K Number
K182052

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-02-26

(210 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Omni Instrument Tray is intended to enclose, protect, and organize the Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage. The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles. The Omni Instrument Tray has been validated for the following sterilization cycles: Prevacuum Steam 132°C (270°F) 4 minutes exposure 35 minutes dry time Contents: - 1 Omni Hysteroscope - 1 Removable Outflow Channel - 1 XL Removable Outflow Channel - 1 Omni 3.7mm Diagnostic Sheath - 1 Omni 5.5mm Operative Sheath - 1 Omni 6mm Operative Sheath - 3 Seal Caps - 2 Light Guide Adapters STERRAD ® - STERRAD 100S Short - STERRAD NX Standard - STERRAD 100NX Standard Contents: - 1 Omni Hysteroscope - 1 Removable Outflow Channel - 1 XL Removable Outflow Channel - 1 Omni 3.7mm Diagnostic Sheath - 1 Omni 5.5mm Operative Sheath - 1 Omni 6mm Operative Sheath - 3 Seal Caps - 2 Light Guide Adapters

Device Description

The Omni Instrument Tray is intended to enclose, protect, and organize The Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage. The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles. The Omni Instrument Tray is made of a stainless steel, santoprene, and silicone and is supplied non-sterile and is designed for multiple use for the sterilization of the Omni Hysteroscope set prior to each use of the Hysteroscope.

AI/ML Overview

The Omni Instrument Tray is intended to enclose, protect, and organize the Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage. It must be used with an FDA-cleared sterilization wrap. The device was validated for Prevacuum Steam at 132°C (270°F) for 4 minutes exposure and 35 minutes dry time, and STERRAD® cycles (100S Short, NX Standard, 100NX Standard).

Here is a summary of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria / Performance AspectAcceptance CriteriaReported Device Performance
Sterilization EfficacySterilitySuccessfully sterilized following validated cycles for listed contents.
BiocompatibilityNon-cytotoxicUnder the conditions of the study, the device extract was found not to be cytotoxic to the testing assay (ISO 10993-5:2009).
UsabilitySafe and effective useTesting to IEC 62366-1:2015 demonstrated usability.
Packaging IntegrityWithstand parcel deliveryPackaging testing to ISTA 3A (for 70kg or less) was performed.
Holding Capability (Instruments)Securely hold instrumentsDesign verification testing confirmed it holds the Omni hysteroscope, sheaths, and removable outflow channels (ANSI/AAMI ST77:2013).
Component FitCorrect fit of componentsAttribute verification of component fit was performed.
Durability (Latch cycle)Withstand repeated useLatch cycle durability testing was performed (AAMI TIR12:2010).
Durability (Drop test)Withstand dropsSimulated table height drop test was performed (AAMI TIR12:2010).
Durability (Handle strength)Withstand sustained loadHandle pull strength to sustained load was performed (AAMI TIR12:2010).
Cleaning EffectivenessEffective cleaningValidation testing confirmed the methods and parameters for cleaning as specified in the labeling (AAMI TIR30:2011).

2. Sample Size Used for the Test Set and Data Provenance:

The document describes non-clinical bench testing and simulated use testing. Specific sample sizes for each test are not explicitly provided in the summary. The data provenance is described as being generated through "non-clinical bench testing and simulated use testing," implying in-house or contract lab testing, not involving human subjects or real-world patient data from specific countries. The testing is and can be considered retrospective to the device design.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. The testing described is for a sterilization tray, focusing on engineering and biological performance. Ground truth for these types of tests is established by adherence to recognized standards and objective measurements (e.g., microbial kill, cytotoxicity, mechanical strength), not expert consensus on medical images or clinical outcomes.

4. Adjudication Method for the Test Set:

Not applicable. This device is not an AI diagnostic tool and does not involve human interpretation or adjudication of results in the traditional sense. Acceptance is based on meeting pre-defined criteria from relevant standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Not applicable. This is a medical device (sterilization tray), not an AI-assisted diagnostic tool for human readers.

6. Standalone Performance Study:

Yes, a standalone performance study was done in the form of non-clinical bench testing and simulated use testing. The document states that "non-clinical bench testing and simulated use testing demonstrate the Omni Instrument Tray is similar to the predicate device." This refers to the testing of the device itself (algorithm-only equivalent for a physical device) against established performance criteria, independent of human interaction during the test process beyond setting up the tests.

7. Type of Ground Truth Used:

The ground truth used for the performance testing is based on:

  • Adherence to recognized industry standards: e.g., ISO 10993-5:2009 for biocompatibility, IEC 62366-1:2015 for usability, ISTA 3A for packaging, ANSI/AAMI ST77:2013 for containment, AAMI TIR12:2010 for durability, AAMI TIR30:2011 for cleaning.
  • Objective measurements and criteria: such as the absence of cytotoxicity, successful completion of sterilization cycles (implied by meeting the standards for sterilizers), confirmed fit of components, and ability to withstand specified mechanical forces.

8. Sample Size for the Training Set:

Not applicable. This is a physical medical device (sterilization tray), not a machine learning algorithm. Therefore, there is no "training set" in the context of AI.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 26, 2019

Hologic, Inc Rachelle Fitzgerald Senior Regulatory Affairs Specialist 250 Campus Drive Marlborough, Massachusetts 01752

Re: K182052

Trade/Device Name: Omni Instrument Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: January 29, 2019 Received: January 25, 2019

Dear Rachelle Fitzgerald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)K182052
Device NameOmni Instrument Tray
Indications for Use (Describe)
The Omni Instrument Tray is intended to enclose, protect, and organize the Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.
The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.
The Omni Instrument Tray has been validated for the following sterilization cycles:
Prevacuum SteamSTERRAD ®
132°C (270°F)- STERRAD 100S Short
4 minutes exposure- STERRAD NX Standard
35 minutes dry time- STERRAD 100NX Standard
Contents:Contents:
- 1 Omni Hysteroscope- 1 Omni Hysteroscope
- 1 Removable Outflow Channel- 1 Removable Outflow Channel
- 1 XL Removable Outflow Channel- 1 XL Removable Outflow Channel
- 1 Omni 3.7mm Diagnostic Sheath- 1 Omni 3.7mm Diagnostic Sheath
- 1 Omni 5.5mm Operative Sheath- 1 Omni 5.5mm Operative Sheath
- 1 Omni 6mm Operative Sheath- 1 Omni 6mm Operative Sheath
- 3 Seal Caps- 3 Seal Caps
- 2 Light Guide Adapters- 2 Light Guide Adapters
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)Page 1 of 1
K182052/S001Attachment 3M
PSC Publishing Services (301) 443-6740
EF
1 of 1

{3}------------------------------------------------

510(K) SUMMARY K182052

Date: February 20, 2019

510(k) Submitter:

Hologic, Inc. 250 Campus Drive Marlborough, MA 01752 Attn: Rachelle D. Fitzgerald P: 508.263.8631 F: 877.793.6434

Establishment Reqistration Number: 1222780

Trade Name: Omni Instrument Tray

Common/Usual Name: Sterilization Tray

Regulation Name: Sterilization wrap

Regulation Number: 21 CFR 880.6850

Product Code: KCT

Classification: Class II

Panel: General Hospital

PREDICATE DEVICE

Tradename: MyoSure Instrument Tray Submitter/510(k) Holder: Hologic, Inc. 510(k) #: K121280 Product Code: KCT Regulation: 21CFR 880.6850

DEVICE DESCRIPTION

The Omni Instrument Tray is intended to enclose, protect, and organize The Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.

The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.

The Omni Instrument Tray is made of a stainless steel, santoprene, and silicone and is supplied non-sterile and is designed for multiple use for the sterilization of the Omni Hysteroscope set prior to each use of the Hysteroscope.

The Omni Instrument Tray has been validated for the following sterilization cycles:

{4}------------------------------------------------

Pre-vacuum SteamSTERRAD®
132°C (270°F)STERRAD NXTM Standard
4 minutes exposureSTERRAD 100NXTM Standard
35 minutes dry timeSTERRAD 100S Short
Contents:Contents:
- 1 Omni Hysteroscope- 1 Omni Hysteroscope
-1 Removable Outflow Channel-1 Removable Outflow Channel
-1 XL Removable Outflow Channel-1 XL Removable Outflow Channel
-1 Omni 3.7mm Diagnostic Sheath-1 Omni 3.7mm Diagnostic Sheath
-1 Omni 5.5mm Operative Sheath-1 Omni 5.5mm Operative Sheath
- 1 Omni 6mm Operative Sheath- 1 Omni 6mm Operative Sheath
- 3 Seal Caps- 3 Seal Caps
- 2 Light Guide Adapters- 2 Light Guide Adapters

INDICATIONS FOR USE:

The Omni Instrument Tray is intended to enclose, protect, and organize The Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.

The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.

The Omni Instrument Tray has been validated for the following sterilization cycles:
-----------------------------------------------------------------------------------------
Pre-vacuum SteamSTERRAD®
132°C (270°F)STERRAD NX™ Standard
4 minutes exposureSTERRAD 100NXT™ Standard
35 minutes dry timeSTERRAD 100S Short
Contents:Contents:
- 1 Omni Hysteroscope- 1 Omni Hysteroscope
-1 Removable Outflow Channel-1 Removable Outflow Channel
-1 XL Removable Outflow Channel-1 XL Removable Outflow Channel
-1 Omni 3.7mm Diagnostic Sheath-1 Omni 3.7mm Diagnostic Sheath
-1 Omni 5.5mm Operative Sheath-1 Omni 5.5mm Operative Sheath
- 1 Omni 6mm Operative Sheath- 1 Omni 6mm Operative Sheath
- 3 Seal Caps- 3 Seal Caps
- 2 Light Guide Adapters- 2 Light Guide Adapters

{5}------------------------------------------------

Technological Omni Instrument Trav Predicate Device Reference Device Characteristic (K121280) (K171692) The tray is intended The tray is intended to The tray is intended Intended Use enclose, protect, and to enclose, protect, to enclose, protect, and organize and organize organize medical medical devices. medical devices. devices. Steam Autoclave Steam Autoclave; Pre-Steam Autoclave; Biocompatibility Vacuum 132°C (270°F), Pre-Vacuum 132°C Pre-Vacuum 132°C 4 min Exposure, 30 min (270°F), 3 min (270°F); 4 min Exposure; Exposure, 15-30 min dry time; 35 min dry time dry time; STERRAD® 100S Short STERRAD NX Standard STERRAD: NX cycle STERRAD 100NX Standard and Standard Hemolytic Test 10993-4 Express, 100S Short STERRAD 100S Short Biological evaluation of Cytotoxicity ISO 10993medical devices-Part: Cytotoxicity ISO 5:2009 Biological Selection of tests for 10993-5:2009 evaluation of medical interactions with blood; Biological evaluation devices-Part 5: Test for in Irritation 10993-10: Tests of medical devicesvitro cytotoxicity. for irritation and skin Part 5: Test for in vitro cytotoxicity. sensitization. Under the condition of the study, the device Under the conditions of Under the condition extract was found not to the study, device extract of the study, the be cytotoxic to the was found not to be device extract was testing assay. hemolytic to the testing found not to be assay. cytotoxic to the testing assay. Under the conditions of the study, the polar and non-polar device extracts were found not to be a sensitizer nor an irritant to the animal model.

TECHNOLOGICAL CHARACTERISTICS:

{6}------------------------------------------------

TechnologicalCharacteristicOmni Instrument TrayPredicate Device(K121280)Reference Device(K171692)
Prescription OnlyNoNoNo
Device DescriptionStainless Steeltray and lidwith metal lid lockingmechanism and handlesAccessories include:siliconegrommets, and stainlesssteelpostsFenestrated plastic trayandlid with metal lid lockingmechanism and handlesAccessories include:silicone grommets, andstainless steel postsRigid containmentdevice consisting ofa base with a lidwhich can befastened by alatchingmechanism. Thedevice is perforatedin order to enablereprocessing ofenclosed medicaldevices held inplace by siliconeretainers.
Materials ofconstructionTrayLidHandles/LatchingGrommetsStainless Steel/SantopreneStainless SteelStainless Steel/Silicone/SantopreneSiliconeRadel®R resinRadel®R resinStainless SteelSiliconeStainless steelStainless steelStainless SteelSilicone
DimensionsLengthWidthHeight16.20"10.11"2.27"13.45"8.41"2.00"22.83"3.03"10.63"
Sterilization cyclesSteam Autoclave:Pre-Vacuum 132°C(270°F);4 min Exposure;35 min dry timeSTERRAD NXStandardSTERRAD 100NXStandard STERRAD100S ShortSteam AutoclavePre-Vacuum 132°C(270°F),4 min Exposure, 30 mindry time;STERRAD® 100S ShortcycleSteam AutoclavePre-Vacuum 132°C(270°F), 3 minExposure, 15-30 mindry time;STERRAD: NXStandard andExpress, 100SShort
Capacity(Maximum totalweight)25 lbs13 lbs8.708 lbs
ReusableYesYesYes
Perforations(Vent to VolumneRatio).435 in²/in³.020 in²/in³72%
TechnologicalCharacteristicOmni Instrument TrayPredicate Device(K121280)Reference Device(K171692)
Patient ContactingMaterialsNoneNoneNone

{7}------------------------------------------------

PERFORMANCE TESTING

Clinical Testing

No clinical testing was performed, submitted, referenced or relied upon for the Omni Instrument Tray.

Bench Testing

Non-clinical bench testing and simulated use testing demonstrate the Omni Instrument Tray is similar to the predicate device. Design verification testing included:

  • . Biocompatibility testing to ISO 10993-5:2009 (Biological evaluation of medicaldevices-Part 5: Test for in vitro cytotoxicity)
  • Usability testing to IEC 62366-1:2015 (Medical devices-Part 1: Application of usability . engineering to medical devices)
  • . Packaging testing to ISTA 3A (Packaged-Products for Parcel Delivery System Shipment70 kq (150 lb or less)
  • Design verification testing to hold the Omni hysteroscope. Sheaths, and Removable ● Outflow Channels (ANSI/AAMI ST77:2013 Containment devices for reusable medical device sterilization)
    • Attribute verification of component fit ●
  • Reliability testing including durability, drop test, and handle strength (AAMI TIR12:2010 Design, testing, and labeling reusable medical devices for reprocessing in health care facilities: A quide for medical device manufacturers)
    • . Latch cycle durability
    • Simulated table height drop test ●
    • Handle pull strength to sustained load

Cleaning and Sterilization

The Omni Instrument Tray is supplied non-sterile and required processing through cleaning and sterilization at the health care facility by the end user. Validation testing confirmed the methods and parameters of these processes as specified in the device labeling.

{8}------------------------------------------------

The following standards were used during the validation process:

Standard ReferenceStandard Title
AAMI/ANSI ST79:2017Comprehensive guide to steam sterilization and sterility assurance inhealth care facilities
ANSI/AAMI ST77:2013Containment devices for reusable medical device sterilization
ISO 17664:2017Processing Of Health Care Products - Information To Be Provided By TheMedical Device Manufacturer For The Processing Of Medical Devices
ISO 17665-1:2006Sterilization Of Health Care Products - Moist Heat - Part 1:Requirements For The Development, Validation And Routine Control OfA SterilizationProcess For Medical Devices
ISO TS 17665-2:2009Sterilization Of Health Care Products - Moist Heat - Part 2: Guidance OnThe Application Of ISO 17665-1
ANSI/AAMI/ISO14937:2009/(R)2013Sterilization of Health Care Products-General Requirements forcharacterization of a sterilizing agent and the development, validation androutine control of a sterilization process for medical devices
AAMI TIR12:2010Designing, testing, and labeling reusable medical devices for reprocessingin health care facilities: A guide for medical device manufacturers
AAMI TIR30:2011A compendium of processes, materials, test methods, acceptance criteriafor cleaning reusable medical devices

The above referenced design verification testing and cleaning and sterilization validations demonstrate the subject device met the design specifications.

CONCLUSION

The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device MyoSure Instrument Tray (K121280).

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).