(62 days)
The OsseoScrew™ System (for use with the Zodiac™ Spinal Fixation System and Illico™ MIS Posterior Fixation System) is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The OsseoScrew is a pedicle screw system that consists of pedicle screws and associated general instruments. Implant components are available in a variety of sizes to suit the individual pathology and anatomical conditions of the patient.
The subject cannulated polyaxial pedicle screws are in diameters of 6.5mm and 7.5mm with lengths ranging from 40 to 55mm. The OsseoScrew is designed to connect to 5.5mm diameter rods and the associated components contained within the Zodiac Spinal Fixation System and the Illico MIS Posterior Spinal Fixation Systems.
The Alphatec Spine OsseoScrew System is an implantable pedicle screw device whose core is manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and the expandable screw shank is manufactured from CP2 Titanium conforming to ASTM F67 which are both industry recognized standards.
The provided text is a 510(k) summary for the Alphatec Spine, Inc. OsseoScrew System. It describes the device, its intended use, and claims substantial equivalence to predicate devices based on performance data and engineering analysis. However, it does not contain specific acceptance criteria, detailed study results, or the other requested information related to human performance, ground truth establishment, or sample sizes beyond the general statement that "Clinical data on the subject device has been provided to support a substantial equivalence determination."
Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, nor can I provide information regarding sample sizes for training/test sets, data provenance, expert adjudication, MRMC studies, or standalone algorithm performance, as these details are not present in the provided document.
The document focuses on demonstrating substantial equivalence through:
- Technological Comparison: Comparing intended use, indications for use, design, function, and technology to predicate devices.
- Performance Data and Engineering Analysis: Mentioning static shank torsion, static shank tensile, dynamic shank bending, dynamic construct compression, and static construct compression testing.
- Clinical Data: Stating that clinical data was provided to support substantial equivalence.
Without further information about the specific "Clinical data" and "Performance data" mentioned, it's impossible to
address most of your questions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services-USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
September 25, 2018
Alphatec Spine, Inc. Mr. Jeremy Markovich Senior Manager, Regulatory and Clinical Affairs 5818 El Camino Real Carlsbad, California 92008
Re: K181980
Trade/Device Name: OsseoScrew System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: July 24, 2018 Received: July 25, 2018
Dear Mr. Markovich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ronald P. Jean -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K181980
Device Name OsseoScrew System
Indications for Use (Describe)
The OsseoScrew™ System (for use with the Zodiac™ Spinal Fixation System and Illico™ MIS Posterior Fixation System) is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
| Type of Use (Select one or both, as applicable) |
|---|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Atec. The logo consists of a green square with a white curved line running through it, followed by the word "atec" in white sans-serif font. A trademark symbol is located to the upper right of the word "atec". The background of the logo is a dark blue color.
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. | SUBMITTER: | Alphatec Spine, Inc. |
|---|---|---|
| 5818 El Camino Real | ||
| Carlsbad, CA 92008 | ||
| Phone: (760) 431-9286 | ||
| Fax: (760) 431-0289 | ||
| Contact Person: | Jeremy Markovich | |
| Senior Manager, Regulatory and Clinical AffairsContact Phone: (760) 494-6893 | ||
| Date Summary Prepared: | September 24, 2018 |
II. DEVICE
| Name of Device: | OsseoScrew System |
|---|---|
| Common or Usual Name: | Orthosis, Spinal Pedicle Fixation |
| Classification Name: | Thoracolumbosacral pedicle screw system |
| Regulation Number: | 21 CFR 888.3070 |
| Regulatory Class: | Class II |
| Product Code: | NKB |
III. LEGALLY MARKETED PREDICATE DEVICES
Primary Predicate:
| 510(k) | Product Name | Clearance Date |
|---|---|---|
| K123623 | Illico™ MIS Posterior Fixation System | February 14, 2013 |
Additional Predicates:
| 510(k) | Product Name | Clearance Date |
|---|---|---|
| K160879 | CONFIDENCE™ High Viscosity Spinal Cement,VIPER® And EXPEDIUM® Fenestrated Screw Systems | December 20, 2016 |
IV. DEVICE DESCRIPTION
The OsseoScrew is a pedicle screw system that consists of pedicle screws and associated general instruments. Implant components are available in a variety of sizes to suit the individual pathology and anatomical conditions of the patient.
The subject cannulated polyaxial pedicle screws are in diameters of 6.5mm and 7.5mm with lengths ranging from 40 to 55mm. The OsseoScrew is designed to connect to 5.5mm
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ATEC. The logo consists of a green square with a white curved line inside, followed by the word "atec" in white letters. The letters are lowercase and have a modern, sans-serif font. The background of the logo is a dark blue color.
diameter rods and the associated components contained within the Zodiac Spinal Fixation System and the Illico MIS Posterior Spinal Fixation Systems.
The Alphatec Spine OsseoScrew System is an implantable pedicle screw device whose core is manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and the expandable screw shank is manufactured from CP2 Titanium conforming to ASTM F67 which are both industry recognized standards.
V. INDICATIONS FOR USE
The OsseoScrew™ System (for use with the Zodiac™ Spinal Fixation System and Illico™ MIS Posterior Fixation System) is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
VI. TECHNOLOGICAL COMPARISON TO PREDICATES
The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. The different technological characteristics do not raise different questions of safety and effectiveness.
VII. PERFORMANCE DATA
Performance data and engineering analysis demonstrate that the subject OsseoScrew System is substantially equivalent to the predicates. Testing and analysis include static shank torsion, static shank tensile, dynamic shank bending, dynamic construct compression and static construct compression.
Clinical data on the subject device has been provided to support a substantial equivalence determination.
VIII. CONCLUSION
Based upon the information provided in this 510(k) submission, it has been determined that the subject devices are substantially equivalent to legally marketed devices in regards to indications for use, intended use, design, technology, and performance,
N/A