K Number
K181659

Validate with FDA (Live)

Device Name
Avologi ENEO
Date Cleared
2018-07-26

(31 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Avologi ENEO is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.

Device Description

Avologi ENEO is a battery operated device that uses low power light spectrum at red and infrared LED, at wavelength of 633±5nm, 830±5nm emitting optical power in a uniform distribution with no hot spots. It is a hand held light emitting diode(LED) device for the treatment of periorbital wrinkles designed for home-use.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Avologi ENEO device, which is an over-the-counter light-emitting diode (LED) device for the treatment of periorbital wrinkles. However, the submission explicitly states that "Clinical data was not including in this submission." This means there is no study presented in this document that proves the device meets specific acceptance criteria based on clinical performance.

Instead, the submission relies on demonstrating substantial equivalence to legally marketed predicate devices through non-clinical testing and comparison of technical specifications.

Here's a breakdown of the requested information based on the provided text, highlighting where information is absent due to the nature of this 510(k) submission:

1. Table of Acceptance Criteria and Reported Device Performance

Since no clinical study was conducted for this 510(k) submission to demonstrate performance against specific clinical acceptance criteria, such a table cannot be provided from this document. The "acceptance criteria" here refers to the device meeting various safety and technical standards for substantial equivalence.

Acceptance Criteria (Non-Clinical Standards)Reported Device Performance (Compliance)
Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R)2012)Complies
Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Test (IEC 60601-1-2 Edition 4.0 2014-02)Complies
Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment (IEC 60601-1-11 Edition 2.0 2015-01)Complies
Photobiological Safety of Lamps and Lamp Systems (IEC 62471 First Edition 2006-07)Complies
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (ISO 10993-1:2009)Complies
Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity (ISO 10993-5:2009/(R)2014)Complies
Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity (ISO 10993-10 Third Edition 2010-08-01)Complies

Note: The performance reported is compliance with the listed standards, not clinical effectiveness for wrinkle reduction, as no clinical study was presented.

2. Sample size used for the test set and the data provenance

Not applicable. No clinical test set data from human subjects was used for this submission. The "test set" in this context refers to the device itself being tested against non-clinical standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. No clinical study with a "ground truth" established by experts was performed or included in this submission.

4. Adjudication method for the test set

Not applicable. No clinical study requiring adjudication was performed or included in this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an LED therapy device, not an AI-assisted diagnostic or interpretive tool. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device does not involve an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. No clinical ground truth was established as no clinical study was conducted. The "ground truth" for the non-clinical tests would be the specifications and requirements of the various international standards.

8. The sample size for the training set

Not applicable. This device does not use an algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This device does not use an algorithm that requires a training set.


Summary of the Document's Conclusion:

The 510(k) submission for Avologi ENEO establishes substantial equivalence based on:

  • Non-clinical bench tests: Verifying the device met design specifications and complied with electrical safety, EMC, home healthcare, photobiological safety, and biocompatibility standards (ISO 10993 suite).
  • Comparison to predicate devices: Demonstrating similar indications for use, product code, device class, technology (LED, wavelengths, energy density, temperature), materials, and use environment (OTC, handheld, for periorbital wrinkles).

The submission explicitly states that no clinical performance data was included. Therefore, the acceptance criteria and study proving performance in the traditional sense of clinical efficacy are not detailed in this document. The "proof" of meeting acceptance criteria is through compliance with recognized technical and safety standards and demonstration of substantial equivalence to existing legally marketed devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 26, 2018

Premier North America Inc. % Doris Dong Consultant Shanghai CV Technology Co., Ltd. Room 903 of Dongbao Building No. 19 Dongbao Road, Songjiang Area Shanghai, 201613 Cn

Re: K181659

Trade/Device Name: Avologi ENEO Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: OHS Dated: June 13, 2018 Received: June 28, 2018

Dear Doris Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Krause -S

  • for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K181659

Device Name Avologi ENEO

Indications for Use (Describe)

Avologi ENEO is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Information:

510(k) Number:K__________
Date:June 13th, 2018
Type of 510(k) Submission:Traditional
Basis for 510(k) Submission:New device
Applicant / Owner:Premier North America Inc.3301 SW 42ND ST., FORT LAUDERDALE, FL 33312-6828, USA
Manufacturer:Premier Dead Sea Cosmetic Laboratories Ltd.Shaked Street No. 21 Shoham, Israe
Submitter / Contact:Doris Dong (Consultant)Shanghai CV Technology Co., Ltd.Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 ChinaE-mail: doris_d@126.comTel: 86 21-31261348

2. Device Description:

Proprietary Name:Avologi ENEO
Common Name:Light Emitting Diode (LED) device
Classification Name:Laser surgical instrument for use in general and plastic surgery and in dermatology
Product Code:OHS
Device Class:II
Regulation Number:21 CFR 878.4810
Review Panel:General & Plastic Surgery
Indications for use:Avologi ENEO is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
Device Description:Avologi ENEO is a battery operated device that uses low power light spectrum at red and infrared LED, at wavelength of 633±5nm, 830±5nm emitting optical power in a uniform distribution with no hot spots. It is a hand held light emitting diode(LED) device for the treatment of periorbital wrinkles designed for home-use.

3. Predicate Device Identification

K110301--Silk'n FX (Home Skinovations Ltd.) -- August 19, 2011 K 152332--Perfectio LED infrared device (OMM IMPORTS INC DBA ZERO GRAVITY) -- January 27, 2015

4. Non-Clinical Test Conclusion

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R)2012

{4}------------------------------------------------

(Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance

  • IEC 60601-1-2 Edition 4.0 2014-02, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Tests

  • IEC 60601-1-11 Edition 2.0 2015-01, Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment * IEC 62471 First Edition 2006-07, Photobiological Safety of Lamps and Lamp Systems

The patient contact materials in Avologi ENEO are the body housing material of ABS and the head housing material of Stainless steel 304. Both the two materials were tested and found to meet the biocompatibility standards of:

  • ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

  • ISO 10993-5:2009/(R)2014, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity, and

  • ISO 10993-10 Third Edition 2010-08-01, Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity

5. Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data was not including in this submission.

New DevicePredicate DevicePredicate Device
510(k) Number:------K152332K110301
Product Code:OHSOHSOHS
Proprietary Name:Avologi ENEOPerfectio LED infrared deviceSilk'n FX
Manufacturer:Premier Dead Sea CosmeticLaboratories Ltd.OMM IMPORTS INC DBAZERO GRAVITYHome Skinovations Ltd.
Indications for use:Avologi ENEO is an over thecounter device indicated toemit energy in the red and IRregion of the spectrum for usein dermatology for thetreatment of periorbitalwrinkles.Perfectio LED infrared deviceis an over the counter deviceindicated to emit energy in thered and IR region of thespectrum for use indermatology for the treatmentof periorbital wrinkles.Silkn FX is an over the counterdevice indicated to emit energyin the red and IR region of thespectrum for use indermatology for the treatmentof periorbital wrinkles.
Product pictures:Image: Avologi ENEOImage: Perfectio LED infrared deviceImage: Silk'n FX
HandheldYesYesYes
3301 SW 42ND ST., FORT LAUDERDALE, FL 33312-6828, USA
MaterialsABS and stainless steelABS and stainless steelABS and stainless steel
product compositionthe CHARGING CRADLE, anAPPLICATOR and a POWERSUPPLYthe BASE UNIT, anAPPLICATOR and anADAPTERthe BASE UNIT, anAPPLICATOR and anADAPTER
Wavelengths633 ±5nm, 830 ±5nm633 ±5nm, 830 ±5nm633 ±5nm, 830 ±5nm
WaveformConstantConstantConstant
Light sourceLight emitting diode(LED)Light emitting diode(LED)Light emitting diode(LED)
light spectrum regionred and IR region of thespectrumred and IR region of thespectrumred and IR region of thespectrum
Energy source24 pcs LEDs(12pcs 633nmLEDs+ 12pcs 830nm LEDs )over 15cm²25 pcs LEDs(12pcs 633nmLEDs+ 13pcs 830nm LEDs)over 17 cm²24 LEDs(12pcs 633nm LEDs+12pcs 830nm LEDs) over 12cm²
Energy density69mW/cm² for 633nm55mW/cm² for 830nm70mW/cm² for 633nm55mW/cm² for 830nm70mW/cm² for 633nm55mW/cm² for 830nm
Therapeutictemperature range41±2°C / 105.8±35.6°F41±2°C / 105.8±35.6°F41°C
Power supplyAdaptor:100~240V AC50/60Hz 2.4ALithium battery: 3.6Vdc, 2600mAhAdaptor:100~240V AC50/60Hz 0.2ALithium battery: 7.4Vd.c.900mAhAdaptor:100~240V AC50/60Hz 0.4ALithium battery: 2x3.7V,750±50mAh
Handpiece chargingmethodIn CHARGING CRADLEIn Base UnitIn Base Unit
Initial treatment courseFor the first month (4 weeks),treatment should be performed3 times a week for 15-20minutes each time (5-7 minuteson each treatment zone).For the first month (4 weeks),treatment should be performed3 times a week for 15-20minutes each time.(5-7 minuteson each treatment zone).Week 1, first day (day 1): 20min IR light on each areaTwo days later (day 3): 20 minRED light on each areaWeek 2, first day (day 8): 20min IR light on each areaTwo days later (day 10): 20min RED light on each areaWeek 3, first day (day 15): 20min IR light on each areaTwo days later (day 17): 20min RED light on each areaWeek 4, first day (day 22): 20min IR light on each areaTwo days later (day 24): 20min RED light on each area
Maintenance regimeOnce a week for 15-20 minutesOnce a week for 15-20 minutes/
Target PopulationIndividuals with periorbitawrinklesIndividuals with periorbitalwrinklesIndividuals with periorbitallines and wrinkles
Anatomical SitesPeriorbital AreaPeriorbital AreaPeriorbital Area
Location for useOTCOTCOTC
Standard meetIEC60601-1IEC60601-1-2IEC60601-1IEC60601-1-2IEC60601-1IEC60601-1-2
IEC60601-1-11IEC62471IEC60601-1-11IEC62471IEC62471
Label and LabelingMeet FDA's RequirementsMeet FDA's RequirementsMeet FDA's Requirements

6. Substantially Equivalent Comparison Conclusion

{5}------------------------------------------------

Premier North America Inc.
3301 SW 42ND ST., FORT LAUDERDALE, FL 33312-6828, USA

{6}------------------------------------------------

Premier North America Inc. 3301 SW 42ND ST., FORT LAUDERDALE, FL 33312-6828, USA

The Conclusions:

Taking into consideration the table for substantial equivalence, an analysis of safety, indications, intended uses, performance, and technological properties, the proposed device raises no new issues of safety or effectiveness and has been found to be substantially equivalent to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.