(116 days)
The SpineEX Sagittae® Lateral Lumber Interbody Fusion Devices are indicated for interbody fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be sketally mature and have completed six months of non-operative treatment. Supplemental fixation is required with SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices. Additionally, the SpineEX devices are intended to be used with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g. posterior pedicle screw and rod systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
The SpineEX Sagittae® Lateral Lumber Interbody Fusion Devices are manufactured out of medical grade Ti-6A1-4V (Grade 5) and Ti-6A-4V (ELI) alloy that conforms to ASTM F1472 and ASTM F136 respectively.
This document, K181531, is a 510(k) Premarket Notification for the SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices. It focuses on demonstrating substantial equivalence to predicate devices, primarily through mechanical performance testing, rather than clinical study data involving AI or human-in-the-loop performance.
Therefore, many of the requested categories related to acceptance criteria for AI performance, clinical study design (sample size for test/training sets, ground truth establishment, expert adjudication, MRMC studies), and "how human readers improve with AI" are not applicable to this specific document as it describes a mechanical implant device, not an AI diagnostic tool.
Here's an analysis based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a quantitative table for clinical performance. Instead, it outlines the types of non-clinical (mechanical) tests performed and a qualitative conclusion regarding their sufficiency and equivalence.
Non-Clinical (Mechanical) Performance Data:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Sufficient strength for intended use and substantial equivalence to predicate devices under various mechanical loads. | The devices were tested according to: - Static axial compression per ASTM 2077 - Static compressive shear per ASTM 2077 - Dynamic axial compression per ASTM 2077 - Dynamic compressive shear per ASTM 2077 - Subsidence per ASTM F2267 Conclusion: "The results of this non-clinical testing show that the strength of the SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices are sufficient for their intended use and is substantially equivalent to legally marketed predicate devices." |
2. Sample sizes used for the test set and the data provenance
- Sample Size for Test Set: Not applicable in the context of clinical images/data for AI. For mechanical testing, the specific number of devices tested is not provided in this summary, but would typically follow ASTM standards.
- Data Provenance: Not applicable in the context of clinical images/data for AI. The device itself is manufactured from medical grade Ti-6A1-4V (Grade 5) and Ti-6A-4V (ELI) alloy.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This document pertains to a mechanical device, not an AI or diagnostic tool requiring expert interpretation of medical images.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. This pertains to expert review of medical data for ground truth establishment, which is not relevant for this device.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is a mechanical implant device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This device does not involve an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Ground Truth for Mechanical Testing: The "ground truth" here is adherence to mechanical performance standards (ASTM 2077, ASTM F2267) and demonstration of "sufficient strength for intended use" and "substantial equivalence" to predicate devices based on these physical properties.
8. The sample size for the training set
- Not applicable. This device does not involve a "training set" in the machine learning sense.
9. How the ground truth for the training set was established
- Not applicable. This device does not involve a "training set" or ground truth establishment in the machine learning sense. The "ground truth" for the device's design and material selection would be established through engineering principles, material science, and regulatory standards for implantable medical devices.
{0}------------------------------------------------
October 5, 2018
Image /page/0/Picture/1 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
SpineEX Inc. % Ms. Meredith May Vice President Empirical Consulting LLC 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K181531
Trade/Device Name: SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: September 17, 2018 Received: September 20, 2018
Dear Ms. May:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name SpineEX Sagittae (R) Lateral Lumbar Interbody Fusion Devices
Indications for Use (Describe)
The SpineEX Sagittae® Lateral Lumber Interbody Fusion Devices are indicated for interbody fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be sketally mature and have completed six months of non-operative treatment. Supplemental fixation is required with SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices. Additionally, the SpineEX devices are intended to be used with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g. posterior pedicle screw and rod systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
5.510(K) SUMMARY
| Submitter's Name: | SpineEX Inc. |
|---|---|
| Submitter's Address: | 4046 Clipper CourtFremont, CA 94538 |
| Submitter's Telephone: | (510) 573-6165 |
| Company Contact Person: | Andrew Rogers |
| Contact Person: | Meredith Lee May MS, RACEmpirical Consulting LLC719-337-7579MMAy@EmpiricalConsulting.com |
| Date Summary was Prepared: | 08 June 2018 |
| Trade or Proprietary Name: | SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices |
| Common or Usual Name: | Intervertebral Fusion Device With Bone Graft, Lumbar |
| Classification: | Class II per 21 CFR §888.3080 |
| Product Code: | MAX |
| Classification Panel: | Division of Orthopedic Devices |
DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION
The SpineEX Sagittae® Lateral Lumber Interbody Fusion Devices are manufactured out of medical grade Ti-6A1-4V (Grade 5) and Ti-6A-4V (ELI) alloy that conforms to ASTM F1472 and ASTM F136 respectively.
INDICATIONS FOR USE
The SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices are indicated for interbody fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have completed six months of non-operative treatment. Supplemental fixation is required with SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices. Additionally, the SpineEX devices are intended to be used with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g. posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
TECHNOLOGICAL CHARACTERISTICS
The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:
{4}------------------------------------------------
- Indications for use ●
- Materials of manufacture ●
- Structural support mechanism ●
- Principles of operation ●
Table 5-1: Predicate Devices
| 510k Number | Trade or Proprietary or ModelName | Manufacturer | PredicateType |
|---|---|---|---|
| K123752 | AccuLiF® | CoAlignInnovation | Primary |
| K123231 | Caliber-L | Globus | Additional |
| K133813 | FLXfit IBFD | ExpandingOrthopedics | Additional |
| K123045 | Brigade® Hyperlorditic System | NuVasive® | Additional |
| K073144 | Timberline IBFD | Lanx | Additional |
Performance Data
The SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices have been tested in the following test modes:
- . Static axial compression per ASTM 2077
- Static compressive shear per ASTM 2077 .
- Dynamic axial compression per ASTM 2077 ●
- Dynamic compressive shear per ASTM 2077 ●
- Subsidence per ASTM F2267 .
The results of this non-clinical testing show that the strength of the SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices are sufficient for their intended use and is substantially equivalent to legally marketed predicate devices.
CONCLUSION
The overall technology characteristics and mechanical performance data lead to the conclusion that the SpineEX Sagittae® Lateral Lumbar Interbody Fusion Devices are substantially equivalent to the predicate devices.
§ 888.3080 Intervertebral body fusion device.
(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.