reference and subject devices are inserted into the intramedullary canal of the long bones and secured with locking screws. The predicate, reference and subject devices are adjusted non-invasively by the External Remote Controller (ERC). The design modifications for the subject device, PRECICE® Bone Transport are as follows:

  • . The subject nail is a slotted intramedullary (IM) nail. The housing tube of this IM nail has longitudinal transport slot.
  • . The stroke length/transport length is independent of rod length since the distraction rod excursion is accommodated within the housing tube making the subject device a fixed length device.
  • The distraction rod spans both the defect site and osteotomy site.
  • Additional size of locking screw available with the subject device .
  • Assembly of the gear train and distraction rod is reversed to incorporate the transport slot ● although the components remain the same as the predicate device

The labeling changes includes added contraindication and warning based on the design modifications of the subject device.

There are no changes being made to the ERCs or PRECICE® surgical instruments as a result of this submission.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic on the left, followed by the word "NUVASIVE" in gray, sans-serif font. Below "NUVASIVE" is the text "SPECIALIZED ORTHOPEDICS, INC." in a smaller, sans-serif font, also in gray.

Conclusions can be drawn from these comparisons that the subject PRECICE® Bone Transport System is substantially equivalent to the predicate device.

G. Performance Data

Nonclinical performance verification testing was performed to demonstrate that the subject PRECICE® Bone Transport System is substantially equivalent to the predicate PRECICE® device. The following testing shows that the subject device conforms to the following standards, practices and guidance:

Testing DescriptionApplicable Standard
Static CompressionBending StrengthASTM F1264 - Standard Specificationand Test Methods for IntramedullaryFixation Devices
Dynamic CompressionBending strength
Torsion
Torque ResistanceASTM F543 - Standard Specification andTest Methods for Metallic Medical BoneScrews
Axial Pullout
BiocompatibilityISO 10993- Biological evaluation ofmedical devices
SterilizationGamma Irradiation Sterilization Testing:VDmax 25 kGy method per IS/EN/ISO11137-1:2015 and IS/EN/ISO 11137-2:2015
Package DistributionASTM D4169-16 Standard Practice forPerformance Testing of ShippingContainers and Systems

In addition, the device also underwent the following performance testing to show that all requirements have been met:

  • Distraction Force Testing

  • Retention Force Testing A
  • A Fluid Ingress Testing
  • Biomechanical Axial Compression Testing A

The results demonstrate that the subject PRECICE® Bone Transport System is substantially equivalent to the predicate device.

H. Conclusions

The subject PRECICE® Bone Transport System has been shown to be substantially equivalent to the legally marketed predicate device for its intended use.

§ 888.3020 Intramedullary fixation rod.

(a)
Identification. An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.(b)
Classification. Class II.