(110 days)
icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR or NCCT images. This software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR or NCCT images.
icobrain consists of two distinct image processing pipelines: icobrain cross and icobrain long.
icobrain cross is intended to provide volumes from MR or NCCT images acquired at a single time point. icobrain long is intended to provide changes in volumes between two MR images that were acquired on the same scanner, with the same image acquisition protocol and with same contrast at two different timepoints. The results of icobrain cross cannot be compared with the results of icobrain long.
The input images can be MR images (current icobrain software - K161148 and K180326) or CT images. During the pre-processing, the modality and/or sequence of each scan is detected and each scan is converted from DICOM format to NIFTI format. The image processing then performs the actual segmentation and calculates the measurements of the brain structures and abnormalities. Finally, the computed measurements are summarized into an electronic report and (some) segmentations are overlaid on the input images, generating output images in DICOM format.
Since the processing of MR images remains unchanged compared to the currently approved icobrain software (see KI 6 I 148 and K180326), the remainder of this file will focus on the design of the software that processes CT images. We refer to the overall architecture focused on (pre)processing CT images as the CT pipeline.
Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Accuracy: Lesions, basal cisterns, lateral ventricles, and midline shift compared to manually segmented ground truth. | Accuracy: For all experiments, the Pearson correlation coefficient between compared measurements was 0.95. |
| Accuracy: Lateral ventricles and whole brain volumes compared to MR images segmented by the cleared icobrain 3.0 software (taken as ground truth). | Accuracy: For all experiments, the Pearson correlation coefficient between compared measurements was 0.95. |
| Reproducibility: Tested on CT images produced in the same scanning session. | Reproducibility: For all experiments, the intraclass correlation coefficient was 0.94. |
| All experiments passed the acceptance criteria set by literature review. | All experiments passed the acceptance criteria. |
Study Details
-
Sample sizes used for the test set and data provenance:
- Test Set Sample Size: 544 subject datasets.
- Data Provenance: The subjects included TBI patients and potential dementia patients. The specific country of origin is not explicitly stated, but the submission is from a Belgian company (icometrix NV). The study appears to be retrospective, using existing subject datasets.
-
Number of experts used to establish the ground truth for the test set and their qualifications:
- The document states that some ground truth was "manually segmented." However, it does not specify the number of experts, their qualifications (e.g., radiologist with X years of experience), or the specific process for this manual segmentation.
- For lateral ventricles and whole brain volumes, the ground truth was "MR images segmented by the cleared icobrain 3.0 software." This implies a form of software-generated ground truth, rather than human expert ground truth for these specific metrics.
-
Adjudication method for the test set:
- The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). It only mentions that lesions, basal cisterns, lateral ventricles, and midline shift were compared to "manually segmented ground truth." Without further detail, it's impossible to determine the adjudication method for the manual segmentation.
-
If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or described. The study focuses on the standalone performance of the device against established ground truth and reproducibility, not on comparing human readers with and without AI assistance.
-
If a standalone performance (algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance study was done. The performance testing "demonstrate[s] the performance of the CT pipeline of icobrain 4.0" by validating its accuracy and reproducibility against ground truth. This is a direct assessment of the algorithm's performance without a human in the loop during the measurement generation.
-
The type of ground truth used:
- Expert Consensus/Manual Segmentation: For lesions, basal cisterns, lateral ventricles, and midline shift, the ground truth was established by "manually segmented ground truth." This implies human expert input, but the specifics of consensus are not detailed.
- Software-generated Ground Truth (Predicate Device): For lateral ventricles and whole brain volumes, the ground truth was "MR images segmented by the cleared icobrain 3.0 software."
-
The sample size for the training set:
- The document does not provide the sample size for the training set. It only mentions the test set size of 544 subject datasets.
-
How the ground truth for the training set was established:
- The document does not specify how the ground truth for the training set was established, as the size and details of the training set are not provided. The technical characteristics state "segmentation by classical machine learning and deep learning (in our case supervised voxel classification with Convolutional Neural Networks)," which implies that a training set with established ground truth would have been necessary for supervised learning.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in blue and white. The logo is commonly used to identify the FDA and its role in regulating food and drugs in the United States.
November 6, 2018
icometrix NV Jan Verheyden VP Kolonel Begaultlaan 1b/12 Leuven, 3012 BELGIUM
Re: K181939
Trade/Device Name: icobrain Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: October 3, 2018 Received: October 3, 2018
Dear Jan Verheyden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K181939
Device Name icobrain
Indications for Use (Describe)
icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR or NCCT images. This software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR or NCCT images.
Icobrain consists of two distinct image processing pipelines: icobrain cross and icobrain long.
icobrain cross is intended to provide volumes from MR or NCCT images acquired at a single time point. icobrain long is intended to provide changes in volumes between two MR images that were acquired on the same scanner, with the same image acquisition protocol and with same contrast at two different timepoints. The results of icobrain cross cannot be compared with the results of icobrain long.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K181939
Doc. 134873199.21 (Idisplay/MSMET/CT_Section+5.+510%28k%29+Summary), 2018-10-09 14:56 UTC
- · 5.1 Submitter
- 5.2 Device
- 5.3 Predicate Device
- Intended Use
- . 2.4
2.5 Device Description - 5.6 Comparison with predicate device
- 5.7 Performance testing
5.1 Submitter
| Name: | icometrix NV |
|---|---|
| Address: | Kolonel Begaultlaan lb/123012 LeuvenBelgium |
| Contact Person: | Jan Verheyden |
| Telephone number: | +32 16 369 000 |
| Fax Number: | N.A. |
| E-mail: | jan.verheyden@icometrix.com |
| Date Prepared: | 10 Sep 2018 |
5.2 Device
| Device Trade Name: | icobrain |
|---|---|
| Common Name | Medical Image Processing Software |
| Classification Name | System, Image processing, Radiological |
| Number | 892.2050 |
| Product Code: | LLZ |
| Classification Panel: | Radiology |
{4}------------------------------------------------
5.3 Predicate Device
| Device | ico brain |
|---|---|
| 510(k) Number | K180326 |
| Manufacturer | icometrix NV |
5.4 Intended Use
icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR or NCCT images. This software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR or NCCT images.
icobrain consists of two distinct image processing pipelines: icobrain cross and icobrain long.
- · icobrain cross is intended to provide volumes from MR or NCCT images acquired at a single time point.
- · icobrain long is intended to provide changes in volumes between two MR images that were acquired on the same scanner, with the same image acquisition protocol and with same contrast at two different timepoints.
The results of icobrain cross cannot be compared with the results of icobrain long.
5.5 Device Description
The following flowchart illustrates the overall architecture of icobrain.
Image /page/4/Figure/10 description: The image shows a flowchart of an image processing pipeline. The pipeline starts with input images in DICOM format, followed by preprocessing and image processing steps. Next, output generation creates reports and images. Finally, the pipeline produces a final report in PDF or DICOM format, along with output images in DICOM format.
The input images can be MR images (current icobrain software - K161148 and K180326) or CT images. During the pre-processing, the modality and/or sequence of each scan is detected and each scan is converted from DICOM format to NIFTI format. The image processing then performs the actual segmentation and calculates the measurements of the brain structures and abnormalities. Finally, the computed measurements are summarized into an electronic report and (some) segmentations are overlaid on the input images, generating output images in DICOM format.
Since the processing of MR images remains unchanged compared to the currently approved icobrain software (see KI 6 I 148 and K180326), the remainder of this file will focus on the design of the software that processes CT images. We refer to the overall architecture focused on (pre)processing CT images as the CT pipeline.
outputs of CT pipeline
Report I:
{5}------------------------------------------------
- · normalized volume of the whole brain (sum of white and grey matter)
- · normalized volume of the lateral ventricles
This report is useful to be applied in (potential) dementia patients.
Report 2:
- · measurement of the midline shift, i.e. the shift of the brain past its center line.
- · normalized volume of basal cisterns (suprasellar, quadrigeminal, prepontine)
- · volume (total, highest) of hyperdensities
This report is useful to be applied in (potential) TBI patients.
5.6 Comparison with predicate device
Likewise our clinical product icobrain 3.0 (K180326), icobrain 4.0 intends for automatic labeling, visualization and volumetric quantification of segmentable brain structures based on 3-dimensional medical images. The devices both take 3D images of the brain as input and generate an electronic report with similar quantitative information.The main difference is that icobrain 4.0 will use CT images in addition to MR images to start from. The table below compares the device to market with the proposed predicate device. The main differences are underlined.
| Predicate device | Device to market | |
|---|---|---|
| Device TradeName | icobrain | icobrain |
| Version | 3.0 | 4.0 |
| CommonName | Medical Image Processing Software | Medical Image Processing Software |
| 510(k)Number | K180326 | |
| Manufacturer | icometrix NV | icometrix NV |
| Kolonel Begaultlaan Ib / 123012 LeuvenBELGIUM | Kolonel Begaultlaan Ib / 123012 LeuvenBELGIUM | |
| RegulationNumber | 21 CFR 892.2050 | 21 CFR 892.2050 |
| DeviceClassificationName | System, Image processing, Radiological | System, Image processing, Radiological |
| Product Code | LLZ | LLZ |
| RegulatoryClass | II | II |
| ClassificationPanel | Radiology | Radiology |
| Function | Automatically identifying and quantifying the volumes ofbrain segmentable structures, automatic labeling andvisualization. | Automatically identifying and quantifying the volumes of brainsegmentable structures, automatic labeling and visualization. |
| Intended use | icobrain is intended for automatic labeling, visualization andand volumetric quantification of segmentable brain structures fromstructures from a set of MR images. This software isintended to automate the current manual process ofidentifying, labeling and quantifying the volume ofsegmentable brain structures identified on MR images.icobrain consists of two distinct image processingpipelines: icobrain cross and icobrain long.icobrain cross is intended to provide volumesfrom images acquired at a single timepoint.icobrain long is intended to provide changes involumes between two images that were acquiredon the same scanner, with the same imageacquisition protocol and with same contrast at twodifferent timepoints.The results of icobrain cross cannot be compared withthe results of icobrain long. | icobrain is intended for automatic labeling, visualization andvolumetric quantification of segmentable brain structures froma set of MR or NCCT images. This software is intended toautomate the current manual process of identifying, labelingand quantifying the volume of segmentable brain structuresidentified on MR or NCCT images.icobrain consists of two distinct image processing pipelines:icobrain cross and icobrain long.icobrain cross is intended to provide volumes fromMR or NCCT images acquired at a single time point.icobrain long is intended to provide changes involumes between two MR images that were acquiredon the same scanner, with the same image acquisitionprotocol and with same contrast at two differenttimepoints.The results of icobrain cross cannot be compared with theresults of icobrain long. |
| Technicalcharacteristics | Software packageOperates on off-the-shelf hardware (multiplevendors)DICOM compatiblesegmentation by classical machine learning(unsupervised voxel classication with GaussianMixture Models)Input: TI-weighted and fluid-attenuated inversionrecovery (FLAIR) MR images from a single ormultiple time pointsOutput:multiple electronic reports with volumetricinformation of brain structuresannotated DICOM images | Software packageOperates on off-the-shelf hardware (multiple vendors)DICOM compatiblesegmentation by classical machine learning and deeplearning (in our case supervised voxel classication withConvolutional Neural Networks)Input:TI-weighted and fluid-attenuated inversionrecovery (FLAIR) MR images from a single ormultiple time pointsnon-contrast CT from a single time pointOutput:multiple electronic reports with volumetricinformation of brain structures and midline shiftannotated DICOM images |
| Performancemeasurementtesting | Accuracybrain segmentable structure volumes / volumechanges compared to simulated and/or manuallylabeled ground truthReproducibilitybrain segmentable structure volumes / volumechanges compared on test-retest images | AccuracyMR measurementsbrain segmentable structure volumes / volumechanges compared to simulated and/ormanually labeled ground truthCT measurementslesions and midline shift: compared to manuallylabeled ground truthlateral ventricles and whole brain: MR imagessegmented by cleared icobrain 3.0 softwaretaken as ground truthReproducibilityMR measurementsbrain segmentable structure volumes / volumechanges compared on test-retest imagesCT measurementssimulation study |
| Environmentof use | icobrain is used by trained professionals in hospitals,imaging centers or in image processing labs. | icobrain is used by trained professionals in hospitals, imagingcenters or in image processing labs. |
| Testing | Product Risk AssessmentSoftware verification testsSoftware validation tests | Product Risk AssessmentSoftware verification testsSoftware validation tests |
{6}------------------------------------------------
5.7 Performance testing
To demonstrate the performance of the CT pipeline of icobrain 4.0, the measurements are validated for accuracy and reproducibility. The subjects upon whom the CT software was tested include TBI patients and potential dementia patients.
In the accuracy experiments, the lesions, basal cisterns, lateral ventricles and midline shift are compared
{7}------------------------------------------------
to manually segmented ground truth, while the lateral ventricles and whole brain volumes are compared to MR images segmented by the cleared icobrain 3.0 software. Reproducibility was also tested on CT images produced in the same scanning session. Literature review has been performed to set relevant acceptance criteria for each type of experiment. All experiments passed the acceptance criteria.
The experiments encompassed 544 subject datasets in total. Averaged over all experiments, the Pearson correlation coefficient between the compared measurements was 0.95 and the intraclass correlation coefficient was 0.94.
Besides the verification experiments, validation tests demonstrate the system as a whole provides all the capabilities necessary to operate according to its intended use.
5.8 Conclusions
The performance testing presented above establishes that the icobrain is safe and effective for its intended use. The comparison above demonstrates that the icobrain device is substantially equivalent to the predicate device.
| Declarations: | This summary includes only information that is also covered in the body of the 510(k). This summary does not contain any puffery or unsubstantiated labeling claims. This summary does not contain any raw data, i.e., contains only summary data. This summary does not contain any trade secret or confidential commercial information. This summary does not contain any patient identification information. |
|---|---|
| --------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
This document is reviewed and approved by Jan Verheyden, Vice President Traumatic Brain Injury of ico metrix, based on the present data and information.
Image /page/7/Figure/7 description: The image shows a signature and a date. The signature is a blue ink scribble. The date is October 1, 2018.
N/A