K Number
K181903

Validate with FDA (Live)

Date Cleared
2018-10-03

(79 days)

Product Code
Regulation Number
892.5700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BRAVOS Afterloader System is intended for use in the treatment of both benign and malignant disease or other conditions, for both curative and palliative intent, in the delivery of remote-controlled High Dose Rate (HDR) brachytherapy.

The Transfer Guide Tubes are intended to connect between the BRAVOS Remote Afterloader system and its range of Applicators. This connection creates a conduit for the source cable to travel through and allows the radioactive isotopes to be positioned within the patient's tumor site.

The Length Assessment Device is intended to allow the user to establish an approximate length channel prior to the afterloader performing definitive length verification.

Device Description

BRAVOS Afterloader System: The BRAVOS Afterloader System is a computer controlled remote electro/mechanical system used for automatically placing a cable incorporating an irradiated iridium pellet (sealed source) internally or close by a malignant tumor or tumor bed in a practice known as Brachytherapy.

Transfer Guide Tubes (TGT): Transfer Guide Tubes (TGT) are Brachytherapy applicator accessories. They are designed to provide a path for the dummy and source cable from the BRAVOS Afterloader System to the Applicator. The applicator end of a Transfer Guide Tube can vary in design to accommodate a range of Applicators.

Length Assessment Device (LAD): The Length Assessment Device (LAD) is a Brachytherapy applicator accessory and is used to determine the approximate length of the inner lumen of the Transfer Guide Tubes (TGT) and applicator assembly.

AI/ML Overview

The provided document is a 510(k) Premarket Notification Summary for the Varian BRAVOS Afterloader Family. It focuses on establishing substantial equivalence to predicate devices rather than proving device performance against specific acceptance criteria in a detailed study.

Therefore, much of the requested information (like specific acceptance criteria values, reported performance, sample sizes for test sets, expert involvement, adjudication methods, MRMC studies, standalone performance, and detailed ground truth establishment for training) is not present in this document.

However, based on the information provided, here's what can be extracted and inferred:

1. A table of acceptance criteria and the reported device performance:

The document states that "Verification testing was performed to demonstrate that the performance and functionality of the new BRAVOS Afterloader Family met the initial design input requirements. Verification testing was performed to verify the integrity of any changes. Validation testing was performed on production equivalent devices, under clinically representative conditions by qualified personnel."

It also states, "The non-clinical data for BRAVOS Afterloader Family supports the safety of the devices and the software verification and validation demonstrate that the subject devices should perform as intended in the specified use conditions."

The document does not provide a specific table of acceptance criteria or quantitative performance metrics. Instead, it broadly states that the device met "initial design input requirements" and that "performance and functionality" were verified. The acceptance criteria for a device like this would likely be related to safety and functionality, such as:

Acceptance Criterion (Inferred from documentation)Reported Device Performance (Inferred from documentation)
Meets design input requirementsPerformance and functionality met initial design input requirements.
Verification of changes integrityVerification testing performed to verify integrity of changes.
Clinically representative validationValidation testing performed on production equivalent devices under clinically representative conditions.
Safety of devicesNon-clinical data supports the safety of the devices.
Software performs as intendedSoftware verification and validation demonstrate anticipated performance.
Substantial Equivalence to PredicateConcluded to be safe, effective, and substantially equivalent to predicate devices.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size: Not specified. The document only mentions "production equivalent devices" being used for validation testing.
  • Data Provenance: Not specified. The document does not indicate the country of origin or whether the data was retrospective or prospective. It does state that "No animal studies or clinical tests have been included in this pre-market submission," implying that the testing was likely bench or lab-based.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

Not specified. The document mentions "qualified personnel" performing validation testing, but does not provide details on their number or qualifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not specified.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device (BRAVOS Afterloader System, Transfer Guide Tubes, and Length Assessment Device) is a standalone medical device used for brachytherapy, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The document describes the performance of the device itself (BRAVOS Afterloader System, Transfer Guide Tubes, and Length Assessment Device) and its software. The testing performed ("Verification testing" and "Validation testing") is essentially assessing the standalone performance of the device without explicit human-in-the-loop as a variable in the performance assessment, beyond operators running the tests.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

Not explicitly stated. Given that the testing involved "initial design input requirements" and "clinically representative conditions," the "ground truth" for verification and validation would likely be defined by engineering specifications, physical measurements, and expected functional output as determined by design documents and industry standards for brachytherapy devices.

8. The sample size for the training set:

Not applicable / Not specified. This device is described as a "computer controlled remote electro/mechanical system" and related accessories. While it has "console software," the document does not suggest it is an AI/ML device that requires a distinct "training set" in the conventional sense of machine learning. The term "training set" is usually associated with AI models that are trained on data. This documentation describes typical software verification and validation for a medical device.

9. How the ground truth for the training set was established:

Not applicable. (See point 8)

{0}------------------------------------------------

January 7, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304

Re: K181903

Trade/Device Name: BRAVOS Afterloader Family: BRAVOS Afterloader System, Transfer Guide Tubes, and Length Assessment Device Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAQ

Dear Mr. Coronado:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated 10/03/2018. Specifically, FDA is updating this SE Letter as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Sullivan, OHT7: Office of In Vitro Diagnostics and Radiological Health, (240) 402-4973, Julie.Sullivan@fda.hhs.gov.

Sincerely.

Michael D. O'Hara

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, which features a stylized human figure. To the right of that is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Varian Medical Systems, Inc. c/o Peter J. Coronado Director, Regulatory Affairs 3100 Hansen Way PALO ALTO, CA 94304

October 3, 2018

Re: K181903

Trade/Device Name: BRAVOS Afterloader Family: BRAVOS Afterloader System, Transfer Guide Tubes, and Length Assessment Device Regulation Number: 21 CFR 892.5700 Regulation Name: Remote Controlled Radionuclide Applicator System Regulatory Class: Class II Product Code: JAQ Dated: August 20, 2018 Received: August 21, 2018

Dear Mr. Coronado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{2}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and

Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K181903

Device Name

BRAVOS Afterloader Family: BRAVOS Afterloader System, Transfer Guide Tubes, and Length Assessment Device

Indications for Use (Describe)

The BRAVOS Afterloader System is intended for use in the treatment of both benign and malignant disease or other conditions, for both curative and palliative intent, in the delivery of remote-controlled High Dose Rate (HDR) brachytherapy.

The Transfer Guide Tubes are intended to connect between the BRAVOS Remote Afterloader system and its range of Applicators. This connection creates a conduit for the source cable to travel through and allows the radioactive isotopes to be positioned within the patient's tumor site.

The Length Assessment Device is intended to allow the user to establish an approximate length channel prior to the afterloader performing definitive length verification.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "varian" in a bold, sans-serif font. The letters are all lowercase and evenly spaced. The word is likely a logo or brand name.

Varian Medical Systems

3100 Hansen Way Palo Alto, CA 94304

650.493.4000 800.544.4636

varian.com

Image /page/4/Picture/5 description: The image contains three blue circles in a horizontal row. The circles are evenly spaced and appear to be the same size. The background is white, which makes the blue circles stand out. There is nothing else in the image.

K181903 Premarket Notification 510(K) Summary

The following information is provided according to 21 CFR 807.92.

Submitter:Varian Medical Systems, Inc.3100 Hansen Way Palo Alto, CA 94304Contact Name: Peter J. CoronadoPhone: 650.424.6320Fax: 650.646.9200E-mail: submissions.support@varian.comDate Prepared: October 2, 2018
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Trade/ProprietaryNames:BRAVOS Afterloader Family BRAVOS Afterloader System Transfer Guide Tubes (TGT) Length Assessment Device (LAD)
BRAVOS Afterloader System, Transfer Guide Tubes (TGT), Length Assessment Device (LAD)Classification Name:Remote controlled radionuclideapplicator system, 21 CFR §892.5700Common/Usual Name:Afterloader SystemSource Guide TubesBrachytherapy AccessoryRegulatory Class:Class IIProduct Code:JAQPredicate Devices:(K120993)GammaMedplus iX Series(K141336)GammaMedplus Source Guide Tubes(K952913)Applicators for Varian VariSource Remote HDRAfterloader➤ Accessory: Measurement Ruler, Marker Wire& Clip Set
Device
Description:BRAVOS Afterloader SystemThe BRAVOS Afterloader System is a computer controlled remote electro/mechanicalsystem used for automatically placing a cable incorporating an irradiated iridium pellet (sealedsource) internally or close by a malignant tumor or tumor bed in a practice known asBrachytherapy.Transfer Guide Tubes (TGT)Transfer Guide Tubes (TGT) are Brachytherapy applicator accessories. They are designed toprovide a path for the dummy and source cable from the BRAVOS Afterloader System to theApplicator. The applicator end of a Transfer Guide Tube can vary in design to accommodate arange of Applicators.Length Assessment Device (LAD)The Length Assessment Device (LAD) is a Brachytherapy applicator accessory and is usedto determine the approximate length of the inner lumen of the Transfer Guide Tubes (TGT)and applicator assembly.
Intended/IndicationsFor UseStatement:The indications for use and the intended use are substantially the same in the subject devices as for the predicate.
BRAVOS Afterloader System
Intended UseIndications for Use
The BRAVOS Afterloader System is intended foruse in the treatment of both benign and malignantdisease or other conditions, for both curative andpalliative intent, in the delivery of remote-controlledHigh Dose Rate (HDR) brachytherapy.The Bravos Afterloader System is indicated for usein the treatment of both benign and malignantdisease or other conditions, for both curative andpalliative intent, in the delivery of remote-controlledHigh Dose Rate (HDR) brachytherapy.
Transfer Guide Tubes (TGT)
Intended UseIndications for Use
The Transfer Guide Tubes are intended to connectbetween the BRAVOS Remote Afterloader systemand its range of Applicators. This connectioncreates a conduit for the source cable to travelthrough and allows the radioactive isotopes to bepositioned within the patient's tumor site.The Transfer Guide Tubes are intended to connectbetween the Bravos Remote Afterloader systemand its range of applicators. This connectioncreates a conduit for the source cable to travelthrough and allows the radioactive isotopes to bepositioned within the patient's tumor site.
Length Assessment Device (LAD)
Intended UseIndications for Use
The Length Assessment Device is intended toallow the user to establish an approximate lengthof an unknown length channel prior to theafterloader performing definitive length verification.The Length Assessment Device is intended toallow the user to establish an approximate length ofan unknown length channel prior to the afterloaderperforming definitive length verification.
SourceActivitySubject Device: BRAVOS Afterloader SystemMaximum shielding activity: 555 GBq / 15 CiMaximum treatment activity: 555 GBq / 15 CiPredicate Device: GammaMedplus iX SeriesMaximum shielding activity: 555 GBq / 15 CiMaximum treatment activity: 555 GBq / 15 Ci

{5}------------------------------------------------

The three devices are used together during a diagnostic procedure and are bundled in this Traditional 510(k) submission so they can be addressed during one review; BRAVOS Afterloader System, Transfer Guide Tubes (TGT), and Length Assessment Device (LAD). The indications for use and the intended use are similar in the subject devices as for the predicate. The source activity was not changed since the predicate.

{6}------------------------------------------------

Comparison of Technological Characteristscs with the Predicate Device

At a high level, the subject and predicate devices are based on the following similar technological elements:

BRAVOS Afterloader System:

  • Similar Design and Technology ●
  • o Similar Console Software (subject device updated to version 2.0)

Transfer Guide Tubes (TGT):

  • Similar Design
  • Composed of same materials ●
  • Non-patient contacting (N/A Biocompatible) o
  • Provided Non-sterile

Length Assessment Device (LAD):

  • Similar Compatibility with its respective Afterloader System ●
  • Non-patient contacting (N/A Biocompatible)
  • Provided Non-sterile

Significant Difference

BRAVOS Afterloader System: The significant difference compared to the predicate devices are

  • The intended and indications for use changed to exclude Pulsed Dose Rate (PDR) mode ●
  • Updated Design and Technology of Remote-controlled Afterloader .
  • BRAVOS Control Software (version 2.0) ●
  • BRAVOS CamScale Device ●
  • BRAVOS Source Cable

The BRAVOS Afterloader System and its predicate are remote-controlled afterloading devices for brachytherapy. The systems both utilize a small, high activity Iridium-192 source that is encapsulated in a steel capsule, which is fixed to a stainless-steel cable and moves into applicator(s) inserted into the patient.

Transfer Guide Tubes (TGT):

  • . Updated Intended and Indications for Use statements
  • New Guide Tubes and lengths
Transfer Guide Tubes (TGT) for APPLICATORS
Product NumberDescription
GM11010800Transfer Guide Tube for 320 mm applicators
GM11010810Coded Transfer Guide Tube Channel 1, for 320 mm applicators
GM11010820Coded Transfer Guide Tube Channel 2, for 320 mm applicators
GM11010830Coded Transfer Guide Tube Channel 3, for 320 mm applicators
GM11010840Transfer Guide Tube for 113 mm applicators
GM11010850Transfer Guide Tube for 200 mm applicators
GM11010860Transfer Guide Tube for 250 mm applicators
GM11010870Transfer Guide Tube, 1000 mm length for applicators

{7}------------------------------------------------

Transfer Guide Tubes (TGT) for CATHETERS
Product NumberDescription
GM11011860Transfer Guide Tube for 113 mm catheters
GM11011870Transfer Guide Tube for 200 mm catheters
GM11011880Transfer Guide Tube for 250 mm catheters
GM11011890Transfer Guide Tube for 320 mm catheters
GM11011910Transfer Guide Tube, 1000 mm length for catheters
LUER Transfer Guide Tube (TGT)
Product NumberDescription
GM11010880Luer Transfer Guide Tube for MammoSite (1500 mm treatment length)
Optional Accessory
Product NumberDescription
GM11001430Wall holder for source and transfer guide tubes
GM11001390Source and transfer guide tube support

Length Assessment Device (LAD):

  • Updated Intended and Indications for Use statements ●
  • New design ●
  • New materials ●
Length Assessment Device (LAD)
Product NumberDescription
GM11011800Length assessment device

Performance Data

Software verification and validation was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for BRAVOS Afterloader System was considered as a "major" level of concern.

Varian's BRAVOS Afterloader Family which consists of the BRAVOS Afterloader System and two accessories; Transfer Guide Tubes (TGT) and Length Assessment Device (LAD) are substantially equivalent to the predicate devices K120993, K141336, and K952913 respectively. Compared with the predicate device, the basic operation and technological characteristics are substantially the same. The indications for use and the intended use for each device are substantially the same as the predicate.

No animal studies or clinical tests have been included in this pre-market submission.

Verification testing was performed to demonstrate that the performance and functionality of the new BRAVOS Afterloader Family met the initial design input requirements. Verification testing was performed to verify the integrity of any changes. Validation testing was performed on production equivalent devices, under clinically representative conditions by qualified personnel.

{8}------------------------------------------------

varie

Standards Conformance

The subject devices conform in whole or in part with the following standards:

  • AAMI/ ANSI 60601-1: 2005 ●
  • IEC 60601-1-2: 2014
  • IEC 60601-1-8:2006 + A1:2012 ●
  • . IEC 60601-2-17:2013
  • IEC 62304: 2006
  • IEC 60601-1-6: 2013-10
  • IEC 60825-1: 2007
  • EN ISO 10993-1:2009
  • EN ISO 10993-5:2009 ●
  • EN ISO 10993-10: 2010
  • EN ISO 10993-18:2009
  • AAMI TIR-12:2010
  • . AAMI TIR-30: 2011
  • EN ISO 17664:2004
  • IEC 62366:2007/(R)2014
  • EN ISO 14971:2012 ●

The subject devices also comply with the following non-FDA recognized standard:

  • EN ISO 13485:2016 .

Conclusion

The non-clinical data for BRAVOS Afterloader Family supports the safety of the devices and the software verification and validation demonstrate that the subject devices should perform as intended in the specified use conditions. Varian therefore considers the BRAVOS Afterloader Family which consists of BRAVOS Afterloader System and two accessories; Transfer Guide Tubes (TGT) and Length Assessment Device (LAD) to be safe and effective and are substantially equivalent to the predicate.

§ 892.5700 Remote controlled radionuclide applicator system.

(a)
Identification. A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.(b)
Classification. Class II.