K Number
K181656

Validate with FDA (Live)

Date Cleared
2019-03-05

(256 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADORE (polymacon) Spherical Soft Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia or hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 0.75 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye. Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

Frequent/Planned Replacement Wear:
Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.

Disposable Wear:
Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When Prescribed for Daily Disposable Wear the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.

Device Description

The ADORE soft contact lenses are hemispherical shells with molded spherical base curves and molded front surfaces. The ADORE soft contact lenses are fabricated from polymacon, which has been adopted by the United States Adopted Names Council (USAN). The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The nonionic lens material, (polymacon) is a hydrophilic polymer of 2-Hydroxyethyl methacrylate (2-HEMA) and cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 62% polymacon and 38% water by weight when immersed in saline solution.

ADORE lenses are available clear, tinted for visibility/handling, and tinted to enhance or alter the apparent color of the eye. Tinted or color lens designs may be distributed under unique or "private label" trade names. ADORE lenses are available in gray, green, yellow, hazel, blue, white or black which are tinted with one or a combination of one or more of the following 'listed' color additives: D&C Yellow 10, Titanium Dioxide, Iron Oxide (Red), D&C Green No. 6, Reactive Blue 19, C.I. Reactive Black 5.

Lenses that contain a unique tinting pattern are subsequently processed to incorporate the 'listed' color additives, and contain only the amount of color additive needed to accomplish the intended coloring effect. When producing the tinted or color lenses, the manufacturing process alters and/or changes the specifications to the clear version of a contact lens by entrapment of the listed color additives in the center of the contact lens (surrounded by layers of contact lens material) in a location that corresponds to the iris. The color additives used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens. The color lens tinting pattern has a minimum Clear Pupil diameter of 6.0 mm.

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a transparent or colored optical surface. The (polymacon) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

The physical properties of the lens are: Refractive Index 1.435, Light Transmission (clear) greater than 98%, Light Transmission (tinted/color) greater than 98% (at clear region corresponding to pupil, minimum 6.0 mm diameter); Opaque or 0-10% (at tinted region corresponding to iris), Surface Character hydrophilic, Water Content 38±2%, Specific Gravity 1.17 (hydrated), Oxygen Permeability 11.15x10-11 (cm²/sec)(ml O₂)(ml x mm Hg @ 35°C) (revised Fatt method).

The hydrophilic characteristics allow aqueous solution to enter the lens, and in its fully hydrated state the lens is approximately 38% water by weight. ADORE (polymacon) Spherical Soft Contact lenses are available in the spherical configuration with the following features and properties: Chord Diameter 14.00 mm, Center Thickness 0.08 mm (to – 3.00D), Base Curve 8.6 mm, Power Range From D. 0.00 to D. -16.00 (from D. -5.00 a D. -16.00 with 0.50 steps) From D. +0.25 to D. +10.00 (from D. +5.00 to D. +10.00 with 0.50 steps), Colors available Grey, Hazel, Yellow, Blue, White, Green.

ADORE lenses are supplied sterile in saline solution containing blister packages with a base made from polypropylene and a laminated foil seal on top. Blister package labeling is printed with the appropriate lot numbering, expiration dating, and lens parameter identification. Expiration dating has been established based on studies of product stability, package integrity, and validation of the sterilization process.

AI/ML Overview

The provided text describes the ADORE (polymacon) Daily Wear Soft (hydrophilic) Contact Lenses and its substantial equivalence to a predicate device, ChicView (Polymacon) Daily Wear Soft (Hydrophilic) Contact Lenses. The submission focuses on non-clinical testing and refers to the previously established clinical performance of the lens material.

Here's an analysis of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a pass/fail quantifiable manner for the device performance. Instead, it compares the physical and material properties of the new device (ADORE) to its predicate device (ChicView) to demonstrate substantial equivalence. The "reported device performance" refers to the measured properties of the ADORE lenses.

PropertyAcceptance Criteria (Predicate Device - ChicView)Reported Device Performance (ADORE)
Water content38±2%38±2%
Light transmittance (clear)>98%>98%
Refraction Index1.4351.435
Oxygen permeability$11.15 \times 10^{-11}$ (cm²/sec)(ml O₂)(ml x mm Hg @ 35°C) (revised Fatt method)$11.15 \times 10^{-11}$ (cm²/sec)(ml O₂)(ml x mm Hg @ 35°C) (revised Fatt method)
Specific Gravity (hydrated)1.171.17
SterilizationSteam Validated AutoclaveSteam Validated Autoclave
Production methodFully-moldedFully-molded
USAN namePolymaconPolymacon
Indications for UseEquivalentCorrective, cosmetic use
FunctionalityEquivalentRefractive medium

The acceptance rationale for these properties appears to be that the ADORE device properties are identical to those of the legally marketed predicate device, ChicView, thus demonstrating substantial equivalence. The document states, "The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the ADORE contact lens material is equivalent to the predicate devices identified previously."

2. Sample size used for the test set and the data provenance

The document states: "A series of in vitro and in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the ADORE (polymacon) Daily Wear Soft (hydrophilic) Contact Lenses." However, it does not specify the sample size for these tests.

Regarding data provenance: "All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols." The document does not explicitly state the country of origin of the data, nor does it specify if the non-clinical tests were retrospective or prospective. Given the nature of pre-clinical in vitro and in vivo tests for a new device, they are typically prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The document describes non-clinical testing (toxicology, biocompatibility, physical properties) which usually involves laboratory measurements and standardized protocols rather than expert ground truth establishment in the same way clinical studies would. For the comparison to the predicate device, it is based on documented properties of the predicate device rather than expert review of the new device against a ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This information is not provided in the document. Adjudication methods are typically relevant for clinical studies where subjective assessments might be involved, or for establishing ground truth in diagnostic studies. The testing performed for this device appears to be primarily objective non-clinical measurements.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is a contact lens, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a contact lens, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical testing, the "ground truth" would be established by the results of standardized laboratory tests and measurements, following valid scientific protocols and GLP regulations. For example, sterility tests, material property measurements (water content, oxygen permeability), and toxicology assays provide objective data.

For the clinical aspect, the document states: "The clinical performance of the (polymacon) lens material has been previously established, and therefore was not required for this 510(k)." This implies that the clinical "ground truth" for the material's safety and effectiveness was based on prior clinical studies and outcomes data for the polymacon material, likely involving patient wear experience and clinical evaluations.

8. The sample size for the training set

Not applicable. This device is a contact lens; it does not involve machine learning or a "training set."

9. How the ground truth for the training set was established

Not applicable. As stated above, there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 5, 2019

Eyemed Technologies S.r.l. Mr. Franco Ceravolo Managing Director Via al Boscaccio 3 21011 Casorate Sempione (VA) - Italy

Re: K181656

Trade/Device Name: ADORE (polymacon) Daily Wear Soft (hydrophilic) Contact Lenses (Tinted/color) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: December 27, 2018 Received: January 24, 2019

Dear Mr. Ceravolo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

J Angelo Green -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181656

Device Name

ADORE (polymacon) Daily Wear Soft (hydrophilic) Contact Lenses (Tinted/color)

Indications for Use (Describe)

The ADORE (polymacon) Spherical Soft Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia or hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 0.75 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye. Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

Frequent/Planned Replacement Wear:

Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.

Disposable Wear:

Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When Prescribed for Daily Disposable Wear the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CER 801 Subpart D) Over-The-Counter Use (21 CER 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) SUMMARY

SUBMITTER INFORMATION

A. Company Name:Eyemed Technologies S.r.l.
B. Company Address:Via al Boscaccio 321011 Casorate Sempione (VA) - Italy
C. Company Phone:+39-0331201024
Company Fax:+39-0331287154
Company e-mail:affair@eyemed.it
D. Contact person:Mr. Franco CeravoloManaging DirectorEyemed Technologies S.r.l.
E. Date Summary Prepared:May 18, 2018

FICE IDENTIFICATION# DEVICE IDENTIFICATION

A. Common Name:Lenses, Soft Contact, Daily Wear
B. Trade/Proprietary Name:ADORE (polymacon) Daily Wear Soft(hydrophilic) Contact Lenses (Tinted/color)
C. Device Classification:Class II
D. Product Code:LPL; MVN
E. Classification Name:Soft (hydrophilic) Contact Lens
E. Regulation Number:21 CFR 886.5925

LEGALLY MARKETED PREDICATE DEVICE

Predicate device510 (k) Holder510 (k) No.Date cleared
ChicView(Polymacon) DailyWear Soft(Hydrophilic)Contact Lenses(Tinted/Color)JOOWONINNOVATION CO.,LTD.K161098August 22, 2016

{4}------------------------------------------------

DEVICE DESCRIPTION

The ADORE soft contact lenses are hemispherical shells with molded spherical base curves and molded front surfaces. The ADORE soft contact lenses are fabricated from polymacon, which has been adopted by the United States Adopted Names Council (USAN). The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The nonionic lens material, (polymacon) is a hydrophilic polymer of 2-Hydroxyethyl methacrylate (2-HEMA) and cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 62% polymacon and 38% water by weight when immersed in saline solution.

ADORE lenses are available clear, tinted for visibility/handling, and tinted to enhance or alter the apparent color of the eye. Tinted or color lens designs may be distributed under unique or "private label" trade names. ADORE lenses are available in gray, green, yellow, hazel, blue, white or black which are tinted with one or a combination of one or more of the following 'listed' color additives:

Name of ColorantListing
D&C Yellow 1021 CFR 74.3710
Titanium Dioxide21 CFR 73.3126
Iron Oxide (Red)21 CFR 73.3125
D&C Green No. 621 CFR 74.3206
Reactive Blue 1921 CFR 73.3127
C.I. Reactive Black 521 CFR 73.3121

Lenses that contain a unique tinting pattern are subsequently processed to incorporate the 'listed' color additives, and contain only the amount of color additive needed to accomplish the intended coloring effect. When producing the tinted or color lenses, the manufacturing process alters and/or changes the specifications to the clear version of a contact lens by entrapment of the listed color additives in the center of the contact lens (surrounded by layers of contact lens material) in a location that corresponds to the iris. The color additives used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens. The color lens tinting pattern has a minimum Clear Pupil diameter of 6.0 mm.

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a transparent or colored optical surface. The (polymacon) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

{5}------------------------------------------------

The physical properties of the lens are:

Refractive Index1.435
Light Transmission (clear)greater than 98%
Light Transmission (tinted/color)greater than 98% (at clear region corresponding to pupil,
minimum 6.0 mm diameter);
Opaque or 0-10% (at tinted region corresponding to iris)
Surface Characterhydrophilic
Water Content38±2%
Specific Gravity1.17 (hydrated)
Oxygen Permeability11.15x10-11 (cm²/sec)(ml O₂)(ml x mm Hg @ 35°C) (revised
Fatt method)

The hydrophilic characteristics allow aqueous solution to enter the lens, and in its fully hydrated state the lens is approximately 38% water by weight. ADORE (polymacon) Spherical Soft Contact lenses are available in the spherical configuration with the following features and properties:

Chord Diameter14.00 mm
Center Thickness0.08 mm (to – 3.00D)
Base Curve8.6 mm
Power RangeFrom D. 0.00 to D. -16.00 (from D. -5.00 a D. -16.00 with 0.50 steps)From D. +0.25 to D. +10.00 (from D. +5.00 to D. +10.00 with 0.50 steps)
Colors availableGrey, Hazel, Yellow, Blue, White, Green

ADORE lenses are supplied sterile in saline solution containing blister packages with a base made from polypropylene and a laminated foil seal on top. Blister package labeling is printed with the appropriate lot numbering, expiration dating, and lens parameter identification. Expiration dating has been established based on studies of product stability, package integrity, and validation of the sterilization process.

INDICATIONS FOR USE STATEMENT

The ADORE (polymacon) Spherical Soft Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 0.75 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.

Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

Frequent/Planned Replacement Wear:

Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.

Disposable Wear:

Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When Prescribed for Daily Disposable Wear the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.

{6}------------------------------------------------

TESTING

Non-clinical Testing A series of in vitro and in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the ADORE (polymacon) Daily Wear Soft (hydrophilic) Contact Lenses. All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols. Test results of the non-clinical testing on the ADORE (polymacon) Daily Wear Soft (hydrophilic) Contact Lenses demonstrate that:

  • Lenses supplied in blister packages are sterile for the indicated shelflife,
  • . The packaging material and extracts are not toxic and not irritating, and
  • Lens physical and material properties are consistent with currently marketed lenses.
  • Clinical Data The clinical performance of the (polymacon) lens material has been previously established, and therefore was not required for this 510(k).

CONCLUSIONS DRAWN FROM STUDIES

Validity of Scientific Data

Several laboratories under Good Laboratory Practice regulations conducted toxicology studies, microbiology, chemistry, shelf-life stability studies and followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7

Substantial Equivalence

Information presented in this Premarket Notification establishes that the ADORE, (polymacon) Soft Contact Lenses for Daily Wear is as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the requested indication.

Risks and Benefits

The risks of the subject device are the same as those normally attributed to the wearing of daily wear soft contact lenses. The benefits to the patient are the same as those for other daily wear soft contact lenses.

SUBSTANTIAL EQUIVALENCE

The ADORE Soft Contact Lens will be manufactured according to specified process controls and a cGMP quality assurance program currently in place. The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the ADORE contact lens material is equivalent to the predicate devices identified previously. Being similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate devices, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate device identified above.

The following matrix illustrates the equivalencies of the ADORE Soft (hydrophilic) Contact Lens, as well as the substantial equivalent predicate devices.

{7}------------------------------------------------

PREDICATE DEVICES COMPARISON CHART

Table 1

EYEMED TECHNOLOGIES S.R.L.ADORENew deviceJOOWON INNOVATION CO., LTD.CHICVIEWPredicate device
"K" NUMBERSK_K161098
Proprietary nameADORECHICVIEW
CFR Section886.5925886.5925
Pro-codeLPL; MVNLPL; MVN
Classification nameDaily Wear, Soft (hydrophilic)Contact Lens (Class II)Daily Wear, Soft (hydrophilic)Contact Lens (Class II)
Intended /Indications for useIndicated for daily wear for the correction ofrefractive ametropia (myopia and hyperopia)in aphakic and not aphakic persons withnon-diseased eyes and/or possessesrefractive astigmatism not exceeding 0.75diopters. The lens is available clear orcolored and may be used to enhance or alterthe apparent color of the eye.Indicated for daily wear for the correctionof visual acuity in aphakic and not aphakicpersons with nondiseased eyes withmyopia or hyperopia and/or possessesrefractive astigmatism not exceeding 0.75diopters. The lens is available clear orcolored and may be used to enhance or alterthe apparent color of the eye.
FunctionalityThe contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retinaThe contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retina
Production methodFully-moldedFully-molded
USAN namePolymaconPolymacon
Water content38±2%38±2%
Light trasmittance>98%>98%
Refraction Index1,4351,435
Oxygenpermeability$11.15x10^{-11}$ (cm²/sec)(ml O₂)(ml x mm Hg@ 35°C) (revised Fatt method)$11.15x10^{-11}$ (cm²/sec)(ml O₂)(ml x mm Hg@ 35°C) (revised Fatt method)
SterilizationSteam Validated AutoclaveSteam Validated Autoclave
Specific Gravity1.17 (hydrated)1.17 (hydrated)

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.