K Number
K181708

Validate with FDA (Live)

Date Cleared
2018-07-31

(33 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Liofilchem® MTS™ (MIC Test Strip) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS™ consists of specialized paper impregnated with a pre-defined concernment of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in ug/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. The MTS™ Plazoni.cin at concentrations of 0.016-256 ug/mL should be interpreted at 16-20 hours of incubation.

MTSTM PLZ can be used to determine the MC of plazomicin against the microorganisms listed in the table below:

Plazomicin Activity According to the FDA Label

Clinical and in vitro

Enterobacter cloacae Escherichia coli Klebsiella pneumoniae Proteus mirabilis

in vitro only Citrobacter freundii Citrobacter koseri Klebsiella (Enterobacter) aerogenes Klebsiella oxytoca Morganella morganii Proteus vulgaris Providencia stuartii Serratia marcescens

Device Description

MTS™ consists of specialized paper impregnated with a pre-defined concernment of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in ug/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures.

AI/ML Overview

This document describes the FDA 510(k) clearance for the Liofilchem MTS Plazomicin 0.016-256 ug/mL device, which is an antimicrobial susceptibility test. The information provided focuses on the regulatory clearance process rather than a detailed study demonstrating acceptance criteria and performance against an AI model.

Therefore, I cannot provide a response that directly addresses the user's request for:

  • A table of acceptance criteria and reported device performance related to an AI model.
  • Sample sizes, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth details specifically for an AI model.

The provided document describes a traditional in vitro diagnostic device, not an AI-powered one. The "acceptance criteria" discussed in the document are related to the substantial equivalence determination for regulatory clearance against a predicate device, and the "performance" refers to the ability of the MTS Plazomicin strips to accurately determine MICs of antimicrobial agents, as established by standard microbiological methods.

Based on the provided text, none of the requested information regarding AI model performance, studies, or ground truth establishment is available. The document is a 510(k) clearance letter for an antimicrobial susceptibility test strip, not a study report for an AI-powered diagnostic device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 31, 2018

Liofilchem s. r. 1. % Laura Koeth President Laboratory Specialists, Inc. 26214 Center Ridge Road Westlake, Ohio 44145

Re: K181708

Trade/Device Name: MTS Plazomicin 0.016-256 ug/mL Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: Class II Product Code: JWY Dated: June 27, 2018 Received: June 28, 2018

Dear Ms. Koeth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Steven R. Gitterman -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K181708

Device Name

MTSTM Plazomicin 0.016-256 ug/mL

Indications for Use (Describe)

The Liofilchem® MTS™ (MIC Test Strip) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS™ consists of specialized paper impregnated with a pre-defined concernment of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in ug/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. The MTS™ Plazoni.cin at concentrations of 0.016-256 ug/mL should be interpreted at 16-20 hours of incubation.

MTSTM PLZ can be used to determine the MC of plazomicin against the microorganisms listed in the table below:

Plazomicin Activity According to the FDA Label

Clinical and in vitro

Enterobacter cloacae Escherichia coli Klebsiella pneumoniae Proteus mirabilis

in vitro only Citrobacter freundii Citrobacter koseri Klebsiella (Enterobacter) aerogenes Klebsiella oxytoca Morganella morganii Proteus vulgaris Providencia stuartii Serratia marcescens

Type of Use (Select one or both, as applicable)

O Prescription Use (Part 21 CFR 801 Subpart D)

O Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"Aft agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: 0MB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).