K Number
K181831

Validate with FDA (Live)

Date Cleared
2018-11-15

(129 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Philips Multi-Measurement Module (MMX) is for use with Philips IntelliVue Patient Monitors.

The Multi-Measurement Module, with its host patient monitor, is intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare facilities.

It sends processed measurement data (for example, measurement waves and numerics) to the monitor screen, generates alerts, and supports the transfer of patient data between monitors.

The Multi-Measurement Module is intended for use by trained healthcare professionals in a hospital environment. The Multi-Measurement Module is additionally intended for use in transport situations within hospital environments together with its host patient monitor.

The Multi-Measurement Module is only for use on one patient at a time. It is not intended for home use. It is not a therapeutic device. The Multi-Measurement Module is for prescription use only.

The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused.

Device Description

The newly introduced IntelliVue Multi-Measurement Module MMX acquires multiple physiological patient signals, sends processed measurement data (for example, measurement waves and numerics) to the host monitor screen, generates alerts, and supports the transfer of patient data between monitors.

The device, together with its compatible host monitor, offers a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. It can be located in the patient vicinity at the bedside or can be used mobile, during patient transport inside hospitals.

The measurement sensors are applied at diverse body locations, depending on the physiological parameters monitored.

The new device has the same range of functions as the legally marketed predicate devices; it provides the same measurement parameters as the predicates.

The Multi-Measurement Module MMX provides multiple non-invasive and invasive measurements: ECG (including arrhythmia and ST), respiration, SpO2, NBP, dual invasive pressure, temperature, and CO2.

AI/ML Overview

The Philips IntelliVue Multi-Measurement Module MMX is a device intended for monitoring, recording, and alarming multiple physiological parameters in adults, pediatrics, and neonates within healthcare facilities. The following details describe its acceptance criteria and the study that proves the device meets those criteria.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes a validation and verification (V&V) study, but it does not present a single consolidated table of acceptance criteria with corresponding performance statistics. Instead, it refers to fulfilling requirements of various standards and guidance documents. Based on the "Summary of V&V Activities" section, the acceptance criteria are generally defined as meeting all applicable requirements of recognized consensus standards and FDA guidance documents, and that "All specified pass/fail criteria have been met."

Since specific numerical acceptance criteria and reported device performance values are not explicitly listed in a table, here's a conceptual representation of what such a table would contain, based on the types of testing mentioned:

Acceptance Criterion (Based on Standards/Guidance)Reported Device Performance (Implied by "All requirements met")
Software Life Cycle Processes: AAMI/ANSI/IEC 62304:2006 complianceDemonstrated compliance with all requirements of AAMI/ANSI/IEC 62304:2006.
Basic Safety & Essential Performance: AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 complianceDemonstrated compliance with all requirements of AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1).
Electromagnetic Compatibility: IEC 60601-1-2:2014 complianceDemonstrated compliance with all requirements of IEC 60601-1-2:2014 (Ed. 4.0).
Alarms: IEC 60601-1-8:2012 complianceDemonstrated compliance with all requirements of IEC 60601-1-8:2012 (Ed. 2.1).
Electrocardiographs: IEC 60601-2-25:2011 complianceDemonstrated compliance with all requirements of IEC 60601-2-25:2011 (Ed. 2.0).
ECG Monitoring Equipment: IEC 60601-2-27:2011 complianceDemonstrated compliance with all requirements of IEC 60601-2-27:2011 (Ed. 3.0).
Automated Noninvasive Sphygmomanometers: ISO 80601-2-30:2013 complianceDemonstrated compliance with all requirements of ISO 80601-2-30:2013 (Ed. 1.1).
Invasive Blood Pressure Monitoring Equipment: IEC 60601-2-34:2011 complianceDemonstrated compliance with all requirements of IEC 60601-2-34:2011 (Ed. 3.0).
Respiratory Gas Monitors: ISO 80601-2-55:2011 complianceDemonstrated compliance with all requirements of ISO 80601-2-55:2011 (Ed. 1).
Clinical Thermometers: ISO 80601-2-56:2009 complianceDemonstrated compliance with all requirements of ISO 80601-2-56:2009 (Ed. 1).
Pulse Oximeters: ISO 80601-2-61:2011 complianceDemonstrated compliance with all requirements of ISO 80601-2-61:2011 (Ed. 1).
Pulse Oximeter Accuracy: Per FDA guidanceMet accuracy requirements for SpO2 and pulse rate values.
Pulse Oximeter Communication Integrity: Per FDA guidanceSpO2 and pulse rate values not corrupted during communication to host device.
Pulse Oximeter Display, Outputs, Indicators: Per FDA guidanceDisplay values, outputs, and indicators met requirements.
Non-Invasive Blood Pressure (NBP) Intra-Device Variability: Per FDA guidance (minimum 3 devices)Demonstrated acceptable intra-device variability between NBP devices.
NBP Comparison to Intra-Arterial Reference: Per FDA guidance (mean blood pressure)Demonstrated acceptable comparison to intra-arterial reference standard for mean blood pressure.
Hazard Analysis Effectiveness: All specified pass/fail criteria metTest results confirmed effectiveness of implemented design risk mitigation measures.
Environmental Testing (Temperature, Humidity): Specified requirements met for general hospital, handling, transport, storageDemonstrated safe and according to specifications under simulated environmental conditions.
Mechanical Testing (7M1, 7M2, 7M3): Specified requirements metDemonstrated safe and according to specifications under simulated mechanical stress conditions.
Integration Tests: Per specified criteriaPerformed according to specified criteria.
Functional Tests for Physiological Parameters: Per specified criteriaPerformed according to specified criteria.
Regression Tests for Patient Monitor Compatibility and Operating System: Per specified criteriaPerformed according to specified criteria, confirming modified and previously available functions work correctly.

2. Sample Size and Data Provenance for the Test Set

The document does not explicitly state the sample size for any specific test set beyond "a minimum of three devices" for non-invasive blood pressure intra-device variability testing.
The origin of the data (country, retrospective/prospective) is not specified. The testing appears to be conducted by the manufacturer, Philips Medizin Systeme Boeblingen GmbH, based in Germany. The nature of the V&V activities suggests prospective testing against established standards and internal specifications.

3. Number of Experts and Qualifications for Ground Truth

The document does not mention the use of experts to establish ground truth for the test set. The validation relies on meeting criteria defined by recognized consensus standards and FDA guidance rather than expert consensus on individual cases. For parameters like SpO2 and NBP, ground truth would typically be established by validated reference methods (e.g., co-oximetry for SpO2, invasive arterial pressure for NBP).

4. Adjudication Method

No adjudication method is mentioned, as the validation strategy is based on direct measurement against technical standards and reference methods, not subjective expert review of cases.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC comparative effectiveness study was not conducted. This device is a physiological monitoring module, and its performance is evaluated against technical standards and reference measurements, not by human reader interpretation of outputs requiring AI assistance.

6. Standalone (Algorithm Only) Performance

A standalone performance evaluation was done. The entire "Summary of V&V Activities" section describes the testing of the device itself (the IntelliVue Multi-Measurement Module MMX) against various technical standards without human-in-the-loop intervention for the core physiological measurements. The measurements (ECG, respiration, SpO2, NBP, invasive pressure, temperature, CO2) are automatically processed and displayed by the device.

7. Type of Ground Truth Used

The ground truth used for testing is primarily based on:

  • Reference Standards: Performance specifications defined by international standards (e.g., IEC, ISO, AAMI/ANSI) for physiological measurement devices.
  • Reference Methods: For parameters like SpO2 and NBP, the ground truth would typically involve comparison against highly accurate and validated reference measurement techniques (e.g., co-oximetry for SpO2, intra-arterial catheter for NBP).
  • Internal Specifications: Philips' own defined specifications for the device's behavior and performance.

The document does not mention pathology, outcomes data, or expert consensus on patient cases as ground truth for the device's technical validation.

8. Sample Size for the Training Set

The document does not mention a training set. This document describes the validation of a medical device with established algorithms for physiological parameter measurement, not a machine learning or AI-based system that typically requires a distinct training phase with a large dataset. The underlying algorithms for most physiological monitors are well-established and are typically validated through engineering and clinical testing rather than "training" in the machine learning sense.

9. How the Ground Truth for the Training Set Was Established

As no training set is described or implied for this type of medical device's validation, no information is provided on how its ground truth would have been established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 15, 2018

Philips Medizin Systeme Boeblingen GmbH Stefan Breuer Senior Regulatory Affairs Engineer Hewlett-Packard-Str. 2 Boeblingen, 71034 De

Re: K181831

Trade/Device Name: IntelliVue Multi-Measurement Module MMX Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DSJ, DSK, DXN, DXG, DRQ, DRT, DPS, MLC, DRJ, DQA, DSB, DSA, MSX, CCK, BZO, FLL Dated: October 11, 2018 Received: October 15, 2018

Dear Stefan Breuer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Arielle Drummond -S

For

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K181831

Device Name IntelliVue Multi-Measurement Module MMX

Indications for Use (Describe)

The Philips Multi-Measurement Module (MMX) is for use with Philips IntelliVue Patient Monitors.

The Multi-Measurement Module, with its host patient monitor, is intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare facilities.

It sends processed measurement data (for example, measurement waves and numerics) to the monitor screen, generates alerts, and supports the transfer of patient data between monitors.

The Multi-Measurement Module is intended for use by trained healthcare professionals in a hospital environment. The Multi-Measurement Module is additionally intended for use in transport situations within hospital environments together with its host patient monitor.

The Multi-Measurement Module is only for use on one patient at a time. It is not intended for home use. It is not a therapeutic device. The Multi-Measurement Module is for prescription use only.

The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused.

Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to
review instructions, search existing data sources, gather and maintain the data needed and complete and review the
collection of information. Send comments regarding this burden estimate or any other aspect of this information
collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM EDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443 6740
EE

FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

5.1 Submitter

The submitter of this Premarket Notification is:

Stefan Breuer Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 71034 Boeblingen Germany Telephone: +49 7031 463-2321 +49 7031 463-2442 Fax: Email: stefan.breuer@philips.com This summary was prepared on July 04, 2018.

5.2 Name and Classification of the Devices

Trade names: IntelliVue Multi-Measurement Module MMX Common name: Multifunction Patient Monitor Module

Classification:

Device PanelClassificationProCodeDescription
CardiovascularDevices§870.1025, IIMHXMonitor, Physiological, Patient (witharrhythmia detection or alarms)
CardiovascularDevices§870.1025, IIDSIDetector and alarm, arrhythmia
CardiovascularDevices§870.1025, IIMLDMonitor, ST Segment with Alarm
CardiovascularDevices§870.1100, IIDSJAlarm, Blood Pressure
CardiovascularDevices§870.1110, IIDSKComputer, Blood Pressure
CardiovascularDevices§870.1130, IIDXNSystem, Measurement, Blood-Pressure, Non-Invasive
CardiovascularDevices§870.1435, IIDXGComputer, Diagnostic, Pre-Programmed, Single-Function
CardiovascularDevices§870.2060, IIDRQAmplifier and Signal Conditioner,Transducer Signal
CardiovascularDevices§870.2300, IIDRTMonitor, Cardiac (incl.Cardiotachometer & Rate Alarm)
CardiovascularDevices§870.2340, IIDPSElectrocardiograph
CardiovascularDevices§870.2340, IIMLCMonitor, ST Segment
CardiovascularDevices§870.2600, IDRJSystem, Signal Isolation
CardiovascularDevices§870.2700, IIDQAOximeter
CardiovascularDevices§870.2770, IIDSBPlethysmograph, Impedance
CardiovascularDevices§870.2900, IDSACable, Transducer and Electrode,incl. Patient Connector
-MSXSystem, Network andCommunication, PhysiologicalMonitors

{4}------------------------------------------------

Device PanelClassificationProCodeDescription
AnesthesiologyDevices§868.1400, IICCKAnalyzer, Gas, Carbon Dioxide,Gaseous-Phase
§868.2375, IIBZQMonitor, Breathing Frequency
General Hospital andPersonal UseDevices§880.2910, IIFLLThermometer, Electronic, Clinical

5.3 Substantial Equivalence

The newly introduced IntelliVue Multi-Measurement Module MMX is substantially equivalent to

  • . the legally marketed IntelliVue Multi-Measurement Module X1 with Capnography Extension M3014A, marketed pursuant to K161531 and K050762
  • the legally marketed IntelliVue Multi-Measurement Module X3, marketed pursuant ● to K171801.

5.4 Description of the Device

The newly introduced IntelliVue Multi-Measurement Module MMX acquires multiple physiological patient signals, sends processed measurement data (for example, measurement waves and numerics) to the host monitor screen, generates alerts, and supports the transfer of patient data between monitors.

The device, together with its compatible host monitor, offers a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. It can be located in the patient vicinity at the bedside or can be used mobile, during patient transport inside hospitals.

The measurement sensors are applied at diverse body locations, depending on the physiological parameters monitored.

The new device has the same range of functions as the legally marketed predicate devices; it provides the same measurement parameters as the predicates.

The Multi-Measurement Module MMX provides multiple non-invasive and invasive measurements: ECG (including arrhythmia and ST), respiration, SpO2, NBP, dual invasive pressure, temperature, and CO2.

5.5 Intended Use

The Philips Multi-Measurement Module MMX is for use with Philips IntelliVue Patient Monitors.

The Multi-Measurement Module, with its host patient monitor, is intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in health care facilities.

It sends processed measurement data (for example, measurement waves and numerics) to the monitor screen, generates alerts, and supports the transfer of patient data between monitors.

{5}------------------------------------------------

The Multi-Measurement Module is intended for use by trained healthcare professionals in a hospital environment. The Multi-Measurement Module is additionally intended for use in transport situations within hospital environments together with its host patient monitor.

The Multi-Measurement Module is only for use on one patient at a time. It is not intended for home use. It is not a therapeutic device. The Multi-Measurement Module is for prescription use only.

The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused.

5.6 Technological Characteristics

The new IntelliVue Multi-Measurement Module MMX has the same technological characteristics as its predicate device IntelliVue Multi-Measurement Module X1 with Capnography Extension M3014A. It has the same operating principles, mechanism of action, energy sources, portability, user interface, and comparable robustness. Dimensions, weight and environmental specifications are very similar.

The new IntelliVue Multi-Measurement Module MMX also has the same technological characteristics as its predicate device IntelliVue Multi-Measurement Module X3. It has the same operating principles, mechanism of action, portability and robustness. The implementation of the physiological measurements are identical in hardware and software. Dimensions, weight and environmental specifications are very similar.

5.7 Summary of V&V Activities

The new devices have been subject to a series of V&V activities:

  • Testing according to the recognized consensus standards: ●
    • o AAMI/ANSI/IEC 62304:2006 (Software life cycle processes)
    • o AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1) (Basic safety and essential performance)
    • IEC 60601-1-2:2014 (Ed. 4.0) (Electromagnetic Compatibility) O
    • IEC 60601-1-8: 2012 (Ed. 2.1) (Alarms) O
    • IEC 60601-2-25: 2011 (Ed. 2.0) (Electrocardiographs) O
    • IEC 60601-2-27: 2011 (Ed. 3.0) (ECG monitoring equipment) O
    • IEC 80601-2-30: 2013 (Ed. 1.1) (Automated noninvasive O sphygmomanometers)
    • O IEC 60601-2-34: 2011 (Ed. 3.0) (Invasive blood pressure monitoring equipment)
    • ISO 80601-2-55: 2011 (Ed. 1) (Respiratory gas monitors) O
    • ISO 80601-2-56: 2009 (Ed. 1) (Clinical thermometers) O
    • ISO 80601-2-61: 2011 (Ed. 1) (Pulse Oximeters) O

{6}------------------------------------------------

All applicable requirements have been met.

  • Additional relevant testing according to the device-specific FDA guidance ● documents:
    • o "Pulse Oximeters Premarket Notification Submissions [510(k)s] -Guidance for Industry and Food and Drug Administration Staff"
      • Accuracy of Pulse Oximeters ■
      • Testing that demonstrates that SpO₂ and pulse rate values calculated . by the OEM system are not corrupted during communication to the host device
      • . Display values, outputs, and indicators.
      • All applicable requirements have been met.
    • "Non-Invasive Blood Pressure (NBP) Monitor Guidance" O
      • The intra-device variability between a minimum of three devices
      • . Comparison to the intra-arterial reference standard for mean blood pressure.

All applicable requirements have been met.

  • Testing as required by the Hazard Analysis ● All specified pass/fail criteria have been met. The test results have confirmed the effectiveness of the implemented design risk mitigation measures.
  • Additional environmental testing (temperature, humidity) and mechanical testing ● (mechanical classes 7M1, 7M2, 7M3) All specified test requirements have been met. The tests have confirmed that the new devices work safely and according to their specifications and indicated claims during tests simulating general hospital conditions, handling and transport in hospital environments and storage.
  • Integration tests, functional system level tests, and regression system level tests ● The new devices perform according to the specified criteria:
    • o Integration tests
    • o Functional tests for physiological parameters

Regression tests for patient monitor compatibility and operating system o All specified test requirements have been met. The test results demonstrate that modified and previously available device functions work correctly according to the specifications and labeling claims.

5.8 Conclusion

Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicates.

Testing comprised electrical and mechanical safety tests, EMC tests, environmental tests, performance tests, functional and regression tests.

Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence.

The results demonstrate that the new devices meet all defined reliability requirements and performance claims.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.