K Number
K181982

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-08-21

(27 days)

Product Code
Regulation Number
888.1100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MIDAScope™ and Introducer Kit, and MIDASystem™ are indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.

Device Description

The MIDAScope™ is a miniaturized, sterile, single-use disposable scope. The image is acquired by an image sensor located at the distal tip. Illumination is provided by an LED and fibers when the disposable scope is connected to the reusable MIDASystem™ via an electrical cable. The scope is available in different lengths to accommodate different treatment locations. The scope is introduced into the joint space using the IntraVu disposable Introducer Kit, consisting of a cannula, trocar and stopcock, which are provided sterile in the same package as the scope.

The MIDASystem™ is a reusable device consisting of an integrated LCD display, and software that receives, processes and displays the images collected by the scope.

AI/ML Overview

This FDA 510(k) summary describes the substantial equivalence of the MIDAScope and Introducer Kit, and MIDASystem to predicate devices. It does not contain information about a study comparing AI vs. human readers or a standalone algorithm performance study. The performance testing section primarily focuses on engineering and regulatory compliance rather than clinical performance or diagnostic accuracy.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document lists various standards followed and states that "All tests met the pre-determined acceptance criteria at baseline and after aging." The specific acceptance criteria within each standard (e.g., specific thresholds for image resolution, strength of materials) are not detailed in this summary. Instead, it indicates that the device passed these standards.

Acceptance Criteria CategoryReported Device PerformanceReferencing Standard(s)
Scope performanceMet pre-determined criteria at baseline and after aging.ISO8600-1, ISO8600-3, and ISO8600-4
BiocompatibilityMet pre-determined criteria.ISO10993-1
Packaging validationMet pre-determined criteria and assured devices perform as intended.ISO11607-1, ISO11607-2, ASTM D4169, ASTM F2096, ASTM F88/F88M
Sterilization validationMet pre-determined criteria.ISO11135, ISO10993-7, ISO11737-1
AgingMet pre-determined criteria.ASTM F1980
Environmental conditionsMet pre-determined criteria.ASTM D4332
Electrical and EMC testingMet pre-determined criteria for electrical safety and electromagnetic compatibility.IEC60601-1 and IEC60601-1-2
Software validationPassed validation.Not explicitly listed, but mentioned for performance testing
Structural and functional performancePassed.Not explicitly listed, but mentioned for performance testing

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This document does not provide information regarding sample sizes for clinical test sets or data provenance. The performance testing mentioned is primarily related to engineering standards (e.g., material testing, electrical safety, sterilization validation) rather than clinical accuracy or diagnostic performance evaluated on patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This document does not provide information on expert involvement for establishing ground truth, as the focus is on device engineering and safety standards, not diagnostic accuracy studies involving human experts interpreting images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This document does not provide information on any adjudication methods.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor any information about AI assistance or its effect size on human reader performance. The device described, MIDAScope, is an arthroscope, which is a visualization and surgical tool, not an AI-powered diagnostic system.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This document does not mention any standalone algorithm performance, as the device is not an AI algorithm but a medical instrument for direct visualization.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Since the document focuses on engineering and regulatory compliance rather than diagnostic accuracy, it does not specify a type of ground truth related to clinical findings. The "ground truth" in this context would refer to meeting the specified parameters of the engineering standards (e.g., sterilization effectively killed all microorganisms, electrical leakage was within limits).

8. The sample size for the training set

This document does not provide information about a training set, as it does not describe an AI or machine learning component.

9. How the ground truth for the training set was established

This document does not provide information about a training set or its ground truth establishment, as it does not describe an AI or machine learning component.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency's name in a lighter shade of blue.

August 21, 2018

IntraVu. Inc. % Mr. Dave Yungvirt Third Party Review Group, LLC The Old Station House 24 Lackawanna Place Millburn, New Jersey 07041

Re: K181982

Trade/Device Name: MIDAScope and Introducer Kit, and MIDASystem Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, GCJ Dated: July 23, 2018 Received: July 25, 2018

Dear Mr. Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jennifer R. Stevenson

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181982

Device Name

MIDAScope™ and Introducer Kit, and MIDASystem™

Indications for Use (Describe)

The MIDAScope™ and Introducer Kit, and MIDASystem™ are indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

510(k) Applicant

IntraVu, Inc. 610 Price Avenue Redwood City, CA 94063 Phone: (650) 395-6067

Contact Person

Plamena Entcheva-Dimitrov, PhD Preferred Regulatory Consulting Inc. Phone: (650) 533-2253 Email: entcheva@preferredregulatoryconsulting.com

Date of Summary

20 Aug 2018

Device Overview

Device Proprietary NameMIDAScope™ and Introducer Kit, and MIDASystem™
Product ClassII
Regulation NameArthroscope
Regulation Number888.1100
Product CodeHRX
GCJ
PanelOrthopedic

Predicate Device

510(k) NumberProduct NameManufacturer
K162475mi-eye 2 and mi-eye 2 Monitor (primary predicate)Trice Medical, Inc.
K141326NeedleView CHTM Scope KitsBioVition Technologies, LLC
K152511Neo-ArthroscopeTM - Single Use Digital VideoArthroscopic System (reference device)Prosurg, Inc.

{4}------------------------------------------------

Device Description

The MIDAScope™ is a miniaturized, sterile, single-use disposable scope. The image is acquired by an image sensor located at the distal tip. Illumination is provided by an LED and fibers when the disposable scope is connected to the reusable MIDASystem™ via an electrical cable. The scope is available in different lengths to accommodate different treatment locations. The scope is introduced into the joint space using the IntraVu disposable Introducer Kit, consisting of a cannula, trocar and stopcock, which are provided sterile in the same package as the scope.

The MIDASystem™ is a reusable device consisting of an integrated LCD display, and software that receives, processes and displays the images collected by the scope.

Intended Use

The device is intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Indications for Use

The MIDAScope™ and Introducer Kit, and MIDASystem™ are indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.

Comparison to Predicate Device

The MIDASystem and Introducer Kit and MIDAScope and predicate devices have the same intended use and thus fulfill the requirements of 513(i) of the FD&C Act (21 U.S.C. $ 360c(i)). Subject device and predicates are similar in design, principle of operation, components, users and use environment. The subject scope and its predicates are sterilized with ethylene oxide to yield SAL106. The subject system is reusable as are the predicate systems. The packaging of the MIDAScope (with the Introducer Kit), and MIDASystem and its predicates is similar and passed testing following the same internationally recognized standards. The subject device and the predicate devices were tested according to the same internationally recognized standards for electromagnetic compatibility, for electrical safety, and for biocompatibility and were shown to be safe and passed the standards' requirements.

Risks associated with the subject device and the predicates are similar. There are no significant technological differences between MIDASystem and MIDAScope (with the Introducer Kit) and the predicate devices.

{5}------------------------------------------------

Therefore, in conclusion, the subject devices MIDAScope (with the Introducer Kit), and MIDASystem are substantially equivalent to the predicates.

Performance Testing

Performance testing of the MIDAScope and Introducer Kit, and MIDASystem included structural and functional performance, software validation, electrical and EMC, biocompatibility, packaging and sterilization studies. Devices were subjected to environmental conditioning and simulated distribution to assure devices will perform as intended during marketing distribution. All tests met the pre-determined acceptance criteria at baseline and after aging. The following FDA-recognized standards were followed and the acceptance criteria set forth in them were met:

Scope performance:ISO8600-1, ISO8600-3, and ISO8600-4
Biocompatibility:ISO10993-1
Packaging validation:ISO11607-1, ISO11607-2, ASTM D4169, ASTMF2096, ASTM F88/F88M
Sterilization validation:ISO11135, ISO10993-7, ISO11737-1
Aging:ASTM F1980
Environmental conditions:ASTM D4332
Electrical and EMC testing:IEC60601-1 and IEC60601-1-2

Safety and Effectiveness and Conclusion

Based on the evaluation of intended use, performance characteristics and meeting FDArecognized standards for safety and performance, the MIDAScope™ and Introducer Kit, and the MIDASystem™ are substantially equivalent to the predicate device. The labeling of the MIDAScope and Introducer Kit, and MIDASystem contains instructions for use and any necessary cautions and warnings to assure safe and effective use of the device.

§ 888.1100 Arthroscope.

(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.