(76 days)
The Piccolo Composite foot and ankle plates are indicated for fixation of osteotomies, fusions, fractures, nonunions, malunions and replantations of small bone fragments in adult and adolescent (12 - 21 years) patients, including the foot and ankle, and including in osteopenic bone.
The Lapidus plates are indicated for fusion and arthrodesis of the 1st tarsometatarsal joint (Lapidus Fusions).
The Piccolo Composite Lapidus Plate System comprises implants (plates and screws), and a set of instruments.
The plates are made of carbon fiber reinforced polyetheretherketone (CFR-PEEK), and are marked with a tantalum thread, to provide for their visualization under fluoroscopy.
The screws are made of titanium alloy. Both non-locking and locking 2.7mm screws are available, in various lengths. Cannulated Lag Screws are also provided.
This document is a 510(k) premarket notification for the "Piccolo Composite® Plate System" and primarily focuses on establishing substantial equivalence to predicate devices, rather than an AI/ML device study. Therefore, most of the requested information regarding acceptance criteria for an AI device, sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance is not applicable to this submission.
However, I can extract the relevant performance criteria and summary from the document:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Evaluation Type) | Reported Device Performance |
|---|---|
| Static Bending (per ASTM F 382) | Comparable to predicate devices |
| Dynamic Bending (per ASTM F 382) | Comparable to predicate devices |
| Bacterial Endotoxin Testing | Conducted (implied acceptable) |
2. Sample size used for the test set and the data provenance: Not applicable. This document describes the mechanical testing and material compatibility of an orthopedic plate system, not an AI device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical testing is established by engineering standards and measurements, not expert consensus.
4. Adjudication method for the test set: Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI device.
7. The type of ground truth used:
- Mechanical Performance: Established through standardized engineering tests (ASTM F 382) for static and dynamic bending. The "ground truth" here is the measured mechanical properties of the device and its predicate devices.
- Biocompatibility: Established through bacterial endotoxin testing.
8. The sample size for the training set: Not applicable, as this is not an AI device.
9. How the ground truth for the training set was established: Not applicable, as this is not an AI device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services, which features a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA). The FDA logo includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" written below in a smaller font.
CarboFix Orthopedics Ltd. Yael Rubin Director of Regulatory Affairs 11 Ha'hoshlim St. Herzeliya, 4672411 Il
October 11, 2018
Re: K182015
Trade/Device Name: Piccolo Composite® Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: July 24, 2018 Received: July 27, 2018
Dear Yael Rubin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir - S 2018.10.11 19:07:40 -04'00'
For: Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K182015
Device Name Piccolo Composite® Plate System
Indications for Use (Describe)
Piccolo Composite Lapidus Plate System:
The Piccolo Composite foot and ankle plates are indicated for fixation of osteotomies, fusions, fractures, nonunions, malunions and replantations of small bone fragments in adult and adolescent (12 - 21 years) patients, including the foot and ankle, and including in osteopenic bone.
The Lapidus plates are indicated for fusion and arthrodesis of the 1st tarsometatarsal joint (Lapidus Fusions).
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary
CarboFix Orthopedics Ltd.
Piccolo Composite® Lapidus Plate System
Applicant Name
CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel
Contact Person
Yael Rubin CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939
Date Prepared
July 2018
Trade/Proprietary Name
Piccolo Composite® Plate System
Common Name
Bone Plating System
Classification Name
Single/multiple component metallic bone fixation appliances and accessories; (21 CFR §888.3030; Class II; Product Code HRS).
{4}------------------------------------------------
Predicate Devices
Primary
- ... Piccolo Composite® Plate System (CarboFix Orthopedics Ltd.; K102597, K120409, K130061, K143496, K160002, K170401)
Additional
- . ORTHOLOC® 3Di Foot Plating Reconstruction System (Wright Medical Technology, Inc .; K152974, and more)
- 2.4mm/2.7mm Variable Angle LCP Forefoot/Midfoot System - First TMT Fusion Plates (Synthes; K100776)
Indications for Use
The Piccolo Composite foot and ankle plates are indicated for fixation of osteotomies, fusions, fractures, nonunions, malunions and replantations of small bones and small bone fragments in adult and adolescent (12 - 21 years) patients, including the foot and ankle, and including in osteopenic bone.
The Lapidus plates are indicated for fusion and arthrodesis of the 1st tarsometatarsal joint (Lapidus Fusions).
System Description
The Piccolo Composite Lapidus Plate System comprises implants (plates and screws), and a set of instruments.
The plates are made of carbon fiber reinforced polyetheretherketone (CFR-PEEK), and are marked with a tantalum thread, to provide for their visualization under fluoroscopy.
The screws are made of titanium alloy. Both non-locking and locking 2.7mm screws are available, in various lengths. Cannulated Lag Screws are also provided.
{5}------------------------------------------------
Substantial Equivalence
The Piccolo Composite Lapidus Plate System intended use, design, materials, technological characteristics, and principles of operation are substantially equivalent to those of the predicate devices, as applicable.
Performance characteristics included static and dynamic bending (evaluated per ASTM F 382) and are comparable to those of predicate devices (as applicable), thus demonstrating that the device is safe and effective for its intended use.
In addition, bacterial endotoxin testing was conducted for the Piccolo Composite Plate System.
N/A